K Number
K970947

Validate with FDA (Live)

Date Cleared
1997-05-14

(61 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To support the Diagnostic Source Assembly & the Image Receptor used in performing general purpose diagnostic fluoroscopic procedures.

Device Description

The B-350 Ceiling Mounted C-Positioner is a device intended to hold an X-Ray tube and an image intensifier 180° apart in a "C" configuration for the purpose of performing general purpose diagnostic fluoroscopic procedures.

The C-Positioner allows variable positioning of the X-Ray tube/Image Intensifier from 30 degrees Left Anterior Oblique (LAO) to 120 degrees Left Posterior Oblique (LPO). The device will maintain isocenter throughout its full range of motorized movement. The entire assembly is mounted to a set of ceiling rails with a detent provided to lock the unit into place for use in creating a common isocenter between the Anterior/Posterior (A/P) and lateral planes. A microswitch controlled interlock system independently prevents X-Ray exposure anywhere outside the detent position.

AI/ML Overview

This document is a 510(k) summary for a medical device called the B-350 Isocentric Ceiling Mounted C-Positioner. It describes the device's function and claims substantial equivalence to previously marketed devices. However, it does not contain any information regarding acceptance criteria, device performance studies, or clinical trial data.

Therefore, I cannot provide the requested information. The document focuses on establishing substantial equivalence based on technological characteristics and intended use, rather than presenting performance metrics against predefined acceptance criteria.

{0}------------------------------------------------

MAY | 4 1997

K970947

510(k) Summarv

March 25, 1997

Submitter: Omega Medical Imaging, Inc. 675 Hickman Circle, Sanford, FL 32771. Douglas Coon, President.

Device Name: B-350 Isocentric Ceiling Mounted C-Positioner.

Classification Name: System, Fluoroscopic, Image Intensified, All Types.

Common/Usual Name: C-Arm

Substantially Equivalent Devices: Fisher's PEGASUS Ceiling Mounted Cardiac Positioner, Philips DIAGNOST Lateral C-Arm, Fisher's ARIEL Ceiling Mounted C-Arm, and Fisher's CPS Ceiling Mounted Imaging System.

Description: The B-350 Ceiling Mounted C-Positioner is a device intended to hold an X-Ray tube and an image intensifier 180° apart in a "C" configuration for the purpose of performing general purpose diagnostic fluoroscopic procedures.

The C-Positioner allows variable positioning of the X-Ray tube/Image Intensifier from 30 degrees Left Anterior Oblique (LAO) to 120 degrees Left Posterior Oblique (LPO). The device will maintain isocenter throughout its full range of motorized movement. The entire assembly is mounted to a set of ceiling rails with a detent provided to lock the unit into place for use in creating a common isocenter between the Anterior/Posterior (A/P) and lateral planes. A microswitch controlled interlock system independently prevents X-Ray exposure anywhere outside the detent position.

Intended Use: To support the Diagnostic Source Assembly & the Image Receptor used in performing general purpose diagnostic fluoroscopic procedures.

Summary of Technological Characteristics: The B-350 Ceiling Mounted C-Positioner has the same technological characteristics as the predicate devices. The B-350 and the predicate devices all have the same intended use: to support the X-Ray tube (Diagnostic Source Assembly) and the Image Intensifier (Image Receptor) directly opposite each other with some variation in the distance from the X-Ray tube to the Image Receptor (SID), for diagnostic fluoroscopic procedures. The only differences in the devices are market driven variables such as the SID, arc rotation for different procedures, and the ability to rotate about the ceiling axis. The above devices are essentially all ceiling suspended C-Arms with some variations depending on user preference. The Omega B-350 Ceiling Mounted C-Positioner (C-Arm) is substantially equivalent to the predicate devices as well as any number of other Ceiling Mounted C-Arms that have been legally marketed in the U.S. for many years.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.