K Number
K971128

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-04-28

(32 days)

Product Code
Regulation Number
892.1980
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Output:
The CAP-35BIII and the CINE 275 are intended to view 35 mm cineangiogram film.

Device Description

The CAP-35BII cineangiogram projector is a continuous film transport 16-facet rotating prism projectors. The CINE 275 is a simpler version of the CAP-35BIII, with fewer options. The CAP-35BIII consists of a 35 mm film projector with built-in viewing hood, a movable cart, a sheet film duplicator, a 35 mm slide camera, a remote control, and an optional TV camera assembly. The CINE 275 consists of a 35 mm film projector with built-in viewing hood and a movable cart.

AI/ML Overview

This document describes a 510(k) submission for two cineangiogram projectors, the CAP-35BIII and CINE 275. It focuses on demonstrating substantial equivalence to a predicate device, the Vanguard XR-350, rather than proving performance against specific acceptance criteria through a clinical study in the way modern AI/ML devices might.

Therefore, many of the requested categories are not applicable to this type of submission. However, I can extract the relevant information and indicate where information is not available or not required for this specific device type and regulatory pathway.

Here's a breakdown based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

For this type of device (cineangiogram projector), the "acceptance criteria" are primarily based on demonstrating substantial equivalence to a predicate device. This involves comparing technical characteristics and intended use. Performance is typically assessed by ensuring the device can perform its intended function (viewing 35mm cineangiogram film) safely and effectively, demonstrating equivalence in key operational and design parameters.

CharacteristicAcceptance Criteria (based on Predicate Device: Vanguard XR-350)CAP-35BIII Reported PerformanceCINE 275 Reported Performance
Intended UseViewing 35 mm angiographic filmViewing 35 mm angiographic film (Equivalent)Viewing 35 mm angiographic film (Equivalent)
Film Capacity35 mm X 500 ft.35 mm X 656 ft. (Equivalent, with larger capacity)35 mm X 500 ft. (Equivalent)
Image Size11.25 in. X 9.50 in.11 in. X 8.1 in (Comparable, within typical viewing requirements)10 in. X 8 in (Comparable, within typical viewing requirements)
Modes of OperationVariable speed (1, 4, 7, 10, 15, 30 fps), Single-frame advance, Forward/reverse, Rewind, Fast forwardVariable speed (3-100 fps), Single-frame advance, Forward/reverse, Rewind, Fast forward (Equivalent, with broader range)Variable speed (3-80 fps), Single-frame advance, Forward/reverse, Rewind, Fast forward (Equivalent, with broader range)
Light SourceXenon arc lampHalogen lamp (Different technology, but asserted as equivalent for performance)Halogen lamp (Different technology, but asserted as equivalent for performance)
Film TransportIntermittent claw advanceContinuous film advance, 16-facet rotating prism (Different technology, but asserted as equivalent for performance)Continuous film advance, 16-facet rotating prism (Different technology, but asserted as equivalent for performance)
Safety/EffectivenessDemonstrated by substantial equivalence to predicate device and compliance with applicable standards (Not explicitly detailed in this summary, but implied by 510(k) process).Demonstrated by substantial equivalence and compliance.Demonstrated by substantial equivalence and compliance.

Note: For a 510(k) for a device like this, "acceptance criteria" are not usually defined as quantitative performance metrics with specific targets (e.g., AUC > 0.9). Instead, the primary criterion is demonstrating that the new device is as safe and effective as a legally marketed predicate device, typically through a comparison of technological characteristics and intended use. Differences are acceptable if they do not raise new questions of safety or effectiveness.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not applicable. This submission is for a medical device (cineangiogram projector), not an AI/ML diagnostic algorithm. There is no "test set" of patient data in the context of evaluating a diagnostic algorithm's performance. The "test" is primarily the comparison of technical specifications and features to the predicate device.
  • Data Provenance: Not applicable. The "data" here refers to the device's design specifications and performance under engineering tests/benchmarking, not patient-derived data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not applicable. This is not a study requiring expert-established ground truth for a diagnostic outcome.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. No "test set" in this context.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

  • MRMC Study: No. This type of study is relevant for evaluating the impact of an AI/ML diagnostic tool on human reader performance, which is not applicable to a cineangiogram projector.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done

  • Standalone Study: Not applicable. This device is an instrument for viewing film, not an algorithm.

7. The Type of Ground Truth Used

  • Type of Ground Truth: The "ground truth" for this submission is based on the technical specifications and functional performance of the predicate device (Vanguard XR-350), and engineering verification of the new devices (CAP-35BIII and CINE 275) against their own design specifications and relevant safety standards. It does not involve patient outcomes, pathology, or expert consensus on a diagnostic finding.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • Ground Truth for Training Set: Not applicable.

Summary of Key Takeaways for this Specific Document:

The provided document is a 510(k) Summary for medical device clearance. The primary focus is on demonstrating substantial equivalence to a predicate device. This is achieved by comparing the new devices' (CAP-35BIII and CINE 275) technical characteristics and intended use with those of the predicate device (Vanguard XR-350). The "acceptance criteria" are implicitly met if the sponsor can convince the FDA that the new devices are as safe and effective as the predicate, despite some technological differences, and do not raise new questions of safety or effectiveness. Clinical trials or extensive performance studies with patient data, as required for AI/ML algorithms, are not typically part of this specific type of 510(k) submission for a non-AI hardware device like a projector.

{0}------------------------------------------------

APR 28 1997

K971128

Summary of Safety and Efficacy

510(k) Summary for the CAP-35BIII and CINE 275 Cineangiogram Projectors Prepared February 6, 1997

: .

This summary of 510(k) safety and efficacy

information is being submitted in compliance with the requirements of SMDA 1990 and CFR 807.92.

Submitter's Name

Jamieson Film Company 10425 Olympic Dr. Dallas, TX 75220-4427

Contact: Gary Fuller, President Phone: (214) 350-1283 Fax: (214) 357-2170

{1}------------------------------------------------

Device Name and Classification

Trade name:CAP-35BIII and CINE 275 Cineangiogram Projectors
Common name:35 mm Cineangiogram Projector
Classification:Angiographic X-Ray System, 21 CFR 892.1600

Predicate Device

The CAP-35BIII and CINE 275 projectors are substantially equivalent to the Vanguard XR-350 cineangiogram projector manufactured which was declared substantially equivalent on October 5, 1995.

Description of the CAP-35BIII and CINE 275 Cineangiogram Projectors

The CAP-35BII cineangiogram projector is a continuous film transport 16-facet rotating prism projectors. The CINE 275 is a simpler version of the CAP-35BIII, with fewer options. The CAP-35BIII consists of a 35 mm film projector with built-in viewing hood, a movable cart, a sheet film duplicator, a 35 mm slide camera, a remote control, and an optional TV camera assembly. The CINE 275 consists of a 35 mm film projector with built-in viewing hood and a movable cart.

Intended Use

The CAP-35BIII and the CINE 275 are intended to view 35 mm cineangiogram film.

{2}------------------------------------------------

CharacteristicCAP-35BIIICINE 275Vanguard XR-350
Intended useViewing 35 mm angiographic filmViewing 35 mm angiographic filmViewing 35 mm angiographic film
Major systemcomponentsProjectorCartRemote control35 mm slide film camera8" X 10" cameraProjectorProjectorViewing hoodPowered mobile cart
OptionsCCD TV cameraCartNTSC TV camera
Light sourceHalogen lampHalogen lampXenon arc lamp
Film transportContinuous film advance, 16-facet rotatingprismContinuous film advance, 16-facetrotating prismIntermittent claw advance
Projector dimensionsWidth 24 in.Depth 27 in.Height 24.4 in.Weight 132 lb.Width 20 in.Depth 21.3 in.Height 20.3 in.Weight 66 lb.Width 26 in.Depth 35.5 in.Height 29 in.Weight 147 lb.
Power120 V 60 Hz 5 A120 V 60 Hz 4.5 A115 V 60 Hz 10 A
Film capacity35 mm X 656 ft.35 mm X 500 ft.35 mm X 500 ft.
Projection optics57-92 mm lensf/4.5 varifocal50 mm lensf/4.565 mm lensf/3.0
Image size11 in. x 8.1 in10 in. x 8 in11.25 in. x 9.50 in.
Modes of operationVariable speed from 3-100 fpsSingle-frame advanceForward/reverseRewindFast forwardVariable speed from 3-80 fpsSingle-frame advanceForward/reverseRewindFast forwardVariable speed: 1, 4, 7, 10, 15, 30 fpsSingle-frame advanceForward/reverseRewindFast forward
FeaturesPush-button focusPush-button framingPush-button trimmingPush-button zoomLED frame counterSelectable auto-stopManual focusManual framingSelectable auto-stopPush-button focusManual framingLED frame counterAuto-stopImage rotation

Table of Comparisons between the CAP-35BIII, CINE 275, and Vanguard XR-350

N/A