(430 days)
The balloon pump is an electromechanical system used to drive intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation. The intra-aortic balloon is placed in the descending aorta, just distal to the left subclavian artery. Once the balloon is positioned, the pump is adjusted to trigger in synchrony with the ECG or arterial pressure waveform to ensure that inflation and deflation occur at the appropriate points during the cardiac cycle.
The Belmont NGPBP System is a light and compact portable system, designed to facilitate balloon pumping at the patient's bedside. The portable balloon pump is a reliable, safe, and effective system designed especially for ease of operation and transportation. The Belmont NGPBP is made up of eight major functional elements: Operator Interface, Signal Acquisition and Display, Triggering and Assist Timing, Pneumatics: Pump Drive System, Pneumatics: Volume Control System, Pneumatics: Shuttle Gas System, System surveillance, Power System.
The provided text describes a Premarket Notification 510(k) Summary for the Belmont Model NGPBP Intra-Aortic Balloon Pump, comparing it to a predicate device, the Belmont Model PBP Intra-Aortic Balloon Pump. The document outlines the device's technical characteristics, intended use, and nonclinical tests conducted to demonstrate substantial equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Inferred from testing) | Reported Device Performance (Belmont NGPBP) |
|---|---|
| Ability to fully inflate and deflate commercially available intra-aortic balloons. | Compared well with the Belmont PBP. Fully inflated and deflated balloons from Bard Cardiac Assist Products, Arrow International, Bard (St. Jude Medical), and Datascope Corporation (nominal volumes 30, 34, 40, 50cc) at static pressure 90 mm Hg and heart rates of 60 and 125 BPM. |
| Appropriate triggering and timing of balloon inflation and deflation from ECG waveform. | Accurately and properly synchronized the inflate/deflate cycle with the patient's heart for a variety of normal and abnormal rhythms with variable heart rate. Maintained proper triggering from ECGs with normal sinus rhythms (30-200 BPM). |
| Performance in presence of ECG artifacts, pacing spikes, and arrhythmias with variable heart rate. | Responded by deflating during PVCs. Pumped acceptably during Atrial Fibrillation (holding inflation until the next R-wave). Triggered properly during Ventricular Tachycardia, erratic rates with Ventricular Fibrillation, and A-Paced rhythms. Responded well to baseline shift, baseline noises, wide QRS, and tall T-wave. |
| Safe operation in the presence of artifacts/interference (fail-safe mechanism). | If the unit could not track the ECG and trigger appropriately in the presence of artifact and interference, the balloon remained deflated, and an alarm was activated. |
| Activation of alarms and cessation of pumping for unsafe conditions. | The Surveillance System activates audible tones, deflates the balloon, and stops further inflation until the source of the alarm is eliminated. An alarm message is displayed. Most alarm conditions require user to restart pumping. |
| Automatic restart for specific conditions (loss of trigger, noisy ECG). | Pumping restarts automatically when the problem (loss of trigger, noisy ECG signal) is eliminated. |
| Ability to trigger from arterial pressure waveform in synchrony with cardiac cycle. | Capable of appropriate trigger balloon inflation and deflation from arterial pressure waveform. (This is mentioned in the conclusion as a common capability with the predicate device, but specific test results for AP triggering aren't detailed as extensively as ECG in section 11.b.) |
| Identification of unsafe balloon and internal conditions (disconnection, overfilling, leakage, etc.). | The Gas Surveillance System uses helium pressure, air drive pressure, and vacuum to determine unsafe balloon and internal conditions including balloon disconnection, overfilling, gas leakage, kinked catheter, and slow deflation. Internal Self Checks determine system malfunctions. |
| Functional equivalence to the predicate device (Belmont PBP) in core functional elements and operation. | The two systems perform identical functions and contain identical eight major functional elements. The NGPBP compared well with the PBP in inflation/deflation tests and triggering/timing under various clinical conditions. The conclusion explicitly states the NGPBP is "substantially equivalent" to the PBP in intended use and core performance capabilities. |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Size: Not explicitly stated as a numerical sample size of patients or recordings. The tests were performed with:
- "Commercially available Balloons" from various manufacturers (Bard, Arrow, Datascope) covering nominal volumes of 30, 34, 40, and 50cc.
- "ECGs with normal sinus rhythms with heart rates of 30-200 BPM."
- Simulated "PVCs, Atrial Fibrillation, Ventricular Tachycardia, erratic rate with Ventricular Fibrillation, and A-Paced" rhythms.
- Simulated "baseline shift, baseline noises, wide QRS, and tall T-wave."
- Data Provenance: The study appears to be retrospective or simulated in nature, utilizing various types of commercially available balloons and simulated/pre-recorded ECG waveforms representing different clinical conditions. There is no mention of prospective patient data or data from specific countries of origin. The testing involves "compar[ing] well with the current IABP," suggesting controlled testing environments rather than direct patient trials.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- This information is not provided in the document. The testing described focuses on the device's functional performance against known or simulated physiological conditions. There is no mention of expert review or labeling of a test set.
4. Adjudication Method for the Test Set
- Not applicable/Not mentioned. Since there's no explicitly stated "test set" in terms of clinical images or patient data that would require interpretation or labeling by human experts, there is no mention of an adjudication method. The testing largely involves objective measurements of device response to simulated inputs (e.g., triggering accuracy, inflation/deflation cycles).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This document describes the nonclinical (benchtop) performance evaluation of an intra-aortic balloon pump, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study or AI assistance effect size is not applicable to this submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, indirectly. The entire nonclinical testing described (Section 11) assesses the "algorithm only" performance of the Belmont NGPBP in an automated or simulated environment. The device's ability to "appropriately trigger balloon inflation and deflation" and "synchronize the inflate/deflate cycle" based on various inputs (ECG, AP waveforms) is essentially its standalone algorithmic performance in detecting and responding to physiological signals. The "automatic mode" of the NGPBP also represents an "algorithm only" operation where the system analyzes signals and determines optimal timing with minimum user input.
7. The Type of Ground Truth Used
- The ground truth for the nonclinical tests appears to be known physiological parameters and simulated clinical conditions.
- For inflation/deflation: The ground truth is the expected full inflation/deflation of balloons of known volumes at specified heart rates and pressures.
- For triggering/timing: The ground truth is the accurate and synchronized response to simulated "normal" ECG rhythms (e.g., normal sinus rhythm within 30-200 BPM) and "abnormal" rhythms or artifacts (e.g., PVCs, Atrial Fibrillation, T-waves, baseline noise) where the appropriate device action (trigger, deflate, alarm) is known.
- For safety alarms: The ground truth is the presence of simulated unsafe conditions (e.g., balloon disconnection, gas leakage, kinked catheter, artifact/interference) that should correctly activate alarms.
8. The Sample Size for the Training Set
- Not applicable / Not mentioned. This device does not appear to utilize machine learning or AI in a way that requires a separate "training set" in the conventional sense (i.e., for a supervised learning model). The device's operational logic and control algorithms would be based on established physiological principles and engineering design, not on "learning" from a dataset.
9. How the Ground Truth for the Training Set Was Established
- Not applicable / Not mentioned. As there is no explicit training set for AI/ML, the concept of establishing ground truth for it does not apply. The device's functional parameters and safety mechanisms are designed and verified against established medical knowledge and engineering standards.
{0}------------------------------------------------
JAN - 6 2000
Attachment 5
Image /page/0/Picture/3 description: The image shows a line graph plotted on a grid. The graph has several peaks and valleys, indicating fluctuations in the data it represents. The line starts from the middle left, rises sharply, then falls, and continues to fluctuate across the grid.
Belmont Instrument Corporation
80 Boston Road, Billerica, MA 01821 (978)663-0212, Fax: (978)663-0214
Registered In Accordance with ISO-9001 (Certificate # 041007407)
Premarket Notification 510(k) Summary
[As Required By 21 CFR 807.92(c)]
-
- Submitter & Belmont Instrument Corporation Manufacturing Site: 780 Boston Road Billerica, MA 01821
Establishment Registration Number: 1219702
- Submitter & Belmont Instrument Corporation Manufacturing Site: 780 Boston Road Billerica, MA 01821
-
- Uraiwan P. Labadini, Quality Assurance/Regulatory Affairs Contact Person: Manager
(978) 663-0212 Ext. 28 Fax: (978) 663-0214 Telephone:
- Uraiwan P. Labadini, Quality Assurance/Regulatory Affairs Contact Person: Manager
-
- Trade Name: Belmont Model NGPBP Intra-Aortic Balloon Pump
-
- Common name: Intra-aortic Balloon Pump
-
ട്. Classification name: Intra-aortic Balloon Pump Control System.
-
ଚ. 74DSP Product Code: Device Class: Classified as a Class III device per Federal Register July, 1978.
-
- The Belmont Model NGPBP IABP is similar in design, function, user interface, principles of operation, and technology to the Belmont Model PBP Intra-Aortic Balloon Pump, which has been commercially available since 1992 as Bard TransAct H-8000, and was the subject of Premarket Notification #K915580 submitted in November 1991.
-
- Brief Description: The Belmont NGPBP System is a light and compact portable system, designed to facilitate balloon pumping at the patient's bedside. The portable balloon pump is a reliable, safe, and effective system designed especially for ease of operation and transportation.
The Belmont NGPBP is made up of eight major functional elements:
- l. Operator Interface
- ii. Signal Acquisition and Display
- iii. Triggering and Assist Timing
- iv. Pneumatics: Pump Drive System
- V. Pneumatics: Volume Control System
- vi. Pneumatics: Shuttle Gas System
- vii. System surveillance
- Power System viii.
{1}------------------------------------------------
Attachment 5 510(k) Summary Page 2 of 5 October 26, 1998
- Intended Use: The balloon pump is an electromechanical system used to drive 9. intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation. The intra-aortic balloon is placed in the descending aorta, iust distal to the left subclavian artery. Once the balloon is positioned, the pump is adjusted to trigger in synchrony with the ECG or arterial pressure waveform to ensure that inflation and deflation occur at the appropriate points during the cardiac cycle.
-
- Summary of the technological characteristics of the Belmont PBP compared to the Belmont NGPBP.
The two systems perform the identical function, both contain the identical eight major function elements as described above. This Belmont NGPBP will be differentiated from the existing Belmont PBP, by reduced operator input (i.e., automated operation) and having reduced size and weight. It will also feature improved ergonomics and more intuitive user interface.
- Operator Interface: The Belmont NGPBP provides for automatic ECG a. amplitude scaling for triggering, automatic gas surveillance, same as the Belmont PBP. However, the Belmont NGPBP provides two modes of operation - an automatic mode and a manual mode. In automatic mode, the system will operate the pump with minimum user input. Once the automatic mode is ど chosen, the system will analyze all three ECG leads, ECG monitor signal (if connected), and arterial pressure (AP) signal (both transducer or monitor, if connected) and choose the best available signal for triggering, and the system will determine the optimal timing of inflation and deflation based on an analysis of the AP, BP, and ECG waveforms.
- b. Signal Acquisition and display: Both systems utilize ECG signals from skin leads via a patient cable and arterial pressure signals from a direct transducer which is connected to an arterial cannula or the central lumen of the intra-aortic balloon. Both systems also utilize amplified signals from an external monitor.
Both systems display 3 signals: the ECG, arterial pressure, and balloon gas pressure waveforms. Both systems also display digital values of the AP and BP signals, and heart rate. Both systems provide alarm messages in case of operational difficulties with the balloon/catheter or with the system and also status messages to indicate AC/DC status, battery charge, and helium gas status
- C. Triggering and Assist Timing: Both systems detect the ECG R-Wave, the V-Pacemaker spike or the upstroke of systolic arterial pressure and use for triggering.
In the manual mode of operation for both pumps, the operator selects from the following modes: a) ECG, b) Arterial pressure trigger, and c) Internal (asynchronous) trigger. The operator sets inflation timing, and deflation timing. In the automatic mode (for Belmont NGPBP only), the Trigger Mode and Timing
{2}------------------------------------------------
Attachment 5 510(k) Summary Page 3 of 5 October 26, 1998
settings are determined by the svstem by analyzing the ECG and arterial pressure waveforms, no user action is required. During total bypass when the activated.
- Pneumatics: Pump Drive System: Same technology for Belmont PBP and ರ. Belmont NGPBP. .
For both systems:
Balloon inflation and deflation is actuated by the Pump Drive System, which activates the Volume Control System which meters the correct volume displacement to inflate the balloon properly, using helium shuttle gas from the Shuttle Gas System. The Pump Drive System consists of a compressor/vacuum pump, 2 ballast tanks with air pressure transducers to monitor pressure and vacuum, and pulse valves which control application of air pressure or vacuum to the Volume Control System and provide for relief of pressure to make the vacuum cycle more efficient. A third pressure transqucer monitors pressure on volume control side of the valves.
- e. Pneumatics: Volume Control System: Same technology for Belmont PBP and Belmont NGPBP.
For both systems:
The Volume Control System consists of the Adiustable Volume Limiter (AVL), which isolates air pressure and vacuum from the balloon and controls balloon inflation volume, two valves for filling and venting helium gas from the Shuttle Gas System, and a pressure transducer for monitoring shuttle gas pressure.
- f. Pneumatics: Shuttle Gas System: Same technology for Belmont PBP and Belmont NGPBP.
For both systems:
The Shuttle Gas System delivers helium gas to the AVL fill valve at low pressure. The helium source is a disposable helium cylinder, which is pressurized to greater than 2000 psi. The system contains a piercing mechanism for puncturing the cylinder, and 2 regulators to reduce the helium pressure to a safe level. The new system also contains a pressure transducer for reading helium source pressure, which is done by a pressure switch in the old system, and a blow off valve as a safety measure.
- Surveillance System: Same technology for Belmont PBP and Belmont NGPBP. g.
For both systems:
The Gas Surveillance System uses the helium pressure, air drive pressure, and vacuum to determine unsafe balloon and internal conditions including balloon
{3}------------------------------------------------
Attachment 5 510(k) Summary Page 4 of 5 October 26, 1998
disconnection, overfilling, gas leakage, kinked catheter, and slow deflation, Internal Self Checks determine system malfunctions, and indicate battery charge and helium status. While pumping, all surveillance systems are active. An override mode allows operation without a balloon and disables the balloon disconnection and gas leakage alarms.
The Surveillance system is used to activate alarms at all unsafe conditions. At an alarm condition, an audible tone is activated, the balloon is deflated, with no further inflation until the source/cause of the alarm is eliminated. To aid the Operator, an alarm message is displayed on the monitor screen indicating the source of the problem. When the source of the problem is eliminated, most alarm conditions require the user to restart pumping. In two conditions, loss of triager and noisy ECG signal, pumping restarts automatically when problem is eliminated. The recorder printout will show the problem and indicated alarm status, if the recorder is activated.
-
Power System h.
Both systems obtain its power directly from the wall receptacle, during AC power operation. When the AC power is discontinued, full operation of the system is automatically transferred to the internal battery if it is adequately charged. The rechargeable battery provides power to the unit during transportation. -
- Summary of Nonclinical Tests and Results
In order to verify performance of the Belmont NGPBP in support of substantial equivalence, the following tests were carried out:
- Summary of Nonclinical Tests and Results
-
Inflation and Deflation Tests with Commercially available Balloons at 60 and a. 125 BPM. The ability of the Belmont NGPBP to fully inflate and deflate all commercially available balloons. Balloons from all of the known manufacturers: Bard Cardiac Assist Products, Arrow International. Bard (St. Jude Medical), and Datascope Corporation balloons were tested with the static pressure 90 mm Hq and heart rates of 60 and 125 BPM. including balloons having nominal volumes of 30, 34, 40, 50cc. The Belmont NGPBP compared well with the Belmont PBP.
-
b. Triggering and Timing under a Variety of Clinical Conditions
The ability of the Belmont NGPBP to appropriately trigger balloon inflation and deflation from ECG waveform, even in the presence of artifacts, pacing spikes, and arrhythmias with variable heart rate was tested.
The Belmont Model NGPBP, same as the Belmont PBP, accurately and properly synchronized the inflate/deflate cycle of the balloon with the patient's heart for a variety of normal and abnormal rhythms with variable heart rate. If the unit could not track the ECG and trigger appropriately in the presence of
{4}------------------------------------------------
Attachment 5 510(k) Summary Page 5 of 5 October 26, 1998
artifact, and interference, the balloon remained deflated and alarm was activated.
- i. Belmont Model NGPBP compared well with the current IABP. Each of the pumps was able to trigger properly from ECGs with normal sinus rhythms with heart rates of 30-200 BPM.
- ii. Each of the pumps responded by deflating during PVCs. The NGPBP and the current IABP pumped acceptably during Atrial Fibrillation (hold inflation until the next R-wave). When the pumps were tested using Ventricular Tachycardia, and erratic rate with Ventricular Fibrillation, and A-Paced, they both triggered properly.
Each of the pumps responded well to the baseline shift, baseline noises, wide QRS, and tall T-wave.
-
- Conclusion: The Belmont NGPBP is substantially equivalent to the Belmont PBP which received 510(k) approval in October 1992. Both system has the same intended use, are capable of appropriate trigger balloon inflation and deflation from arterial pressure waveform and ECG waveforms, even in the presence of artifacts, pacing spikes, and arrhythmias. Both systems alarm, display alarm message, and stop at all unsafe conditions.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three overlapping human profiles, with a design of three curved lines above them, resembling a bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 6 2000
Ms. Uraiwan P. Labadini Ouality Assurance/Requlatory Affairs Manager Belmont Instrument Corporation 780 Boston Road Billerica, MA 01821
Re: K983866 Belmont Model AUTO-CAT Intra-Aortic Balloon Pump Trade Name: Regulatory Class: III Product Code: DSP Dated: October 8, 1999 Received: October 15, 1999
Dear Ms. Labadini:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements concerning action. your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation
{6}------------------------------------------------
Page 2 - Ms. Uraiwan P. Labadini
you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-Also, please note the regulation entitled, "Misbranding by 4639. reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Celia M. Witten, Ph.D. Acting Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Attachment 1
Belmont Instrument Corporation
Page 1 Of 1
510(k) number: _______________________________________________________________________________________________________________________________________________________________
Device Name: Belmont NGIABP
Indications For Use:
The balloon pump is an electromechanical system used to drive intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation. The intra-aortic balloon is placed in the descending aorta, just distal to the left subclavian artery. Once the balloon is positioned, the pump is adjusted to trigger in synchrony with the ECG or arterial pressure waveform to ensure that inflation occur at the appropriate points during the cardiac cycle.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE QN ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Zele & Amper Ci
Division Sign-Off Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K983564
Prescription Use _ سما (Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
§ 870.3535 Intra-aortic balloon and control system.
(a)
Identification. An intra-aortic balloon and control system is a prescription device that consists of an inflatable balloon, which is placed in the aorta to improve cardiovascular functioning during certain life-threatening emergencies, and a control system for regulating the inflation and deflation of the balloon. The control system, which monitors and is synchronized with the electrocardiogram, provides a means for setting the inflation and deflation of the balloon with the cardiac cycle.(b)
Classification. (1) Class II (special controls) when the device is indicated for acute coronary syndrome, cardiac and non-cardiac surgery, or complications of heart failure. The special controls for this device are:(i) Appropriate analysis and non-clinical testing must be conducted to validate electromagnetic compatibility and electrical safety of the device;
(ii) Software verification, validation, and hazard analysis must be performed;
(iii) The device must be demonstrated to be biocompatible;
(iv) Sterility and shelf-life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components;
(v) Non-clinical performance evaluation of the device must demonstrate mechanical integrity, durability, and reliability to support its intended purpose; and
(vi) Labeling must include a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) when the device is indicated for septic shock and pulsatile flow generation.
(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 31, 2014, for any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation that was in commercial distribution before May 28, 1976. Any other intra-aortic balloon and control system indicated for septic shock or pulsatile flow generation shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.