(256 days)
The Silicone Antibacterial Foley Catheter is intended for short term use to provide continuous urinary bladder drainage in adult males and females requiring catheterization for surgical procedures, monitoring uring output, management of incontinence, and voiding dysfunction. The catheter has been shown to provide a statistically significant redution in the incidence of catheter acquired bacterial urinary tract infection during the first 5 days of catheterization. This device is not intended to be used as a treatment for active urinary tract infection.
The catherer consists of a standard configuration Foley catheter with a drainage lumen that exits through two drainage eyes on the proximal end. A retention balloon is located near the proximal catheter tip. The retention balloon is filled thru a second lumen with a Luer valve that extends from a side arm on the distal end of the catherer to the balloon. The catherer is coated with nitrofurazone impregnated siliconc on the outside surface and inner lumen. It is available in 12, 18 and 20 Fr sizes. All French sizes are provided in a 16 inch length with a 5 or 10 cc size balloon.
Here's a breakdown of the acceptance criteria and study information for the Rochester Medical Corporation Silicone Antibacterial Foley Catheter, based on the provided 510(k) summary:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Reduction in incidence of catheter-acquired bacterial urinary tract infection (CAUTI) | - Statistically significant reduction in CAUTI during the first 5 days of catheterization. (2.4% incidence vs. 6.9%incidence for conventional catheter group). - Statistically significant difference in actuarial survival curves for time until CAUTI between the two groups (p = 0.007, Kaplan-Meier survival, Breslow logrank statistic) for the first 5 days of use. |
| Material Biocompatibility | Passed Cytotoxicity, USP Systemic Toxicity, Subchronic Toxicity, Intracutaneous Injection, Muscle Implant Study, Sensitization Study, Pyrogenicity, and Hemolysis. |
| Mutagenicity | Some mutagenic changes indicated (Note: This is a reported finding, not explicitly stated as an "acceptance criterion" that was passed, but part of the testing results). |
| Antibacterial Activity (In-vitro) | Nitrofurazone-impregnated catheter segments showed antibacterial activity against clinical isolates of uro-pathogens including Coagulase-negative staphylococci, S. aureus, Enterococcus spp., C. Minutissimum, E. Coli, Enterobacter spp., and Klebstella spp. |
| Conformity to Standards | Meets the requirements of ASTM 623-89 Standard Specification for Foley Catheters. |
| Elution Profile & Bacteriocidal Levels | Elution profile reproducible, and antibacterial agent released at bacteriocidal levels. |
| Material Degradation with Aging | No evidence of degradation with aging. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: A total of 344 patients were enrolled in the clinical trial.
- Data Provenance: The study was a "single site randomized, double-blind, controlled clinical trial." This indicates it was a prospective study. The country of origin is not explicitly stated, but given the 510(k) submission to the FDA (USA), it is highly probable the study was conducted in the USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- The document does not explicitly state the number of experts or their qualifications used to establish the ground truth (diagnosis of CAUTI) for the clinical trial. Clinical trials usually rely on defined diagnostic criteria and sometimes independent clinical adjudication committees, but this specific detail is not provided.
4. Adjudication Method for the Test Set
- The document does not explicitly describe an adjudication method for the diagnosis of CAUTI in the clinical trial. It was a "double-blind" study, meaning neither the patients nor the clinicians administering the treatment knew which type of catheter was being used, which helps reduce bias in assessing outcomes.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, an MRMC comparative effectiveness study was not done. This study was a clinical trial comparing the device to a conventional catheter in human patients, not a study evaluating human readers' diagnostic performance with or without AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Yes, a standalone study was done, in the sense that the device itself (the catheter) was tested for its inherent antibacterial properties in an in-vitro capacity and its clinical effectiveness without direct human interaction. The "algorithm" in this context is the inherent antibacterial mechanism of the device.
- In-vitro studies specifically showed the antibacterial activity of the nitrofurazone-impregnated catheter segments against various uro-pathogens. This is a standalone performance assessment of the device's mechanism.
7. The Type of Ground Truth Used
- For the clinical trial, the ground truth was clinical outcomes data, specifically the "incidence of catheter associated urinary tract infection (CAUTI)." This would have been determined through clinical diagnosis based on established medical criteria (e.g., urine culture results, clinical symptoms).
8. The Sample Size for the Training Set
- This device is a physical medical device (catheter) with an antibacterial coating, not an AI or software-based device that explicitly requires a "training set" in the machine learning sense. Therefore, the concept of a "training set" is not applicable to this submission. The "training" for the device's performance is embedded in its manufacturing process and design, not in a data-driven training regimen.
9. How the Ground Truth for the Training Set was Established
- As explained above, the concept of a "training set" for an AI algorithm is not applicable to this physical medical device.
{0}------------------------------------------------
K97162 1-12
510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
JAN 13 1998
Richard D. Fryar Submitter and Contact Person Rochester Medical Corporation
Classification Name: Urological catherer Name of the Device Common/Usual Name: Foley catheter Proprietary Name: Rochester Medical Corporation Silicone Antibacterial Foley Catheter
Predicate Device
The predicate devices for purposes of substantial equivalence are the C.R. Bard Corp. Bardex® I.C. Foley Catheter (K910318), the American Pharmascal Co. Antimicrobial Foley Catheter (K871429) and the Rochester Medical All Silicone Foley Catheter (K896053).
Intended Use of the Device
The Silicone Antibacterial Foley Catheter is intended for short term use to provide continuous urinary bladder drainage in adult males and females requiring catheterization for surgical procedures, monitoring uring output, management of incontinence, and voiding dysfunction. The catheter has been shown to provide a statistically significant redution in the incidence of catheter acquired bacterial urinary tract infection during the first 5 days of catheterization. This device is not intended to be used as a treatment for active urinary tract infection.
Device Description
The catherer consists of a standard configuration Foley catheter with a drainage lumen that exits through two drainage eyes on the proximal end. A retention balloon is located near the proximal catheter tip. The retention balloon is filled thru a second lumen with a Luer valve that extends from a side arm on the distal end of the catherer to the balloon. The catherer is coated with nitrofurazone impregnated siliconc on the outside surface and inner lumen. It is available in 12, 18 and 20 Fr sizes. All French sizes are provided in a 16 inch length with a 5 or 10 cc size balloon.
Technological Characteristics
The catheter described in the 510(k) has similar technological and performance characteristics to the predicate devices. The catheter is manufactured entirely from silicone elastomer. The predicate devices are manufactured from silicone elastomer or latex. The device is supplied in French sizes from 12 to 20 with 5 or 10 cc capacity balloons. The predicate devices are available in French sizes ranging from 8 to 30 French with 3, 5 10 and 30 cc balloons. All of the devices are supplied sterile for single use. The C.R. Bard Corp. Bardex® I.C. Foley Catheter and American Pharmaseal Co. Antimicrobial Foley Catheter provide antibacterial activity through the use of a silver or silver oxide coating or impregnation of the catheter
Rochester Medical Corporation Silicone Antibacterial Foley Catheter Section 510(k) Notification Summary of Safety and EMcacy Puge 1 of 2 Junuary 8, 1998
Image /page/0/Picture/18 description: The image shows a logo with the text "ISO 9001" in the center. The text is surrounded by a globe-like design with dashed lines. Below the text, the word "CERTIFIED" is written in a bold, curved font. The logo appears to be a certification mark, possibly indicating compliance with ISO 9001 standards.
Tel: (507) 533-9600
{1}------------------------------------------------
K971627 2 - 2
materials. The Rochester Medical Silicone Antibacterial Foley Catheter provides antibacterial activity by coating the catheter with nitrofurazone.
Testing and Test Results
Test results indicated that the Silicone Antibacterial Foley Catheter meets the requirements of ASTM 623-89 Standard Specification for Foley Catheters. Simulated use testing shows that the elution profile of the antibacterial agent from the catheter is reproducible and the agent is released at bacteriocidal levels. There has been no evidence that the catheter materials or performance are degraded with ageing. Samples of Silicone Antibacterial Foley Catheter passed Cytotoxicity, USP Systemic Toxicity, Subchronic Toxicity, Intracutancous Injection, Muscle Implant Study, Sensitization Study. Pyrogenicity and Hemolysis. Mutagenicity testing indicated some mutagenic changes.
In-vitro studies
In vitro studies have shown that nitrofurazone impregnated catherer segments have antibacterial activity against clinical isolates of uro-pathogens including Coagulase-negative staphylococci, S. aureus, Enterococcus spp., C. Minutissimum, E. Coli, Enterobacter spp., and Klebstella spp. Gram-negative bacilli including Serratia spp., Proteus spp., and Pseudomonas spp. were not inhibited by nitrofurazone impregnated catheter segments. Nitrofurazone is not effective against yeast and fungi.
Clinical studies
A single site randomized, double-blind, controlled clinical trial was conducted to compare the antibacterial properies of the Silicone Antibacterial Catheter (nitrofurazone impregnated) to a Conventional Silicone Catheter. During the eleven month trial, a total of 344 patients (68% men, 32% women) mecting the study evaluation criteria were enrolled (mean age = 56 years, range 18 to 91) and the incidence of catheter associated urinary tract infection (CAUTI) during catheterization was recorded.
For patients with catheters inserted for 7 days or less (91.9% of population, median = 3 days), the group receiving the Silicone Antibacterial Catherer had an approximate threefold reduction in the rate of bacterial CAUTI compared to the Conventional Silicone Catheter group (2.4% vs. 6.9% incidence). The difference in actuarial survival curves for time until the occurence of bacterial CAUTI between the two groups was found to be statistically significant (p = 0.007, Kaplan-Mcicr survival, Breslow logrank statistic) for the first 5 days of use.
However, the trend in the hazard ratio over time to bacterial CAUTI between the antibacterial catheter and standard catherer was found to be statistically significantly increasing (P=0.03 Wald Chi Square Statistic) for the 7 days of use, resulting in the reversal of the hazard ratio in favor of the standard catheter from the 6th day.
Rochester Medical Corporation Silicone Antibacterial Foley Catheter Section 510(k) Notification Summary of Safety and Efficacy Page 2 of 2 January 8, 1998
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" surrounding the top half of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 3 1998
Re: K971627
Rochester Medical Silicone Antibacterial Foley Catheter Dated: December 5, 1997 Received: December 8, 1997 Regulatory class: II 21 CFR §876.5130/Product code: 78 MJC
Mr. Richard D. Fryar Vice President of Research and Development Rochester Medical Corporation One Rochester Medical Drive Stewartville, Minnesota 55976
Dear Mr. Fryar:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration. "
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Pederal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions; or _ other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, -permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (2) CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041-01-(301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely vours.
William Yu
Lillian Yiu, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
Page 1 of 1
510(k) Number (if Known):
Silicone Antibacterial Foley Catheter Device Name: _ Rochester Medical Corporation
Indications for Use:
INTENDED USE
The Silicone Antibacterial Foley catheter is intended for short term use to provide continuous urinary bladder drainage in adult males and females requiring catheterization for surgical procedures, monitoring urine output, management of incontinence, and voiding dysfunction. The catheter has been shown to provide a statistically significant reduction in the incidence of catheter acquired bacterial urinary tract infection during the first 5 days of catheterization. This device is not intended to be used as a treatment for active urinary tract infections.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert D. Sattler
(Division Sign Off)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K971627
OR
Prescription Use _ (Per 21 CFR 801.109)
Over-The-Counter-Use
(Optional Formal 1-2-96)
§ 876.5130 Urological catheter and accessories.
(a)
Identification. A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).(b)
Classification. (1) Class II (performance standards).(2) Class I for the ureteral stylet (guidewire), stylet for gastrourological catheter, ureteral catheter adapter, ureteral catheter connector, and ureteral catheter holder. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.