K Number
K983706
Date Cleared
1998-11-12

(22 days)

Product Code
Regulation Number
888.3050
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TOWNLEY Pedicle Screw Plating System is intended to stabilize the spine as an aid to fusion. After first making drill holes in the pedicles, a DYNA-LOK® Plate is positioned over the pedicles. TOWNLEY Pedicle screws are then inserted through the plate, down the center of the pedicles, and into the vertebral body. Bone graft must be used with each procedure.

This system is indicated for the treatment of any or all of the following at the C2 to S1 (inclusive) spinal levels: (a) Trauma, including spinal fractures and/or dislocations, (b) Spondylolisthesis, all grades and types, (c) Spondylolysis, (d) Pseudarthrosis, (e) Degenerative disc disease and/or degenerative diseases which include: (1) degenerative disc disease (ddd) as defined by neck and/or back pain of discogenic origin as confirmed by patient history with degeneration of the disc as confirmed by radiographic studies and/or (2) degenerative disease of the facets with instability.

For all these indications, bone graft must be used and the system is limited to screw fixation from C2 to S1.

Device Description

This system consists of a broad-headed, partially threaded screw designed to compress bone grafts. Screw lengths range from 0.5 to 2.5 inches. Both cortical and cancellous screw threads are available. The screw comes in two configurations: flat-head or hex. The hex screws are wider in diameter. The DYNA-LOK® plates are used to interconnect two or more vertebrae together via screw fixation through the pedicles. The components are made of titanium alloy such as described by ASTM F-136. or ISO 5832-3. Alternatively, the entire system may be made out of medical grade stainless steel such as described by ASTM F-138 or ISO 5832-1 or ISO 5832-9. This 510(k) submission seeks to add the titanium alloy components to the TOWNLEY Pedicle Screw Plating System.

AI/ML Overview

This document describes the TOWNLEY Pedicle Screw Plating System but does not contain information regarding acceptance criteria or a study proving the device meets said criteria.

The document is a 510(k) summary from 1998, focused on demonstrating substantial equivalence to a predicate device. For medical devices, "substantial equivalence" means that the new device is as safe and effective as a legally marketed device (the predicate). This typically involves comparing features, materials, indications for use, and performance data, often relying on existing data or widely accepted engineering principles rather than a new clinical study with specific acceptance criteria in the way a clinical trial for a drug or a novel AI software might.

Therefore, I cannot populate the table or answer the specific questions about acceptance criteria and a study as the provided text does not contain that information. The document explicitly states:

"Documentation was provided which demonstrated the TOWNLEY Pedicle Screw Plating System to be substantially equivalent to itself."

This phrase, while perhaps unusual in its wording ("to itself"), implies that the submission focused on changes to an existing device already on the market (in this case, adding titanium alloy components to a system previously described). The FDA's response confirms that the device is "substantially equivalent... to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified." This reaffirms the focus on equivalence rather than meeting new, specific performance metrics from a dedicated study.

{0}------------------------------------------------

NOV 1 2 1998

TOWNLEY Pedicle Screw Plating System 510(k) Summary November 4th, 1998

  • I. Company: Sofamor Danek USA 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133

II. Proprietary Trade Name: TOWNLEY Pedicle Screw Plating System

Product Description III.

This system consists of a broad-headed, partially threaded screw designed to compress bone grafts. Screw lengths range from 0.5 to 2.5 inches. Both cortical and cancellous screw threads are available. The screw comes in two configurations: flat-head or hex. The hex screws are wider in diameter. The DYNA-LOK® plates are used to interconnect two or more vertebrae together via screw fixation through the pedicles. The components are made of titanium alloy such as described by ASTM F-136. or ISO 5832-3. Alternatively, the entire system may be made out of medical grade stainless steel such as described by ASTM F-138 or ISO 5832-1 or ISO 5832-9. This 510(k) submission seeks to add the titanium alloy components to the TOWNLEY Pedicle Screw Plating System.

IV. Indications

The TOWNLEY Pedicle Screw Plating System is intended to stabilize the spine as an aid to fusion. After first making drill holes in the pedicles, a DYNA-LOK® Plate is positioned over the pedicles. TOWNLEY Pedicle screws are then inserted through the plate, down the center of the pedicles, and into the vertebral body. Bone graft must be used with each procedure.

This system is indicated for the treatment of any or all of the following at the C2 to S1 (inclusive) spinal levels: (a) Trauma, including spinal fractures and/or dislocations, (b) Spondylolisthesis, all grades and types, (c) Spondylolysis, (d) Pseudarthrosis, (e) Degenerative disc disease and/or degenerative diseases which include: (1) degenerative disc disease (ddd) as defined by neck and/or back pain of discogenic origin as confirmed by patient history with degeneration of the disc as confirmed by radiographic studies and/or (2) degenerative disease of the facets with instability.

For all these indications, bone graft must be used and the system is limited to screw fixation from C2 to S1.

V. Substantial Equivalence

Documentation was provided which demonstrated the TOWNLEY Pedicle Screw Plating System to be substantially equivalent to itself.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract symbol resembling a bird in flight, composed of three curved lines. The symbol is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". The text is arranged around the circumference of the circle.

NOV 1 2 1998

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Richard W. Treharne, Ph.D. Vice President, Research and Regulatory Affairs Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132

Re: K983706 Townley Pedicle Screw Plating System Regulatory Class: II Product Codes: MNI, MNH, and KWP October 20, 1998 Dated: Received: October 21, 1998

Dear Dr. Treharne:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Richard W. Treharne, Ph.D.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other genera Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

tcoelke

Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): Jeg 8370 6

TOWNLEY Pedicle Screw Plating System Device Name:_

Indications for Use:

The TOWNLEY Pedicle Screw Plating System is intended to stabilize the spine as an aid to fusion. After first making drill holes in the pedicles, a DYNA-LOK® Plate is positioned over the pedicles. TOWNLEY Pedicle screws are then inserted through the plate, down the center of the pedicles, and into the vertebral body. Bone graft must be used with each procedure.

This system is indicated for the treatment of any or all of the following at the C2 to S1 (inclusive) spinal levels: (a) Trauma, including spinal fractures and/or dislocations, (b) Spondylolisthesis, all grades and types, (c) Spondylolysis, (d) Pseudarthrosis, (e) Degenerative disc disease and/or degenerative diseases which include: (1) degenerative disc disease (ddd) as defined by neck and/or back pain of discogenic origin as confirmed by patient history with degeneration of the disc as confirmed by radiographic studies and/or (2) degenerative disease of the facets with instability.

For all these indications, bone graft must be used and the system is limited to screw fixation from C2 to S1.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) (Optional 1-2-96)

OR

Over-the-counter Use

c

(Division Sign-Off)
Division of General Restorative Devices K983706
510(k) Number. 000021

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.