(171 days)
Not Found
No
The device is a chemical solution for sterilization and the description focuses on its chemical composition, intended use in specific sterilizers, and performance in sterilization tests. There is no mention of AI or ML.
No.
Vapo Solution™ is an alcohol solution intended for use in sterilizers to sterilize medical and dental surgical instruments, not to treat or diagnose patients.
No
Explanation: Vapo Solution™ is an alcohol solution used to sterilize medical and dental surgical instruments, not to diagnose diseases or conditions.
No
The device description clearly states it is a "specific solution of various alcohols, acetone, ketone, formaldehyde, and distilled water," which is a chemical substance, not software.
Based on the provided information, no, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use of Vapo Solution™ is to sterilize medical and dental surgical instruments within a specific type of sterilizer (Harvey Chemiclave). This is a process for making instruments safe for use on patients, not for performing diagnostic tests on biological samples.
- Device Description: The device is described as a chemical solution used in a sterilization process. It does not involve the analysis of biological specimens.
- Performance Studies: The performance studies focus on the efficacy of the solution in killing bacterial spores (Bacillus stearothermophilus) on instruments, which is a measure of sterilization effectiveness, not diagnostic accuracy.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), detecting specific analytes, or providing information for the diagnosis, treatment, or prevention of disease based on the analysis of such samples.
In summary, Vapo Solution™ is a chemical sterilant used in a medical device (a sterilizer) to prepare instruments for use. It does not fit the definition of an In Vitro Diagnostic device, which is used to examine specimens derived from the human body to provide information for diagnostic purposes.
N/A
Intended Use / Indications for Use
"Vapo Solution™ is un alcohol solution intended to be used only in Harvey Chemiclave sterifizes to sterilizes to sterilize medical and dental surgical instrument. Varo Solution™ is intential to be a direct substitute for Vapo-Steril®. Vapo Solution™ is recommended for use in the following Chemiclave models: 4000, 5500, 6000 and models covered under K924380 and K943654.
Vapo Solution™ is compatible with Chemicluye cycle parameters as tollows:
- Model 4000, SOON, SSOO, 6000, EMSOON, EMSSOO, und EM6000
- Operating pressure: 20-40 psi (1.41 2.82 kg/cm²) ◆
- Operating Temperature: 132 +2 ℃ (270 + 5 0r) ●
Maximum load: Bagged instruments: | ||||
---|---|---|---|---|
6" Chamber (models 4000, 5000, EM5000) | 0.6 Kg | (1.3 lbs) | ||
8" Chamber (models 5500, EM5500) | 1.2 Kg | (2.6 lbs) | ||
10" Chamber (models 6000, EM6000) | 2.4 Kg | (5.3 lbs) | ||
● | Maximum load: Unwrapped instruments: | |||
6" Chamber (models 4000, 5000, EM5000) | 0.75 Kg | (1.7 lbs) | ||
8" Chamber (models 5500, EM5500) | 1.4 Kg | (3.1 lbs) | ||
10" Chamber (models 6000, EM6000) | 2.8 Kg | (6.2 lbs) | ||
Model 7000 and 8000 | ||||
● | Operating pressure: approximately 25 psi | |||
● | Operating Temperature: 132 ± 2°C (270 ±5°F) | |||
● | Maximum load: Bagged instruments: | |||
8" Chamber (model 7000) | 1.2 Kg | (2.6 lbs) | ||
10" Chamber (model 8000) | 2.4 Kg | (5.3 lbs) | ||
● | Maximum load: Unwrapped instruments: | |||
8" Chamber (model 7000) | 1.4 Kg | (3.1 lbs) | ||
10" Chamber (model 8000) | 2.8 Kg | (6.2 lbs) | ||
Vapo Solution™ is compatible with but not limited to the following instruments and material types:
- Surgical and cutting instruments, handpieces, and similar items .
- Plastics that can tolerate in excess of 132 °C. i.c. Polysultione, Phonolic, Nylon, Teflor, . Polycthylene, Polycact, Polypropylenc. Note: Polycarbonate is not chemically compaithe.
- . Rubber components made of Silicone, Fluorosilicone"
"Vapo Solution™ is not compatible with the following matcrials:
- Lincn, toxtiles or fabrics These are incompatible with the chemical vapor process. .
- Liquids and agars These are incompatible with high temperature processing (132°C). .
- Items contained in tightly woven packs or wraps, or sealed containcrs These cause poor . vapor penetration.
- Plastics which cannot tolerate temperatures in excess of 132ºC. .
- . Polycarbonate - Chemically incompatible.
- Nylon tubing and bags Inhibits penctration of chemical vapors. .
Vapo Solution™ is not recommended for use for "flash" sterilization cycles."
Product codes (comma separated list FDA assigned to the subject device)
80MLR, MED
Device Description
"A specific solution of various alcohols, acetone, ketone, formaldehyde, and distilled water for use only in MDT Harvey Chemiclave® sterilizers."
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
"dental and surgical instruments"
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
"Performance Data: 4,000 independent tests were performed using 'fresh' solutions; 600 additional tests using one-year old solutions; 600 tests using two-year old solutions.
Model 5000-6"
tests completed +growth
Vapo Solution™ 680 219
Vapo-Steril solution® 670 222
Model 5500-8"
tests completed +growth
Vapo Solution™ 650 119
Vapo-Steril solution® 640 183
Model 6000-10"
tests completed +growth
Vapo Solution™ 680 261
Vapo-Steril solution® 680 328
Statistical analysis indicates that the Vapo Solution™ is equal to or better than the predicate. Vapo-Steril® solution in killing spore strips of Bacillus stearothermophilus when tested according to method used for such testing.
Conclusion: From non-clinical tests, the Model 5000 shows no significant difference in survivors for the two solutions; the Models 5500 and 6000 show a significant difference in survivors at shorter cycle times (Vapo Solution™ has fewer survivors), but no significant difference at longer cycle times. The two solutions are substantially the same in efficacy.
One-year old solution: Same comparative results.
Two-year old solution: Same comparative results."
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Not Found
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 880.6885 Liquid chemical sterilants/high level disinfectants.
(a)
Identification. A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.(b)
Classification. Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.
0
MAY 20 1999
510(K) SUMMARY
| [2] | Roger L. Goodman, D.D.S., P.C.
October 1, 1997
200 Temple Street
Mason, Michigan 48854-1837
Contact Person: Roger L. Goodman | | | | | (517) 676-5200 | FAX: (517)676-5442 |
|------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----------------|---------|-----------------|--------------------|
| [3] | Name of device: Vapo Solution™ for use only in Chemiclave Sterilizers
Trade name: Vapo Solution™ for use only in Chemiclave Sterilizers
Common name: Alcohol Solution for use only in Chemiclave Sterilizers
Classification name: 21 CFR 880.6890
Product code: 80MLR
Device classification: Class II | | | | | | |
| [4] | Equivalence to: MDT Harvey Vapo-Steril® solution for use only in Harvey Chemiclave® sterilizers | | | | | | |
| [5] | Description of device: A specific solution of various alcohols, acetone, ketone, formaldehyde, and distilled water for use only in MDT Harvey Chemiclave® sterilizers. | | | | | | |
| [6] | Intended use of Device: Vapo Solution™, a generic substitute for Harvey Vapo-Steril® solution, is used in sterilizing dental and surgical instruments in MDT Harvey Chemiclave® sterilizers. The unsaturated chemical vapor is analogous to steam vapor in autoclave sterilizers. | | | | | | |
| [7] | Technological characteristics: Vapo Solution™ and Harvey Vapo-Steril® solution have identical active ingredients of 72.38% ethyl alcohol and 0.23% formaldehyde: and substantially the same inert ingredients of 27.39% | | | | | | |
| [8] | Performance Data: 4,000 independent tests were performed using 'fresh' solutions; 600 additional tests using one-year old solutions; 600 tests using two-year old solutions. | | | | | | |
| | | Model 5000-6" | | Model 5500-8" | | Model 6000-10" | |
| | | tests completed | +growth | tests completed | +growth | tests completed | +growth |
| | Vapo Solution™ | 680 | 219 | 650 | 119 | 680 | 261 |
| | Vapo-Steril solution® | 670 | 222 | 640 | 183 | 680 | 328 |
| | Statistical analysis indicates that the Vapo Solution™ is equal to or better than the predicate. Vapo-Steril® solution in killing spore strips of Bacillus stearothermophilus when tested according to method used for such testing. | | | | | | |
| [9] | Conclusion: From non-clinical tests, the Model 5000 shows no significant difference in survivors for the two solutions; the Models 5500 and 6000 show a significant difference in survivors at shorter cycle times (Vapo Solution™ has fewer survivors), but no significant difference at longer cycle times. The two solutions are substantially the same in efficacy.
One-year old solution: Same comparative results.
Two-year old solution: Same comparative results. | | | | | | |
| [10] | Toxicology: | A study conducted in dental offices using an analogous alcohol solution containing 0.23% formaldehyde reported the following: 15 min STEL ave. 0.81 ppm (OSHA 2.0 ppm); 8-hour average 0.029 ppm (OSHA TWA 1.0 ppm). 29 CFR 1910.1048 >0.1 ppm considered a Health Hazard. | | | | | |
@MDT. Harvey, and Chemiclave are registered trademarks of MDT Corp. @Vapo-Stcril is a registered trademark of MDT Biologic Company.
This summury of 510(k) safety und effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
1
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
MAY 20 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Roger L. Goodman, D.D.S., P.C. 200 Temple Street Mason, MI 48854
Re: K984270
Trade Name: Vapo-Solution for use in MDT Harvey Chemiclave Sterilizers Models 5000, 5500 and 6000 Regulatory Class: Unclassified Product Code: MED Dated: November 25, 1998 Received: November 30, 1998
Dear Dr. Goodman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
2
Page 2 - Dr. Goodman
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Philip J. Phillips
Philin J Deputy Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
3
Device Namo: VAPO SQLUTION™
Indications for Use:
Vapo Solution™ is un alcohol solution intended to be used only in Harvey Chemiclave sterifizes to sterilizes to sterilize medical and dental surgical instrument. Varo Solution™ is intential to be a direct substitute for Vapo-Steril®. Vapo Solution™ is recommended for use in the following Chemiclave models: 4000, 5500, 6000 and models covered under K924380 and K943654.
Vapo Solution™ is compatible with Chemicluye cycle parameters as tollows:
- Model 4000, SOON, SSOO, 6000, EMSOON, EMSSOO, und EM6000
- Operating pressure: 20-40 psi (1.41 2.82 kg/cm²) ◆
- Operating Temperature: 132 +2 ℃ (270 + 5 0r) ●
Maximum load: Bagged instruments: | ||||
---|---|---|---|---|
6" Chamber (models 4000, 5000, EM5000) | 0.6 Kg | (1.3 lbs) | ||
8" Chamber (models 5500, EM5500) | 1.2 Kg | (2.6 lbs) | ||
10" Chamber (models 6000, EM6000) | 2.4 Kg | (5.3 lbs) | ||
● | Maximum load: Unwrapped instruments: | |||
6" Chamber (models 4000, 5000, EM5000) | 0.75 Kg | (1.7 lbs) | ||
8" Chamber (models 5500, EM5500) | 1.4 Kg | (3.1 lbs) | ||
10" Chamber (models 6000, EM6000) | 2.8 Kg | (6.2 lbs) | ||
Model 7000 and 8000 | ||||
● | Operating pressure: approximately 25 psi | |||
● | Operating Temperature: 132 ± 2°C (270 ±5°F) | |||
● | Maximum load: Bagged instruments: | |||
8" Chamber (model 7000) | 1.2 Kg | (2.6 lbs) | ||
10" Chamber (model 8000) | 2.4 Kg | (5.3 lbs) | ||
● | Maximum load: Unwrapped instruments: | |||
8" Chamber (model 7000) | 1.4 Kg | (3.1 lbs) | ||
10" Chamber (model 8000) | 2.8 Kg | (6.2 lbs) | ||
Vapo Solution™ is compatible with but not limited to the following instruments and material types:
- Surgical and cutting instruments, handpieces, and similar items .
- Plastics that can tolerate in excess of 132 °C. i.c. Polysultione, Phonolic, Nylon, Teflor, . Polycthylene, Polycact, Polypropylenc. Note: Polycarbonate is not chemically compaithe.
- . Rubber components made of Silicone, Fluorosilicone
(PLEASE IN NOT WRITE BELOW THIS LINE . CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Fivaluation (ODF)
Phi Phit
(Division Sign-off)
Division of Dental, Infection Control,
and General Hospital Devices
Proscription 113c (Per 21 CFR 801.109)
Over-the-counter 1.50
(Optional livemat 1-2.96)
4
510(k) Number (if known) K984270
Device Name: VAPO SOLUTION™
Indications for Use:
Vapo Solution™ is not compatible with the following matcrials:
- Lincn, toxtiles or fabrics These are incompatible with the chemical vapor process. .
- Liquids and agars These are incompatible with high temperature processing (132°C). .
- Items contained in tightly woven packs or wraps, or sealed containcrs These cause poor . vapor penetration.
- Plastics which cannot tolerate temperatures in excess of 132ºC. .
- . Polycarbonate - Chemically incompatible.
- Nylon tubing and bags Inhibits penctration of chemical vapors. .
Vapo Solution™ is not recommended for use for "flash" sterilization cycles.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Ollice of Device Evaluation (ODF) Cal, Infection Control. General Mospital Devices 510(k) Numbe
Prescription I lso (Per 21 CFR 801.109)
Over-the-counter I Jses_______________________________________________________________________________________________________________________________________________________