(244 days)
The intended use of the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision) is the acquisition, amplification, display, recording, and transmission of electrical signals of biological origin obtained during electrophysiology studies and related procedures. Such signals include ECG, intracardiac ECG, and blood pressure. Physiological parameters as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from these signals and displayed numerically, recorded, and/or transmitted.
Additionally, the EPXpress System (Timberwolf Revision) may acquire, transpose, amplify, display, record and transmit measurement data received from other medical devices typically used during these procedures, such as oximeters, RF generators, or electronic thermometers.
The EPXpress System (Timberwolf Revision) is an integrated catheter switching amplification and display system that can be installed at the patient table in the electrophysiology lab or moved to other locations where particular procedures may require its use. The system is made up of the EPAmp, Real-Time Controller, and monitor that are usually mounted on a cart. The EPAmp unit amplifies and conditions electrical signals from intracardiac catheters and displays these signals on a real-time monitor and/or prints these results to a chart recorder. The Timberwolf Revision added the capability of numerically displaying diastolic, systolic and mean blood pressures, as well as heart rate, cycle length, and certain measurement data received from other instruments.
Here's a summary of the acceptance criteria and study details for the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision), based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria (Equivalence Region) | Reported Device Performance |
|---|---|
| Mean difference between each pair of diastolic, systolic, and mean measurements < +/- 2 mmHg (at 0.05 confidence level) | The study results showed that the Timberwolf can be regarded as equivalent to Q-Cath at the 95% confidence level (i.e., 0.05 significance level) for phasic IVBP systolic, diastolic, and mean pressure measurements in the range of 0-300 mmHg, using a +/- 5% equivalence region. |
| The device must cover the specified range for the EPXpress System (0-300 mmHg systolic and mean, -30 to 300 mmHg diastolic) | Measurement values were selected from the range of available, pre-set values provided by a simulator, covering the specified range for the EPXpress System (0-300 mmHg systolic and mean, -30 to 300 mmHg diastolic) using dynamic IVBP waveforms. |
Note: The acceptance criteria were initially stated as "+/- 2 mmHg," but the conclusion references a "+/- 5% equivalence region." This suggests that the final assessment of equivalence was made using the 5% equivalence region, and the device met this criterion. If the intent was for both to be met, the documentation is slightly ambiguous, but it explicitly states the device met equivalence using the 5% region.
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated as a number of measurements. The study used "measurement values... selected from the range of available, pre-set values provided by a simulator" and applied various dynamic IVBP waveforms (left ventricular, aortic, left arterial, pulmonary capillary wedge, and pulmonary arterial waveforms). This suggests a variety of data points were generated but not a specific count.
- Data Provenance: The data was generated through a controlled, repeatable simulated input using a simulator. This is prospective data, as it was generated specifically for the testing. The country of origin is not explicitly stated for the simulated data, but the submitter is from Canada.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Number of Experts: Not applicable. The ground truth was established by a "simulator" which provided "pre-set values." This is a technical, engineered ground truth rather than expert-derived.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set:
- Adjudication Method: Not applicable. The ground truth was established by direct output from a simulator, eliminating the need for human adjudication of the ground truth. The comparison was a direct measurement comparison between the test device and the predicate device's measurements against the simulator's known values.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This study focused on the equivalence of a physiological monitoring system's blood pressure measurement capabilities, not an AI-assisted diagnostic or interpretative system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- Yes, this was effectively a standalone performance study. The "EPXpress System (Timberwolf Revision)" was compared directly against a predicate device using simulated inputs. The performance being evaluated was the device's ability to measure and display blood pressure, which is an algorithmic/hardware function without human interpretation in the loop for the core measurement task.
7. The Type of Ground Truth Used:
- Ground Truth Type: Simulator-generated, pre-set values. This can be categorized as a type of technical/device-generated ground truth or known input ground truth, where the simulator provided the reference standard.
8. The Sample Size for the Training Set:
- Not applicable. The EPXpress System (Timberwolf Revision) is a physiological monitoring system, not an AI/machine learning device that requires a training set. Its "capabilities" are built through hardware and software design rather than learned from data.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there was no training set.
{0}------------------------------------------------
K971570
DEC 29 1997
510(k) SUMMARY
| Submitter's Name: | Quinton Electrophysiology Corporation |
|---|---|
| Submitter's Address: | 200 West Beaver Creek RoadRichmond Hill, OntarioCanadaL4B 1B4 |
| Submitter's Phone Number: | (Canada) 905-709-1726 |
| Submitter's Fax Number: | (Canada) 905-709-1736 |
| Contact Person: | Dr. Harold Wodlinger |
| Date Summary Prepared: | July 16, 1997 |
A. Device Name and Classification
-
Device Trade Name: EPXpress System (Timberwolf Revision)
-
Device Common Name: Physiological Monitoring System
-
- Device Classification Names
- 870.2340 Electrocardiograph
- 870.2050 Biopotential Amplifier and Signal Conditioner
870.1425 Programmable Diagnostic Computer
- 870.1110 Blood Pressure Computer
- 870.2450 Medical Cathode Ray Tube Display
- 870.2810 Paper Chart Recorder
- B. Predicate Device
The legally marketed device to which we claim equivalence is the PE Electrophysiology Lab System and CardioLab Amplifier Module both manufactured by Prucka Engineering, Inc. . The 510(k) numbers for these devices are K902716 and K910307, respectively. Additionally, we claim equivalence to the Quinton Q-Cath System manufactured by Quinton Instrument Company. The 510(k) number for this device is K862740.
{1}------------------------------------------------
C. Device Description
The EPXpress System (Timberwolf Revision) is an integrated catheter switching amplification and display system that can be installed at the patient table in the electrophysiology lab or moved to other locations where particular procedures may require its use. The system is made up of the EPAmp, Real-Time Controller, and monitor that are usually mounted on a cart. The EPAmp unit amplifies and conditions electrical signals from intracardiac catheters and displays these signals on a real-time monitor and/or prints these results to a chart recorder. The Timberwolf Revision added the capability of numerically displaying diastolic, systolic and mean blood pressures, as well as heart rate, cycle length, and certain measurement data received from other instruments.
D. Intended Use of Device
The intended use of the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision) is the acquisition, amplification, display, recording, and transmission of electrical signals of biological origin obtained during physiological studies and related procedures. Such signals include ECG, intracardiac ECG, and blood pressure. Physiological parameters as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from these signals and displayed numerically, recorded, and/or transmitted. Additionally, the EPXpress System may acquire, transpose, amplify, display, record and transmit measurement data received from other medical devices typically used during these procedures, such as oximeters, RF generators, or electrical thermometers.
E. Summary of Technological Characteristics Compared with the Predicate Device
The EPXpress System (Timberwolf Revision) and the predicate devices described above have amplifiers which condition incoming physiological signals. The physiological signals acquired by the EPXpress and the predicate devices include ECG. intracardiac ECG. and blood pressure. Both the EPXpress and the predicate devices may display this physiological data on a monitor or the data may be printed out on chart recorder. All systems use a mobile cart for transport and have similar power and environmental requirements. The predicate devices have hardware and software which allow for the analysis of the physiological signals. The EPXpress System does not have this capability other than the ability to transform certain waveform data into numerial displays (Timberwolf Revision). (The analysis function is supplied by the EPLab System working in conjunction with the EPXpress System)
{2}------------------------------------------------
F. Performance Testing and Conclusions
1) Performance Testing
Performance testing was an equivalence study comparing blood pressure measurements taken and numerically displayed by the EPXpress System (Timberwolf Revision) with blood pressure measurements obtained and displayed in a similar fashion by a predicate device, the Quinton Q-Cath Monitoring and Analysis System. The study was a prospective paired comparison of measurements done using a controlled, repeatable simulated input. Measurement values were selected from the range of available, pre-set values provided by a simulator. The measurement values covered the specified range for the EPXpress System ( i.e. 0-300 mmHg systolic and mean, -30 to 300 mmHg diastolic) using dynamic IVBP waveforms - left ventricular, aortic, left arterial, pulmonary capillary wedge, and pulmonary arterial waveforms. This range of waveforms provided low, medium, and high values for systolic, diastolic, and mean pressures.
It was formulated that the EPXpress System (Timberwolf Revision) would be considered subtantially equivalent to its predicate device with respect to IVBP measurements if it could be shown at the 0.05 confidence level that the mean difference between each pair of diastolic, systolic, and mean measurements is less than +/- 2 mmHq.
2) Conclusions
The study results met the above criteria for substantial equivalence. Using a +/- 5% equivalence region, the study results showed that the Timberwolf can be regarded as equivalent to Q-Cath at the 95% confidence level (i.e., 0.05 significance level) for phasic IVBP systolic, diastolic, and mean pressure measurements in the range of 0-300 mmHq.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Quinton Electrophysioloqy Corporation c/o Mr. Matt Hedlund Regulatory Coordinator Quinton Instrument Company 3303 Monte Villa Parkway Bothell, Washington 98021-8906
DEC 2 9 1997
Re : K971570 The EPXpress System (Timberwolf Revision) Regulatory Class: II (two) Product Code: 74 DRR Dated: September 26, 1997 Received: October 2, 1997
Dear Mr. Hedlund:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) , it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" .
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K971570 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The intended use of the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision) is the acquisition, amplification, display, recording, and transmission of electrical signals of biological origin obtained during electrophysiology studies and related procedures. Such signals include ECG, intracardiac ECG, and blood pressure. Physiological parameters as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from these signals and displayed numerically, recorded, and/or transmitted.
Additionally, the EPXpress System (Timberwolf Revision) may acquire, transpose, amplify, display, record and transmit measurement data received from other medical devices typically used during these procedures, such as oximeters, RF generators, or electronic thermometers.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---|---|
| (Division Sign-Off) | |
| Division of Cardiovascular, Respiratory, | |
| and Neurological Devices | |
| 510(k) Number | K971570 |
| Prescription Use | OR | Over-The-Counter Use |
|---|---|---|
| ------------------ | ---- | ---------------------- |
(Optional Format 1-2-96)
000008
§ 870.2050 Biopotential amplifier and signal conditioner.
(a)
Identification. A biopotential amplifier and signal conditioner is a device used to amplify or condition an electrical signal of biologic origin.(b)
Classification. Class II (performance standards).