K Number
K974521
Date Cleared
1998-02-20

(80 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VNUS Closure™ System is indicated for use in the coagulation of blood vessels during general surgery.

Device Description

The VNUS Closure™ System consists of three main components: The VNUS Closure Probe, the VNUS RF Generator and the VNUS Instrument Cable. The Closure Probe is provided Sterile, and is a single-use, disposable device. The RF Generator is non-sterile. The Instrument Cable is autoclave sterilized by the user. An optional Footswitch for RF ON/RF OFF is provided for use at the physician's discretion.

The RF Generator is a high frequency electronic, bipolar, software controlled instrument. It allows the user to set Power, Temperature and Time values, and provides user displays of Power. Temperature and Time (setpoints and measured values) as well as measured impedance and other messages. The RF Generator works in a temperature controlled, power limited manner, based on operator settings and temperature feedback provided by a thermocouple in the Closure Probe.

The Closure Probe is used to carry RF energy to the desired treatment site and provide temperature feedback to the RF Generator. It is designed to deliver the RF energy in a bipolar manner.

The Instrument Cable is used to connect the Closure Probe to the RF Generator.

AI/ML Overview

The provided document is a 510(k) premarket notification for the VNUS Closure System. It focuses on demonstrating substantial equivalence to predicate devices rather than conducting a detailed performance study with acceptance criteria and specific statistical hypothesis testing as might be found in a PMA (Premarket Approval) submission or a dedicated clinical trial report for an AI/ML device.

Therefore, much of the requested information regarding "acceptance criteria," "study proving device meets acceptance criteria," "sample sizes," "expert ground truth," "adjudication methods," "MRMC studies," "standalone performance," and "training set details" cannot be found within this document. These studies are typically performed for new or significantly modified devices, or for AI/ML devices where performance claims against a defined ground truth need to be established.

The document states:

  • "Non-clinical tests performed by VNUS have demonstrated the substantially equivalent performance of the Closure System with predicate electrosurgery systems used for substantially equivalent indications."
  • "Based upon the design, materials, function, intended use, comparison with currently marketed devices and the non-ciinical testing performed by VNUS, it is concluded that the Closure System is substantially equivalent to the noted predicate devices in safety and effectiveness."

This indicates that non-clinical testing (bench testing, perhaps animal studies) was conducted to demonstrate equivalence, but the specific details of "acceptance criteria" and how they were "proven" in a quantitative sense are not disclosed here. The basis for approval is substantial equivalence to existing devices, not a direct demonstration of meeting detailed performance metrics in human studies.


Given these limitations, here is what can be extracted and what cannot:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Not specified."Demonstrated substantially equivalent performance" to predicate devices based on non-clinical tests.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size (Test Set): Not specified. The document refers to "non-clinical tests," which typically involve bench testing and possibly animal models, not human subject test sets with specific sample sizes as would be relevant for AI/ML performance.
  • Data Provenance: Not specified. "Non-clinical tests performed by VNUS" suggests internal testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Number of Experts: Not applicable/not specified. This type of expert assessment for ground truth is typically relevant for medical imaging or diagnostic AI/ML devices, not for an electrosurgery system described in this 510(k). The "ground truth" for this device would likely be based on physical measurements and functional performance in a laboratory setting.
  • Qualifications of Experts: Not applicable/not specified.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: Not applicable/not specified, as there is no mention of expert ground truth establishment for a test set in the context of this device and submission type.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, an MRMC study was not done, and is not applicable to an electrosurgical device as described here. This type of study is specifically designed for evaluating diagnostic devices, especially those involving human interpretation of medical images or data (e.g., AI in radiology).

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Standalone Performance: Not applicable. An electrosurgical system is inherently a device used with human intervention. The concept of "standalone performance" as typically applied to an AI algorithm without human input doesn't fit this device. The system has automated features (temperature control, power limiting), but these are integral to its operation in human hands.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth: Not explicitly stated in terms of common medical imaging 'ground truth' types. For non-clinical tests of an electrosurgical device, ground truth would relate to physical parameters and outcomes of energy delivery (e.g., precise temperature readings, coagulation efficacy in tissue models, impedance measurements, safety parameters like absence of unintended tissue damage).

8. The sample size for the training set

  • Sample Size (Training Set): Not applicable/not specified. This is not an AI/ML device in the modern sense that would require a "training set" for model development. The system's software controls are likely based on engineering principles and embedded logic, not statistical machine learning from data.

9. How the ground truth for the training set was established

  • Ground Truth (Training Set): Not applicable/not specified, as there is no mention of a training set for an AI/ML model.

{0}------------------------------------------------

Summary of Safety and Effectiveness

  • Determination of Substantial Equivalence A. VNUS Closure™ System
    FEB 2 0 1998

в. Common Name

Electrosurgery System/Electrosurgical Coagulator

C. Predicate Device(s)

ArthroCare Electrosurgery System (K971532) Boston Surgical Products, Inc. Model 50 Bipolar Coagulator (K972350) Somnus Model 215 Electrosurgical Generator and Accessories (K971711) Valley Forge Scientific Mini-Symm Bipolar Coagulator (K964143) Bergen Model 500 Electrosurgery Generator (K972299) Bergen 610 Coagulator (K964736) Kirwan 28 1000 Coagulator (K971341)

K974521

D. Device Description

The VNUS Closure™ System consists of three main components: The VNUS Closure Probe, the VNUS RF Generator and the VNUS Instrument Cable. The Closure Probe is provided Sterile, and is a single-use, disposable device. The RF Generator is non-sterile. The Instrument Cable is autoclave sterilized by the user. An optional Footswitch for RF ON/RF OFF is provided for use at the physician's discretion.

The RF Generator is a high frequency electronic, bipolar, software controlled instrument. It allows the user to set Power, Temperature and Time values, and provides user displays of Power. Temperature and Time (setpoints and measured values) as well as measured impedance and other messages. The RF Generator works in a temperature controlled, power limited manner, based on operator settings and temperature feedback provided by a thermocouple in the Closure Probe.

The Closure Probe is used to carry RF energy to the desired treatment site and provide temperature feedback to the RF Generator. It is designed to deliver the RF energy in a bipolar manner.

The Instrument Cable is used to connect the Closure Probe to the RF Generator.

ய் Intended Use

The VNUS Closure™ Catheter System is intended for use in the coagulation of blood vessels during general surgery.

This premarket notification is being submitted as the VNUS Closure System is a new device. Electrosurgical Coagulation Devices and Accessories are Class 2 per 21 CFR 878.4400.

{1}------------------------------------------------

F. Intended Use of Predicate Devices

The specified predicate devices are indicated for "soft tissue resection", "soft body tissue coagulation", "the coagulation of tissue", and/or for "ablation and coagulation of blood vessels during general surgical procedures."

Technological Comparison G.

The technological characteristics and principals of operation of the VNUS Closure System are substantially equivalent to the noted predicate devices. All devices rely on the delivery of RF energy to achieve their intended use.

H. Discussion of Non-Clinical Tests and Conclusions

Non-clinical tests performed by VNUS have demonstrated the substantially equivalent performance of the Closure System with predicate electrosurgery systems used for substantially equivalent indications.

.. Summary of Safety and Effectiveness

Based upon the design, materials, function, intended use, comparison with currently marketed devices and the non-ciinical testing performed by VNUS, it is concluded that the Closure System is substantially equivalent to the noted predicate devices in safety and effectiveness.

Tom M. Holdych

Vice President, Regulatory and Clinical Affairs VNUS Medical Technologies, Inc. December 1, 1997

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure, with three overlapping profiles facing to the right. The profiles are black against a white background.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 0 1998

Mr. Tom Holdych Vice President, Regulatory and Clinical Affairs VNUS Medical Technologies, Inc. 238 East Caribbean Drive Sunnyvale, California 94089

Re: K974521 VNUS™ Closure™ System Trade Name: Regulatory Class: II Product Code: GEI Dated: December 1, 1997 December 2, 1997 Received:

Dear Mr. Holdych:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act · include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with most with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Holdych

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your . . device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Ph.D., M.D. a M. Witten, Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Device Name: Closure™ System

510(k) Number (if known):

974521

Indications for Use:

The VNUS Closure™ System is indicated for use in the coagulation of blood vessels during general surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

CONCURRENCE OF CORH, OFFICE OF DEVICE EVALUATION (ODE)

Over the Counter Use: (Per 21 CFR 801.109)

1

or

Prescription Use:

coello

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.