K Number
K982266
Manufacturer
Date Cleared
1998-09-25

(88 days)

Product Code
Regulation Number
888.1100
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ArthroJet system is intended to resect damaged tissue and remove extraneous matter found in articular body cavities. The system may be used open or arthroscopically to resect tears and other defects, remove loose fragments, shave away debris and perform appropriate synovectomy procedures.

Device Description

The HydroCision ArthroJet system consists of an electric motor used to power a drive train inside a console. The drive train converts the rotational motion of a cam to a transverse motion of a plunger which in turn drives a piston contained in a disposable cartridge. The repeated motion of the piston pressurizes saline in the cartridge which is then carried through tubing to a handpiece. The handpiece delivers the high pressure saline to the joint, resulting in the cutting action of the waterjet. The result is a thin waterjet stream capable of cutting and ablating tissue which is then removed from the surgical site.

AI/ML Overview

The provided text details a 510(k) submission for the HydroCision ArthroJet system, focusing on its substantial equivalence to a predicate device. It describes the device, its intended use, and performance data. However, the document does not present "acceptance criteria" and "reported device performance" in the structured format typically seen for medical device evaluations against specific performance metrics with quantitative results.

Based on the provided text, here's an analysis of the information requested:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly state quantitative acceptance criteria or detailed reported device performance in a tabular format. Instead, it describes performance in qualitative terms aimed at demonstrating equivalence to the predicate device and functionality for intended use.

Acceptance CriteriaReported Device Performance
Tissue Resection and Ablation CapabilityDemonstrated in porcine knee evaluation. (Qualitative assessment based on observation)
Efficacy in Resecting MenisciDemonstrated in in-vivo rabbit testing, comparable to the predicate device. Confirmed adequate function for intended use. (Qualitative assessment and comparative observation)
Handpiece Bending StrengthEvaluated for ability to withstand arthroscopic applications. (Implies a successful evaluation, but without specific metrics or thresholds)
Substantial Equivalence to Predicate Device (Dyonics Articular Shaver System)Indications for use are identical. Performance data supports the claim of substantial equivalence.
Meeting Intended UseThe device functions adequately to meet its intended use of resecting damaged tissue and removing extraneous matter in articular body cavities.

2. Sample Size for the Test Set and Data Provenance

  • Porcine Evaluation: Implies use of porcine knee tissue. The exact number of porcine knees or tissue samples is not specified.
  • In-Vivo Testing: Used rabbits. The exact number of rabbits used is not specified.
  • Bending Strength: Refers to evaluation of "the handpiece." The number of handpieces tested is not specified.
  • Data Provenance:
    • In-Vivo Testing: "resecting rabbit menisci" suggests a pre-clinical animal model.
    • The evaluations appear to be prospective studies conducted by the manufacturer for the 510(k) submission.
    • The country of origin is not explicitly stated but can be inferred to be the United States, given the submitter's location (Wilmington, MA) and the FDA submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

The document does not mention the use of experts to establish "ground truth" in the context of diagnostic accuracy, as this is a surgical device. The "performance data" focuses on tissue interaction and mechanical integrity. Therefore, the concept of "experts" for ground truth as typically understood for diagnostic AI/imaging devices is not directly applicable here. The evaluations were likely performed by the manufacturer's engineers and/or animal study researchers.

4. Adjudication Method for the Test Set

Not applicable. The described tests are not diagnostic in nature, nor do they involve subjective interpretations by multiple observers requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study was conducted or reported. This type of study is typically relevant for diagnostic imaging devices where human readers interpret medical images with and without AI assistance. The ArthroJet is a surgical cutting system, not a diagnostic device.

6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

Not applicable. The ArthroJet is a human-operated surgical device; it does not operate autonomously as an algorithm or AI system. Its "performance" inherently involves interaction with a human surgeon.

7. Type of Ground Truth Used

The "ground truth" in this context refers to the observable effects and outcomes of the device's action, rather than an independent diagnostic confirmation.

  • Porcine Evaluation: The ability to "resect and ablate various types of tissue" implies direct visual assessment and potentially histological examination of the resected tissue. The ground truth is the physical evidence of tissue removal and alteration.
  • In-Vivo Testing (Rabbit Menisci): Demonstrated "efficacy... in resecting rabbit menisci" and confirmed "device functions adequately." The ground truth would be the observed state of the menisci post-procedure, possibly confirmed by histology or gross observation, and comparison to the predicate device's effects.
  • Bending Strength: The "ability to withstand arthroscopic applications" suggests engineering tests against defined forces or conditions, where "ground truth" is the structural integrity or failure point of the handpiece.

8. Sample Size for the Training Set

Not applicable. This device is not an AI/ML system that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set was used.

{0}------------------------------------------------

SEP 2 5 1998

510(K) SUMMARY

大982266

1. SUBMITTER:

SEP 2 5 1998

HydroCision, Inc. 220 Ballardvale Street Wilmington, MA 01887 Telephone: 978-657-0020

Contact: Don Freeman, President Date Prepared: June 17, 1998

2. DEVICE:

Arthroscopic Cutting System Classification Name: Arthroscope and Accessories Trade Name: HydroCision ArthroJet

3. PREDICATE DEVICE:

The predicate device used to determine substantial equivalence for the HydroCision ArthroJet was the Dyonics Articular Shaver System, marketed by Smith & Nephew Dyonics, Andover, MA

4. DEVICE DESCRIPTION:

The HydroCision ArthroJet system consists of an electric motor used to power a drive train inside a console. The drive train converts the rotational motion of a cam to a transverse motion of a plunger which in turn drives a piston contained in a disposable cartridge. The repeated motion of the piston pressurizes saline in the cartridge which is then carried through tubing to a handpiece. The handpiece delivers the high pressure saline to the joint, resulting in the cutting action of the waterjet. The result is a thin waterjet stream capable of cutting and ablating tissue which is then removed from the surgical site.

5. INTENDED USE:

The Arthrojet system is intended to resect damaged tissue and remove extraneous matter found in articular body cavities. The system may be used open or arthroscopically to resect tears and other defects, remove loose fragments, shave away debris and perform appropriate synovectomy procedures.

{1}------------------------------------------------

6. COMPARISON OF CHARACTERISTICS:

The HydroCision ArthroJet system is electrically powered and consists of a console unit, pump cartridge, tubing and handpiece. It uses a drive train to pressurize saline in a pump cartridge which is delivered to a hand piece resulting in a thin high pressure waterjet stream capable of cutting and ablating tissue, which is then removed from the site.

The Dyonics shaver is also an electrically powered unit which is used to power a hand held motor. The motor is used to exert rotational force to a disposable cutter. The cutter resects tissue by rotating at high rates of speed resulting in the tissue being cut and removed from the area.

The indications for use of the two devices are identical.

7. PERFORMANCE DATA:

The following performance data is provided in support of the substantial equivalence determination:

    1. Porcine Evaluation: The ability of the ArthroJet to resect and ablate variuos types of tissue was demonstrated in a porcine knee.
    1. In-Vivo Testing: The testing demonstrated the efficacy of the HydroCision ArthroJet in resecting rabbit menisci when compared to the predicate device. The testing confirmed that the device functions adequately to meet its intended use.
    1. Bending Strength: The bending strength of the handpiece was evaluated for its ability to withstand arthroscopic applications.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

P 25 1998

D.C. Freeman, Jr. Ph.D. President HydroCision, Inc. 220 Ballardvale Street Wilmington, Massachusetts 01887

Re: K982266 Trade Name: HydroCision Arthrojet System Regulatory Class: II Product Code: HRX Dated: June 25, 1998 Received: June 29, 1998

Dear Dr. Freeman:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ಗಿ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Dr. Freeman, Jr.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

982266

INDICATION FOR USE

The ArthroJet system is intended to resect damaged tissue and remove extraneous matter found in articular body cavities. The system may be used open or arthroscopically to resect tears and other defects, remove loose fragments, shave away debris and perform appropriate synovectomy procedures.

Prescription Use
(Per 21 CFR 801.109)

(Divis

§ 888.1100 Arthroscope.

(a)
Identification. An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.(b)
Classification. (1) Class II (performance standards).(2) Class I for the following manual arthroscopic instruments: cannulas, currettes, drill guides, forceps, gouges, graspers, knives, obturators, osteotomes, probes, punches, rasps, retractors, rongeurs, suture passers, suture knotpushers, suture punches, switching rods, and trocars. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 888.9.