K Number
K990082
Date Cleared
1999-04-09

(88 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reusable sensor for continuous non-invasive functional arterial oxygen saturation and pulse rate monitoring for patients weighing more than 30 kg. with the Datex Satlite and AS/3 monitors.

Device Description

The Epic Medical Equipment Services. Inc. E403-09 SpO2Finger Sensor is a reusable clip-on finger sensor with the optical components, protected in a hard plastic case. It functions without skin penetration, electrical contact, or heat transfer. This electrooptical sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The E403-09 Finger Sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector.

The E403-09 is three feet in length and is terminated in a DB9 style connector as is the predicate sensor. The E403-09 is used with E710-09 10 foot cable which both extends the length of the E403-09 as well as adapts the DB9 connector to a Nicolay style connector which connects to the oximeter.

The E412-09 is equivalent to the E403-09 (3 foot) plus E710-09 (10 foot). The E412-09 is 12 feet long and terminates with a Nicolay style connector.

AI/ML Overview

This document describes the 510(k) submission for the Epic Medical Equipment Services E403-09 SpO2 Finger Sensor. The information provided outlines the device's technical characteristics, non-clinical tests, and limited details on clinical tests.

Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria for the E403-09 SpO2 Finger Sensor. However, it indicates that the device's accuracy was established through comparison with blood samples analyzed on a multiwavelength hemoximeter. The conclusion states that "Clinical validation studies demonstrates the successful use of the sensor and its ability to provide accurate information."

Therefore, based on the provided text, the acceptance criterion implicitly relates to demonstrating accurate SpO2 measurements. No specific numerical performance metrics (e.g., accuracy range, bias, precision) are given in the summary.

Acceptance Criteria (Implied)Reported Device Performance
Device provides accurate functional arterial oxygen saturation.Clinical validation studies demonstrate "successful use of the sensor and its ability to provide accurate information."

Study Information

The document provides very limited details about the clinical study.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not specified.
  • Data Provenance: Not specified, but the study was described as "Controlled hypoxia studies."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Not specified. The ground truth method was "blood samples analyzed on a multiwavelength hemoximeter," implying laboratory analysis rather than expert consensus on images or interpretations.

4. Adjudication Method for the Test Set

  • Not applicable. The ground truth was established by instrumental analysis ("multiwavelength hemoximeter") of blood samples, not by human expert review requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This device is an SpO2 finger sensor, not an AI-assisted diagnostic imaging device. No MRMC study was conducted or is relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Yes, this was a standalone performance study. The device's SpO2 measurements were compared directly against blood gas analysis, without human interpretation as part of the primary measurement pathway.

7. The Type of Ground Truth Used

  • Clinical Ground Truth: Blood samples analyzed on a multiwavelength hemoximeter. This is a common and accepted method for establishing the ground truth for arterial oxygen saturation.

8. The Sample Size for the Training Set

  • Not applicable. This device is a sensor and does not utilize machine learning or AI algorithms requiring a training set in the conventional sense. Its "training" or calibration would be inherent to its physical design and factory calibration processes rather than a data-driven training set.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set mentioned for this device.

{0}------------------------------------------------

Epic Medical Equipment Services

Page
940082

510(k) Summary of Safety and Effectiveness Information 10.0

Date:Establishment:7 January, 1999Epic Medical Equipment Services1800 10th Street Suite 300Plano, TX 75074
Official Correspondent:Jeffrey SecundaVP R&D972-801-9854972-801-9859 (fax)
Model Number /Name:E403-09 SpO2 Finger Sensor
Classification Name:Oximeter - 74 DQA (CV)21 CFR 820.2700 (Class II)
Predicate Devices:Epic Series E100 SpO2 Finger Sensor (K970098)Satlite Plus Pulse Oximeter (K905140)

Description of the Device:

The Epic Medical Equipment Services. Inc. E403-09 SpO2Finger Sensor is a reusable clip-on finger sensor with the optical components, protected in a hard plastic case. It functions without skin penetration, electrical contact, or heat transfer. This electrooptical sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The E403-09 Finger Sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector.

The E403-09 is three feet in length and is terminated in a DB9 style connector as is the predicate sensor. The E403-09 is used with E710-09 10 foot cable which both extends the length of the E403-09 as well as adapts the DB9 connector to a Nicolay style connector which connects to the oximeter.

The E412-09 is equivalent to the E403-09 (3 foot) plus E710-09 (10 foot). The E412-09 is 12 feet long and terminates with a Nicolay style connector.

Statement of the Intended Use:

"Reusable sensor for continuous non-invasive functional arterial oxygen saturation and pulse rate monitoring for patients weighing more than 30 kg. with the Datex Satlite and AS/3 monitors."

{1}------------------------------------------------

510(k) Summary of Safety and Effectiveness Information - continued 10.0

Technological Characteristics Summary:

The E403-09 utilizes a red and infrared LED of the same or tighter specifications as the predicate device. The designated oximeter energizers the LED's which transmit light through a pulsatile arterial bed. The photodiode is of comparable design as the predicate device. The photodiode senses the signal strength of the two LED's which vary with the amount of energy transmitted through the tissue. The oximeter receives the signal from the photodiode and calculates a value for functional oxygen saturation (SpO2).

Discussion of Non-Clinical Tests:

Testing has been carried out to ensure that the proposed devices meets the requirements of the following standards: Mechanical, electrical, thermal, and environmental conditions: EN 60601-1 Electromagnetic compatibility: EN 60601-1-2 Biocompatibility: ISO 10993-1 (EN 30993-1). The device also meets relevant performance and safety requirements from the standard for Pulse Oximeters, EN 865 (ASTM F1415-92).

Discussion of Clinical Tests:

Controlled hypoxia studies were conducted in which device data was compared to blood samples analyzed on a multiwavelength hemoximeter. These tests were conducted to establish the accuracy of the E403-09.

Conclusions Demonstrating Safety Effectiveness and Performance:

The testing carried out for the E403-09 indicates that it meets its design and performance requirements. Clinical validation studies demonstrates the successful use of the sensor and its ability to provide accurate information.

Signed:

Sa

Date: 1/7/99

Jeffrey'S Epic Medical Equipment Services

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine. The caduceus is depicted with a staff entwined by two snakes and topped with wings.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 19 1999

Mr. Jeffrey Secunda Vice President R & D Epic Medical Equipment Services 1800 10th Street, Suite 300 Plano, TX 75074

Re: K990082 E403-09 SpO2 Finger Sensor Requlatory Class: II (two) Product Code: DQA Dated: January 7, 1999 Received: January 11, 1999

Dear Mr. Secunda:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any

{3}------------------------------------------------

Page 2 - Mr. Edward F. Waddell

obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, ² Misbranding by reference to premarket notification² (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D Director Division of Cardiovascular, Respiratory, and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

11.0 Intended Use

Statement of the Indicated Use:

"Reusable sensor for continuous non-invasive functional arterial oxygen saturation and - reasure benever in for patients weighing more than 30 kg. with the Datex Satlite and AS/3 monitors."

Atl. A. Giall.

(Division Sign-Off) Division of Cardiovascular, Re and Neurological Devices 510(k) Number

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).