K Number
K972223

Validate with FDA (Live)

Date Cleared
1997-09-10

(89 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bone Plate with Cable's device is indicated for use where wire, cable, or band cerclage is used in combination with bone plates. The Bone Plate with Cable's device is intended for long bone fractures and is to be used with commercially available bone screws of the same general material type as the bone plate.

Device Description

The Bone Plate with Cable's device consists of a bone plate with transverse holes for passage of multifilament cables. A cable is passed through the transverse bone plate hole, and then through a crimp. The crimp is typically positioned adjacent to the plate. The cable is locked into a tensioning device and tightened which applies compression to the plate and bone. The crimp and cable are then crimped with a crimping tool to lock the cable in place. Excess cable is cut and removed. Additional cables of the same system type maybe applied as required.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Pioneer Laboratories Bone Plate with Cables device.

It's important to note that this 510(k) summary is from 1997, and the information presented is typical for the regulatory standards of that era. Modern medical device submissions often include more detailed performance studies and statistical analyses.

Acceptance Criteria and Study for Pioneer Laboratories Bone Plate with Cables Device

Based on the provided 510(k) Summary, the "Bone Plate with Cables" device underwent performance testing to demonstrate substantial equivalence to predicate devices, primarily focusing on mechanical strength and loading capabilities.

1. Table of Acceptance Criteria and Reported Device Performance

Criteria CategorySpecific CriteriaAcceptance Criteria (Implied)Reported Device Performance
Mechanical StrengthLoading Values (for Cable/Plate System)"Exceed the loading values for wire systems" and "equivalent to the values of the Pioneer Laboratories Bone Plate with Cerclage Cable's system."The Bone Plate with Cable's device was found to "exceed the loading values for wire systems" and its loading values were "equivalent to the values of the Pioneer Laboratories Bone Plate with Cerclage Cable's system."

Observation: The document does not explicitly state numerical acceptance criteria (e.g., "must withstand X Newtons of force"). Instead, the performance is reported relative to existing systems. The "acceptance criteria" are implied to be "at least as good as or better than" the predicate devices.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state the numerical sample size (e.g., number of plates or cables tested). It refers to "tests of each of the systems."
  • Data Provenance: The data is generated from performance testing conducted by the manufacturer, Pioneer Laboratories. The Country of Origin is the USA (where Pioneer Laboratories is based). The study is prospective in the sense that the testing was performed specifically to support the 510(k) submission for this new device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This section is not applicable to this type of device and study. The "ground truth" for a mechanical device like a bone plate with cables is typically established through engineering principles, material science, and mechanical testing standards. It isn't based on expert consensus on interpretations as would be the case for diagnostic imaging. The "truth" in this context is the physical properties measured in a lab.

4. Adjudication Method for the Test Set

This section is not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving discrepancies in human expert assessments (e.g., in clinical trials or image interpretation studies). For mechanical testing, if there were any discrepancies in test results, standard engineering practices for repeatability, reproducibility, and retesting would be employed, not an adjudication process by experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. MRMC studies are typically used to assess the performance of diagnostic imaging devices or software where human readers interpret medical images. This 510(k) is for a mechanical orthopedic implant, not a diagnostic device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This question is not applicable as the device is a mechanical implant, not an algorithm or AI system. Its performance is inherent in its physical and mechanical properties, not in an "algorithm only" mode.

7. The Type of Ground Truth Used

The "ground truth" for this device's performance study was based on mechanical testing results and comparison to established performance of predicate devices. The "truth" refers to the measured loading values and the device's ability to withstand forces, as determined by laboratory methods.

8. The Sample Size for the Training Set

This section is not applicable. The device is not an AI/ML algorithm requiring a "training set." Its design and performance are based on engineering principles and material properties, validated through mechanical testing.

9. How the Ground Truth for the Training Set Was Established

This section is not applicable for the reasons stated above (not an AI/ML device).

{0}------------------------------------------------

510(K) Summary ડ.

Pioneer Laboratories 510(K) Notification Summary SEP 1 0 1997 For Bone Plate with Cables

Administrative Information

Manufacturer Identification and Sponsor:

Pioneer Laboratories 375 River Park Circle Marquette, MI 49855-1781 Telephone: 906-226-9909 FAX: 906-226-9932

Official Contact:

Burns Severson Vice President, Regulatory Affairs/Quality Assurance

6/9/97 Date Prepared:

Device Identification

Proprietary Name: Bone Plate with Cables

Common Name Bone Plate

Classification Name and Reference:
Plate, Fixation, Bone:
Regulation Number:CFR 888.3030
Classification Number:87 HRS
Cerclage, Bone Fixation:
Regulation Number:CFR 888.3010
Classification Number:87 JDQ II

Devices on which substantial equivalence is claimed: Dall-Miles Cable Grip System, Broad Bone Plate and Pioneer Laboratories Bone Plate with Cerclage Cables.

ﻓﻲ ﺍﻟﻤﺴﺘﻘﻠﻴﻞ ﺍﻟﻤﺘ

1 of 2

{1}------------------------------------------------

දෙකු පොලි 2 of 2

Device Description

The Bone Plate with Cable's device consists of a bone plate with transverse holes for . passage of multifilament cables. A cable is passed through the transverse bone plate hole, and then through a crimp. The crimp is typically positioned adjacent to the plate. The . cable is locked into a tensioning device and tightened which applies compression to the plate and bone. The crimp and cable are then crimped with a crimping tool to lock the cable in place. Excess cable is cut and removed. Additional cables of the same system type maybe applied as required.

Intended Use

The Bone Plate with Cable's device is indicated for use where wire, cable, or band cerclage is used in combination with bone plates. The system is designed to provide increased compression as compared to only a bone plate in situations where there is inadequate bone stock, multiple fractures or butterfly fragments. The Bone Plate with Cable's device is intended for long bone fractures and is to be used with commercially available bone screws of the same general material type as the bone plate.

Technological Characteristic Compared to Predicate Device

All the devices use cables, crimps, and a bone plate as a system used for long bone fracture fixation. Each system uses a tensioner and crimp to apply compression, and lock the cable. The Dall-Miles Broad Bone Plate device places the rectangular crimp and cable over the bone plate. The Pioneer Laboratories Bone Plate with Cerclage Cables device has the crimp internal to the Bone Plate with Cable's device has the oval crimp adjacent to the plate with the cable passing through a hole in the plate.

Performance Data

The Bone Plate with Cable's device was predicated on the use of stainless steel, wire, Dall-Miles Broad Bone Plate, and the Pioneer Laboratories Bone Plate with Cerclage Cables. Based on tests of each of the systems the Bone Plate with Cable's device and the Pioneer Laboratories Bone Plate with Cerclage Cables exceed the loading values for wire systems. The loading values indicated that the Bone Plate with Cable's device is equivalent to the values of the Pioneer Laboratories Bone Plate with Cerclage Cable's system.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing line that forms the shape of a bird or wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 0 1997

Mr. Burns Severson 'Vice President, Regulatory Affairs and Ouality Assurance Pioneer Laboratories 375 River Park Circle Marquette, Michigan 49855

K972223 Re: Bone Plate with Cable's Device Trade Name: Regulatory Class: II Product Codes: HRS and JDQ Dated: August 19, 1997 Received: August 22, 1997

Dear Mr. Severson:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory

{3}------------------------------------------------

Page 2 - Mr. Burns Severson

In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

[signature]

Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Pioneer Laboratories

Bone Plate with Cables Device

Indications for Use

The Bone Plate with Cable's device is indicated for use where wire, cable, or band cerclage is used in combination with bone plates. The Bone Plate with Cable's device is intended for long bone fractures and is to be used with commercially available bone screws of the same general material type as the bone plate.

Prescription Use X
(Per 21 CFR 801.109)

signature

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.