(35 days)
The Anatomica Humeral Stem/Head is intended for cemented use in treatment of the following:
- Advanced wear and tear of the shoulder joint resulting from degenerative, posttraumatic or rheumatoid arthritis.
- Omarthrosis.
- Rheumatoid arthritis.
- Revision of shoulder prosthesis.
- Traumatology: the only cone to be used in traumatological indications is the traumatology cone.
The Anatomica Humeral Stem Component is intended for replacement of the proximal portion of the humerus. The Anatomica Humeral Stem is a metallic humeral component manufactured from forged stainless steel alloy (ISO 5832-9). It is available in four sizes which are determined by the diameter of the stem. The proximal aspect of the stem is "trumpet shaped" to match the metaphysis of the humeral shaft. Rotational stability is ensured by a proximal fin on the lateral aspect. Two holes in the fin and one in the medial stem body allow for attachment of the tuberosities via sutures/wires. The stem features a slit-ball head fixed at an angle of 135°. An impact screw and expansion cone are inserted through a shaft in the lateral aspect of the stern and into the slit-ball head at the time of surgery. This forces the slit-ball feature to spread open allowing fixation of one of the humeral heads used with the system. The use of the slit ball has the advantage of allowing fixation of the head in a variety of angles. A traumatology cone identical to the expansion cone, except with positioning pegs for head placement, is also available. The heads are manufactured from cast CoCrMo alloy (ISO 5832-4). They are available in 10 sizes with diameters ranging from 40-52mm (in 2mm increments). Sizes 48, 50 and 52mm are offered in two head heights. The spherical female recess for attachment to the slit-ball head of the stem is situated eccentrically to the radius of the humeral head. This recess also features two internal holes for defining position in fracture cases (used in conjunction with the traumatology cone).
The provided text is a 510(k) summary for the "Sulzer Orthopedics Anatomica Humeral Stem/Heads." This document describes a medical device and its intended use, but it does not contain information about acceptance criteria for device performance or a study demonstrating that the device meets such criteria.
The document primarily focuses on:
- Product Description: Detailing the materials, design, and sizes of the humeral stem and heads.
- Specific Diagnostic Indications: Listing the conditions for which the device is intended (e.g., arthritis, omarthrosis, revision of shoulder prosthesis, traumatology).
- Substantial Equivalence: Comparing the device to other legally marketed predicate devices. It states that "Static and Dynamic Testing indicated that the device would survive physiologic loading," but does not provide details of this testing, specific acceptance criteria, or the results.
- FDA Review Letter: The letter indicates that the FDA has determined the device to be substantially equivalent to devices marketed prior to May 28, 1976.
Therefore, I cannot populate the table or answer most of the questions as the required information is not present in the provided text.
Here is what can be inferred or explicitly stated from the document, though it falls short of the detailed request:
1. Table of Acceptance Criteria and Reported Device Performance
- Acceptance Criteria: Not specified in the document.
- Reported Device Performance: "Static and Dynamic Testing indicated that the device would survive physiologic loading." No specific performance metrics (e.g., fatigue life, strength thresholds) or quantitative results are provided.
2. Sample size used for the test set and the data provenance:
- Sample size: Not specified.
- Data provenance: Not specified, but implied to be laboratory testing (static and dynamic).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable as no clinical study or expert-based ground truth is described. The "testing" referred to is mechanical.
4. Adjudication method for the test set:
- Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:
- No, this document describes a mechanical orthopedic implant, not an AI or imaging device that would typically undergo an MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable as this is not an algorithm or AI device.
7. The type of ground truth used:
- For the engineering tests, the "ground truth" would be established mechanical engineering principles and benchmarks for orthopedic implants (e.g., ISO standards for an implant or similar specific standards). The document only states the device "would survive physiologic loading," implying it met some pre-defined mechanical performance criteria.
8. The sample size for the training set:
- Not applicable as this is not a machine learning/AI device.
9. How the ground truth for the training set was established:
- Not applicable as this is not a machine learning/AI device.
{0}------------------------------------------------
2/18/99
510(k) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this is to serve as a Summary of Safety and Effectiveness for the Sulzer Orthopedics Anatomica Humeral Stem/Heads.
| Manufacturer: | Sulzer Orthopedics Ltd.Grabenstrasse 25CH 6341 Baar, Switzerland |
|---|---|
| US Designated Agent: | Sulzer Orthopedics Inc.9900 Spectrum DriveAustin, Texas 78717(512) 432-9900 |
| Date: | January 15, 1998 |
| Contact Person: | Mitchell A. Dhority, RACManager, Regulatory Affairs |
| Classification Name: | Shoulder joint metal/polymer semi-constrained cementedprosthesis - 21 CFR 888.3660 |
| Common/Usual Name: | Humeral Stem and Head Components |
| Trade/Proprietary Name: | Sulzer Orthopedics Anatomica Humeral Stem/Head |
PRODUCT DESCRIPTION
Humeral Stem
The Anatomica Humeral Stem Component is intended for replacement of the proximal portion of the humerus.
The Anatomica Humeral Stem is a metallic humeral component manufactured from forged stainless steel alloy (ISO 5832-9). It is available in four sizes which are determined by the diameter of the stem. The proximal aspect of the stem is "trumpet shaped" to match the metaphysis of the humeral shaft. Rotational stability is ensured by a proximal fin on the lateral Two holes in the fin and one in the medial stem body allow for attachment of the aspect. tuberosities via sutures/wires.
The stem features a slit-ball head fixed at an angle of 135°. An impact screw and expansion cone are inserted through a shaft in the lateral aspect of the stern and into the slit-ball head at the time of surgery. This forces the slit-ball feature to spread open allowing fixation of one of the humeral heads used with the system. The use of the slit ball has the advantage of allowing fixation of the head in a variety of angles. A traumatology cone identical to the expansion cone, except with positioning pegs for head placement, is also available.
Humeral Heads
The heads are manufactured from cast CoCrMo alloy (ISO 5832-4). They are available in 10
{1}------------------------------------------------
sizes with diameters ranging from 40-52mm (in 2mm increments). Sizes 48, 50 and 52mm are offered in two head heights. The spherical female recess for attachment to the slit-ball head of the stem is situated eccentrically to the radius of the humeral head. This recess also features two internal holes for defining position in fracture cases (used in conjunction with the traumatology cone).
SPECIFIC DIAGNOSTIC INDICATIONS
The Anatomica Humeral Stem/Head is intended for cemented use in treatment of the following:
-
- Advanced wear and tear of the shoulder joint resulting from degenerative, posttraumatic or rheumatoid arthritis.
-
- Omarthrosis.
-
- Rheumatoid arthritis.
- Revision of shoulder prosthesis. 4.
-
- Traumatology: the only cone to be used in traumatological indications is the traumatology cone.
SUBSTANTIAL EQUIVALENCE
The Anatomica Humeral Prosthesis is similar to the Sulzer Orthopedics Select Shoulder Titanium Humeral Prosthesis, the Sulzer Orthopedics Select Shoulder CoCr Humeral Stem Prosthesis, the Orthomet/3M Modular Neer II Shoulder System, the Zimmer Fenlin Total Shoulder, the Smith & Nephew Richards Cofield Shoulder, the Kirschner/Biomet Atlas Shoulder, the Kirschner/Biomet Mod II-C Shoulder, the Biomet Bio-Modular Total Shoulder, and the Depuv Global Total Shoulder System.
Static and Dynamic Testing indicated that the device would survive physiologic loading.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles that are connected to each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 8 1999
Mr. Mitchell A. Dhority Manager, Regulatory and Clinical Affairs Sulzer Orthopedics Inc. 9900 Spectrum Drive Austin, Texas 78717
K990137 Re: Anatomica Humeral Stem/Head Components Trade Name: Regulatory Class: III KMS Product Code: January 12, 1999 Dated: Received: January 14, 1999
Dear Mr. Dhority:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory 1 In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{3}------------------------------------------------
Page 2 - Mr. Mitchell A. Dhority
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| 510(k) Number (if known): | K990137 |
|---|---|
| --------------------------- | --------- |
Device Name: Anatomica Humeral Stem/Heads
Indications for Use:
The Anatomica Humeral Stem/Head is intended for cemented use in treatment of the following:
and the control of the comments of the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the contrib
- Advanced wear and tear of the shoulder joint resulting from degenerative, 1. posttraumatic or rheumatoid arthritis.
- Omarthrosis. 2.
-
- Rheumatoid arthritis.
- Revision of shoulder prosthesis. 4.
- Traumatology: the only cone to be used in traumatological indications is the 5. traumatology cone.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---|---|
| (Division Sign-Off) | |
| Division of General Restorative Devices | |
| 510(k) Number | K990137 |
Prescription Use
OR
Over-the Counter Use
(Optional Format 1-2-96)
§ 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.
(a)
Identification. A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”