K Number
K983052
Device Name
PERITRON PERINEOMETER MODEL 9300V OR 9300A
Date Cleared
1998-10-19

(48 days)

Product Code
Regulation Number
884.1425
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Peritron is used in assessing the strength of pelvic floor muscles, teaching pelvic floor muscle exercises and for providing feedback during pelvic floor muscles exercises.
Device Description
The Peritron Perineometer comprises a vaginal or anal Sensor, Readout Unit and interconnecting tube. Peritron incorporates a pressure transducer and microprocessor to interpret the air pressure in the sensor and display it either numerically in cm water column pressure or as an analogue bar graph. Hand-held Peritron weighs 220 gm and is powered by a standard 9 volt battery housed in an integral, user-accessible compartment separated physically from the electronic components. All operating functions are controlled by one push-button. The case of Peritron is of injection moulded plastic.
More Information

Not Found

No
The device description details a pressure transducer and microprocessor interpreting air pressure and displaying it numerically or as a bar graph. There is no mention of AI or ML algorithms for analysis, interpretation, or feedback generation beyond simple pressure measurement and display. The "Mentions AI, DNN, or ML" section also explicitly states "Not Found".

No.
The device is used for assessment and teaching/feedback for exercises, but does not directly treat a condition. While it aids in improving muscle strength, it is not a therapeutic device itself.

Yes

The Peritron is used for "assessing the strength of pelvic floor muscles," which involves evaluating a physiological condition.

No

The device description explicitly states that the Peritron Perineometer comprises a vaginal or anal Sensor, Readout Unit, interconnecting tube, pressure transducer, microprocessor, and is powered by a battery. These are all hardware components, indicating it is not a software-only device.

Based on the provided information, the Peritron device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Peritron's Function: The Peritron measures the strength of pelvic floor muscle contractions using a sensor inserted into the body (vaginal or anal). It does not analyze biological samples taken from the body.
  • Intended Use: The intended use is focused on assessing muscle strength, teaching exercises, and providing biofeedback for pelvic floor muscles. This is a physiological measurement and training tool, not a diagnostic test performed on a sample.

Therefore, the Peritron falls under the category of a medical device used for physiological assessment and therapy, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

Peritron is used in assessing the strength of pelvic floor muscles, teaching pelvic floor muscle exercises and for providing feedback during pelvic floor muscles exercises.

Product codes

85 HIR

Device Description

The Peritron Perineometer comprises a vaginal or anal Sensor, Readout Unit and interconnecting tube. Peritron incorporates a pressure transducer and microprocessor to interpret the air pressure in the sensor and display it either numerically in cm water column pressure or as an analogue bar graph.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

pelvic floor muscles

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

Contimed II™

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 884.1425 Perineometer.

(a)
Identification. A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/1 description: The image is a black and white graphic. The graphic contains a thick, black, semi-circle shape on the left side of the image. To the right of the semi-circle are several short, horizontal lines. The lines are also black and thick.

K983052

OUT 19 1998

Safety and Effectiveness Summary

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Background:

The Peritron Perineometer comprises a vaginal or anal Sensor, Readout Unit and interconnecting tube.

Perineometers are used in the diagnosis and treatment of stress incontinence in patients and for providing exercise feedback.

In the clinic Peritron is used to assess the condition of the pelvic floor muscles, (strength of contraction), teach pelvic floor exercises with the view to improving muscle strength and may be used at home as a biofeedback unit by patients performing pelvic floor exercises.

In operation, air displaced from the sensor caused by a voluntary contraction of the muscles of the pelvic floor travel to the readout units via the interconnecting tube. Peritron does not cause the contraction. These muscles are under the voluntary control of the user at all times.

1 Safety

None of these products diminish the safety level set by the substantially equivalent Contimed II™ (Hollister, Illinois) of Appendix 4(a) due to the following

1.1 Sensors

ﻣﻌﺘﺴﺴﺴﺴﺴﺴﺴﺴﺴﺴﺴﺴ

The key element concerning patient/user safety and the only contact component is the sensor.

As a principal manufacturer of catheters and tubing packs for Cardiac Surgery (Registered in Australia under the Therapeutic Goods Act as a GMP manufacturer but with no goods registered or offered for sale in the USA) we believe we have a good understanding of appropriate materials and manufacturing procedures necessary to produce a product of the highest biocompatibility and patient safety.

Accordingly we designed the sheath of the sensor (the only patient-contact surface) to be moulded in one piece without joins or crevices to potentially harbour bacteria. (In this respect it is safer than the vaginal sensor of the substantially equivalent Contimed unit, the several elements of which join together on the patient contact surface.) The sheath is injection moulded from liquid silicon rubber, Wacker Elastosil cat LR 3003/50A and LR/50B, that meets the requirements of FDA regulation 177.2600. Refer to documents in appendix 6(a)

As the sheath of the sensor contains 600ppm of blue pigment (Wacker cat RAL 5010 ref enclosure 6(c)) of unknown biocompatibility the FDA previously advised that we should conduct testing of the finished product. Refer to appendix 6 (a) for the test results.

1

1.2 Readout Unit

Peritron incorporates a pressure transducer and microprocessor to interpret the air pressure in the sensor and display it either numerically in cm water column pressure or as an analogue bar graph.

Hand-held Peritron weighs 220 gm and is powered by a standard 9 volt battery housed in an integral, user-accessible compartment separated physically from the electronic components.

All operating functions are controlled by one push-button.

The case of Peritron is of injection moulded plastic. There are no external parts that are of metal or are sharp and no electronic parts are accessible without the use of a tool.

At the time of the previous 510(k) application we understand electrical testing of the device was not required. However, since the previous approval we have had Peritron tested for Emission and Safety. The testing shows that Peritron is safe for the intended use. Refer to appendix 6(b) for test results.

1.3 Summary of Safety Features of Peritron

    • Peritron is a passive device - it measures the result of a voluntary action of the user.
    • There are no external wires, electrodes or electrical items of any kind.
    • Internal user adjustments are neither required nor possible.
    • A PVC tube is the only connection between the vaginal probe of the patient/user and the Readout unit.
    • There are no liquids or dangerous items incorporated.
    • All tests show that Peritron is safe for the intended use and meets standards where they exist.

Safety Experience

Since its inception in 1991, over 1200 units have been put into operation and at no time during the development or usage of the Peritron has there been an incident that in any way threatened the safety of any individual.

Effectiveness 2.

Peritron does not alter the effectiveness compared to substantially equivalent Contimed II™ device as both devices measure the increase in air pressure in a sensor caused by a voluntary contraction of pelvic floor muscles.

Like Peritron, Contimed 11 conveys air from its sensor to the readout unit by means of a plastic tube.

Peritron is tested during manufacture for compliance with the performance criteria set out in the Brochure and Handbook of Appendix five. We would be pleased to make test results available on request.

2

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus symbol. The text is in all capital letters.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 19 1998

Mr. Neville Motley Manager cardio design pty ltd P.O. Box 5407 BHBC Baulkham Hills NSW 2153 AUSTRALIA

Re: K983052

Peritron, Model 9300A with Anal Sensor Model 9300V with Vaginal Sensor Dated: August 12, 1998 Received: September 1, 1998 Regulatory Class: II 21CFR 884.1425/Procode: 85 HIR

Dear Mr. Motley:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note; this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".

Sincerely yours

Lillian Yin, Ph.D.

Director, Division of Reproducti Abdominal, Ear, Nose and Thro and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

3

Page

K983052 510(k) Number (if known):_

PERITRON PERINEOMETER- Model 9300V or 9300A Device Name:_

Indications For Use:

ns For Use.
Peritron is used in assessing the strength of pelvic floor muscles, r entroll is about in associety
teaching pelvic floor muscle exercises and for providing feedback during pelvic floor muscles exercises. **********

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Whind h. Segen

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number .

Prescription Jise_V (Per 21 CFR 801.109)

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)