K Number
K974325
Date Cleared
1997-12-08

(20 days)

Product Code
Regulation Number
866.5680
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Chiron Diagnostics ACS:180 Myoglobin Assay is used for the quantitative determination of Myoglobin in serum or plasma and as an aid in the diagnosis of acute myocardial infarction using the Chiron Diagnostics ACS Automated Chemiluminescence Systems.

Device Description

Myoglobin is a oxygen-binding, heme protein, found in cardiac and skeletal muscle. Myoglobin is noted for its rapid release into the circulation following tissue injury. Elevated levels of myoglobin can be found in conditions of muscle damage, such as acute and chronic skeletal muscle disease, renal failure, myocarditis, open-heart surgery, and after heavy exercise. Myoglobin releases into the circulation as early as 2 to 4 hours after cell damage, peaks at between 9 and 12 hours, and returns to normal within 24 to 36 hours. In the absence of skeletal muscle trauma, myoglobin has been used as an early indicator of myocardial infarction, and therefore as a rule out indicator. Myoglobin has a negative predictive value of 99%, which improves the rule out capabilities of the emergency department and helps reduce the number of patients inappropriately admitted to the Coronary Care Units with symptoms atypical of acute myocardial infarction. When used in combination with other cardiac markers such as CK-MB or cTnl, the ACS myoglobin assay is a valuable diagnostic tool to be used in the early evaluation of the potential acute myocardial infarction patient. The Chiron Diagnostics ACS:180 Myoglobin assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies. The first antibody, in the Lite Reagent, is a polyclonal goat anti-myoglobin antibody labeled with acridinium ester. The second antibody, in the Solid Phase, is a monoclonal mouse anti-myoglobin antibody covalently coupled to paramagnetic particles.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the KCN4325 device, based on the provided document:

Acceptance Criteria and Device Performance

CriteriaAcceptance Standard (Implicit from results)Reported Device Performance
Reference RangeEstablish 95% interval for apparently healthy individualsMale: 5.1–90.7 ng/mL (µg/L) (N=353) Female: 10–65.3 ng/mL (µg/L) (N=350)
Analytical Sensitivity (LOD)Detect myoglobin at low concentrations.≤10 ng/mL (µg/L)
Assay Reportable RangeMeasure myoglobin across a relevant concentration range.Up to 1000 ng/mL (µg/L)
Accuracy (Method Comparison)Strong correlation with an alternate established EIA method.Correlation coefficient (r) = 0.99 with alternate EIA method. Equation: ACS:180 Myoglobin = 1.06 x (alternate EIA method) - 4.6 ng/mL(µg/L)
PrecisionDemonstrate consistent and reproducible results.Total precision (% CV) range between 4.2 to 5.3.

Note: The document does not explicitly state "acceptance criteria" with predefined thresholds. Instead, it presents the results of performance studies that are implicitly accepted as demonstrating the device's suitability for its intended use, especially given the FDA's 510(k) clearance. The "Acceptance Standard" column is derived from the fact that the reported performance was deemed sufficient for regulatory approval.

Study Information

  1. Sample Size Used for the Test Set and Data Provenance:

    • Reference Range: 703 serum and plasma samples from "apparently healthy individuals."
    • Method Comparison: 123 samples in the range of 27 to 957 ng/mL (μg/L).
    • Precision: 8 samples.
    • Data Provenance: Not specified (e.g., country of origin). The studies appear to be prospective as they were conducted to establish performance characteristics for the device's submission.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • This information is not provided in the document. For an immunoassay like this, the "ground truth" for method comparison would typically be the results from the predicate or established method itself, which would have its own validation. For reference ranges, the "truth" is the biological measurement in a healthy population.
  3. Adjudication Method for the Test Set:

    • Not applicable or not specified. This type of adjudication is more common in image-based diagnostic systems or clinical trials requiring expert consensus on complex findings. For a quantitative assay, the comparison is against another assay or a statistically derived range.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is an automated immunoassay for Myoglobin detection, not an AI-assisted diagnostic tool that involves human readers interpreting images or complex data. Therefore, an MRMC study with human readers assisting AI is irrelevant.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • The studies described (Reference Range, Analytical Sensitivity, Reportable Range, Method Comparison, Precision) are all standalone performance evaluations of the ACS:180 Myoglobin assay as an algorithm/device only. The "ACS Automated Chemiluminescence Systems" implies an automated, standalone operation without direct human intervention in the result generation after sample loading.
  6. The type of ground truth used:

    • Reference Range: Biological measurements from a "healthy individual" population.
    • Analytical Sensitivity/Reportable Range: Defined by the assay's ability to statistically differentiate from a zero standard and measure within a calibrated range.
    • Method Comparison: Results from an "alternate myoglobin EIA method" (a predicate or established laboratory method).
    • Precision: Intrinsic variability of the assay itself, calculated from replicate measurements.
  7. The sample size for the training set:

    • Not specified. This document describes the performance studies for a diagnostic device prior to marketing. These assays are typically developed and calibrated by the manufacturer based on internal R&D, not a "training set" in the machine learning sense. The reported studies are more akin to validation or test sets.
  8. How the ground truth for the training set was established:

    • Not applicable/Not specified. As mentioned above, a "training set" in the machine learning context is not directly referred to or implied for this type of immunoassay. The development and establishment of the assay's calibration curve and parameters (which could be considered analogous to a 'training phase' in a broad sense) would be based on the scientific principles of immunoassay and extensive in-house research and development using various known concentration samples and standards.

{0}------------------------------------------------

KCN4325

Company Confidential

DEC = 8 1997

Summary of Safety and Effectiveness

As required by 21 CFR 807.92, the following 510(k) Summary is provided:

1. Submitters Information

Contact person:William J. PignatoDirector of Regulatory Affairs
Address:Chiron Diagnostics Corporation63 North StreetMedfield, MA 02052
Phone:Fax(508) 359-3825(508) 359-3885
Date Summary Prepared:November 14, 1997

2. Device Information

Proprietary Name:Chiron Diagnostics ACS:180 Myoglobin
Common Name:Myoglobin Immunological test system
Device Classification:Class II, 21 CFR 866.5680

3. Predicate Device Information

Name:Stratus Myoglobin Fluorometric Enzyme
Immunoassay
Manufacturer:Dade International Inc.

4. Device Description

Myoglobin is a oxygen-binding, heme protein, found in cardiac and skeletal muscle. Myoglobin is noted for its rapid release into the circulation following tissue injury. Elevated levels of myoglobin can be found in conditions of muscle damage, such as acute and chronic skeletal muscle disease, renal failure, myocarditis, open-heart surgery, and after heavy exercise.

Myoglobin releases into the circulation as early as 2 to 4 hours after cell damage, peaks at between 9 and 12 hours, and returns to normal within 24 to 36 hours. In the absence of skeletal muscle trauma, myoglobin has been used as an early indicator of myocardial infarction, and therefore as a rule out indicator. Myoglobin has a negative predictive value of 99%, which improves the rule out capabilities of the emergency department and helps reduce the number of patients inappropriately admitted to the Coronary Care Units with symptoms atypical of acute

{1}------------------------------------------------

myocardial infarction. When used in combination with other cardiac markers such as CK-MB or cTnl, the ACS myoglobin assay is a valuable diagnostic tool to be used in the early evaluation of the potential acute myocardial infarction patient.

5. Statement of Intended Use

The intended use of Chiron Diagnostics ACS:180 Myoglobin Assay is for the quantitative determination of Myoglobin in serum or plasma and as an aid in the diagnosis of acute myocardial infarction using the Chiron Diagnostics ACS Automated Chemiluminescence Systems.

6. Summary of Technological Characteristics

The Chiron Diagnostics ACS:180 Myoglobin assay is a two-site sandwich immunoassay using direct chemiluminometric technology, which uses constant amounts of two antibodies. The first antibody, in the Lite Reagent, is a polyclonal goat anti-myoglobin antibody labeled with acridinium ester. The second antibody, in the Solid Phase, is a monoclonal mouse anti-myoglobin antibody covalently coupled to paramagnetic particles.

6. Performance Characteristics

Expected Results

To confirm the ACS:180 Myoglobin reference range 703 serum and plasma samples from apparently healthy individuals were analyzed. Based on a central 95% interval, the following reference ranges were established:

SampleCategoryNAverage(ng/mL)(µg/L)Range(ng/mL)(µg/L)
Male35347.95.1–90.7
Female35032.710–65.3

As with all in vitro diagnostic asavs, each laboratory should determine its own reference range(s) for the diagnostic evaluation of patient results.

Sensitivity and Assay Reportable Range

The ACS:180 Myoglobin assay measures myoglobin concentrations up to 1000 ng/mL(ug/L) with a minimum detectable concentration (analytical sensitivity) of ≤10

{2}------------------------------------------------

ng/mL(ug/L). Analytical sensitivity is defined as the concentration of myoglobin that corresponds to the RLUs that are two standard deviations greater than the mean RLUs of 20 replicate determinations of the Myoglobin zero standard.

Method Comparison

For 123 samples in the range of 27 to 957 ng/mL(μg/L) the relationship between the ACS:180 Myoglobin assay and an alternate myoglobin EIA method is described by the equation:

ACS:180 Myoglobin = 1.06 x (alternate ElA method) - 4.6 ng/mL(µg/L) Correlation coefficient (r) = 0.99

Precision

Eight samples were assayed 6 times in 6 assays, on each of 4 systems (n =144 for each sample), over a period of 3 days. Total precision (% CV) range between 4.2 to 5.3.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle faces right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC - 8 1997

William J. Pignato Director of Regulatory Affairs Chiron Diagnostics 63 North Street 02052-1688 Medfield, Massachusetts

K974325 Re : Chiron Diagnostics ACS:180® Myoglobin Assay Product Code: DDR Requlatory Class: II Dated: November 14, 1997 Received: November 18, 1997

Dear Mr. Pignato:

requlations.

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or

{4}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity (CDIA-667, Chita device may require at it does, you should contact a the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA described in four in four alence of your device to a legally Finding or bubbeansaaries in a classification for your marketed produce acits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact Also, please note the Office of Compliance at (301) 594-4639. the regulation entitled, "Misbranding by reference to che regaration Cherchon" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a close-up shot of a dark, textured surface. The texture appears rough and uneven, with variations in light and shadow creating a sense of depth. The overall impression is one of a natural or weathered material, possibly stone or a similar substance. The image lacks any distinct objects or features, focusing solely on the surface texture.

Company Confidential

Page Lof

R 9743 85 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Chiron Diagnostics ACS:180 Myoglobin Assay

Indications for Use:

The Chiron Diagnostics ACS:180 Myoglobin Assay is used for the quantitative rric Onliner Blaghoodies in serum or plasma and as an aid in the diagnosis of acute myocardial infarction using the Chiron Diagnostics ACS Automated Chemiluminescence Systems.

(PLEASE DO NOT WRITE BELOW THIS LINE---CONTINUE ON ANOTHER PAGE, IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use(Per 21 CFR 801.109)
--------------------------------------------

OR

Over-The-Counter Use(Optional Format 1-2-96)
----------------------------------------------------
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number974328

§ 866.5680 Myoglobin immunological test system.

(a)
Identification. A myoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the myoglobin (an oxygen storage protein found in muscle) in serum and other body fluids. Measurement of myoglobin aids in the rapid diagnosis of heart or renal disease.(b)
Classification. Class II (performance standards).