AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CODMAN Neurotrend Cerebral Tissue Monitoring System measures Intracranial oxygen, carbon dioxide, pH and temperature, and is intended as an indicallarial oxygon, and in these parameters, indicating the perfusion and metabolic acidosis/alkalosis status of cerebral tissue local to sensor placement. metabolic academand values are relative within an individual, the Neurotrend should not be used as the sole basis for decisions as to diagnosis or therapy. It is intended to provide data additional to that obtained by current clinical practice, in cases where hypoxia/ischaemia is a concem.

Device Description

The Neurotrend Sensor (C7004S) is a modified version of the Fluorescent Paratrend 7 sensor which has been cleared for the US Market (K953893). The measurement parameters for Neurotrend are essentially the same as the Fluorescent Paratrend 7. The sensor does not require a heparin treatment, as a consequence of the application - to monitor cerebral tissue and fluid gas parameters.

AI/ML Overview

The provided text describes the Neurotrend Multiparameter Sensor, a device intended to measure intracranial oxygen, carbon dioxide, pH, and temperature. The submission focuses on demonstrating substantial equivalence to a predicate device by showing that the modification (absence of heparin treatment) does not negatively impact performance.

Here's an analysis of the acceptance criteria and study information based on the provided text:

Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in terms of specific thresholds for bias, precision, or drift. Instead, the "acceptance criteria" appear to be met by demonstrating comparable performance between the modified device (non-heparin treated) and the predicate device (heparin treated) in terms of accuracy characteristics.

Parameter MeasuredAcceptance Criteria (Implied)Reported Device Performance
Accuracy (pO₂, pCO₂, pH, Temp)Demonstrated comparable bias and precision to the predicate heparin-treated sensor."The data satisfy product accuracy claims and show comparable results for heparin-treated and non-heparin treated sensors."
90% Step Response TimeComparable step response times to the predicate heparin-treated sensor.Step response times were calculated and used in the comparison. No specific values provided, but results were part of the "comparable results."
Drift (over 72 hours)Comparable mean drift to the predicate heparin-treated sensor.Mean drift was calculated and used in the comparison. No specific values provided, but results were part of the "comparable results."

Study Information

  1. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size:
      • Heparin-treated sensors (predicate): n = 8
      • Non-heparin treated sensors (device under review): n = 7
    • Data Provenance: The study was an in vitro test, not involving human data. The devices were manufactured in the UK (based on the submitter's address).
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • The ground truth for the in vitro test set was established using precision gas mixtures and the Henderson-Hasselbalch equation for pH in tonometers maintained at 37°C. This is a laboratory-based, objective method, so no human experts were used to establish the ground truth for this in vitro study.
  3. Adjudication Method for the Test Set:

    • No adjudication method was mentioned as the ground truth was objectively determined by precision instruments and physical/chemical equations.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, a multi-reader, multi-case comparative effectiveness study was not done. This was an in vitro performance study comparing device characteristics.
  5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • This is not applicable in the context of this device. The Neurotrend Sensor directly measures physiological parameters. The "standalone" performance described here is the in vitro performance of the sensor itself, without human interpretation of the raw measurements.
  6. The Type of Ground Truth Used:

    • For the in vitro study, the ground truth was established by precision gas mixtures for partial pressures of O2 and CO2, the Henderson-Hasselbalch equation for pH, and a reference thermometer for temperature within equilibrated tonometers. This is an objective, laboratory-based physical/chemical ground truth.
  7. The Sample Size for the Training Set:

    • The document does not describe the development or training of an algorithm in the traditional sense. It's a sensor device providing direct measurements. Therefore, there's no "training set" for an AI algorithm as typically understood. The sensors themselves are calibrated according to the Instructions For Use (IFU), but this is part of device manufacturing and operational procedure, not an algorithm training process.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable as no algorithm training set is described. The "calibration" of the sensors is done against primary standards in a manufacturing/testing environment.

{0}------------------------------------------------

K980380

Section 11 : 510(k) Summary

This section satisfies the 510(k) summary as required by 21 CFR 807-92 (C).

1. Submitter

Miss Jan Walters Diametrics Medical Limited Short Street High Wycombe Bucks HP11 2QH UK Tel: +44 1494 471671 Fax: +44 1494 474890 Date of Summary : 30th January 1998

2. Device Name

Proprietary : Neurotrend Multiparameter Sensor, with Paratrend 7+ Multiparameter Senior & Satellite Monitor System

Common : Multiparameter Catheter

3. Device Classification

Panel: Neurology Devices Panel
Classification: Class II
FDA CFR: 21 CFR 882.1620
Description: Device, Intracranial Pressure Monitoring
Classification No.: 84GWM

4. Statement of Substantial Equivalence

Diametrics Medical Ltd claims substantial equivalence to products which have received FDA clearance by the Premarket Notification process.

{1}------------------------------------------------

న్న Device Description

The Neurotrend Sensor (C7004S) is a modified version of the Fluorescent Paratrend 7 sensor which has been cleared for the US Market (K953893). The measurement parameters for Neurotrend are essentially the same as the Fluorescent Paratrend 7. The sensor does not require a heparin treatment, as a consequence of the application - to monitor cerebral tissue and fluid gas parameters.

6. Technological Characteristics

Measurement Technology

pO₂ :fibre optic, fluorescence quenching
Temp :thermocouple
pCO₂ :fibre optic, photometric absorption
pH :fibre optic photometric absorption

Monitor Technology

Analogue/digital

7. Intended Use

The Neurotrend Multiparameter Sensor measures intracranial oxygen, carbon dioxide, pH, and temperature. The Sensor is used in conjunction with the Paratrend 7+ Multiparameter Senior and/or Satellite Monitor Systems.

8. Predicate Devices

.

K953893Fluorescent Paratrend 7 Multiparameter Blood Gas MonitoringSensor
K962928Neurocare Intracranial Pressure & Temperature Monitoring Kit
K914497Microsensor Intracranial Pressure Transducer
K896515Laserflo Blood Perfusion Monitor 8PM₂

588

{2}------------------------------------------------

9. Comparison of Accuracy to the Predicate Fluorescent Paratrend 7 Device

The device under review was manufactured under standard process conditions and calibrated on the Senior Monitor.

Sensors were tested in vitro to determine whether the absence of the heparin treatment affected the performance of the sensors. Sensors, both heparin treated (n=8) and untreated (n=7), were calibrated as detailed in the Instructions For Use (IFU).

The tonometers were maintained at 37°C and were pre-equilibrated with precision gas mixtures to precise barometric pressure which allowed the calculation of "actual" partial pressures. The Henderson-Hasselbalch equation was used to determine the "actual" pH values in the different tonometers.

The 90% step response times were calculated from the data collected when moving the sensors from one equilibrated tonometer to another. Sensors were then allowed to run for 72 hours and the bias at the end of this time was averaged over the period to give mean drift.

The data were analysed by calculating the bias (the mean of the differences between the measured Paratrend 7 value and that the actual value of the equilibrated solution) and the precision (sample standard deviation of the differences). The data were collected over the relevant range of gases.

The results from the sensors were then collated and used to determine the performance characteristics of both sub-sets of sensors versus the primary standard.

10. Conclusion

The data satisfy product accuracy claims and show comparable results for heparin-treated and non-heparin treated sensors.

The device under review is considered substantially equivalent.

589

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Public Health Service

JUL 1 4 1999

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Jan Walters Diametrics Medical, Ltd. Short Street High Wycombe, Bucks UNITED KINGDOM HP11 2QH

Re: K980380

Neurotrend Multiparameter Sensor C7004S Dated: February 23, 1999 Received: February 25, 1999

Dear Ms. Walters:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OO) for Medical Devices: General regulation (21 CFR Part 820) and that, through veriodic (QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

3

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally parketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Philip J. Phillips
Philip J. Phillips

Deputy Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Attachment 1

510(k) Number : K980380

Page 1 of 1

Device Name: CODMAN Neurotrend™ Multiparameter Monitoring System

Indications For Use:

"The CODMAN Neurotrend Cerebral Tissue Monitoring System measures Intracranial oxygen, carbon dioxide, pH and temperature, and is intended as an indicallarial oxygon, and in these parameters, indicating the perfusion and metabolic acidosis/alkalosis status of cerebral tissue local to sensor placement. metabolic academand values are relative within an individual, the Neurotrend should not be used as the sole basis for decisions as to diagnosis or therapy. It is intended to provide data additional to that obtained by current clinical practice, in cases where hypoxia/ischaemia is a concem."

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER RAGE IF NECESSARY) 7/14/99

Concurrence of CDRH, Office of Office of Device Evaluation (ODE)
ADE

Prescription Use
(Per 21 CFR §801.109)

OR

Over-the-Counter Use (Optional Format 1-2-96)

$\zeta$

§ 882.1620 Intracranial pressure monitoring device.

(a)
Identification. An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.(b)
Classification. Class II (performance standards).