K Number
K984492
Manufacturer
Date Cleared
1999-03-15

(88 days)

Product Code
Regulation Number
888.3358
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tigran TM Hip Prosthesis for total hip arthoplasty consists of three basic parts: femoral heads, femoral stems, and the instrumentation necessary for proper implantation of these components. The Tigran TM Hip Prosthesis is designed for implantation into the human femur in total or hip replacement and is indicated for the following:

  1. Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur
  2. Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis;
  3. Patients suffering from disability due to previous fusion;
  4. Patients with previously failed endoprostheses and/or total hip components in the affected extremity:
  5. Patients with acute femoral neck fractures.
Device Description

The Tigran™ Femoral Prosthesis for hip arthroplasty consists of three basic parts: femoral heads, femoral stems, and the instrumentation necessary for proper implantation of these components. The femoral stem is made from Titanium Ti-6A1-4V Alloy, a high fatigue strength material with a long history of clinical use and exceptional The surface of the femoral stem is prepared with a surface finish and biocompatibility. pattern to provide optimal initial and lasting fixation in bone. The femoral stem has a collar and is modular in design, incorporating a Morse-type tapered neck and utilizing modular femoral heads. The Femoral stem is straight and with a rectangular crosssectional shape to provide optimal, rotational stability. The Tigran™ Prosthesis Femoral Stem is available in four neck sizes (short, medium, long and extra long).

The femoral head is fabricated from Cobalt-Chromium-Molybdenum Alloy and comes in three neck configurations (short, medium and long) and in one head diameter (28mm). The various neck lengths provided allow this prosthesis to fit different variations found in the human femora.

The titanium and chrome-cobalt alloy components can be used together, although they are made from dissimilar alloys. In vitro electrochemical corrosion studies have shown that no detrimental changes occur when the two alloys are coupled.

Special instruments are available to assist in mating, disassembling, and the accurate installation of the implant components.

The Tigran™ Femoral Prosthesis (femoral stem and head) may be used with any acetabular prosthesis with an articulating surface made of ultra high molecular weight polyethylene (UHMWP) and designed to be used with a 28 mm CoCrMo Alloy femoral head.

AI/ML Overview

The provided text describes a 510(k) summary for the Tigran™ Hip Prosthesis. However, it does not include detailed information regarding specific acceptance criteria, studies proving device performance against those criteria, sample sizes for test sets, data provenance, expert qualifications, or adjudication methods for ground truth, which would typically be found in a more comprehensive clinical trial report or a detailed premarket approval (PMA) submission.

The document focuses on demonstrating substantial equivalence to predicate devices for regulatory clearance rather than providing detailed performance data against specific acceptance criteria.

Here's an analysis of what information is available related to your request:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not explicitly stated as quantifiable metrics. The document implies the acceptance criteria are met if the device demonstrates "substantial equivalence" to predicate devices and passes "appropriate testing standards / requirements" for mechanical testing, and shows "safety and performance" through clinical testing. No specific performance thresholds, such as a minimum tensile strength, fatigue life, or clinical success rate, are provided.
  • Reported Device Performance:
    • Non-Clinical (Mechanical): "Non-clinical mechanical testing was performed to demonstrate that the femoral stem met or exceed the appropriate testing standards / requirements." No specific quantitative results (e.g., actual fatigue life, material properties) are reported.
    • Clinical: "Long-term clinical testing was performed and reported on to demonstrate the safety and performance of the Tigran™ Hip Prosthesis." Again, no specific quantitative outcomes (e.g., revision rates, pain scores, functional outcome measures) are reported.
    • Electrochemical Corrosion: "In vitro electrochemical corrosion studies have shown that no detrimental changes occur when the two alloys are coupled." This is a qualitative statement of performance.
Acceptance Criteria (Implied)Reported Device Performance
Substantial equivalence to predicate devices (K964769, K915594)"The Tigran ™ Hip Prosthesis is substantially equivalent to the Zimmer Versys Hip System (510(k) Number K 964769) and the DePuy Stability™ Hip System (K 915594)." This is the primary regulatory acceptance criterion for a 510(k). The document further states differences between the proposed device and predicate devices are not significant.
Compliance with appropriate mechanical testing standards/requirements for femoral stem"Non-clinical mechanical testing was performed to demonstrate that the femoral stem met or exceed the appropriate testing standards / requirements." (Specific standards or results not provided).
Safety and performance shown through clinical testing"Long-term clinical testing was performed and reported on to demonstrate the safety and performance of the Tigran™ Hip Prosthesis." (Specific outcomes or results not provided).
No detrimental electrochemical changes between dissimilar alloys"In vitro electrochemical corrosion studies have shown that no detrimental changes occur when the two alloys [titanium and chrome-cobalt alloy] are coupled." (Specific study details not provided).
Compliance with general controls provisions of the Act (e.g., annual registration, listing, GMP, labeling)The FDA letter (page 3) confirms the device can be marketed subject to these provisions, indicating that the submission was deemed compliant in principle.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: Not specified.
  • Data Provenance: Not specified. The document mentions "Long-term clinical testing was performed and reported on," but provides no details about the study design, number of patients, or geographic location.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • This question is not applicable as the document does not describe a study involving expert review for establishing ground truth, such as in image analysis or diagnostic AI. Instead, it refers to clinical testing for a physical orthopedic implant.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • This question is not applicable as the document does not describe a study involving expert review for establishing ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This question is not applicable. The device is a hip implant, not an AI diagnostic or assistance tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This question is not applicable. The device is a hip implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • For the clinical testing, the "ground truth" would likely be based on patient outcomes data, considering the nature of a hip prosthesis. This would typically involve measures like revision rates, pain relief, functional scores, and radiographic evidence of proper fixation and absence of complications over time. However, the document only states that the testing demonstrated "safety and performance" without detailing the specific outcome measures or the "ground truth" criteria.
  • For mechanical testing, the "ground truth" would be established by engineering standards and specifications (e.g., ISO standards for orthopedic implants), which define acceptable performance limits for properties like fatigue strength, wear, and corrosion resistance.

8. The sample size for the training set

  • This question is not applicable. The device is a physical hip implant. The concept of a "training set" is typically associated with machine learning or AI models, which is not relevant here.

9. How the ground truth for the training set was established

  • This question is not applicable for the same reason as above.

In summary, the provided 510(k) summary is a regulatory document focused on demonstrating "substantial equivalence" to predicate devices, rather than a detailed scientific publication outlining specific performance criteria and comprehensive study results. Many of the questions about detailed study design, sample sizes, and ground truth establishment are not addressed in this type of document.

{0}------------------------------------------------

-Ponmed Limited

Confidential

Tigran 510(k)

P981117

Date of current draft 226/99

Section 16.0 510(k) Summary

K984492

510K(k) SUMMARY

SUBMITTER:

Ponmed Limited

DATE PREPARED: November 17th, 1998 Tigran™ Hip Prosthesis DEVICE NAME: CLASSIFICATION NAME: Femoral Hip Prosthesis

PREDICATE DEVICES:

The Tigran TM Hip Prosthesis is substantially equivalent to the Zimmer Versys Hip System (510(k) Number K 964769) and the DePuy Stability™ Hip System (K 915594).

Device Description:

The Tigran™ Femoral Prosthesis for hip arthroplasty consists of three basic parts: femoral heads, femoral stems, and the instrumentation necessary for proper implantation of these components. The femoral stem is made from Titanium Ti-6A1-4V Alloy, a high fatigue strength material with a long history of clinical use and exceptional The surface of the femoral stem is prepared with a surface finish and biocompatibility. pattern to provide optimal initial and lasting fixation in bone. The femoral stem has a collar and is modular in design, incorporating a Morse-type tapered neck and utilizing modular femoral heads. The Femoral stem is straight and with a rectangular crosssectional shape to provide optimal, rotational stability. The Tigran™ Prosthesis Femoral Stem is available in four neck sizes (short, medium, long and extra long).

The femoral head is fabricated from Cobalt-Chromium-Molybdenum Alloy and comes in three neck configurations (short, medium and long) and in one head diameter (28mm). The various neck lengths provided allow this prosthesis to fit different variations found in the human femora.

The titanium and chrome-cobalt alloy components can be used together, although they are made from dissimilar alloys. In vitro electrochemical corrosion studies have shown that no detrimental changes occur when the two alloys are coupled.

Special instruments are available to assist in mating, disassembling, and the accurate

000003

{1}------------------------------------------------

· Ponned Limited

ConfidentialFilenameTigran 510(k)File No.P981112Date of current draft2/26/99
------------------------------------------------------------------------------------------------

installation of the implant components.

The Tigran™ Femoral Prosthesis (femoral stem and head) may be used with any acetabular prosthesis with an articulating surface made of ultra high molecular weight polyethylene (UHMWP) and designed to be used with a 28 mm CoCrMo Alloy femoral head.

Intended Use:

The Tigran TM Hip Prosthesis for total hip arthoplasty consists of three basic parts: femoral heads, femoral stems, and the instrumentation necessary for proper implantation of these components. The Tigran TM Hip Prosthesis is designed for implantation into the human femur in total or hip replacement and is indicated for the following:

    1. Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur
    1. Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis;
    1. Patients suffering from disability due to previous fusion;
    1. Patients with previously failed endoprostheses and/or total hip components in the affected extremity:
    1. Patients with acute femoral neck fractures.

This indication statement is essentially the same as the indication statement for the predicate device.

Comparison to Predicate Devices:

All hip systems listed in the previous table are substantially equivalent to each other and to the Tigran TM Hip Prosthesis which are intended for cementless fixation into the intramedullary canal for pathological or degenerative conditions involving the femur and / or acetabulum. All predicate devices feature a porous or roughened surface and are modular in design. All incorporate pure titanium, a Titanium alloy, or hydroxy apetitie as a surface texture coating or interface for the femoral stem. Each has Morse-type proximal neck taper that mates with a femoral head which, in turn articulates upon the acetabular component. All predicate devices and the Tigran TM Hip Prosthesis are manufactured from metal alloys that have a history of successful clinical use in orthopedic applications.

000004

{2}------------------------------------------------

  • Ponned Limited
ConfidentialFilenameFile No.Date of current draft
Tigran 510(k)P9811122/26/99

The Tigran ™ Hip Prosthesis is substantially equivalent to the Zimmer Versys Hip System (510(k) Number K 964769) and the DePuy Stability™ Hip System (K 915594).

Examination of the information pertaining to the Tigran ™ Hip Prosthesis demonstrates that this device is substantially equivalent in composition, intended use, packaging and labeling to other hip prostheses currently approved for commercial distribution in the United States by the FDA. There are no significant differences between these marketed products and our proposed device.

Summary of Non-Clinical Tests:

Non-clinical mechanical testing was performed to demonstrate that the femoral stem met or exceed the appropriate testing standards / requirements

Clinical Test Results:

Long-term clinical testing was performed and reported on to demonstrate the safety and performance of the Tigran™ Hip Prosthesis.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 5 1999

Jeffrey R. Shideman, Ph.D. Ponmed Limited 7307 Glouchester Drive Edina, Minnesota 55435

Re: K984492 Tigran™ Hip Prosthesis Trade Name: Regulatory Class: II Product Codes: LPH and JDI Dated: November 17, 1998 December 17, 1998 Received:

Dear Dr. Shideman:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{4}------------------------------------------------

Page 2 - Jeffrey R. Shideman, Ph.D.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health

Enclosure

{5}------------------------------------------------

Page / of /

510 (k) NUMBER (IF KNOWN): K984492

DEVICE NAME: Tigran Hip Prosthesis

INDICATIONS FOR USE:

Indications for Use Statement

    1. Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur
  • 2). Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis;
    1. Patients suffering from disability due to previous fusion;
    1. Patients with previously failed endoprostheses and/or total hip components in the affected extremity;
    1. Patients with acute femoral neck fractures.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use 11/1 (Per 21 CFR 801.109)

OR

Over-The-Counter-Use
(Optional Format 1-2-96)

to coely

(Division Sign-Off)
Division of General Restorative Devices K984492
510(k) Number __

§ 888.3358 Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis.

(a)
Identification. A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.(b)
Classification. Class II.