(90 days)
The Sphincter of Oddi Manometry Module is intended for use by Gastroenterologists or Surgeons to determine intraluminal pressures within the pancreaticobiliary ductal system and motor activity within the sphincter of Oddi zone. The InSIGHT Sphincter of Oddi Manometry Module includes analysis software, but requires skilled interpretation by a physician to make a diagnosis.
The Sphincter of Oddi Manometry Module is intended for use by Gastroenterologists or Surgeons to determine intraluminal pressures within the pancreaticobiliary ductal system and motor activity within the sphincter of Oddi zone. The InSIGHT Sphincter of Oddi Manometry Module includes analysis software, but requires skilled interpretation by a physician to make a diagnosis. The SOM Module is intended for use only with the InSIGHT Model S980000 Gastrointestinal Motility system (K984444). The InSIGHT system consists of three basic subsystems: Signal conditioning and recording hardware, Analysis software, and Probes and transducers.
The provided text describes a medical device called the InSIGHT Sphincter of Oddi Manometry Module and its submission for 510(k) clearance. However, it explicitly states that no formal clinical testing was performed. Therefore, there is no study described that proves the device meets specific acceptance criteria based on performance metrics.
Here's a breakdown of the requested information based on the provided text, highlighting the absence of performance-based acceptance criteria and a corresponding study:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not explicitly defined as performance criteria. The submission focuses on substantial equivalence to predicate devices in terms of safety and performance, and compliance with general safety standards. | The device is stated to exhibit "features equivalent in safety and performance to those included on prior legally marketed devices." No specific performance metrics (e.g., sensitivity, specificity, accuracy, precision) are provided. The device "performs as well as or better than the legally marketed predicate devices" but no data is presented to support this. |
| Compliance with UL 2601-1: Standard for Medical Electrical Equipment, Part 1: General Requirements for Safety. | Design certified through internal design testing to written protocols and outside laboratories. |
| Mitigation or elimination of all known health hazards. | Accomplished by rigorous application of a risk management program, including design modifications, detection and alarming of hazard conditions, and clear labeling in the instruction manual. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not applicable. No test set for performance evaluation was used as no formal clinical testing was performed.
- Data Provenance: Not applicable.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not applicable. No test set was used to establish ground truth for performance evaluation.
4. Adjudication Method for the Test Set
- Not applicable. No test set was used for performance evaluation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No and Effect Size: No MRMC comparative effectiveness study was done. The submission explicitly states, "No formal clinical testing has been performed, nor is any believed to be necessary." Therefore, there is no reported effect size regarding human reader improvement with AI assistance. The device includes analysis software, but relies on "skilled interpretation by a physician to make a diagnosis," indicating a human-in-the-loop system, but its comparative effectiveness was not studied.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No: The device "includes analysis software, but requires skilled interpretation by a physician to make a diagnosis." This indicates a human-in-the-loop system, and no standalone algorithm performance was reported or evaluated.
7. The Type of Ground Truth Used
- Not Applicable for Performance: Since no formal clinical testing or performance study was conducted, no ground truth (e.g., expert consensus, pathology, outcomes data) was established for validating device performance against clinical reality. The "ground truth" for the submission's claims revolves around comparison to technical specifications and general safety standards of predicate devices.
8. The Sample Size for the Training Set
- Not Applicable / Not Provided: The nature of the device (a manometry recording system with analysis software, not explicitly an AI/ML algorithm requiring extensive training data in the modern sense) and the stated lack of formal clinical testing mean no training set sample size is provided or appears relevant to this submission.
9. How the Ground Truth for the Training Set Was Established
- Not Applicable / Not Provided: As no training set is described in the context of performance-based ground truth, this information is not available.
{0}------------------------------------------------
MAY 1 2 1999
Summary of Safety and Effectiveness
Introduction
The Safe Medical Device Act (SMDA) of 1990 requires that in addition to other information submitted in a 510(k), medical device manufacturers submit a summary of information regarding safety and effectiveness for the device subject to the 510(k). The summary is to include detailed information regarding adverse health effects of the device. This Summary of Safety and Effectiveness document is intended to comply with the SMDA requirement. FDA will make this summary available to the public within 30 days following a finding of substantial equivalence.
510(k) Submitted by
Sandhill Scientific, Inc. 8955 S. Ridgeline Blvd., #500 Highlands Ranch, CO 80126
USA Contact Person
Linda L. Diederich, MT-ASCP VP of Marketing & Customer Relations Tel: 303-470-7020 / Fax: 303-470-2975
Date Prepared
January 22, 1999
Trade Name of Device
InSIGHT Sphincter of Oddi Manometry Module
Common Name of Device
Sphincter of Oddi Manometry
Classification Name
System, Gastrointestinal Motility (Electrical)
510(k) Classification
Class II
{1}------------------------------------------------
K990441 PAGE 2 OF 3
Comparison to Predicate Devices
The InSIGHT Sphincter of Oddi Manometry (SOM) Module is a manometry recording device with features equivalent in safety and performance to those included on prior legally marketed devices. The InSIGHT SOM Module is Substantially Equivalent to:
K932306 Microinfuser, Manufactured by Sandhill Scientific, Inc.
K900058, Wilson-Cook Biliary Motility Catheter, Manufactured by Wilson-Cook Medical. Inc.
Pre-Amendment, Hewlett-Packard Model 7788-A. Manufactured by Hewlett-Packard Corporation.
Device Description and Intended Use
The Sphincter of Oddi Manometry Module is intended for use by Gastroenterologists or Surgeons to determine intraluminal pressures within the pancreaticobiliary ductal system and motor activity within the sphincter of Oddi zone. The InSIGHT Sphincter of Oddi Manometry Module includes analysis software, but requires skilled interpretation by a physician to make a diagnosis.
Technological Characteristics
The SOM Module is intended for use only with the InSIGHT Model S980000 Gastrointestinal Motility system (K984444). The InSIGHT system consists of three basic subsystems. These are:
Signal conditioning and recording hardware
This subsystem includes the hardware that amplifies the transduced signals for recording. provides electrical isolation for safety, and processes the signals for subsequent analysis.
Analysis software
The software displays the waveforms, shows the analyzed results and incorporates these results into reports.
Probes and transducers
These are used to generate the signal that is recorded. Sandhill does not manufacture the transducers and probes that are intended for use with the system, but these carry the manufacturers' approval to market and meet the manufacturers' safety requirements for their use.
The technological characteristics of the InSIGHT SOM Module are the same as, or perform equivalently to, the predicate devices.
Applicable Standards and Non-Clinical Testing
In compliance with the company's Design Controls procedures, the InSIGHT Model S980000 system has been designed to meet the requirements of the following standard:
{2}------------------------------------------------
Medical Electrical Equipment. Part 1: General Requirements
990441
PAGE 3 OF 3
UL 2601-1: Standard for Medical Electrical Equipment, Part 1: General Requirements for Safety
Certification of the design to the above standards and the device design specifications is through a planned combination of internal design testing to written protocols and outside laboratories. No formal clinical testing has been performed, nor is any believed to be necessary.
Potential Adverse Health Effects
The InSIGHT SOM Module has been designed to either completely eliminate or mitigate all known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program. One or more of the following means (in order of preference) was used to implement mitigation of health hazards identified by the risk management program:
-
- Design modifications.
-
- Detection of hazard conditions and alerting of the user through alarms and visual indications.
-
- Identification of any potentially undetectable health hazard conditions in the instruction manual and other device labeling.
The user must be qualified in biliary diagnostic procedures, trained in the use of the system, and must be familiar with all labeling and instructions for use associated with the equipment. Many device injuries are due to user error and failure to follow the instructions for use. The user of the device is advised to thoroughly understand the use of the equipment, and familiarize themselves with the location and function of all controls and alarms prior to using the equipment.
Sandhill Scientific believes that the InSIGHT SOM Module is safe and effective when used as instructed by knowledgeable and trained personnel, and performs as well as or better than the legally marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
MAY 1 2 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Linda L. Diederich, MT-ASCP Vice-President of Marketing & Customer Relations Sandhill Scientific, Inc. 8955 South Ridgeline Boulevard Unit #500 Highlands Ranch, Colorado 80126
Re: K990441 InSIGHT Sphincter of Oddi Manometry Module, Model S981300 Dated: February 9, 1999 Received: February 11, 1999 Requiatory Class: II 21 CFR 876.1725/Procode: 78 FFX
Dear Ms. Diederich:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number:
InSIGHT Sphincter of Oddi Manometry Module Device Name:
Indications for Use:
The Sphincter of Oddi Manometry Module is intended for use by Gastroenterologists or Surgeons to determine intraluminal pressures within the pancreaticobiliary ductal system and motor activity within the sphincter of Oddi zone. The InSIGHT Sphincter of Oddi Manometry Module includes analysis software, but requires skilled interpretation by a physician to make a diagnosis.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use
:
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Dev 510(k) Number.
OR
§ 876.1725 Gastrointestinal motility monitoring system.
(a)
Identification. A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.(b)
Classification. Class II (performance standards).