K Number
K981255
Date Cleared
1998-04-29

(23 days)

Product Code
Regulation Number
864.5425
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sigma Diagnostics ACCUCLOT™ CONTROL III is a human plasma control that is suitable for use as an abnormal control with patient citrated plasma in the one-stage prothrombin time (PT), in the activated partial thromboplastin time (APTT), the chromogenic Antithrombin III (AT-III) assay and in fibrinogen determinations using a clotting method. Plasma controls are routinely used in the coagulation laboratory as a means of quality control for coagulation testing

Device Description

Sigma Diagnostics ACCUCLOT™ Control III is a lyophilized human plasma based product. After reconstitution with water, ACCUCLOT™ Control III is stable for 48 hours when stored at 2-8°C.

AI/ML Overview

The provided text is a 510(k) summary for the Sigma Diagnostics ACCUCLOT™ Control III, which is a human plasma control. The document states that the safety and effectiveness of this device were demonstrated by its substantial equivalence to a legally marketed predicate device, the Sigma Diagnostics Coagulation Control, Level III (Cat. No. 9916).

Crucially, this document does not detail specific acceptance criteria for the ACCUCLOT™ Control III or a study with performance data against those criteria. Instead, it relies on the concept of substantial equivalence to an existing device.

Therefore, for many of your specific questions, the information is not available in the provided text, as the method of demonstrating effectiveness was comparative to a predicate, not through a study with pre-defined acceptance criteria and reported performance metrics.

Here is what can be extracted and what is not available based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

Not available in the provided text. The submission relies on substantial equivalence to a predicate device rather than presenting specific acceptance criteria for the ACCUCLOT™ Control III and then reporting its performance against those criteria.

2. Sample Size Used for the Test Set and Data Provenance

Not available. No specific test set or study data is presented for the ACCUCLOT™ Control III. Its effectiveness is based on equivalence to the predicate device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable/Not available. No specific test set requiring expert-established ground truth is described for the ACCUCLOT™ Control III.

4. Adjudication Method for the Test Set

Not applicable/Not available.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No. This type of study is typically for evaluating diagnostic imaging or interpretation systems with human readers. The ACCUCLOT™ Control III is a laboratory control, not a diagnostic imaging device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

Not available/Applicable in the way usually meant for devices with algorithms. The ACCUCLOT™ Control III is a laboratory quality control product, not an algorithmic diagnostic tool. Its performance is inherent in its chemical and biological properties when used in assays. The "standalone" performance would refer to its stability and performance in the coagulation assays, but specific data is not provided, only the claim of equivalence.

7. The Type of Ground Truth Used

The concept of "ground truth" as typically applied to diagnostic algorithms (e.g., pathology, outcomes data) does not directly apply here. For a quality control product, the "ground truth" in showing effectiveness is its ability to consistently yield expected results within specified ranges when used with reference methods, and its stability. However, explicit data on this for the ACCUCLOT™ Control III is not provided; rather, its equivalence to a predicate control is asserted.

8. The Sample Size for the Training Set

Not applicable/Not available. This device is a control material, not a machine learning algorithm that requires a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable/Not available.


Summary of Device and Basis for Approval per Provided Text:

  • Device Name: Sigma Diagnostics ACCUCLOT™ Control III
  • Intended Use: A human plasma control suitable for use as an abnormal control with patient citrated plasma in one-stage prothrombin time (PT), activated partial thromboplastin time (APTT), chromogenic Antithrombin III (AT-III) assay, and fibrinogen determinations using a clotting method. Routinely used for quality control in coagulation testing.
  • Demonstration of Safety and Effectiveness: By its substantial equivalence to the Sigma Diagnostics Coagulation Control, Level III (Cat. No. 9916).
  • Key Characteristics: Lyophilized human plasma-based product, stable for 48 hours when stored at 2-8°C after reconstitution with water.

{0}------------------------------------------------

K9812JJ

APR 2 9 1998 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ACCUCLOT™ Control III, Cat. No. A6089

Sigma Diagnostics ACCUCLOT™ Control III is a human plasma control that is suitable for use as a abnormal control with patient citrated plasma in the one-stage prothrombin time (PT), in the activated partial thromboplastin time (APTT), in the chromogenic Antithrombin III (AT-III) assay and in fibrinogen determinations using a clotting method.

The safety and effectiveness of the Sigma Diagnostics ACCUCLOT™ Control III has been demonstrated by its substantial equivalence to the Sigma Diagnostics Coagulation Control, Level III (Cat. No. 9916).

Sigma Diagnostics ACCUCLOT™ Control III is a lyophilized human plasma based product. After reconstitution with water, ACCUCLOT™ Control III is stable for 48 hours when stored at 2-8°C.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

APR 2 9 1998

William Gilbert, Ph.D. .Manager, Scientific Affairs Sigma Diagnostics Inc. 545 South Ewing Avenue St. Louis, Missouri 63103

K981255 Re : ACCUCLOT™ Control III Regulatory Class: II Product Code: GGC, GGN Dated: April 2, 1998 Received: April 6, 1998

Dear Dr. Gilbert:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as-set --forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{2}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Gutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): _

Device Name: Sigma Diagnostics ACCUCLOT™ CONTROL III_

Indications For Use:

Sigma Diagnostics ACCUCLOT™ CONTROL III is a human plasma control that is suitable for use as an abnormal control with patient citrated plasma in the one-stage prothrombin time (PT), in the activated partial thromboplastin time (APTT), the chromogenic Antithrombin III (AT-III) assay and in fibrinogen determinations using a clotting method. Plasma controls are routinely used in the coagulation laboratory as a means of quality control for coagulation testing

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices

510(k) NumberK981258
------------------------
Prescription Use(Per 21 CFR 801.109)OROver-The-Counter Use
--------------------------------------------------------------------

§ 864.5425 Multipurpose system for in vitro coagulation studies.

(a)
Identification. A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.(b)
Classification. Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.