K Number
K981745
Date Cleared
1998-07-17

(60 days)

Product Code
Regulation Number
870.2850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BSI BIOTRANS is used to convert hemodynamic pressure waveforms into electrical signals which can be displayed using separate monitoring equipment. As such, its Indications for Use are identical to those for Medex's MX 860 and MX 960. BIOTRANS cables are used to make an electrical connection between the sensor and monitor. Their use is identical to that of cables manufactured by Utah Medical Products, Inc.

The BSI BIOTRANS Sensor Base is used to convert hemodynamic pressure waveforms via a piezoresistive bridge circuit into electrical signals which can be displayed using separate monitoring equipment.

Device Description

The BIOTRANS Pressure Monitoring System consists of the sterile, single-use BIOTRANS Pressure Monitoring Kit (subject of a separate 510k) and BIOTRANS Reusable Sensor Base and Monitor Adapter Cables. The Base consists of a rugged silicon-based piezoresistive pressure sensor mounted on a rigid plastic support. A standard Wheatstone Bridge circuit etched into the sensor is used to detect pressure changes applied to the sensor and convert them to electronic signals for display on standard monitoring equipment.

The sensor is electrically isolated from the patient by a dielectric gel and two elastomeric diaphragms. The physiologic pressure waveform is conveyed from the patient catheter via a fluid-filled pathway to the single-use dome diaphragm. The dome diaphragm is mechanically coupled to the reusable pressure sensor diaphragm for pressure monitoring. The diaphragms make intimate contact during use permitting transfer of the pressure waveform to the transducer, with fluid contact occurring on the patient side of the disposable dome diaphragm. A 25 cm (10") Sensor cable permanently wired to the sensor transmits the electronic signal from the transducer to the patient monitor via the appropriate Monitor Adapter Cable.

The integral Sensor cables terminate in 5 different electrical adapters compatible with 5 common brands of Monitor Cables (Utah Medical, Baxter International, Medex, Abbott, and Ohmeda.) Monitor Adapter Cables designed for specific use with all major monitor suppliers are available from BSI.

The Base features built-in organizer/mounting systems compatible with standard manifold clamps for pole-mount applications.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the BSI BIOTRANS™ Reusable Sensor Base:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Standard)Reported Device Performance
American National Standard for Blood Pressure Transducers, ANSI/AAMI BP22-1994 compliance"BSI's BIOTRANS meets, or exceeds, the performance requirements indicated in the American Standard for Blood Pressure Transducers, ANSI/AAMI BP22-1994." "The snap fit configuration meets all electrical requirements listed in the ANSI/AAMI BP22-1994 standard."

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a specific sample size for a "test set" in the context of human subjects or cases. The performance evaluation appears to be based on meeting the requirements of the ANSI/AAMI BP22-1994 standard. The data provenance is not specified beyond this.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. The performance is assessed against a technical standard (ANSI/AAMI BP22-1994), not against expert interpretations of cases.

4. Adjudication Method for the Test Set

Not applicable. There's no mention of an adjudication process for a test set, as the evaluation is based on compliance with a technical standard.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No. An MRMC study was not done. The device is a physical transducer, and the evaluation focuses on its technical performance against a standard, not human interpretational accuracy.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in essence. The evaluation provided focuses on the device's inherent technical performance (e.g., electrical requirements, conversion of pressure waveforms) in a standalone capacity, without human intervention in its core function as a transducer.

7. The Type of Ground Truth Used

The "ground truth" for the device's acceptance is based on adherence to the American National Standard for Blood Pressure Transducers, ANSI/AAMI BP22-1994. This is a technical standard that defines performance requirements for such devices.

8. The Sample Size for the Training Set

Not applicable. This device is a hardware component (pressure transducer), not an AI/machine learning algorithm. Therefore, there is no "training set" in the conventional sense.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this hardware device.

{0}------------------------------------------------

7 of 285/09/98
------------------

II. JUL 17 1998

BSI BIOTRANS ™ Reusable Sensor Base

Premarket Notification

510K SUMMARY

Prepared April 28, 1998

K981745
1. Submitted by:John Shulze
Sunscope International, Inc.
20250 Acacia St., Suite 115
Newport Beach, CA 92660
Tel: (714) 553-8300
Fax: (714) 553-9129
2. Contact Person:John Shulze
3. Device Identification:
Trade Name:BIOTRANS Reusable Sensor Base
Common Name:Physiologic Pressure Transducer
Classification Name:Transducer, Blood Pressure, Extravascular

FDA/CORH/ODE/DMC

18 MAY 98 14 09

RECEIVED

4. Predicate Device(s):

Medex Inc.'s (Medex's) MX 860 Novatrans II reusable pressure transducer (K951129) and MX 960 (K961404), and Utah Medical Products, Inc.'s Monitoring Cables (K841788).

5. Device Description:

The BIOTRANS Pressure Monitoring System consists of the sterile, single-use BIOTRANS Pressure Monitoring Kit (subject of a separate 510k) and BIOTRANS Reusable Sensor Base and Monitor Adapter Cables. The Base consists of a rugged silicon-based piezoresistive pressure sensor mounted on a rigid plastic support. A standard Wheatstone Bridge circuit etched into the sensor is used to detect pressure changes applied to the sensor and convert them to electronic signals for display on standard monitoring equipment.

The sensor is electrically isolated from the patient by a dielectric gel and two elastomeric diaphragms. The physiologic pressure waveform is conveyed from the patient catheter via a fluid-filled pathway to the single-use dome diaphragm. The dome diaphragm is mechanically coupled to the reusable pressure sensor diaphragm for pressure monitoring. The diaphragms make intimate contact during use permitting transfer of the pressure waveform to the transducer, with fluid contact occurring on the patient side of the disposable dome diaphragm. A 25 cm (10") Sensor cable permanently wired to the sensor transmits the electronic signal from the transducer to the patient monitor via the appropriate Monitor Adapter Cable.

The integral Sensor cables terminate in 5 different electrical adapters compatible with 5 common brands of Monitor Cables (Utah Medical, Baxter International, Medex, Abbott, and Ohmeda.) Monitor Adapter Cables designed for specific use with all major monitor suppliers are available from BSI.

The Base features built-in organizer/mounting systems compatible with standard manifold clamps for pole-mount applications.

{1}------------------------------------------------

6. Intended Use:

The BSI BIOTRANS is used to convert hemodynamic pressure waveforms into electrical signals which can be displayed using separate monitoring equipment. As such, its Indications for Use are identical to those for Medex's MX 860 and MX 960. BIOTRANS cables are used to make an electrical connection between the sensor and monitor. Their use is identical to that of cables manufactured by Utah Medical Products, Inc.

7. Summary of Technological Characteristics of Device in relation to Predicate Device(s):

The electronic design of the BSI BIOTRANS is functionally identical to that of the predicate devices. Both the BSI BIOTRANS Sensors and predicate devices employ a piezoresistive Wheatstone bridge circuit to convert physiologic pressure waveforms to an electrical signal for monitoring purposes. BIOTRANS and predicate sensors and cables devices have been shown to meet the American National Standard for Blood Pressure Transducers, ANSI/AAMI BP22-1994.

The BIOTRANS Pressure Monitoring System employs a snap fit to secure the singleuse dome to the reusable sensor base, assuring intimate mechanical contact of the dome diaphragm with the sensor diaphragm. The predicate device incorporates a bayonet mechanism to secure the dome to the base. The snap fit configuration meets all electrical requirements listed in the ANSI/AAMI BP22-1994 standard.

8. Assessment of Performance Data used to justify Substantial Equivalence Claim

BSI's BIOTRANS meets, or exceeds, the performance requirements indicated in the American Standard for Blood Pressure Transducers, ANSI/AAMI BP22-1994. The requirements of this standard were likewise met by the predicate devices.

9. Conclusion:

Based on the test data gathered, Biosensors BIOTRANS Reusable Sensor Bases and Cables are substantially equivalent to Medex's reusable blood pressure transducers, Cat. No. MX 860 and MX 960, and Utah Medical Products Monitor Cables, respectively.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The eagle is composed of three curved lines that suggest the feathers of the wings and the head of the eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

JUL 1 1 1950

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. John E. Shulze Sunscope International, Inc. 20250 Acacia Street, Suite 115 Newport Beach, CA 92660

Re: K981745 BIOTRANS™ Reusable Sensor Base and Monitor Adapter Cables Requlatory Class: II (two) Product Code: 74 DRS May 15, 1998 Dated: Received: May 18, 1998

Dear Mr. Shulze:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page__________________________________________________________________________________________________________________________________________________________________________

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ 17.070

Device Name: BIOTRANS REUSABLE SENSOR BASE

Indications For Use:

The BSI BIOTRANS Sensor Base is used to convert hemodynamic pressure waveforms via a piezoresistive bridge circuit into electrical signals which can be displayed using separate monitoring equipment.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark Kram
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) NumberK98 1745
Prescription UseOROver-The-Counter Use
(Per 21 CFR 801.109)(Optional Format 1-2-96)

§ 870.2850 Extravascular blood pressure transducer.

(a)
Identification. An extravascular blood pressure transducer is a device used to measure blood pressure by changes in the mechanical or electrical properties of the device. The proximal end of the transducer is connected to a pressure monitor that produces an analog or digital electrical signal related to the electrical or mechanical changes produced in the transducer.(b)
Classification. Class II (performance standards).