K Number
K984088
Date Cleared
1999-12-06

(385 days)

Product Code
Regulation Number
870.1650
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Optistar MR Injector System is intended for injecting MR contrast agents and flushing solutions for the purpose of enhancing diagnostic imaging of humans.

Device Description

The Optistar MR is a contrast delivery system that is designed to inject image enhancing contrast media into the vascular system of a human during an MRI (Magnetic Resonance Imaging) procedure. The main components of the Optistar MR system are the console, power pack, and powerhead. The console resides in the control room of the MR suite, while the power pack and powerhead remain in the scanning room with the patient. The parameters of the injection, such as volume and flow rate are programmed by the operator using a touch screen on the console. The console is powered by AC line voltage at 110 Volts at 60 Hz. The console communicates with the 24 VDC powerpack which in turn supplies power and the injection parameters to the powerhead. The powerhead performs the injections by driving one or two lead screw rams. The rams push the plunger of a syringe, which expels fluid from the barrel of the syringe.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Optistar MR Injector System, based on the provided document:

Acceptance Criteria and Device Performance

The document states that performance testing was conducted to verify the Optistar MR Injector System's ability to achieve its specified parameters, as outlined in the substantial equivalence chart. The "substantial equivalence chart" itself effectively serves as the acceptance criteria for various parameters by comparing the new device to the predicate device. The performance testing section then reports that the device "successfully delivered" or "fell within the specified range" for these criteria.

Characteristic/FeatureAcceptance Criteria (from predicate or defined by Optistar MR)Reported Device Performance (Optistar MR)
Volume Accuracy"Accurately measured and compared to the volume programmed" (implicit acceptance of close agreement). The predicate device's volume limits are 1-63 ml in 1 ml increments. Optistar MR specifies 1-61 ml (60ml syringe) and 1-27 ml (25ml syringe)."Over the range of volumes programmed, the Optistar MR Injector system successfully delivered the programmed volume within the allowable specification accuracy. This held true regardless of syringe size (25 ml or 60 ml), injection side (A or B) or phase type (single or dual). "
Flow Rate AccuracyFlow rates should be within the specified range (0.1 to 10 ml/sec for 60ml syringe, 0.1 to 8.0 ml/sec for 25ml syringe, 0.1 to 7.0 ml/sec for flushing fluid)."Any injection accurately measured, that was not pressure limited, fell within the specified range. If pressure-limited, flow rate reduces, but the programmed volume is still delivered."
Pressure MonitoringMonitored pressure should be within the upper limit set for each syringe type. The device's displayed pressure should be within an accuracy range of the measured pressure. Predicate max pressure limit is 300 psi. Optistar MR is 200 psi (25ml syringe) and 150 psi (60ml syringe)."In all cases the injector was within the accuracy range" (referring to the comparison between measured and displayed pressure).
Drip Mode AccuracyExpected volume (calculated from programmed increment, time interval, flow rate, and measuring time) should match the actual measured volume within a specified accuracy range."Over the range of various drip mode parameters, the system fell within the specified accuracy range for all cases."
Manual Speeds Accuracy1 ml/s for single key press, 5 ml/s for accelerator key press, within a specified accuracy range."The tested speeds were within a specified accuracy range for both slow and fast speeds, for both A and B side, and for both 25 ml and 60 ml syringes."
General System CapabilityAbility to perform all functions outlined in the substantial equivalence chart, safely and effectively."The Optistar MR Injector System proved its ability to perform within its specified parameters. As a result, its performance is deemed acceptable and substantially equivalent to the predicate device."

Study Information

The document describes "Performance Testing" to verify the device's capabilities, primarily to support the claim of substantial equivalence to the predicate device (Medrad Spectris MR Injection System K935668).

  1. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The document mentions "Over the range of volumes programmed," "During the test injections," "In all cases," "Over the range of various drip mode parameters," and "The tested speeds." This suggests that a range of parameters and configurations were tested, but no specific number of injections or test cases is provided.
    • Data Provenance: The testing was performed internally by Liebel-Flarsheim. The location is Cincinnati, OH, USA, as indicated by the company address. This is a prospective validation of the device's technical specifications.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this is a performance validation of a medical device's physical functions (e.g., flow rate, volume delivery), not an diagnostic imaging study requiring expert interpretation of medical images. The "ground truth" for the test set was defined by the device's design specifications and accurately measured physical quantities (e.g., actual delivered volume, measured flow rate, actual pressure).
  3. Adjudication method for the test set:

    • Not applicable. This was a physical performance test, not a subjective assessment requiring adjudication. Measurements were taken and compared against specifications.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No, an MRMC study was not done. This study is a performance validation of the injector system's mechanical and software functions (flow, volume, pressure, etc.), not a diagnostic imaging study involving human readers and AI for image interpretation.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, implicitly. The performance testing described is a standalone evaluation of the device's mechanical and software performance against its specifications, without a human operator's "performance" being part of the primary measured outcome (beyond programming the desired parameters). The device's ability to inject as programmed is the focus.
  6. The type of ground truth used:

    • Defined Performance Specifications / Measured Physical Quantities: The ground truth was established by the design specifications of the device for parameters like volume, flow rate, pressure limits, and speeds. The "truth" for the comparison was then actual physical measurements taken during the tests (e.g., measured volume, calculated flow rate, physical pressure readings).
  7. The sample size for the training set:

    • Not applicable. This is not a machine learning model, so there is no training set. The device operates based on programmed logic and physical mechanisms.
  8. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this type of device validation.

{0}------------------------------------------------

DEC - 6 1999

Image /page/0/Picture/2 description: The image shows the logo for Liebel-Flarsheim, which is identified as "A MALLINCKRODT COMPANY" in smaller letters below the main text. The logo consists of the letters "LF" in a stylized, bold font on the left. To the right of the letters is the name "Liebel-Flarsheim" in a serif typeface.

Optistar MR Injector System 510(k) Summary

November 4, 1998

Contact Name:Bridget DrakeCompliance AdministratorLiebel-Flarsheim2111 E. Galbraith Rd.Cincinnati, OH 45215Phone: (513) 761-2700Fax: (513) 761-2388
Proprietary Name:Optistar MR Injector System
Common Name:Injector System
Classification Name:Injector and Syringe, Angiographic
Predicate Device:Medrad Spectris MR Injection System (K935668)

Intended Use:

The Optistar MR Injector System is intended for injecting MR contrast agents and flushing solutions for the purpose of enhancing diagnostic imaging of humans.

Description:

The Optistar MR is a contrast delivery system that is designed to inject image enhancing contrast media into the vascular system of a human during an MRI (Magnetic Resonance Imaging) procedure.

The main components of the Optistar MR system are the console, power pack, and powerhead. The console resides in the control room of the MR suite, while the power pack and powerhead remain in the scanning room with the patient. The parameters of the injection, such as volume and flow rate are programmed by the operator using a touch screen on the console. The console is powered by AC line voltage at 110 Volts at 60 Hz. The console communicates with the 24 VDC powerpack which in turn supplies power and the injection parameters to the powerhead. The powerhead performs the injections by driving one or two lead screw rams. The rams push the plunger of a syringe

{1}------------------------------------------------

syringe, which expels fluid from the barrel of the syringe. The console, power pack and powerhead are discussed in greater detail below.

Powerhead: The powerhead provides a means for accepting and identifying a given syringe and then applying a force to the plunger of the syringe via a ram, which ejects contrast from the syringe. The powerhead communicates with the power pack in order to determine the flow rates and volumes designated by the user. This information is then relayed to the injector. The powerhead resides in the imaging room with the scanner.

Console: The console is the main user interface for the iniector system. It allows the user to input the injection parameters via a touch screen as well as view the results of an injection. The console incorporates a remote hand switch for starting or stopping an injection. The console resides in the control room of the MR imaging suite.

Power Pack: The power pack is the interface between the console and the powerhead. The power pack processes the information input by the user from the console and relays this information to the powerhead. The power pack also supplies the powerhead with DC voltage. Two batteries supply the power pack. The power pack resides in the imaging room with the scanner.

The system is designed to deliver two injections per procedure. It can deliver a single or dual phase contrast injection from a user filled syringe and then follow up with an injection of a flushing solution, such as saline. To accomplish this task the powerhead holds two syringes with the ability to inject both syringes. Two syringe sizes can be used with the Optistar MR system, a 60 ml size and a 25 ml size. In addition, a Y tubing set is used to transfer the contrast flushing solution to the patient. One short leg of the Y connects to the contrast filled syringe using a luer lock connector. The other short leg of the Y connects to the syringe with the flushing solution using a luer lock connector. The long single leg of the Y goes to the patient.

The system also includes a battery charger, which is used to recharge the 24 V batteries.

Substantial Equivalence:

A comparison chart is shown on the next page that compares the technological characteristics of the Optistar MR Injection System to the predicate device, the Spectris MR Injection System. The performance and effectiveness of the Optistar MR is substantially equivalent to the predicate device. The new injector safely and successfully injects MR contrast media at user specified flow rates and volumes in an MR environment. The primary differences from the predicate device lie in the area of user preference features. The differences do not adversely affect the safety or efficacy of the device. A summary of the testing that verified the technological characteristics of the Optistar MR follow shall follow the comparison chart.

{2}------------------------------------------------

Characteristic/FeatureSpectris(predicate device)(K935668)Optistar MR(new device)
Multiphasic ContrastInjections2 phases per protocol2 phases per protocol
Protocol Storage20 protocols20 protocols
Injection HistoryStores last 5 injection resultsStores last 20 injection results
Scan Delay00:00 - 00:59 minutes00:00 - 00:59 minutes
Syringe System - A side65 ml60 ml or 25 ml
Syringe System - B side65 ml60 ml or 25 ml
Volume1-63 in 1 ml increments1-61 in 1 ml increments (60 ml syringe)1-27 in 1 ml increments (25 ml syringe)
Flow Rate - Contrast0.1 to 10 ml/secin 0.1 ml/sec increments25 ml syringe - 0.1 to 8.0 ml/sec in 0.1ml/sec increments60 ml syringe - 0.1 to 10 ml/sec in 0.1ml/sec increments
Flow Rate - Flushing fluid0.1 to 10 ml/sec in 0.1 ml/secincrements25 and 60 ml syringe - 0.1 to 7.0 ml/secin 0.1 ml/sec increments
Max Pressure Limit300 psiFor 25 ml syringe - 200 psi(adjustable)For 60 ml syringe - 150 psi(adjustable)
Remote Start SwitchYesYes
Mode for keeping a veinopenYesYes - Drip Mode
Safety Stop MechanismElectrical stop when injectionparameters are out of specElectrical stop when injectionparameters are out of spec
Fill/expel controlpush buttons on headpush buttons on head
Air detectionoperator visual inspectionoperator visual inspection
Programming injectionsTouch screen displayTouch screen display
Injection status displaylights on top of power headlights on top of power head
MaterialsPlastics and nonferrous metalsPlastics and nonferrous metals
Anatomical sitesContrast injection into venous system.Contrast injection into venous system.
Intended useThe injection of MR contrast agentsand flushing solutions for the purposeof enhancing diagnostic imaging ofhumans.The injection of MR contrast agentsand flushing solutions for the purposeof enhancing diagnostic imaging ofhumans.
Target populationHumansHumans
SterilityDisposables are provided sterileDisposables are provided sterile

{3}------------------------------------------------

Performance Testing:

Performance testing was completed in order to verify that the Optistar MR Injector System was capable of achieving the specification parameters for the system as outlined in the substantial equivalence chart. Verification of the system's ability to accurately achieve these values validates the substantial equivalence claims. Injection performance was tested for accuracy of delivered flow rate, volume, pressure, drip mode and manual speeds.

Volume

Fluid delivery of an injector is characterized by the volume delivered and the flow rate. The volume delivered is the most critical parameter in that it has the greatest effect on image quality. During the testing, the volume delivered was accurately measured and compared to the volume programmed. Over the range of volumes programmed, the Optistar MR Injector system successfully delivered the programmed volume within the allowable specification accuracy. This held true regardless of syringe size (25 ml or 60 ml), injection side (A or B) or phase type (single or dual).

Flow Rate

The second most important variable in delivering fluid to a patient is flow rate. During the test injections, the time duration for an injection was recorded along with the volume delivered. The flow rate delivered was then calculated. Any injection accurately measured, that was not pressure limited, fell within the specified range. At higher flow rates, through a restricted orifice, the injector limits the pressure in the syringe by reducing the flow rate. In this case, the flow rate will not be achieved; however, the programmed volume will still be delivered.

Pressure

During an injection, the injector monitors pressure in order to determine if the injection should be pressure limited. An upper limit is set for each syringe type used in the injector, in order to avoid failure of the syringe. During a number of injections pressure was measured and compared to the pressure displayed by the injector. In all cases the injector was within the accuracy range

Drip Mode

Drip mode is used to keep a catheter or a vein open. The injector achieves this by administrating small increments of fluid over short time intervals. The injector allows the programming of the volume increment, a time interval and a flow rate. The performance of drip mode was evaluated by measuring the total volume delivered over a predetermined time period. Given a volume increment, a time interval, a flow rate and a measuring time period an expected volume was calculated and then compared to an actual measured volume to determine the accuracy for those parameters. Over the range of various drip mode parameters, the system fell within the specified accuracy range for all cases.

{4}------------------------------------------------

Manual Retract and Extend Speeds

The Optistar MR Injector has two speeds for manually moving the rams. By depressing either the forward or reverse key, the ram should move 1 ml/s. By pressing the accelerator key in combination with the forward or reverse key, the ram should move 5 ml/s. The tested speeds were within a specified accuracy range for both slow and fast speeds, for both A and B side, and for both 25 ml and 60 ml syringes.

Conclusion

The Optistar MR Injector System proved its ability to perform within its specified parameters. As a result, its performance is deemed acceptable and substantially equivalent to the predicate device.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC - 6 1999

Ms. Bridget Drake Compliance Administrator Mallinckrodt Inc. Liebel-Flarsheim Business 2111 E. Galbraith Rd. P.O. Box 156305 Cincinnati, OH 45215-6305

Re: K984088 Optistar MR Injector System Trade Name: Requlatory Class: II Product Code: DXT September 3, 1999 Dated: September 7, 1999 Received:

Dear Ms. Drake:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act

{6}------------------------------------------------

Page 2 - Ms. Bridget Drake

for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at Also, please note the regulation entitled, (301) 594-4639. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Ohitopmth fr

Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Page 1 of 1

510(k) NUMBER (IF KNOWN): K984088

Optistar MR Injector System DEVICE NAME:

INDICATIONS FOR USE:

The Optistar MR is designed to inject MR contrast agents and flushing solutions for the purpose of enhancing diagnostic imaging of humans. Contraindications for the use of this device are determined by the prescribing physician at the time of use based upon the contrast media Not to be used in a magnetic field greater than 1.5 package inserts. Tesla.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter-Use (Optional Format 1-2-96)

Christopher for Witten

Division of Cardiovascular Respira and Neurological D 510(k) Number

§ 870.1650 Angiographic injector and syringe.

(a)
Identification. An angiographic injector and syringe is a device that consists of a syringe and a high-pressure injector which are used to inject contrast material into the heart, great vessels, and coronary arteries to study the heart and vessels by x-ray photography.(b)
Classification. Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.