K Number
K984372
Date Cleared
1998-12-14

(7 days)

Product Code
Regulation Number
862.1660
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Boehringer Mannheim Elecsys® CalCheck™ Troponin T is used to verify the calibration assignment for the Boehringer Mannheim Elecsys Troponin T assay.

Elecsys CalCheck Troponin T calibration verification solutions comprise three levels - low, mid. and high - each with a defined Troponin T concentration. The low solution concentration is near the lower detection limit of the assay. The middle solution is in the middle or at the clinically critical point of the measuring range. The high solution is near the upper limit of the measuring range.

The Elecsys CalCheck Troponin T is intended for use in periodic verification of the calibration of the Elecsys Troponin T assay.

Device Description

The Boehringer Mannheim Elecsys CalCheck Troponin T is manufactured using bovine serum albumin, human recombinant Troponin T, stabilizers, and preservatives. The analyte is appropriately spiked into the CalCheck matrix to the correct CalCheck concentration levels.

AI/ML Overview

The provided text describes the 510(k) summary for the "Elecsys CalCheck Troponin T" device. This device is a calibration verification material, not a diagnostic device that measures patient samples. Therefore, the acceptance criteria and study details are focused on the performance of the calibration material itself (value assignment and stability), rather than diagnostic accuracy metrics like sensitivity, specificity, or AUC which are typical for devices that provide patient results.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Value AssignmentThe Elecsys® CalCheck™ Troponin T was evaluated for value assignment. (No specific quantitative criteria or results are provided, only that it was evaluated.)
StabilityThe Elecsys® CalCheck™ Troponin T was evaluated for stability. (No specific quantitative criteria or results are provided, only that it was evaluated.)

Note on Acceptance Criteria: The document is a 510(k) summary which focuses on demonstrating substantial equivalence to a predicate device. It states that the device "was evaluated for value assignment and stability." However, it does not explicitly list quantitative acceptance criteria for these evaluations or the specific results obtained. This level of detail is often found in the full 510(k) submission, not typically in the publicly available summary. For a calibration verification material, typical acceptance criteria would involve:

  • Value Assignment: The measured values for each level (low, mid, high) must fall within a predefined range of the assigned target values. This range is usually determined by precision goals and clinical requirements for the assay being calibrated.
  • Stability: The assigned values must remain within predefined acceptable limits over a specified period (e.g., shelf-life, open-vial stability) under various storage conditions (e.g., refrigerated, room temperature). Changes within these limits are usually compared to initial assigned values.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a "test set" in the context of diagnostic performance on patient data. For a calibration verification material, the "test set" would refer to the lots or units of the Elecsys CalCheck Troponin T that were manufactured and evaluated.

  • Sample size: Not specified. The summary states that the material "was evaluated for value assignment and stability," implying that a sufficient number of product lots or units were tested, but no concrete numbers are given.
  • Data Provenance: Not applicable in the traditional sense of patient data. The evaluations would have been conducted internally by Boehringer Mannheim Corporation, likely in their manufacturing and R&D facilities. The country of origin for the data is therefore the USA, where Boehringer Mannheim Corporation is listed. The study would be prospective in the sense of manufacturing and testing new lots of the CalCheck material.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This concept is not applicable for a calibration verification material. "Ground truth" for this type of device relates to the accurately assigned concentrations of Troponin T within the calibration solutions.

  • Establishing Ground Truth: The "ground truth" (i.e., the target concentration values for each level – low, mid, high) would be established through highly controlled analytical methods and reference materials, often involving a hierarchical traceability chain to international standards. This process typically involves:
    • Using reference measurement procedures (RMPs) or highly accurate, validated methods.
    • Comparing against certified reference materials (CRMs).
    • Multiple measurements by skilled analytical chemists or laboratory professionals.
  • The document does not detail the specific number or qualifications of individuals involved in the value assignment process, as it is a standard laboratory procedure for reference material production.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods (like 2+1 or 3+1) are used for resolving discrepancies in expert interpretations of diagnostic images or data in clinical studies. For a calibration material, discrepancies in value assignment would be resolved through re-analysis, investigation of analytical errors, or use of a reference method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC study was not done. MRMC studies are used to evaluate the impact of a diagnostic device (especially imaging AI) on human reader performance. The Elecsys CalCheck Troponin T is a calibration material, not a diagnostic device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done

Yes, in a sense. The "performance characteristics" evaluated (value assignment and stability) represent the standalone performance of the calibration material itself, independent of human interpretation of a diagnostic result. The device's output (its assigned value) is assessed directly through analytical methods.

7. The Type of Ground Truth Used

The ground truth used for this calibration verification material would be assigned values derived from:

  • Reference materials: Highly pure human recombinant Troponin T.
  • Analytical methods: Precisely measured concentrations using validated laboratory techniques, traceable to higher-order reference measurement procedures or international standards where available.
  • Formulation specifications: The material is manufactured by "appropriately spiking" the analyte to achieve specific concentrations.

It is not based on expert consensus, pathology, or outcomes data, as those relate to diagnostic findings in patients.

8. The Sample Size for the Training Set

Not applicable. The Elecsys CalCheck Troponin T is a chemical reagent product, not an AI/machine learning algorithm. Therefore, there is no "training set" in the context of algorithm development.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

DEC 1 4 1998

... . Radisplay or and

K984372

510(k) Summary

IntroductionAccording to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
Submittername, address,contactBoehringer Mannheim Corporation9115 Hague RoadIndianapolis, IN 46250(317) 845 - 3723
Contact Person: Priscilla A. Hamill
Date Prepared: December 3, 1998
Device nameProprietary name: Elecsys CalCheck Troponin T
Common name: Calibration Verification Material
Classification name: Single (specified) analyte controls (assayed + unassayed)
PredicatedeviceThe Boehringer Mannheim Elecsys CalCheck Troponin T is substantially equivalent to the currently marketed Elecsys CalCheck Troponin T.
DevicedescriptionThe Boehringer Mannheim Elecsys CalCheck Troponin T is manufactured using bovine serum albumin, human recombinant Troponin T, stabilizers, and preservatives. The analyte is appropriately spiked into the CalCheck matrix to the correct CalCheck concentration levels.
Continued on next page

{1}------------------------------------------------

510(k) Summary, Continued

Intended useThe Boehringer Mannheim Elecsys® CalCheck™ Troponin T is used toverify the calibration assignment for the Boehringer Mannheim ElecsysTroponin T assay.
Comparison topredicatedeviceThe Boehringer Mannheim Elecsys® CalCheck™ Troponin T is substantiallyequivalent to other products in commercial distribution intended for similaruse. Most notably, it is substantially equivalent to the currently marketedElecsys® CalCheck™ Troponin T.Both products are intended to be used for the verification of calibration foranalytes on the Elecsys immunoassay analyzers.
PerformanceCharacteristicsThe Elecsys® CalCheck™ Troponin T was evaluated for value assignmentand stability.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure.

DEC 1 4 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Priscilla A. Hamill Regulatory Affairs Consultant Boehringer Mannheim Corporation 9115 Hague Road Indianapolis Indiana 46250-0457

Re: K984372 Elecsys CalCheck Troponin T Trade Name: Regulatory Class: I Product Code: JJY Dated: December 3, 1998 Received: December 7, 1998

Dear Ms. Hamill:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The qeneral controls provisions of the Act include requirements for annual registration, listinq of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Paqe 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K984372 510(k) Number (if known): N/A

Device Name: Boehringer Mannheim Elecsys® CalCheck™ Troponin T

Indications For Use:

Elecsys CalCheck Troponin T calibration verification solutions comprise three levels - low, mid. and high - each with a defined Troponin T concentration. The low solution concentration is near the lower detection limit of the assay. The middle solution is in the middle or at the clinically critical point of the measuring range. The high solution is near the upper limit of the measuring range.

The Elecsys CalCheck Troponin T is intended for use in periodic verification of the calibration of the Elecsys Troponin T assay.

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K984372

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.