K Number
K973282
Manufacturer
Date Cleared
1998-03-04

(183 days)

Product Code
Regulation Number
868.1720
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the Ceramatec® Handi oxygen analyzer is to monitor the oxygen concentration in the patient-breathing environment.

Device Description

The Ceramated Handi oxygen analyzer is comprised of a galvanic oxygen sensor and an analyzer module. The galvanic oxygen sensor produces a millivolt output that is proportional to the partial pressure of oxygen in the monitored gas. The analyzer module contains an electronic circuit that converts the millivolt output of the sensor to a digital percent-oxygen reading on a liquid-crystal display (LCD).

AI/ML Overview

Here's an analysis of the Ceramatec Handi Oxygen Analyzer based on the provided text, focusing on acceptance criteria and the study that proves its adherence:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Ceramatec Handi Oxygen Analyzer are implicitly established by demonstrating substantial equivalence to a legally marketed predicate device, the MSA Miniox® IA oxygen analyzer. The device performance is then compared against these criteria, which are the specifications of the predicate device.

Acceptance Criteria (Predicate Device Specification - MSA Miniox® IA)Reported Device Performance (Ceramatec® Handi)
Display Range: 0-100% oxygenDisplay Range: 0-99% oxygen
Display Resolution: 0.1% oxygenDisplay Resolution: 1% oxygen
Warm-up time: none requiredWarm-up time: none required
Operating Temperature Range: 0 - 40°COperating Temperature Range: 10 - 40°C
Operating Humidity Range: 0-95% RH non-condensingOperating Humidity Range: 0-95% RH non-condensing
Accuracy: ±3% full-scaleAccuracy: ±3% full-scale
Linearity: ±2% full-scaleLinearity: ±2% full-scale
Sensor type: galvanic fuel sensor 0-100% O2Sensor type: galvanic fuel sensor 0-100% O2
Sensor Operating Life: 12 months under normal operating conditionsSensor Operating Life: 12 months under normal operating conditions
Sensor Shelf Life: 6 monthsSensor Shelf Life: 6 months
Storage Temperature: -20 - 55 °CStorage Temperature: -15° - 50 °C
90% FS Response Time: <15 Seconds @ 25 °C90% FS Response Time: <15 Seconds @ 25 °C
Interference: ±2.3% full-scaleInterference: ±2% full-scale
Low-Battery Indicator: LO BAT appears on LCDLow-Battery Indicator: instantaneous shut-off
Power Requirement: 9 V alkaline batteryPower Requirement: lithium button-cell battery
Battery Life: 1400 hoursBattery Life: 1100 hours
Instrument Weight: 11 ouncesInstrument Weight: approximately 3 ounces

Note on "Acceptance Criteria" interpretation: In the context of a 510(k) submission, "acceptance criteria" for a new device are primarily defined by its substantial equivalence to a predicate device. The new device must meet or exceed the performance specifications of the predicate, or any differences must not raise new questions of safety or effectiveness. The table above reflects this comparison.

Study Details Proving Device Meets Acceptance Criteria

The document provided does not describe a traditional "study" in the sense of a clinical trial or a formal research study with a test set, experts, and adjudication methods. Instead, the "proof" that the device meets acceptance criteria is based on a demonstration of substantial equivalence to a predicate device (MSA Miniox® IA oxygen analyzer, K935370).

This is a regulatory pathway for medical devices in the United States, where a new device is deemed safe and effective if it is substantially equivalent to a device already legally marketed. The core of this demonstration involves a detailed comparison of the technological characteristics, intended use, and performance specifications between the new device and the predicate.

Based on the provided text, the following information is not applicable or not explicitly detailed as it would be in a typical clinical or performance study:

  1. Sample size used for the test set and the data provenance: Not applicable. The submission relies on a comparison of technical specifications, not a test set of patient data or samples.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a test set is not established in this type of submission.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This device is a measurement instrument, not an AI algorithm.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of a "test set." The ground truth for the device's accuracy and linearity would be established through laboratory calibration and testing against gases of known oxygen concentrations, as implied by the "calibration potentiometer" feature. The reported accuracy and linearity specifications are assumed to be derived from such internal testing by the manufacturer.
  7. The sample size for the training set: Not applicable. This is not an AI/machine learning device.
  8. How the ground truth for the training set was established: Not applicable.

Summary of "Study" and "Proof" for Substantial Equivalence:

The "study" in this context is the detailed technical comparison presented in section 15.0 of the submission, "Comparison of Technological Characteristics." This comparison systematically evaluates:

  • Intended Use: The device shares the same intended use as the predicate: "to monitor the oxygen concentration in the patient-breathing environment."
  • Technological Characteristics: A side-by-side comparison of numerous specifications (display range, resolution, operating temperature, accuracy, linearity, sensor type, battery life, weight, etc.) is provided between the Ceramatec Handi and the MSA Miniox® IA.
  • Performance Data: The manufacturer asserts that for critical performance metrics like accuracy, linearity, sensor operating life, shelf life, and response time, the Ceramatec Handi's performance is equivalent to or better than the predicate device. Minor differences (e.g., display range, resolution, battery type, weight) are discussed and deemed not to raise new safety or effectiveness concerns. For instance, the operating temperature range difference is justified by referring to ANSI standards, stating that the Handi's range is "well within the ANSI specification."

Conclusion: The device's compliance with acceptance criteria is established by demonstrating that its specifications and intended use are substantially equivalent to a legally marketed predicate device, implying that it is as safe and effective. The evidence provided is primarily a technical specification comparison, rather than data from a clinical or performance study with a dedicated "test set" or "training set" as might be seen for more complex or novel devices.

{0}------------------------------------------------

14713282

  • August 28, 1997 2.0 Submitter: Ceramatec, Inc. 2425 South 900 West Salt Lake City, Utah 84119 3.0 Contact Person: Gordon Roth Quality System Manager 4.0 Telephone: (801) 978-2117 Phone Direct: (801) 972-2455 Phone Business: (801) 972-1925 FAX 5.0 Proprietary Device Name: Ceramatec® Handi 6.0 Classification Name: Oxygen Gas Analyzer 7.0 Common Name: Oxygen Analyzer 8.0 Predicate Device: MSA Miniox® IA oxygen analyzer (K935370)

9.0 Device Function:

Date:

1.0

The device function of the Ceramatec Handi oxygen analyzer is to monitor oxygen concentration in the patient environment.

MAR - 4 1998

{1}------------------------------------------------

Method of Operation: 10.0

The Ceramated Handi oxygen analyzer is comprised of a galvanic oxygen sensor and an analyzer module. The galvanic oxygen sensor produces a millivolt output that is proportional to the partial pressure of oxygen in the monitored gas. The analyzer module contains an electronic circuit that converts the millivolt output of the sensor to a digital percent-oxygen reading on a liquid-crystal display (LCD).

The electronic circuit includes an "ON" button, a calibration potentiometer, and a two-digit LCD. The "ON" button is used to power-on the instrument. The LCD is blank when the Handi is in power-off mode. When the "ON" button is pressed, the unit begins to display the percent-oxygen concentration of the monitored gas. The unit is equipped with an auto-off circuit that automatically turns the device off approximately 1.3 minutes after power-on. The device also contains a low-battery detection circuit to safeguard against faulty readings due to a low-battery condition. When a low-battery condition is detected, the low-battery detection circuit prevents operation of the device by immediately powering-down the analyzer upon release of the "ON" button.

The calibration potentiometer allows the device to be calibrated against a gas of a known oxygen concentration, such as air (21% oxygen) or 99% oxygen. The two-digit LCD provides direct readout of percent-oxygen concentration. An over-range indication is included on the LCD to differentiate between 0% oxygen and 100% oxygen.

Please find the attached flow-diagram illustrating the operation of the Ceramatec® Handi oxygen analyzer.

The galvanic oxygen sensor is a lead-oxygen battery consisting of a lead anode, and an oxygen cathode. The oxygen cathode is made up of gold and an aqueous electrolyte solution. The gold electrode is in close proximity to a non-porous fluoropolymer membrane. Oxygen permeating through the membrane is reduced electrochemically at the gold electrode. An electronic network, consisting of resistors and / or thermistors for temperature compensation, is connected between the cathode and anode. This allows the lead-oxygen battery to continually discharge in the presence of oxygen. The voltage across the resistor / thermistor network is proportional to the oxygen partial pressure of the gas in contact with the fluoropolymer membrane. This voltage is converted by the analyzer to a digital output that is displayed on the LCD.

{2}------------------------------------------------

11.0 Materials of Construction:

Corrosion-Resistant Plastic:Housing
LCD:Display
Potentiometer:Calibration
Epoxy-Resin Circuit Board:Printed Circuit Board
Lithium Battery:Power Source
Galvanic Fuel-Cell Oxygen Sensor:Sensor

12.0 Product Specifications:

PRODUCTCERAMATEC® HANDI
Display Range0-99% oxygen
Display Resolution1% oxygen
Warm-up timenone required
Operating Temperature Range10 - 40°C
Operating Humidity Range0-95% RH non-condensing
Accuracy$\pm$ 3% full-scale
Linearity$\pm$ 2% full-scale
Sensor typegalvanic fuel sensor 0-100% O2
Sensor Operating Life12 months under normal operating conditions
Sensor Shelf Life6 months
Storage Temperature-15° - 50 °C
90% FS Response Time<15 Seconds @ 25 °C
Interference$\pm$ 2% full-scale
Low-Battery Indicatorinstantaneous shut-off
Power Requirementlithium button-cell battery
Battery Life1100 hours
Instrument Weightapproximately 3 ounces

13.0 Intended Use:

The intended use of the Ceramatec® Handi oxygen analyzer is to monitor the oxygen concentration in the patient-breathing environment.

Patient Population: 14.0

The patient population of the Ceramatec Handi oxygen analyzer consists of those patients who require the oxygen concentration in their environment to be monitored.

{3}------------------------------------------------

PRODUCTMSA MINIOX® IACERAMATEC® HANDI
Display Range0-100% oxygen0-99% oxygen
Display Resolution0.1% oxygen1% oxygen
Warm-up timenone requiredsame
Operating Temperature Range0 - 40°C10 - 40 °C5
Operating Humidity Range0-95% RH non-condensingsame
Accuracy±3% full-scale6same
Linearity±2% full-scalesame
Sensor typegalvanic fuel sensor 0-100% O2same
Sensor Operating Life12 months under normal operatingconditionssame
Sensor Shelf Life6 monthssame
Storage Temperature-20 - 55 °C-15° - 50 °C
90% FS Response Time<15 Seconds @ 25 °Csame
Interference±2.3% full-scale±2% full-scale
Low-Battery IndicatorLO BAT appears on LCDinstantaneous shut-off
Power Requirement9 V alkaline batterylithium button-cell battery
Battery Life1400 hours1100 hours
Instrument Weight11 ouncesapproximately 3 ounces

Comparison of Technological Characteristics: 15.0

16.0 Conclusion:

In conclusion, based on the information provided, it has been determined that the Ceramatec Handi oxygen analyzer is substantially equivalent to MSA Miniox IA oxygen analyzer is safe and effective for its intended use.

Since the patient breathing-environment is a controlled environment, oxygen analyzers are not typically exposed to temperatures between 0 - 10 °C. ANSI Z79.10 Section 3.2.3.1 specifies that an oxygen analyzer used in a patient breathing environment shall be capable of operating in a temperature range of 15-40 °C. The operating temperature range of the Handi is well within the ANSI specification. Therefore the operating temperature and relative humidity specifications of the Ceramated Handi oxygen analyzer and the predicate device are equivalent.

6 Calculated per temperature effects in MSA Operating Manual, page 5-3.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 4 1998

Mr. Gordon Roth Quality Systems Manager Ceramatec Inc. 2425 South 900 West Salt Lake City, UT 84119

Re: K973282 Handi Oxygen Analyzer Regulatory Class: II (two) Product Code: 73 CCL December 5, 1997 Dated: Received: December 8, 1997

Dear Mr. Roth:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is a man substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will Verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510K Number (if known) :

Device Name: Ceramatec Handi Oxygen Analyzer

Indication for Use:

Purpose: The purpose of the Ceramatec Handi oxygen analyzer is to monitor oxygen concentration in the patient breathing environment.

Function: The Ceramatec" Handi oxygen analyzer measures the oxygen partial pressure of a gas stream, and displays a digital reading of the oxygen concentration of the measured gas on a liquid-crystal display (LCD).

Target Patient Population: The target patient population consists of patients who require the oxygen concentration in their environment to be monitored.

Environment of Use: The Ceramatec Handi oxygen analyzer is used in patient breathing environments whose temperatures range from 10 to 40 ℃, and whose relative humidity ranged from 0 - 95% (non-condensing).

Device Claims: The Ceramatec Handi oxygen analyzer has the claim of monitoring percentoxygen concentration in the patient breathing environment.

Legally Marketed Predicate Device: The legally marketed predicate device is the MSA Miniox® IA oxygen analyzer. The predicate device was assigned 510(k) number K935370 and declared substantially equivalent by FDA.

Safety and Effectiveness: No differences in intended use or application of the Ceramated Handi or MSA Miniox IA have been identified that could affect product safety or effectiveness.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) NumberK973282
Prescription Use(Per 21 CFR 801.109)OROver-The-Counter Use
--------------------------------------------------------------------

(Optional Format 1-2-96)

§ 868.1720 Oxygen gas analyzer.

(a)
Identification. An oxygen gas analyzer is a device intended to measure the concentration of oxygen in respiratory gases by techniques such as mass spectrometry, polarography, thermal conductivity, or gas chromatography. This generic type of device also includes paramagnetic analyzers.(b)
Classification. Class II (performance standards).