K Number
K974901
Manufacturer
Date Cleared
1998-09-29

(272 days)

Product Code
Regulation Number
866.3175
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Digene Hybrid Capture® CMV DNA Assay is a qualitative, in vitro, diagnostic assay intended for the detection of human cytomegalovirus (CMV) DNA in human peripheral white blood cells isolated from whole blood specimens collected in EDTA. It is indicated for use as an aid in diagnosing CMV infection in solid organ transplant, bone marrow transplant and HIV/AIDS patients. This assay has not been cleared by the FDA for blood/plasma donor screening.

Device Description

The Hybrid Capture® CMV DNA Assay is a qualitative in vitro diagnostic assay. It is a nucleic acid, signal enhanced, solution hybridization, antibody capture assay that uses chemiluminescent signaling to detect CMV DNA. The assay kit consists of 13 reagents and two accessories.

Whole blood specimens are treated with an agent that lyses red blood cells, and the specimen is centrifuged, resulting in a pellet of white blood cells. The white blood cell pellet is then used in the Hybrid Capture CMV DNA Assay.

Specimens potentially containing CMV DNA are denatured and then hybridized with a specific CMV RNA probe cocktail. This cocktail contains a probe mixture chosen to eliminate cross-reactivity with human or other herpesvirus sequences. The CMV probe supplied with the Hybrid Capture CMV DNA Assay is complementary to approximately 40,000 base pairs or 17% of the CMV genome (230,000 base pairs).

The RNA:DNA hybrids resulting from hybridization are captured on the surface of a tube coated with affinity-purified polyclonal caprine antibodies specific for RNA:DNA hybrids. The immobilized hybrids are then reacted with alkaline phosphatase-conjugated, murine monoclonal antibody to RNA:DNA hybrids, and are detected with a chemiluminescent substrate. Several alkaline phosphatase molecules are conjugated to each antibody. Multiple conjugated antibodies bind to each captured hybrid, resulting in signal enhancement. As the substrate is cleaved by the bound alkaline phosphatase, light is emitted and is measured in Relative Light Units (RLUs) on a standard commercial luminometer. The RLU value of a specimen is compared to a Positive Cutoff Value and to an Equivocal Cutoff Value to determine if the specimen is positive, equivocal, or negative for the presence of CMV DNA.

AI/ML Overview

Here's a summary of the acceptance criteria and the study details for the Digene Hybrid Capture® System CMV DNA Assay, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

The provided document does not explicitly present a table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, or accuracy derived from a pre-defined threshold. Instead, it describes analytical sensitivity and clinical performance relative to predicate devices and established methods.

The key performance metrics and their reported values are:

MetricAcceptance Criteria/Reported Performance
Analytical Sensitivity (Positive Cutoff)0.98 pg/mL (0.81 - 1.28) of plasmid CMV DNA
Analytical Sensitivity (Equivocal Cutoff)0.48 pg/mL (0.40 - 0.63) of plasmid CMV DNA
SpecificityDemonstrated with a battery of blood-borne microorganisms, viruses, human genomic DNA, and CMV-related viruses. "Specificity" and "agreement in diagnostic outcome" also demonstrated in a population with approximately equal numbers of CMV seropositive and seronegative individuals. Precise performance metrics (e.g., % specificity) are not provided.
ReproducibilityShown to be reproducible over a range of CMV DNA concentrations, with "acceptable" within day, between days, between sites, and overall percent coefficients of variation and standard deviations, as well as agreement in diagnostic outcome. Precise quantitative results (e.g., CV% values) are not provided.
Clinical PerformancePerformed "comparably" to shell vial culture and cell culture in detecting CMV in HIV/AIDS patients and solid-organ or bone marrow transplant patients. It performed "as well or better than cell culture." Precise sensitivity/specificity/PPA/NPA relative to the predicate or composite reference standard are not explicitly quantified with percentages or confidence intervals in the summary, though the claim of "comparable" and "as well or better" constitutes the implied acceptance.

Study Details

  1. Sample size used for the test set and data provenance:

    • Clinical Testing: The summary mentions "clinical testing in HIV/AIDS patients and patients who had undergone a solid-organ or bone marrow transplants" and "parallel testing of specimens from solid organ transplant, bone marrow transplant and HIV/AIDS patients." It also states testing in "a population comprised of approximately equal numbers of CMV seropositive and seronegative individuals."
    • Sample Size: The exact number of patients or specimens in these clinical test sets is not specified in the provided document.
    • Data Provenance: The document does not specify the country of origin of the data. The studies appear to be prospective clinical studies, as indicated by "clinical testing" and "parallel testing of specimens."
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document implies that the ground truth for clinical performance was established by comparing the device's results to "shell vial culture and cell culture" and "expected results based on a combination of the cell culture result, the shell vial culture result, and clinical information."
    • It does not specify the number of individual experts (e.g., clinicians, microbiologists) involved in establishing the "clinical information" component of the ground truth or their specific qualifications (e.g., years of experience). The reference methods (cell culture, shell vial culture) are laboratory-based and generally performed by trained laboratory personnel.
  3. Adjudication method for the test set:

    • The primary adjudication method for clinical performance appears to be a composite reference standard: "expected results based on a combination of the cell culture result, the shell vial culture result, and clinical information."
    • The specific method of combining these different results (e.g., how conflicts were resolved, if a 2+1 or 3+1 rule was used) is not detailed in the summary.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This device is an in vitro diagnostic (IVD) assay for detecting CMV DNA, not an imaging device or an AI-assisted diagnostic system where "human readers" would be involved in interpreting images or other complex data. Therefore, an MRMC study related to human reader improvement with AI assistance is not applicable and was not performed. The assay generates a quantitative RLU value that is compared to predefined cutoffs.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, this is a standalone assay. The entire evaluation described is of the assay itself (Hybrid Capture® System CMV DNA Assay) without human interpretation steps that would be "in-the-loop." The luminometer measures RLU values, and the comparison to cutoff values for positive, equivocal, or negative results is algorithm-driven (or rule-based), not requiring human reader judgment for the final diagnostic outcome from the assay itself.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For analytical performance, the ground truth was derived from known dilutions of plasmid CMV DNA.
    • For clinical performance, the ground truth was a composite reference standard using traditional laboratory methods (shell vial culture, cell culture) combined with "clinical information." This clinical information could encompass patient symptoms, serology, treatment response, and other relevant medical data, potentially involving physician adjudication or consensus, though specific details are not provided.
  7. The sample size for the training set:

    • The document does not explicitly refer to a "training set" in the context of machine learning or AI. This is a traditional IVD assay. The development of the assay (e.g., defining the probe cocktail, optimizing reaction conditions, establishing cutoff values) would involve empirical studies and iterative testing/optimization, but these are not typically referred to as "training sets" in the same way as in AI/ML validation studies. The analytical sensitivity studies on known concentrations of CMV DNA would contribute to establishing cutoff values but aren't described as a "training set."
  8. How the ground truth for the training set was established:

    • As there is no explicitly defined "training set" in the AI/ML sense, this question is not directly applicable. The establishment of assay parameters (like cutoff values for positive/equivocal results) would be based on studies using known CMV DNA concentrations (as mentioned for analytical sensitivity) and potentially retrospective clinical samples alongside the predicate methods.

{0}------------------------------------------------

K974901

SEP 2 9 1998

510(k) SUMMARY

1

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Submitter

Digene Corporation 9000 Virginia Manor Road Beltsville, MD 20705

Phone:(301) 470-6500
Facsimile:(301) 470-2881

Contact Person Mark Del Vecchio Associate Director, Regulatory and Clinical Affairs Digene Corporation (301) 470-6287 Phone: (301) 470-2881 Facsimile:

Date Prepared: September 10, 1998

Name of Device and Name/Address of Sponsor

Name of Device Hybrid Capture® System CMV DNA Assay

Sponsor Digene Corporation 9000 Virginia Manor Road Beltsville, MD 20705 Tel: 301-470-6500 Fax: 301-680-0696

Common or Usual Name

Hybrid Capture® CMV DNA Assay

Classification Name

Cytomegalovirus DNA Reagents

{1}------------------------------------------------

Predicate Devices

CMV Cell Culture (in use prior to May 28, 1976) INCSTAR Corporation's CMV-vue™ Kit ( K921616)

Device Description

The Hybrid Capture® CMV DNA Assay is a qualitative in vitro diagnostic assay. It is a nucleic acid, signal enhanced, solution hybridization, antibody capture assay that uses chemiluminescent signaling to detect CMV DNA. The assay kit consists of 13 reagents and two accessories.

Whole blood specimens are treated with an agent that lyses red blood cells, and the specimen is centrifuged, resulting in a pellet of white blood cells. The white blood cell pellet is then used in the Hybrid Capture CMV DNA Assay.

Specimens potentially containing CMV DNA are denatured and then hybridized with a specific CMV RNA probe cocktail. This cocktail contains a probe mixture chosen to eliminate cross-reactivity with human or other herpesvirus sequences. The CMV probe supplied with the Hybrid Capture CMV DNA Assay is complementary to approximately 40,000 base pairs or 17% of the CMV genome (230,000 base pairs).

The RNA:DNA hybrids resulting from hybridization are captured on the surface of a tube coated with affinity-purified polyclonal caprine antibodies specific for RNA:DNA hybrids. The immobilized hybrids are then reacted with alkaline phosphatase-conjugated, murine monoclonal antibody to RNA:DNA hybrids, and are detected with a chemiluminescent substrate. Several alkaline phosphatase molecules are conjugated to each antibody. Multiple conjugated antibodies bind to each captured hybrid, resulting in signal enhancement. As the substrate is cleaved by the bound alkaline phosphatase, light is emitted and is measured in Relative Light Units (RLUs) on a standard commercial luminometer. The RLU value of a specimen is compared to a Positive Cutoff Value and to an Equivocal Cutoff Value to determine if the specimen is positive, equivocal, or negative for the presence of CMV DNA.

Using a test panel consisting of dilutions of a stock solution of plasmid CMV DNA complementary to the CMV RNA probe (approximately 39,000 base pairs), the analytical sensitivity of the Hybrid Capture® CMV DNA Assay at the Positive Cutoff was determined to be 0.98 pg/mL (0.81 - 1.28). The analytical sensitivity of the assay at the Equivocal Cutoff was determined to be 0.48 pg/mL (0.40 - 0.63). Testing with a battery of blood-borne microorganisms, viruses, and human genomic DNA, as well as a collection of viruses known to be related to CMV, has demonstrated that the CMV probes used in the Hybrid Capture CMV DNA Assay are specific for CMV. A reproducibility study has shown that the Hybrid Capture CMV DNA Assay is reproducible over a range of CMV DNA concentrations, with acceptable within day, between days, between sites,

{2}------------------------------------------------

and overall percent coefficients of variation and standard deviations, as well as agreement in diagnostic outcome.

Clinical testing in HIV/AIDS patients and patients who had undergone a solid-organ or bone marrow transplants has demonstrated that the Hybrid Capture® CMV DNA Assay performs comparably to shell vial culture and cell culture in detecting CMV. Testing in a population comprised of approximately equal numbers of CMV seropositive and seronegative individuals has demonstrated that the assay is specific.

Intended Use

The Digene Hybrid Capture® CMV DNA Assay is a qualitative, in vitro diagnostic assay intended for the detection of human cytomegalovirus (CMV) DNA in human peripheral white blood cells isolated from whole blood specimens collected in EDTA. It is indicated for use as an aid in diagnosing CMV infection in solid organ transplant, bone marrow transplant and HIV/AIDS patients. This assay has not been cleared by the FDA for blood/plasma donor screening.

Technological Characteristics and Substantial Equivalence

The HC CMV DNA Assay is substantially equivalent to cell culture methods that were in use prior to May 28, 1976 for the isolation of CMV. While cell culture detects the replication of the microorganism, the HC CMV DNA Assay is intended to detect a component of the microorganism directly. The HC CMV DNA Assay is similar to cell culture methods in that white blood cells are used as specimens and in that controls are employed to ensure that reagents are working properly. Although the HC CMV DNA Assay is different from cell culture methods in its operating principles and technological characteristics, the effect of these differences can be assessed by comparing the number of CMV positives detected by each to expected results based on a combination of the cell culture result, the shell vial culture result, and clinical information. Parallel testing of specimens from solid organ transplant, bone marrow transplant and HIV/AIDS patients, has demonstrated that the HC CMV DNA Assay performs as well or better than cell culture in detecting CMV.

The Hybrid Capture® CMV DNA Assay is substantially equivalent to INCSTAR Corporation's CMV-vue™ Kit. The two assays have the same intended use and indications for use, detect a component of the cytomegalovirus, use whole blood from which white blood cells are isolated and tested, and include indirect immunoenzymatic procedures in their testing protocols. Although the assays have some differences in technological characteristics, the performance of these assays has been shown to be equivalent through testing clinical specimens and comparing the results to the same standard methods for CMV detection.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines suggesting movement or connection.

Public Health Service

SEP 2 9 1998

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mark Del Vecchio Associate Director. Regulatory and ·Clinical Affairs Digene Corporation 9000 Virginia Manor Road Beltsville, Maryland 20705

Re: K974901

Trade Name: Hybrid Capture System CMV DNA Assav Regulatory Class: II Product Code: LJO Dated: September 10, 1998 Received: September 11, 1998

Dear Mr. Del Vecchio:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

K974901 510(K) Number (if known): ....................................................................................................................................................

Device Name: Digene Hybrid Capture® System CMV DNA Assay

Indications for Use:

The Digene Hybrid Capture® CMV DNA Assay is a qualitative, in vitro, diagnostic assay intended for the detection of human cytomegalovirus (CMV) DNA in human peripheral white blood cells isolated from whole blood specimens collected in EDTA. It is indicated for use as an aid in diagnosing CMV infection in solid organ transplant, bone marrow transplant and HIV/AIDS patients. This assay has not been cleared by the FDA for blood/plasma donor screening.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANTOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubose

(Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use -

§ 866.3175 Cytomegalovirus serological reagents.

(a)
Identification. Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.(b)
Classification. Class II (performance standards).