K Number
K971812
Manufacturer
Date Cleared
1997-08-08

(84 days)

Product Code
Regulation Number
862.1345
Panel
CH
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Precision QID Blood Glucose Testing System is intended for in vitro diagnostic use (i.e., for external use only) for the quantitative measurement of glucose in fresh capillary whole blood. For home or professional use with the Precision QID Blood Glucose Sensor. Compatible with the MediSense 2 Card and Pen Blood Glucose Sensors, and the Companion 2 Card and Pen Blood Glucose Sensors.

The product may also be used by healthcare professionals for the quantitative measurement of glucose in venous, arterial, or neonate whole blood, provided the sample is used within 30 minutes of collection.

The Precision G Blood Glucose Testing System is intended for in vitro diagnostic use (i.e., for external use only) for the quantitative measurement of glucose in fresh whole capillary blood. The Precision G Blood Glucose Testing System is intended for home or professional use.

Device Description

The Precision QID and Precision G Blood Glucose Test Strips are identical in test strip design. Both test strips utilize amperometric biosensor technology to quantitatively measure glucose in whole blood and control solutions. The Precision QID Blood Glucose Test Strips are for use with the Precision QID Blood Glucose Testing System and are also compatible with the MediSense 2 Card and Pen Blood Glucose Testing Systems. The Precision G Blood Glucose Test Strip is only for use with the Precision G Blood Glucose Testing System.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Precision QID® and Precision G Blood Glucose Test Strips:

Overview

The provided document is a 510(k) Summary of Safety and Effectiveness for the Precision QID® and Precision G Blood Glucose Test Strip. It seeks substantial equivalence to previously cleared predicate devices. The primary focus of the performance studies is to demonstrate that the new device is "substantially equivalent" to existing methods for blood glucose measurement.

Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical "acceptance criteria" in the format typically seen with accuracy or precision targets (e.g., "95% of results within ±X% of reference"). Instead, the acceptance criterion is implicitly framed within the context of substantial equivalence to predicate devices. The reported device performance is qualitative:

Acceptance Criteria (Implicit)Reported Device Performance
Substantial equivalence to current methods for blood glucose measurement, including predicate devices: Precision QID Blood Glucose Test Strip (K945887, K962295), Precision G Blood Glucose Testing System (K963676), and Accu-Chek Advantage Test Strips (K951887, K954833). This implies that the new device should provide glucose results comparable in accuracy and reliability to these established devices across various user groups (healthcare professionals and lay users) and sample types (capillary, venous, arterial, neonate whole blood). The statement also notes comparable form, function, material composition, manufacturing process, and intended use."The studies demonstrated that user can obtain blood glucose results that are substantially equivalent to the current methods for blood glucose measurement including the predicate devices named above."

"Results of laboratory and clinical testing demonstrate that the performance of the Precision QID and Precision G Blood Glucose Test Strip when used according to the intended use stated above is acceptable and comparable to the performance of the predicate devices..." |

Study Details

  1. Sample Size used for the test set and the data provenance:

    • The document states that the performance was studied "in the laboratory and in clinical settings by healthcare professionals and lay users."
    • Specific sample sizes are NOT provided.
    • Data Provenance: Not explicitly stated, but clinical settings by "healthcare professionals and lay users" suggests data was collected from human subjects, likely in the US where the submission originated. It is a prospective study as it's evaluating the performance of the new device.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for this type of device. For blood glucose meters, the "ground truth" or reference method is typically an established laboratory-based glucose analyzer, not expert consensus on image interpretation. The document doesn't specify the reference method used, but it would almost certainly be a highly accurate lab instrument.
  3. Adjudication method for the test set:

    • Not applicable. Adjudication is typically used to resolve discrepancies among human readers or to establish ground truth in subjective assessments (e.g., image interpretation). This is a quantitative measurement device (blood glucose), where performance is compared directly to a reference method, not an adjudicated human reading.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a standalone diagnostic device for measuring blood glucose, not an AI-assisted interpretation tool for human readers. Therefore, an MRMC study is not relevant.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, in essence. The evaluation described is of the device itself (the test strip and its associated meter system) against a reference method. While human users (healthcare professionals and lay users) operate the device, the performance being assessed is the quantitative output of the device, which is a "standalone" measurement by the instrument based on its internal algorithm/chemistry. The study is assessing the device's ability to accurately measure glucose, independent of human interpretation of complex data.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The document implies that the ground truth was established by "current methods for blood glucose measurement," which almost certainly refers to a laboratory reference method (e.g., a glucose analyzer like a YSI analyzer), known for its high accuracy and precision.
  7. The sample size for the training set:

    • Not explicitly stated and not directly applicable in the AI/machine learning sense. This device is based on amperometric biosensor technology, which relies on chemical reactions and electrochemical detection, not a machine learning model that requires a "training set" in the computational sense. While the device's development and calibration would involve extensive testing and refinement with many samples, this isn't referred to as a "training set" in the context of this 510(k) summary.
  8. How the ground truth for the training set was established:

    • Not applicable for a "training set" in the AI/ML sense. For the development and calibration of such a biosensor, the "ground truth" for establishing the device's performance characteristics (e.g., sensitivity, linearity, accuracy) would be established using precisely measured glucose solutions and/or blood samples analyzed by highly accurate laboratory reference methods.

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.