K Number
K972859
Device Name
FLUTTER D
Date Cleared
1998-06-05

(305 days)

Product Code
Regulation Number
868.5665
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The efficacy of the Flutter®D as a mucus clearance device for cystic fibrosis patients is based on its ability to 1) vibrate the airways (which loosens mucus from the airway walls), 2) intermittently increase endobronchial pressure (to maintain the patency of airways during exhalation, so that mucus does not become trapped as it moves up the airways), and 3) accelerate expiratory airflow (to facilitate the upward movement of mucus through the airways so that it can be more easily coughed out). Flutter D may also be useful in the removal of mucus from the lungs of patients who have chronic bronchitis or bronchiectasis and in conjunction with a medical need for Positive Expiratory Pressure (PEP) Therapy.

Device Description

The Flutter®D is a small hand held device consisting of a hardened plastic mouthpiece at one end, a plastic cover with an opening at the other end, and a valve on the inside created by a high density steel ball resting in a plastic circular cone.

The Flutter®D is an expiration resisting device that delivers internal vibrations to patients' airways. The moveable ball inside the Flutter®D opposes the patient's exhaled air, creating a resistance to his expiration due to the weight of the steel ball.

The vibrations cause an oscillation effect down in the smallest airway passages within the lungs. With repeated vibrations, mucus secretions are mobilized from deep within the lungs to the throat where they can be huffed up and expectorated.

AI/ML Overview

The provided text does not contain detailed acceptance criteria or a study proving the device meets specific performance metrics in the way that a typical AI/ML medical device submission would. Instead, it's a 510(k) summary for a "Flutter®D Mucus Clearance Device," which is a mechanical device rather than an AI/ML algorithm.

Therefore, many of the requested categories, such as sample size for test/training sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth types, are not applicable in this context. The submission relies on demonstrating substantial equivalence to a previously cleared device.

Here's a breakdown of the information that can be extracted or inferred:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (Implicit)Reported Device Performance
SafetyDevice does not introduce new safety concerns compared to predicate.Not explicitly stated, but assumed to be acceptable as device cleared.
EffectivenessDevice is as effective as the original Flutter® in clearing mucus and facilitating PEP therapy."Flutter®D will be as effective as the original Flutter® in dislodging mucus from airway walls and facilitating mucus clearance and in conjunction with a medical need for positive expiratory pressure (PEP) therapy." Due to minor design and material modifications that did not significantly affect oscillation patterns.
Technological CharacteristicsMinor modifications to design and material do not significantly affect oscillation patterns.Modifications in design and material were minor and did not significantly affect the oscillation patterns generated by a continuous air flow.

2. Sample size used for the test set and the data provenance

  • Not applicable for a clinical test set. The submission explicitly states: "Although no clinical trials have been performed on the Flutter®D disposable device..."
  • The "testing" mentioned is related to engineering analysis (e.g., oscillation patterns) of the device itself rather than clinical performance on patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. No clinical test set with ground truth established by medical experts was performed. The substantiation is based on engineering similarity to a predicate device.

4. Adjudication method for the test set

  • Not applicable. No clinical test set requiring adjudication was performed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable. This is a mechanical device, not an AI/ML algorithm, and no MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a mechanical device, not an AI/ML algorithm.

7. The type of ground truth used

  • Implied Ground Truth for Design Comparison: The "ground truth" used here is the performance characteristics (specifically oscillation patterns) of the original Flutter® device. The new device's engineering characteristics were compared to this established baseline.
  • The original Flutter®'s effectiveness was established through prior clinical evidence, which informally serves as the "ground truth" for the predicate device's performance.

8. The sample size for the training set

  • Not applicable. This is a mechanical device, not an AI/ML algorithm with a training set.

9. How the ground truth for the training set was established

  • Not applicable. No training set was used.

{0}------------------------------------------------

5 1998 JUN

SUMMARY OF SAFETY AND EFFECTIVENESS

August 1, 1997 Date of Preparation:

Device Name: Flutter®D Mucus Clearance Device

Powered Percussors Classification Name:

Manufacturer:

Atrion Medical Products Inc. (formerly Ryder International Corp.) for VarioRaw SA, a subsidiary of Scandipharm, Inc., Birmingham, Alabama.

Contact:

Patricia Bradstreet (Consultant to VarioRaw SA) Bradstreet Clinical Research Associates, Inc. 1588 Route 130 North, Suite C-2 North Brunswick, NJ 08902 (732) 821-0800

Predicate:

Flutter® Mucus Clearance Device which was cleared for market under K940986 and K946083.

Device Description:

The Flutter®D is a small hand held device consisting of a hardened plastic mouthpiece at one end, a plastic cover with an opening at the other end, and a valve on the inside created by a high density steel ball resting in a plastic circular cone.

Intended Use and Technological Characteristics:

The Flutter®D is to be used to assist in the clearance of excessive secretions from the lungs of patients with cystic fibrosis, bronchitis, bronchiectasis and other diseases which cause the lungs to produce excessive amounts of mucus. It may also be used in conjunction with a medical need for positive expiratory pressure (PEP) therapy.

The Flutter®D is an expiration resisting device that delivers internal vibrations to patients' airways. The moveable ball inside the Flutter®D opposes the patient's exhaled air, creating a resistance to his expiration due to the weight of the steel ball.

The vibrations cause an oscillation effect down in the smallest airway passages within the lungs. With repeated vibrations, mucus secretions are mobilized from deep within the lungs to the throat where they can be huffed up and expectorated.

NEC2 C:\MyFiles\fluttor\DISPE10K\summary.w

{1}------------------------------------------------

Clinical and Non-Clinical Similarities and Differences:

The original Flutter® has been shown to be effective in clearing mucus from patients with cystic fibrosis, bronchitis, bronchiectasis and other diseases in which mucus accumulates in the lungs and is effective in dislodging mucus from airway walls and thus facilitating its clearance by subsequent voluntary cough. It may also be useful in conjunction with a medical need for positive expiratory pressure (PEP) therapy.

Although no clinical trials have been performed on the Flutter®D disposable device, the modifications in design and material were minor and did not significantly affect the oscillation patterns generated by a continuous air flow; therefore, Flutter®D will be as effective as the original Flutter® in dislodging mucus from airway walls and facilitating mucus clearance and in conjunction with a medical need for positive expiratory pressure (PEP) therapy.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

ഉ 1698 JUN

Ms. Patricia Bradstreet Bradstreet Clinical Research Associates, Inc. 1588 Route 130 North, Suite C-2 North Brunswick, NJ 08902

Re: K972859 Flutter-D Mucus Clearance Device Requlatory Class: II (two) Product Code: 73 BYI March 16, 1998 Dated: Received: March 17, 1998

Dear Ms. Bradstreet:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic. (QS) . . inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Patricia Bradstreet

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

r, US

(k) Number (if known): .

Device Name: Flutter®D Mucus Clearance Device

Indications for Use:

The efficacy of the Flutter®D as a mucus clearance device for cystic fibrosis patients is based on its ability to 1) vibrate the airways (which loosens mucus from the airway walls), 2) intermittently increase endobronchial pressure (to maintain the patency of airways during exhalation, so that mucus does not become trapped as it moves up the airways), and 3) accelerate expiratory airflow (to facilitate the upward movement of mucus through the airways so that it can be more easily coughed out). Flutter D may also be useful in the removal of mucus from the lungs of patients who have chronic bronchitis or bronchiectasis and in conjunction with a medical need for Positive Expiratory Pressure (PEP) Therapy.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number K972854

Prescription Use (Per 21CFR 801.109)

OR

Over-the-Counter Use

(Optional Format 1-2-96)

NEC2 C:\MyFiles\Uurs:\DISPS10Kltnith:Race statorient wpd

§ 868.5665 Powered percussor.

(a)
Identification. A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.(b)
Classification. Class II (performance standards).