K Number
K974684
Date Cleared
1998-02-19

(65 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SCIMED 7 French Wiseguide guide catheter is intended for use in general intravascular and coronary applications. It provides a pathway through which medical instruments, such as stents, balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. This device is not intended for use in the cerebral vasculature.

Device Description

The shaft of the 7 F Wiseguide guide catheter utilizes common biocompatible materials and consists of the following three layers: 1) the inner layer that provides a low coefficient of friction and facilitates passage of medical devices such as stents, balloon dilatation catheters, guide wires or other therapeutic devices, 2) the middle layer that extends from the shaft to the tip to provide kink resistance and torque control; and 3) the outer layer which provides stiffness, backup support, curve retention and radiopacity.

In addition, the outer primary catheter shaft is constructed of various material durometers providing a distal curve area of the catheter with transitional flexibility.

The distal tip is radiopaque to allow visualization under fluoroscopy during a procedure.

The devices will be provided sterile and are intended for one procedure use only.

AI/ML Overview

The medical device in question is the SCIMED® 7 French Wiseguide™ Guide Catheter, which is similar to Diagnostic Intravascular Catheters (21CFR Part 870.1200).

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Functional Test)Reported Device Performance
Pressure burstTest results verified as adequate for intended use.
Tip bond tensileTest results verified as adequate for intended use.
Shaft tensileTest results verified as adequate for intended use.
Hub tensileTest results verified as adequate for intended use.
Material adhesionTest results verified as adequate for intended use.
Tip coefficient of frictionTest results verified as adequate for intended use.
Force transmitted by catheter tipTest results verified as adequate for intended use.
Torque responseTest results verified as adequate for intended use.
Dye flowTest results verified as adequate for intended use.

2. Sample Size Used for the Test Set and Data Provenance:

The provided document describes non-clinical (in-vitro) testing. Therefore, there were no human patient test sets used. The document does not specify the exact sample sizes (number of units tested) for each functional test mentioned.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. This was an in-vitro functional testing study of a medical device, not a study requiring human expert assessment for ground truth.

4. Adjudication Method for the Test Set:

Not applicable. Adjudication methods are typically relevant for studies involving human interpretation or clinical outcomes, which is not the case for this in-vitro functional testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. The document only describes non-clinical (in-vitro) functional testing, not a comparative effectiveness study involving human readers or cases.

6. Standalone (Algorithm Only) Performance Study:

Not applicable. This is a physical medical device, not an algorithm, so a standalone algorithm performance study is not relevant.

7. Type of Ground Truth Used:

The "ground truth" for this study was established through engineering specifications and performance standards for various physical and mechanical properties of the catheter. The functional tests (e.g., pressure burst, tensile strength, torque response) directly measured whether the device met these predefined engineering requirements.

8. Sample Size for the Training Set:

Not applicable. This document describes the functional testing of a specific medical device, not the development or training of a machine learning algorithm.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there was no training set for an algorithm. The "ground truth" for the device's design and performance expectations would have been established during its development based on engineering principles, predicate device performance, and the intended clinical use.

{0}------------------------------------------------

K974684

Section 4

Summary of Safety and Effectiveness

(Pursuant To Section 12 of the SAFE MEDICAL DEVICES ACT of 1990)

FEB 1 9 1998

I. General Provisions

Submitter's Nameand AddressSCIMED Life Systems, Inc.One SCIMED PlaceMaple Grove, Minnesota 55311
Contact PersonMelanie Raska(612) 494-2962
Classification NameSimilar to Diagnostic IntravascularCatheters (21CFR Part 870.1200)
Common or Usual NameCoronary Guide Catheter
Proprietary NameSCIMED® 7 French Wiseguide™ GuideCatheter
Name of Predicate DevicesSCIMED® 8 French Wiseguide™ and 7French Triguide® -Max Guide Catheters,and Cordis® Corporation 7 French VistaBrite Tip™ Guide Catheters

III. Device Description

II.

The shaft of the 7 F Wiseguide guide catheter utilizes common biocompatible materials and consists of the following three layers: 1) the inner layer that provides a low coefficient of friction and facilitates passage of medical devices such as stents, balloon dilatation catheters, guide wires or other therapeutic devices, 2) the middle layer that extends from the shaft to the tip to provide kink resistance and torque control; and 3) the outer layer which provides stiffness, backup support, curve retention and radiopacity.

In addition, the outer primary catheter shaft is constructed of various material durometers providing a distal curve area of the catheter with transitional flexibility.

The distal tip is radiopaque to allow visualization under fluoroscopy during a procedure.

The devices will be provided sterile and are intended for one procedure use only.

{1}------------------------------------------------

Section 4

Summary of Safety and Effectiveness (cont.)

IV. Intended Use

The 7 F Wiseguide catheter is designed to provide a pathway through which medical instruments, such as stents, balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. This device is not intended for use in the cerebral vasculature.

Summary of Technological Characteristics: V.

The 7 F Wiseguide catheter is similar to SCIMED's currently marketed 8 F Wiseguide catheter.

VI. Non-clinical Test Summary

Functional testing consisted of pressure burst, tip bond tensile, shaft tensile, hub tensile, material adhesion, tip coefficient of friction, force transmitted by catheter tip, torque response and dye flow. Test results verified that the 7 F Wiseguide catheter is adequate for its intended use. The 7 F Wiseguide catheter is considered substantially equivalent to guide catheters currently marketed by SCIMED and Cordis based on a comparison of intended use, the design, and the results of in-vitro testing and evaluation.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is positioned to the right of a circular emblem containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856

FEB 1 9 1998

Ms. Melanie Raska Requlatory Affairs Specialist Scimed Life System, Inc. Boston Scientific Corporation One Scimed Place Maple Grove, MN 55311-1566

Re : K974684 Scimed® 7 French Wiseguide™ Guide Catheter Regulatory Class: II (Two) Product Code: 73 DQY December 15, 1997 Dated: Received: December 16, 1997

Dear Ms. Raska:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. ----------------------------------------------------------------------------------------------------------------------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality and him System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the

{3}------------------------------------------------

Page 2 - Ms. Melanie Raska

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Notification SCIMED® 7 F Wiseguide Guide Catheter

Section 2

Indications for Use

510(k) Number (if known) _

Device Name: SCIMED® 7 French Wiseguide™ Guide Catheter

Indications for Use:

SCIMED 7 French Wiseguide guide catheter is intended for use in general intravascular and coronary applications. It provides a pathway through which medical instruments, such as stents, balloon dilatation catheters, guide wires or other therapeutic devices may be introduced. This device is not intended for use in the cerebral vasculature.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR Over The Counter Use __

(Optional Format 1-2-96)

(Division Sign-Off) Division of Cardinvasc Dar and Neurological Devices 510(k) Number

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).