K Number
K983164
Manufacturer
Date Cleared
1999-03-05

(176 days)

Product Code
Regulation Number
876.5160
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Restore device is indicated for the prevention or decrease of episodes of urine leakage in women with stress incontinence.

Device Description

Restore Female Continence device is a simple, noninvasive suction cup which fits over the urinary meatus. Air is squeezed out of the device's cap while it is positioned over the urinary meatus. When the cap is released, the device self-adheres to the anterior vaginal wall over the meatus via suction. If inadequate suction occurs. Aquaphor or a petroleum based ointment can be applied to the outside rim of the device to enhance suction. The suction or negative pressure created by the device causes coaptation and occlusion of the meatus and, therefore, increased urethral resistance. The amount of negative pressure generated by the device is optimized to counteract increases in intraabdominal and intravesical pressure during valsalva, thereby decreasing or preventing leakage.

AI/ML Overview

Here's an analysis of the NEBL Restore Female Incontinence Device's acceptance criteria and the study used to prove it, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance for NEBL Restore Female Incontinence Device

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary doesn't explicitly state pre-defined acceptance criteria with numerical targets in the same way a device specification document might. Instead, the "acceptance criteria" are implied by the clinical study's objectives: to demonstrate a statistically significant improvement in objective and subjective efficacy measures, and to show the device is safe and well-tolerated.

Here's a table summarizing the reported device performance against these implied criteria:

Acceptance Criterion (Implied)Reported Device Performance
Efficacy:
Decrease or prevent incontinence episodes (objective)Pad Weight Test (PdWt): 97% improvement (from 6.67gm to 0.19gm), p = .0001
Provocative Stress Test (PST): 100% improvement (from 2 to 0), p = .0001
Decrease or prevent incontinence episodes (subjective)Incontinence Episodes Per Day (IEPD): 91% improvement (from 3.4 to 0.3), p = .0001
Reduce impact of incontinence on quality of lifeIncontinence Quality of Life (I-QOL): 45% improvement (from 62.3 to 90.4), p = .0001
Ease of device placementSatisfaction Survey (placement): 90% found easy to place at first visit, 95% at 3 months (Rate of Positive Response)
Ease of device removalSatisfaction Survey (removal): 100% found easy to remove (Rate of Positive Response)
Overall patient satisfaction with performanceSatisfaction Survey (overall): 93% satisfied at first visit, 98% at 3 months (Rate of Positive Response)
Safety:
Low incidence of positive urine culturesUrine cultures: 1.5% prevalence, 3% incidence of positive cultures during device usage (significantly below historic figures of 10-38% for incontinent females)
Minimal vaginal irritationIrritation Questionnaire/Satisfaction Survey (Q8): Mean Response and Rate of Positive Response were "relatively high" (implying low irritation) throughout the study, indicating device was well-tolerated.
Few, minimal, and self-limited adverse events (physical exam)Physical examinations: Adverse events were few, minimal, and self-limited. No therapeutic intervention required, no complications or sequelae. Only 5% withdrew due to vaginal irritation, and even this subgroup showed significant improvement in incontinence while using the device.
Biocompatibility, non-toxicity, well tolerated by tissues (non-clinical)Nonclinical Tests: Materials and product are biocompatible, non-toxic, and well-tolerated by tissues (based on tests like Delayed Contact Sensitization, Cytotoxicity, Irritation Study, Muscle Implantation, etc.).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 100 women
  • Data Provenance: The study was conducted in the U.S. across 8 Investigational Sites. It is a prospective clinical study as women used the device during a 12-week device usage period, with baseline (Control) measurements taken before device use and follow-up measurements at Week 12.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts used to establish the ground truth or their specific qualifications (e.g., "radiologist with 10 years of experience"). For objective tests like Pad Weight Test and Provocative Stress Test, the "ground truth" is inherently quantitative measurement. For subjective measures like the Incontinence Impact Questionnaire and Satisfaction Survey, the patients' own responses constitute the "ground truth." Clinical assessments (physical exams, urine cultures) would have been performed by medical professionals at the investigational sites, but their specific roles or the number involved in establishing a "ground truth" consensus are not detailed.

4. Adjudication Method for the Test Set

The document does not mention a specific adjudication method (e.g., 2+1, 3+1, none) for the test set. Given that the efficacy parameters were statistically analyzed using paired t-test, repeated-measures analysis, and Wilcoxon signed-rank testing, it implies direct measurement and patient reporting rather than a consensus-based adjudication process for the primary endpoints. Clinical safety assessments would have been part of standard clinical practice at each site.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study described is a single-arm, pre-post comparison of device performance in a cohort of patients. It does not involve multiple human readers evaluating cases or comparing human performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This question is not applicable as the NEBL Restore device is a physical medical device (a suction cup) worn by patients, not an AI algorithm. Therefore, there is no "algorithm-only" performance to describe.

7. The Type of Ground Truth Used

The ground truth used for the test set was a combination of:

  • Objective Measurements:
    • Pad Weight Test (PdWt): Direct measurement of urine leakage in grams.
    • Provocative Stress Test (PST): A score representing the severity of leakage during specific maneuvers.
    • Urine Cultures: Laboratory results for the presence of bacteriuria.
  • Subjective Patient-Reported Outcomes:
    • Voiding Diary: Patient-reported number of incontinence episodes per day (IEPD).
    • Incontinence Impact Questionnaire (I-QOL): Patient-reported quality of life scores.
    • Satisfaction Survey: Patient-reported satisfaction with ease of placement/removal, performance, comfort, and tolerability (e.g., vaginal irritation).
  • Clinical Observations:
    • Physical Examinations: Observations by medical professionals regarding adverse events.

8. The Sample Size for the Training Set

The document does not mention a "training set" in the context of device development. This term is primarily relevant for machine learning algorithms. For a physical medical device like the Restore, the "training" typically refers to engineering design iterations and perhaps pilot or feasibility studies. The clinical study of 100 women serves as the primary evaluation of the final device design.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" described for an AI algorithm, this question is not applicable. For the physical device, the "ground truth" during its development phases would have been established through a combination of engineering testing, user feedback on prototypes, and nonclinical studies (e.g., biocompatibility) to refine the design before formal clinical efficacy trials.

{0}------------------------------------------------

3/5/99

510(K) Summary NEBL Restore Female Incontinence Device

    1. Name: NEBL, Inc. Address: 44 Terrace Drive Worcester, MA 01609 Phone: 781-322-4501 Contact: Jeffrey Dann M.D. Date Prepared: September 1, 1998
    1. Device Name:

Proprietary Name: Restore (a.k.a. Re/Stor) Common Name: Female Incontinence Device Classification Name: Device, urethral occlusion for incontinence Classification Code: 78MNG

  • Predicate Devices: Impress Soft Patch (Uromed Inc) K974600 3. Restore (NEBL Inc.) K971359
    1. Device Description

Restore Female Continence device is a simple, noninvasive suction cup which fits over the urinary meatus. Air is squeezed out of the device's cap while it is positioned over the urinary meatus. When the cap is released, the device self-adheres to the anterior vaginal wall over the meatus via suction. If inadequate suction occurs. Aquaphor or a petroleum based ointment can be applied to the outside rim of the device to enhance suction. The suction or negative pressure created by the device causes coaptation and occlusion of the meatus and, therefore, increased urethral resistance. The amount of negative pressure generated by the device is optimized to counteract increases in intraabdominal and intravesical pressure during valsalva, thereby decreasing or preventing leakage.

    1. Intended Use
      The Restore device is indicated for the prevention or decrease of episodes of urine leakage in women with stress incontinence.

{1}------------------------------------------------

6. Indications for Use

The Restore device is indicated for the prevention or decrease of episodes of urine leakage in women with stress incontinence.

7. Substantial Equivalence Companson

The Restore device is substantially equivalent in function, features and indications to the Impress Soft Patch (UroMed, Inc.). Both devices have the same Indications For Use to prevent or decrease episodes of urine leakage in women with stress urinary incontinence. Both devices are worn intralabially over the meatus and occlude the urethral meatus by creating a seal over the urethral opening. Both are applied by patients to the same anatomical position with ease by utilizing similar techniques of device placement. Both are removed to urinate and reapplied after urination is completed.

Restore has several important advantages compared to Impress Soft Patch. First, it does not require a medical adhesive to function; therefore, it should be easier to manipulate into the correct anatomical position or to reapply if initially incorrectly positioned. Secondly, it naturally migrates to the midline of the antenor vaginal wall and, therefore, over the meatus even if slightly malpositioned. Third, there is no risk of mucosal contact or irritation from an adhesive. Fourth, there are fewer restrictions in use such as showering, bathing and swimming since Restore is not composed of hydrophilic material. In summary, Restore appears easier to place, less wasteful and cumbersome to reposition and reapply if inadequately placed, and requires fewer restrictions in use compared to the Impress Soft Patch. Restore also provides protection if inappropriately used by a patient with urgency incontinence secondary to uninhibited bladder contractions since the device easily "props-off" under this clinical condition.

8. Nonclinical Tests

Restore and its component material has been tested for biocompatibility, Delayed Contact Sensitization, Vaginal Cytotoxicity. Irritation Study. Bacteriostatic Fungistasis, 90 Day Muscle Implantation Study, Microbial Limits Screening, Microbial Limits Prepatory Testing, Bioburden Recovery and Zone of Inhibition Testing. The results indicate the materials and product are biocompatible, nontoxic and well tolerated by tissues.

{2}------------------------------------------------

Summary of Clinical Testing റ്റ.

The Restore device has been extensively tested for its safety and efficacy in decreasing or preventing incontinence episodes in women (see K971359). All testing results indicate that the Restore device provides minimal risk yet provides significant benefit for women in controlling urinary leakage.

Effectiveness A.

Clinical testing was completed on 100 women from 8 Investigational Sites in the U.S. Women used the Restore device during a device usage period of 12 weeks to test the efficacy hypothesis that the Restore Continence device will: 1) decrease or prevent incontinence episodes and 2) reduce the impact of incontinence on quality of life. Objective testing included Pad Weight Test (PdWt) and Provocative Stress Test (PST). Subjective testing included a voiding diary documenting the number of incontinence episodes per day (IEPD), an incontinence impact questionnaire and a Satisfaction Survey. Efficacv parameters were statistically analyzed using paired t-test analysis, repeatedmeasures analysis and Wilcoxon signed rank testing. Analysis demonstrates a statistically significant improvement in all objective and subjective efficacy measures. The Table below details the average measurements before device use (Control) and at the Week 12 Device Utilization Visit.

TestControlWeek 12%ImprovementP
PdWt6.67gm.19gm97%.0001
PST20100%.0001
IEPD3.4.391%.0001
I-QOL62.390.445%.0001

Breakdown by Visits demonstrated that the effect was immediate with continued improvement as patients became more proficient with device placement.

The women in the Study were stratified into categories of mild (0-2gm), moderate (2-8gm) and severe (greater than 8gm) urinary incontinence based on Baseline PdWt. Analysis of PdWt results demonstrate a statistically significant improvement in urine loss for each subpopulation of subjects.

{3}------------------------------------------------

The impact of urinary incontinence on quality of life is a measure of the patient's perception of the degree to which leakage has a negative effect on various aspects of daily living. A 22 question incontinence impact questionnaire (I-QOL) with a maximum score of 110 evaluated subjects before and during device usage. Subjects demonstrated a significant improvement in their quality of life during device use.

Patient Satisfaction Surveys specifically evaluated Restore's ease of placement and removal and overall patient satisfaction with performance. The Rate of Positive Response (RPR) was calculated as the percent of responses that were equal to or greater than neutral. i.e.satisfied with that aspect of device function. Question 2 evaluated the ease of device placement. Ninety percent of patients found the device easy to place at the first study visit and 95% at 3 months. Question 3 evaluated the ease of device removal. One hundred percent of patients found the device easy to remove during the study. Noting that correct device placement is critical to performance, Question 7 evaluated overall patient satisfaction with Restore. Ninety three percent of patients were satisfied with device performance at the first study visit and 98% at 3 months. This data confirms that Restore is easy to place and remove and that patients are satisfied with device performance. Furthermore, patients found the device convenient, able to remain in place during activity, improved their enjoyment of life and self confidence. Finally, patients demonstrated a high degree of satisfaction with device comfort.

A six week Post Device Utilization Period assessed subjects degree of urine loss after device discontinuation. PdWt, PST and IEPD testing demonstrated statistically significant improvement in urine loss during this period compared to the Baseline control. This response, however, deserves further investigation before any long term therapeutic claims can be made.

в. Safety

.

To provide clinical safety assurance of the Restore Continence device. objective assessment of Safety included 1) urine cultures 2) imitation questionnaire and 3) periodic physical examinations.

Frequent urine cultures revealed a 1.5% prevalence and 3% incidence of positive urine cultures during device usage. Based on literature review of the age-specific prevalence of bacteriuria in incontinent females, this low prevalence rate is significantly below historic figures of 10-38%.

Satisfaction Survey Question 8 quantitates vaginal irritation. The Mean Response and Rate of Positive Response were relatively high throughout the Study signifying that the Restore device was well tolerated in the vast majority of patients.

{4}------------------------------------------------

Adverse events on physical examination were few, minimal, and self No therapeutic intervention was required in any patients and no limited. complications or sequelae occurred.

Analysis of patients Withdrawing from the Study demonstrated that the most common reason cited was inability to keep scheduled visits. Only 5% withdrew because of vaginal irritation. Furthermore, PdWt and PST demonstrated a statistically significant improvement in urine loss in this subgroup during device use which was equivalent to subjects completing the Study.

். Conclusions

The safety of the Restore device was demonstrated by extensive nonclinical and clinical testing. Technological characteristics do not raise new types of safety and effectiveness questions relative to predicate devices. The Clinical Study demonstrated that the Restore device did not effect the incidence or prevalence of significant bacteriuria. The device was well tolerated by the vast majority of patients. The efficacy of the Restore device was demonstrated by both objective and subjective measures. The clinical data showed that statistically significant improvement was obtained by subjects using Restore. Satisfaction Surveys demonstrated that the device is easy to place and remove with high patient satisfaction with device performance. In summary, these data orovide reasonable assurance that the Restore Female Continence Device is a safe and effective alternative for women requiring stress incontinence management and is statistically equivalent to the predicate device.

{5}------------------------------------------------

Clinical/Market Experience 10.

Confirming the safety data reported in its clinical study, NEBL has received no significant adverse event reports since the product's release.

Physicians associated with the clinical study and/or with clinical expertise using Restore have attested that patients can adequately self select and use Restore on a non-prescription, over-the-counter basis based on its Instructions Furthermore, patients have also supported the concept of non-For Use. prescription OTC access to the product.

11. Conclusions

The safety and efficacy of Restore has been demonstrated by nonclinical and clinical testing. Comparison with Impress Soft Patch shows that both devices are substantially equivalent in function and indications and that Restore appears to be easier to place and utilize. Clinical experience derived from patients and physicians confirm that Restore can be used over-the-counter based on its product labelings. This data, supports NEBL's claim that Restore is equivalent to its predicate device and therefore should receive non-prescription, over-the-counter market approval.

{6}------------------------------------------------

Public Health Service

Image /page/6/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is positioned to the right of a circular text element. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" and is arranged around the circumference of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

APR 2 6 2010

Jeffrey A. Dann, M.D. NEBL, Inc. 44 Terrace Drive WORCESTER MA 01609

Re: K983164

Trade/Device Name: Restore (Re/stor) Female Incontinence Device Regulation Number: 21 CFR& 876.5160 Regulation Name: Urological clamp for males Regulatory Class: I Product Code: MNG Dated: December 9, 1998 Received: December 10, 1998

Dear Dr. Dann:

This letter corrects our substantially equivalent letter of March 5, 1999.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 Dr. Jeffrey A. Dann K983164

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

-10(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Device Name: ____ NEBL, Inc. Restone Device

Indications For Use:

The Restore device is indicated for the prevention or decrease of episodes of urine leakage in women with stress incontinence.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Usc 21 CFR 801.109) }.

OR

Over-The-Counter Use

(Optional Formal 1-2-96)

§ 876.5160 Urological clamp.

(a)
Identification. A urological clamp for males is a device used to close the urethra of a male to control urinary incontinence or to hold anesthetic or radiography contrast media in the urethra temporarily. It is an external clamp.(b)
Classification. Class I (general controls). Except when intended for internal use, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.