K Number
K974661
Manufacturer
Date Cleared
1998-02-11

(58 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 0.2 µ Bacterial Filter can be used with administration sets or extension sets for removal of particulate matter during the administration of fluids or medications.

Device Description

The purpose of this submission is to offer an 0.2 u Bacterial Filter with standard luer connections for use with administration sets or extension sets used with fluid delivery devices. The subject device is intended to be "added-on" to an administration set or extension set.

The filter is an air-eliminating filter and is circular in shape with a center vent. The inlet of the filter is a standard female luer and the outlet of the filter is a standard male luer. The filter membranes are enclosed within a plastic case which is approximately 1.22 in. (width) x 1.85 in. (length).

AI/ML Overview

The provided 510(k) summary for the 0.2 µ Bacterial Filter by SIMS Deltec, Inc. primarily describes a device comparison rather than a detailed study with acceptance criteria and specific performance metrics in the way a diagnostic AI/ML device would be evaluated. This medical device submission is from 1997, and the regulatory requirements and types of studies for devices like this were different than those for modern AI-driven devices.

Here's an analysis based on the provided text, highlighting the information that is present and noting what is absent:

1. Table of Acceptance Criteria and Reported Device Performance:

The submission does not contain a table with explicit acceptance criteria (e.g., a specific bacterial retention rate or flow rate threshold) and reported device performance results against those criteria. Instead, it makes a general statement about functional testing and biocompatibility.

Acceptance Criteria (Implied)Reported Device Performance
Filtration of particulate matter (Implied by intended use)"the filter functions according to specification"
Biocompatibility requirements"the filter meets the biocompatibility requirements"

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not specified. The document mentions "Functional testing was performed on the filter," but does not provide details on the number of filters tested or the amount of fluid/particulate matter used.
  • Data Provenance: Not specified. The testing was presumably conducted by SIMS Deltec, Inc., but the location of the testing or origin of materials is not detailed. It is retrospective, as the testing was completed prior to submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. This device is a filter, and its performance is evaluated through physical and biological testing (e.g., filtration efficiency, biocompatibility), not through expert interpretation of data.

4. Adjudication Method for the Test Set:

Not applicable. As noted above, this device is not evaluated by human experts interpreting data.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC study was not done. This type of study is relevant for diagnostic devices where human readers interpret medical images or data. The 0.2 µ Bacterial Filter is a physical device, not an interpretative diagnostic tool.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

Not applicable. This is not an AI/ML algorithm. The performance of the filter itself, as a physical object, is what was tested.

7. Type of Ground Truth Used:

The "ground truth" for this device would be established by validated laboratory methods for assessing filtration efficiency (e.g., bacterial retention tests, particulate removal tests) and standardized biocompatibility assays. The document states:

  • "Functional testing was performed on the filter to establish its operating parameters."
  • "Biocompatibility testing was conducted on the filter."
    This implies that the ground truth was based on these specific, objective laboratory measurements.

8. Sample Size for the Training Set:

Not applicable. This device is not an AI/ML algorithm, so there is no training set in the conventional sense. The "training" or development of the filter would involve engineering design and material selection, followed by testing.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set. The design of the filter (likely based on material science and fluid dynamics principles) would be iterated and refined through a series of internal tests, but these are not explicitly detailed as "ground truth" establishment for a training set.

Summary of the Study and Device Proof:

The submission outlines that the device meets its intended purpose based on functional testing and biocompatibility testing. The core of the proof is the demonstration that:

  • Functional Testing: The filter "functions according to specification," meaning it effectively removes particulate matter as advertised. While specific percentages or particle sizes are not given, the claim of a "0.2 µ Bacterial Filter" implies it can filter particles down to that size, which would have been demonstrated by the functional tests.
  • Biocompatibility Testing: The filter "meets the biocompatibility requirements," indicating it is safe for human contact and use without adverse biological reactions.

The submission specifically states: "Clinical studies were not deemed necessary regarding the filter due to its similarity in materials, design and function to current commercially available filters and extension sets with in-line filters." This is a key point, as it allowed SIMS Deltec, Inc. to rely on substantial equivalence to predicate devices rather than conducting extensive clinical trials, which would typically be required for novel device types or those with significant differences in performance or safety. The 510(k) process is fundamentally about demonstrating substantial equivalence to a legally marketed predicate device. By showing similar materials, design, and function, they could leverage the established safety and effectiveness of the predicate.

{0}------------------------------------------------

K974661

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

0.2 u Bacterial Filter

FEB | | 1998

December 12, 1997

GENERAL INFORMATION L

Applicant's Name and Address:SIMS Deltec, Inc.1265 Grey Fox RoadSt. Paul, MN 55112
Contact Person:Lisa J. StoneManager, Regulatory AffairsTel. (612) 628-7224
Common/Usual Name:0.2 μ Bacterial Filter
Proprietary Name:0.2 μ Bacterial Filter
Equivalence Device Comparison:Extension Set with microbore tubing and 0.2 μ filter(manufactured by SIMS Deltec, Inc.)Disposable I.V. Filter (0.2 μ)(manufactured by The Medi-Dose® Group)

II. DEVICE DESCRIPTION

The purpose of this submission is to offer an 0.2 u Bacterial Filter with standard luer connections for use with administration sets or extension sets used with fluid delivery devices. The subject device is intended to be "added-on" to an administration set or extension set.

The filter is an air-eliminating filter and is circular in shape with a center vent. The inlet of the filter is a standard female luer and the outlet of the filter is a standard male luer. The filter membranes are enclosed within a plastic case which is approximately 1.22 in. (width) x 1.85 in. (length).

INTENDED USE OF DEVICE III.

The 0.2 µ Bacterial Filter can be used with administration sets or extension sets for removal of particulate matter during the administration of fluids or medications.

{1}------------------------------------------------

510(k) Summary of Safety and Effectiveness Page 2 of 2

IV. DEVICE COMPARISON

0.2 μ Bacterial FilterExtension Set withmicrobore tubing and0.2 μ filterDisposable I.V. Filter(0.2 μ)
MANUFACTURERSIMS Deltec, Inc.SIMS Deltec, Inc.The Medi-Dose® Group
INDICATION FOR USEThe 0.2 μ Bacterial Filtercan be used withadministration sets orextension sets for removalof particulate matterduring the administrationof fluids or medications.The Extension Set withmicrobore tubing attachesto the Micro MedicationReservoir for theadministration of fluids ormedications with theCADD-Micro® pump.
FILTER SIZE0.2 μ0.2 μ0.2 μ
FILTRATIONSURFACE AREA$4.3 cm^2$$0.3 in^2$
ADD-ON FILTER WITHLUER LOCKCONNECTION(S)YESNOYES

V. SUMMARY OF STUDIES

A. Functional Testing

Functional testing was performed on the filter to establish its operating parameters.

Biocompatibility testing was conducted on the filter.

B. Clinical Studies

Clinical studies were not deemed necessary regarding the filter due to its similarity in materials, design and function to current commercially available filters and extension sets with in-line filters.

C. Conclusions Drawn from the Studies

The results of the testing indicated that the filter functions according to specification and the filter meets the biocompatibility requirements. Therefore, the filter is considered acceptable for human use.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's emblem in the center. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Lisa J. Stone Manager, Regulatory Affairs Sims Deltec, Incorporated 1265 Grey Fox Road St. Paul, Minnesota 55112

FEB 11 1998

K974661 Re : 0.2 μ Bacterial Filter Trade Name: Requlatory Class: II Product Code: FPB December 12, 1997 Dated: December 15, 1997 Received:

Dear Ms. Stone:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially.equivalent determination assumes compliance with-------the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{3}------------------------------------------------

Page 2 - Ms. Stone

through 542 of the Act for devices under the Electronic chrough 542 or the nec-ros-rovisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA deborizoof substantial equivalence of your device to a legally rinding of bubbeanelar equiredults in a classification for your markets production of mits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy b. Ulatowski

Timoth Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K974661

510(k) Number (if known):

Device Name: 0.2 u Bacterial Filter

Indications for Use:

"The 0.2 µ Bacterial Filter can be used with administration sets or extension sets for removal of particulate matter during the administration of fluids or medications."

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Paboux Cassante

(Division Sign-C Division of Dental, In n Control, and . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . and General Hospital 510(k) Number -

Prescription Use (Per 21 CFR 801.109)

OR

Over-The Counter Use

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.