K Number
K971570
Device Name
EPXPRESS SYSTEM (TIMBERWOLF REVISION) /000391
Date Cleared
1997-12-29

(244 days)

Product Code
Regulation Number
870.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The intended use of the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision) is the acquisition, amplification, display, recording, and transmission of electrical signals of biological origin obtained during electrophysiology studies and related procedures. Such signals include ECG, intracardiac ECG, and blood pressure. Physiological parameters as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from these signals and displayed numerically, recorded, and/or transmitted. Additionally, the EPXpress System (Timberwolf Revision) may acquire, transpose, amplify, display, record and transmit measurement data received from other medical devices typically used during these procedures, such as oximeters, RF generators, or electronic thermometers.
Device Description
The EPXpress System (Timberwolf Revision) is an integrated catheter switching amplification and display system that can be installed at the patient table in the electrophysiology lab or moved to other locations where particular procedures may require its use. The system is made up of the EPAmp, Real-Time Controller, and monitor that are usually mounted on a cart. The EPAmp unit amplifies and conditions electrical signals from intracardiac catheters and displays these signals on a real-time monitor and/or prints these results to a chart recorder. The Timberwolf Revision added the capability of numerically displaying diastolic, systolic and mean blood pressures, as well as heart rate, cycle length, and certain measurement data received from other instruments.
More Information

Not Found

No
The description focuses on signal acquisition, amplification, display, recording, and transmission, along with derived physiological parameters. There is no mention of AI, ML, or related concepts like deep learning, image processing for analysis, or training/test sets for algorithmic learning. The performance study describes a comparison of measurements against a predicate device using simulated inputs, not an evaluation of an AI/ML model's performance.

No.
The device is used for acquisition, amplification, display, recording, and transmission of electrical signals of biological origin, and does not provide therapy.

Yes

Explanation: The device acquires, amplifies, displays, records, and transmits electrical signals of biological origin and physiological parameters (like blood pressure and heart rate) obtained during electrophysiology studies, which are used for diagnostic purposes to identify and characterize cardiac arrhythmias. The performance studies also focus on the accuracy of blood pressure measurements taken by the device.

No

The device description explicitly states the system is made up of hardware components: EPAmp, Real-Time Controller, and monitor. It also mentions the EPAmp unit amplifies and conditions electrical signals, which is a hardware function.

Based on the provided information, the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision) is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • Intended Use: The intended use clearly states the system is for the acquisition, amplification, display, recording, and transmission of electrical signals of biological origin obtained during electrophysiology studies and related procedures. This involves directly interacting with the patient's electrical activity and blood pressure in vivo.
  • Device Description: The description reinforces this by detailing how the system amplifies and conditions electrical signals from intracardiac catheters and displays these signals on a real-time monitor.
  • Lack of IVD Characteristics: IVD devices are designed to examine specimens derived from the human body (like blood, urine, tissue) in vitro (outside the body) to provide information for diagnostic purposes. The EPXpress System does not perform this function. It is a system for monitoring and analyzing physiological signals from within the body.

Therefore, the EPXpress System is a medical device used for physiological monitoring and analysis during electrophysiology procedures, not an IVD.

N/A

Intended Use / Indications for Use

The intended use of the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision) is the acquisition, amplification, display, recording, and transmission of electrical signals of biological origin obtained during physiological studies and related procedures. Such signals include ECG, intracardiac ECG, and blood pressure. Physiological parameters as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from these signals and displayed numerically, recorded, and/or transmitted. Additionally, the EPXpress System may acquire, transpose, amplify, display, record and transmit measurement data received from other medical devices typically used during these procedures, such as oximeters, RF generators, or electrical thermometers.

Product codes (comma separated list FDA assigned to the subject device)

74 DRR

Device Description

The EPXpress System (Timberwolf Revision) is an integrated catheter switching amplification and display system that can be installed at the patient table in the electrophysiology lab or moved to other locations where particular procedures may require its use. The system is made up of the EPAmp, Real-Time Controller, and monitor that are usually mounted on a cart. The EPAmp unit amplifies and conditions electrical signals from intracardiac catheters and displays these signals on a real-time monitor and/or prints these results to a chart recorder. The Timberwolf Revision added the capability of numerically displaying diastolic, systolic and mean blood pressures, as well as heart rate, cycle length, and certain measurement data received from other instruments.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing was an equivalence study comparing blood pressure measurements taken and numerically displayed by the EPXpress System (Timberwolf Revision) with blood pressure measurements obtained and displayed in a similar fashion by a predicate device, the Quinton Q-Cath Monitoring and Analysis System. The study was a prospective paired comparison of measurements done using a controlled, repeatable simulated input. Measurement values were selected from the range of available, pre-set values provided by a simulator. The measurement values covered the specified range for the EPXpress System (i.e. 0-300 mmHg systolic and mean, -30 to 300 mmHg diastolic) using dynamic IVBP waveforms - left ventricular, aortic, left arterial, pulmonary capillary wedge, and pulmonary arterial waveforms. This range of waveforms provided low, medium, and high values for systolic, diastolic, and mean pressures.

It was formulated that the EPXpress System (Timberwolf Revision) would be considered subtantially equivalent to its predicate device with respect to IVBP measurements if it could be shown at the 0.05 confidence level that the mean difference between each pair of diastolic, systolic, and mean measurements is less than +/- 2 mmHq.

The study results met the above criteria for substantial equivalence. Using a +/- 5% equivalence region, the study results showed that the Timberwolf can be regarded as equivalent to Q-Cath at the 95% confidence level (i.e., 0.05 significance level) for phasic IVBP systolic, diastolic, and mean pressure measurements in the range of 0-300 mmHq.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K902716, K910307, K862740

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.2050 Biopotential amplifier and signal conditioner.

(a)
Identification. A biopotential amplifier and signal conditioner is a device used to amplify or condition an electrical signal of biologic origin.(b)
Classification. Class II (performance standards).

0

K971570

DEC 29 1997

510(k) SUMMARY

Submitter's Name:Quinton Electrophysiology Corporation
Submitter's Address:200 West Beaver Creek Road
Richmond Hill, Ontario
Canada
L4B 1B4
Submitter's Phone Number:(Canada) 905-709-1726
Submitter's Fax Number:(Canada) 905-709-1736
Contact Person:Dr. Harold Wodlinger
Date Summary Prepared:July 16, 1997

A. Device Name and Classification

  1. Device Trade Name: EPXpress System (Timberwolf Revision)

  2. Device Common Name: Physiological Monitoring System

    1. Device Classification Names
  • 870.2340 Electrocardiograph
  • 870.2050 Biopotential Amplifier and Signal Conditioner

870.1425 Programmable Diagnostic Computer

  • 870.1110 Blood Pressure Computer
  • 870.2450 Medical Cathode Ray Tube Display
  • 870.2810 Paper Chart Recorder
  • B. Predicate Device

The legally marketed device to which we claim equivalence is the PE Electrophysiology Lab System and CardioLab Amplifier Module both manufactured by Prucka Engineering, Inc. . The 510(k) numbers for these devices are K902716 and K910307, respectively. Additionally, we claim equivalence to the Quinton Q-Cath System manufactured by Quinton Instrument Company. The 510(k) number for this device is K862740.

1

C. Device Description

The EPXpress System (Timberwolf Revision) is an integrated catheter switching amplification and display system that can be installed at the patient table in the electrophysiology lab or moved to other locations where particular procedures may require its use. The system is made up of the EPAmp, Real-Time Controller, and monitor that are usually mounted on a cart. The EPAmp unit amplifies and conditions electrical signals from intracardiac catheters and displays these signals on a real-time monitor and/or prints these results to a chart recorder. The Timberwolf Revision added the capability of numerically displaying diastolic, systolic and mean blood pressures, as well as heart rate, cycle length, and certain measurement data received from other instruments.

D. Intended Use of Device

The intended use of the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision) is the acquisition, amplification, display, recording, and transmission of electrical signals of biological origin obtained during physiological studies and related procedures. Such signals include ECG, intracardiac ECG, and blood pressure. Physiological parameters as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from these signals and displayed numerically, recorded, and/or transmitted. Additionally, the EPXpress System may acquire, transpose, amplify, display, record and transmit measurement data received from other medical devices typically used during these procedures, such as oximeters, RF generators, or electrical thermometers.

E. Summary of Technological Characteristics Compared with the Predicate Device

The EPXpress System (Timberwolf Revision) and the predicate devices described above have amplifiers which condition incoming physiological signals. The physiological signals acquired by the EPXpress and the predicate devices include ECG. intracardiac ECG. and blood pressure. Both the EPXpress and the predicate devices may display this physiological data on a monitor or the data may be printed out on chart recorder. All systems use a mobile cart for transport and have similar power and environmental requirements. The predicate devices have hardware and software which allow for the analysis of the physiological signals. The EPXpress System does not have this capability other than the ability to transform certain waveform data into numerial displays (Timberwolf Revision). (The analysis function is supplied by the EPLab System working in conjunction with the EPXpress System)

2

F. Performance Testing and Conclusions

1) Performance Testing

Performance testing was an equivalence study comparing blood pressure measurements taken and numerically displayed by the EPXpress System (Timberwolf Revision) with blood pressure measurements obtained and displayed in a similar fashion by a predicate device, the Quinton Q-Cath Monitoring and Analysis System. The study was a prospective paired comparison of measurements done using a controlled, repeatable simulated input. Measurement values were selected from the range of available, pre-set values provided by a simulator. The measurement values covered the specified range for the EPXpress System ( i.e. 0-300 mmHg systolic and mean, -30 to 300 mmHg diastolic) using dynamic IVBP waveforms - left ventricular, aortic, left arterial, pulmonary capillary wedge, and pulmonary arterial waveforms. This range of waveforms provided low, medium, and high values for systolic, diastolic, and mean pressures.

It was formulated that the EPXpress System (Timberwolf Revision) would be considered subtantially equivalent to its predicate device with respect to IVBP measurements if it could be shown at the 0.05 confidence level that the mean difference between each pair of diastolic, systolic, and mean measurements is less than +/- 2 mmHq.

2) Conclusions

The study results met the above criteria for substantial equivalence. Using a +/- 5% equivalence region, the study results showed that the Timberwolf can be regarded as equivalent to Q-Cath at the 95% confidence level (i.e., 0.05 significance level) for phasic IVBP systolic, diastolic, and mean pressure measurements in the range of 0-300 mmHq.

3

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Quinton Electrophysioloqy Corporation c/o Mr. Matt Hedlund Regulatory Coordinator Quinton Instrument Company 3303 Monte Villa Parkway Bothell, Washington 98021-8906

DEC 2 9 1997

Re : K971570 The EPXpress System (Timberwolf Revision) Regulatory Class: II (two) Product Code: 74 DRR Dated: September 26, 1997 Received: October 2, 1997

Dear Mr. Hedlund:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) , it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

4

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" .

Sincerely yours,

Thomas J. Callehan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

K971570 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

The intended use of the Quinton Electrophysiology Corporation's EPXpress System (Timberwolf Revision) is the acquisition, amplification, display, recording, and transmission of electrical signals of biological origin obtained during electrophysiology studies and related procedures. Such signals include ECG, intracardiac ECG, and blood pressure. Physiological parameters as diastolic, systolic, and mean blood pressure, heart rate, and cycle length may be derived from these signals and displayed numerically, recorded, and/or transmitted.

Additionally, the EPXpress System (Timberwolf Revision) may acquire, transpose, amplify, display, record and transmit measurement data received from other medical devices typically used during these procedures, such as oximeters, RF generators, or electronic thermometers.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) NumberK971570
Prescription UseOROver-The-Counter Use
--------------------------------------------

(Optional Format 1-2-96)

000008