K Number
K972950

Validate with FDA (Live)

Device Name
HIPALLAUR 19
Date Cleared
1997-09-05

(77 days)

Product Code
Regulation Number
872.3060
Panel
Dental
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Dental casting alloy used in combination with dental ceramics for fabrication of metallo-ceramic restoration.

Device Description

Gold reduced palladium based casting alloy

AI/ML Overview

The provided document is a 510(k) summary for a dental casting alloy, Hipallaur 19. It focuses on demonstrating substantial equivalence to a predicate device (Aurolite 2A) through a comparison of composition and physical/mechanical properties. This type of submission is not a clinical study report and therefore does not contain the information requested in your prompt regarding acceptance criteria, study design, expert ground truth, or human-AI performance.

The document discusses the technical characteristics of the device and its predicate, including:

  • Comparison of composition (weight %):

    • Legal (Aurolite 2A): Au (13.5), Ag (27.8), Pd (49.5), Sn (4.6), In (4.4), Fe (0), Mn (0), Ru (x < 1%)
    • New (Hipallaur 19): Au (19), Ag (10.4), Pd (57.1), Sn (10.3), In (2.3), Fe (0.3), Mn (0.5), Ru (0.1)
  • Comparison of physical and mechanical properties:

ALLOYPropertyAurolite 2AHipallaur 19
Melting point (solid.)1270°C1210°C
Melting point (liquid)1335°C1250°C
Hardness (Vickers 5/30)170 (soft)270 (hard)
370 (hard)
Yield strength (MPa)462 (soft)640 (hard)
670
Elongation (%)8 (soft)20 (hard)
Density (g/cm³)14.613.7
  • Test Methods Applied: ANSI/ADA 5 and ISO 9693 standards were used for testing the characteristics.

The report concludes that the new alloy (Hipallaur 19) is very similar to Aurolite 2A based on its constitution, with potentially slightly better corrosion resistance due to higher noble metal content.

In summary, the document does not contain the information you requested about acceptance criteria and a study proving device performance in a clinical or AI context because it is a regulatory submission for a dental casting alloy, not a clinical trial or AI performance study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for AUREX. The logo consists of two overlapping triangles, one slightly offset from the other, creating a three-dimensional effect. Below the triangles, the word "AUREX" is written in capital letters. The logo is black and white.

P.O. Box 509 Edenvale 1610 Republic of South Africa Telephone : +27 (11) 609-8640 Telefax : +27 (11) 452-3918

-14-

zasto

510 (k) Summary

Trade name: Common name: Classification name: Class ification number:

HIPALLAUR 19 Dental casting alloy Gold based alloys and precious metal alloys for clinical use EJT

Legally marketed device: Aurium Aurolite 2A Description of the device: Gold rdeuced palladium based casting alloy Intended use of the device: Type IV restoration with porcelain veneering

Summary of the technological characteristics

Test methods applied: as in ANSI/ADA 5 and ISO 9693

Comparison of composition:

ALLOYCOMPOSITION (weight%)
NameAuAgPdSnInFeMnRu
LegalAurolite 2A13.527.849.54.64.400x
NewHipallaur 191910.457.110.32.30.30.50.1

x is less than 1 %

Comparison of physical and mechanical properties

ALLOYMelting pointrange (°C)Hardness(Vickers5/30)Yieldstrength(MPa)Elongation(%)Density(g/cm3)
Namesolid.liquid. soft soft ahardsoft soft softhardsofthard11.7
LegalNewAurolite 2AHipallaur 191270121013351250170270370462640670 -82014.613.712.3

Discussion

Both alloys have the same constituents with the exception of two minor ones, i.e. Mn and Fe in Hipallaur 19. Nevertheless, these are oxidized out of the alloy during melting.

  1. Nevelliedess, aloos are onlines on ent (67.1 % vs. 63 %) and less silver and base metals.

Conclusion

AND A STATE A CALL A CARA A CARA

On the basis of the constitution the new alloy is very similar to Aurolite 2A, and it may be assumed that due to the higher noble metal content Hipallaur may have slightly better corrosion resistance.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract shapes that resemble human figures or stylized birds in flight. The text is arranged in a circular pattern around the symbol.

Food and Drug Administration -9200 Corporate Boulevard Rockville MD 20850

SEP - 5 1997

Mr. Jerome D. Davis Managing Director Aurex Precious Metal Industries. (PTY) LTD. . P.O. Box 509 ················ Edenvale 1610 Republic of South Africa

K972950 Re : Hipallaur 19 Trade Name: Requlatory Class: II EJT Product Code: Dated: June 13, 1997 Received: June 20, 1997

Dear Mr. Davis:

We have reviewed your Section 510.(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: "

{2}------------------------------------------------

Page 2 - Mr. Davis ... .

this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510 k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fya.gov/cdrh/dsmamain.html".

Sincerely Yours

Timothy A. Ulatowski

Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

Device Name: _________________________________________________________________________________________________________________________________________________________________ HIPALLAUR 19

Indications For Use:

Dental casting alloy used in combination with dental ceramics for fabrication of metallo-ceramic restoration.

Image /page/3/Picture/4 description: The image shows a black and white drawing of a sphere. A crescent shape is cut out of the left side of the sphere. The crescent shape is white, while the rest of the sphere is black.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Rivera

(Division Sign-Off) Division of Denial, Infection Control, and General Hospital Devices ાળ 510(k) Number _

Prescription Use_ Per 21 CFR 801.109) OR

Over-The-Counter Use

§ 872.3060 Noble metal alloy.

(a)
Identification. A noble metal alloy is a device composed primarily of noble metals, such as gold, palladium, platinum, or silver, that is intended for use in the fabrication of cast or porcelain-fused-to-metal crown and bridge restorations.(b)
Classification. Class II (special controls). The special control for these devices is FDA's “Class II Special Controls Guidance Document: Dental Noble Metal Alloys.” The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. See § 872.1(e) for availability of guidance information.