(210 days)
The KinAir IV and Therpulse II beds are for supporting bedridden patients in a manner that helps prevent and treat complications of immobility such as skin breakdown and decubitus ulcers. More particularly, it employs a patient supporting mattress overlay having an arrangement of transverse inflatable cushions that can be controlled to provide slowly pulsating pressures beneath the patient, as well as bladders to assist in turning the patient to one side or the other.
The KinAir IV, Therapulse II Air Bed bed is indicated to assist in the prevention and treatment of the complications of immobility, especially for patients who would benefit from a pressure relieving surface with the additional optional benefit of pneumatic pulsation.
The KinAir IV, Therapulse II (a.k.a. KT Air Bed) device is a pulsating low air-loss support mattress replacement system mounted on a commercially-available Stryker 3000 Acute Care Bed Frame, that also includes a Turn Assist feature. The mattress replacement system itself is essentially of the same general construction as the predicate Therapulse (510(k) No. K875321) and the KinAir (510(k) No. K880340 and the substantially equivalent TriaDyne device (510(k) No. K944094). It consists of the fabric support and the various cushions. The fabric support is a Gore-Tex fabric assembly for supporting and containing the twenty five upper cushions. Two additional inflatable cushions are on the underside of the fabric support, to assist the caregiver when turning the patient. All inflatable cushions are made of a vinyl Gore-Tex fabric to provide a low shear, low friction surface that is washable and durable.
The KinAir IV, Therapulse II Air Bed has two general modes of operation: 1) static, and 2) turn assist, with an optional third mode, pulsation, available only on the Therapulse II. The Display membrane panel mounted on the footboard and in each foot siderail, are used to select the static and pulsing modes. For the static mode, four individually adjustable sections of upper cushions are inflated to support the patient with minimal air pressures. For the pulsing mode, over an adjustable cycle of between 2 1/2 and 40 minutes, some cushions are inflated and deflated more or less than others in order to alternate the pressures supporting the patient. The cycle repeats continuously until the pulse mode is deactivated.
The Turn Assist mode is only available on the Home Display membrane panel mounted on the footboard. When the caregiver selects this mode to turn the patient either left or right, either the left or right bladder cushion will inflate and lean the patient approximately 15-20 degrees and stay inflated until the caregiver deactivates the turn assist mode. This feature was added to assist the caregiver when performing routine patient care, such as bathing, bedpan placement, etc.
The blower unit and Air Supply Unit are mounted on the Stryker bed frame, beneath the mattress. They are controlled to achieve the various modes of operation according to user input through switches on the membrane panels mounted on the footboard and on the outside of the foot siderails. All user input is accomplished through these membrane panels. The membrane panels with the various settings and displays are illustrated throughout the KinAir IV and Therapulse II Quick Reference Guides.
The Air Supply Unit's warmer is provided to warm air in the mattress for patient comfort. The blower is mechanically and electrically isolated to help prevent injury and reduce The warming element itself is a conventional resistive silicon heater. The warmer noise. actually warms the air in the Air Supply Unit's manifold, and the air then passively warms the patient to the degree selected by the patient's attendant. There are three settings: 1) LOW, 2) MEDIUM and 3) HIGH. At the HIGH setting, the heating element is capable of achieving maximum temperatures of roughly ten degrees Fahrenheit above ambient. Temperature feedback from sensors adjacent to the blower manifold allows for control of such temperatures. This is the same system that was used in the predicates Therapulse and KinAir.
When first inflated, the KinAir IV and the Therapulse II are automatically set in the static mode. Pulsation therapy, available on the Therapulse II, only, when activated, begins alternating pressure between sets of cushions. When pulsation is first activated, the cushion closest to the footboard decreases its pressure, together with every other cushion along the mattress length, while the other pressures are increased. Once the first set of cushions reaches its lowest point and starts to increase, the second set of cushions starts to decrease. The second set reaches its minimum just as the first set reaches its maximum. The cycle then repeats all over again, resulting in a pulsating effect. The minimum cycle time of 2 1/2 minutes is the it takes for a pulsating cushion to partially deflate and then inflate again.
This document is a 510(k) summary for the KinAir IV/Therapulse II (KT Air Bed) and primarily focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study of its own performance against set acceptance criteria. The document describes the device, its intended use, indications, contraindications, and differences from predicate devices. It does not contain information about specific acceptance criteria or a study designed to prove the device meets those criteria, as typically seen in a clinical trial or performance validation report.
The comparison provided is against existing predicate devices (KinAir, Therapulse, TriaDyne Plus) to establish substantial equivalence, which is a regulatory pathway for medical devices. This type of submission generally relies on demonstrating that a new device is as safe and effective as a legally marketed device, and therefore often focuses on technological characteristics and functional equivalence rather than new performance studies with pre-defined acceptance criteria.
Therefore, many of the requested elements cannot be extracted from this document, as they are not part of a typical 510(k) submission focused on substantial equivalence based on device design and existing predicate devices.
However, based on the provided text, here's what can be gathered, with an emphasis on what is not present:
1. Table of Acceptance Criteria and Reported Device Performance:
This information is not available in the provided text. The document does not describe specific numerical acceptance criteria (e.g., minimum pressure reduction percentage, specific decubitus ulcer prevention rates) or present a study with quantitative performance results against such criteria for the KinAir IV/Therapulse II. Instead, it compares the new device's features to those of predicate devices in a table (see page 4 of the document). This table is for "Comparative Information" of features, not performance criteria.
2. Sample Size Used for the Test Set and Data Provenance:
This information is not available. No test set (e.g., patient cohort for a clinical study) is described. The 510(k) submission is based on demonstrating substantial equivalence through design and functional comparison, not on a new clinical study with a test set.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:
This information is not available. No test set or ground truth establishment process is described.
4. Adjudication Method for the Test Set:
This information is not available. No test set or adjudication method is described.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study and Effect Size:
This information is not available. The document does not describe any MRMC study or any study involving human readers or AI assistance. The device is a medical bed, not an imaging or diagnostic AI system.
6. Standalone (Algorithm Only) Performance Study:
This information is not available. The device is a physical medical bed, not an algorithm.
7. Type of Ground Truth Used:
This information is not applicable/not available as no new performance study generating "ground truth" for the KinAir IV/Therapulse II is presented. The "ground truth" for the 510(k) pathway is essentially the established safety and effectiveness of the predicate devices that the new device is compared against.
8. Sample Size for the Training Set:
This information is not available. No training set (as in machine learning) is described.
9. How the Ground Truth for the Training Set Was Established:
This information is not applicable/not available.
Summary of what the document does provide regarding substantial equivalence:
The document focuses on comparing the KinAir IV/Therapulse II to predicate devices (KinAir, Therapulse, TriaDyne Plus) to establish substantial equivalence.
- Key Differences and Similarities: The primary difference highlighted is the use of a Stryker MPS 3000 frame for the platform, instead of the Hill-Rom frame used by the original KinAir and Therapulse. The low air-loss mattress replacement system itself is stated to be "essentially the same" as the predicates.
- Operational Modes: The new device offers static, turn assist, and an optional pulsation mode (on Therapulse II), similar to the predicate devices' functionalities.
- Intended Use & Indications: The intended use and indications (prevention and treatment of immobility complications like skin breakdown and decubitus ulcers) are consistent with the predicate devices.
- Comparative Table (Page 4): This table provides a detailed feature-by-feature comparison, including dimensions, maximum patient weight, patient surface material, various therapy features (low air-loss, percussion/vibration, rotation, turn assist, pulsation, warmed air), and electrical data. This table implicitly acts as the "proof" of substantial equivalence by showing that the new device's characteristics are either the same or functionally equivalent to the predicate devices. For example, while the original KinAir had "Rotation" with a 45-degree angle, the new KT Air Bed has "Turn Assist" (Yes) but "Rotation" (No), suggesting a functional alternative for patient repositioning. The maximum patient weight is higher for the KT Air Bed (700 lbs) compared to KinAir (300 lbs) and Therapulse (500 lbs), indicating an improvement in capacity.
In conclusion, this 510(k) summary is a regulatory submission demonstrating substantial equivalence through design and functional comparison to existing devices, not a study presenting primary performance data against pre-defined acceptance criteria.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for KCI. The logo consists of a series of curved lines on the left, followed by the letters "KCI" in a bold, sans-serif font. There is a registered trademark symbol to the right of the "I". A thin line runs horizontally beneath the letters.
510 (k) Summary
KinAir IV/Therapulse II (KT Air Bed) I. Name of Device: Bed, Flotation Therapy, Powered II. Classification Name: (per 21 CFR 890.5170) Therapulse, 510(k) No. K875321 III Substantial Equivalence: KinAir, 510(k) No. 880340 TriaDyne Plus, 510(k) No. 944094
IV. Device Description:
The KinAir IV, Therapulse II (a.k.a. KT Air Bed) device is a pulsating low air-loss support mattress replacement system mounted on a commercially-available Stryker 3000 Acute Care Bed Frame, that also includes a Turn Assist feature. The mattress replacement system itself is essentially of the same general construction as the predicate Therapulse (510(k) No. K875321) and the KinAir (510(k) No. K880340 and the substantially equivalent TriaDyne device (510(k) No. K944094). It consists of the fabric support and the various cushions. The fabric support is a Gore-Tex fabric assembly for supporting and containing the twenty five upper cushions. Two additional inflatable cushions are on the underside of the fabric support, to assist the caregiver when turning the patient. All inflatable cushions are made of a vinyl Gore-Tex fabric to provide a low shear, low friction surface that is washable and durable.
The KinAir IV, Therapulse II Air Bed has two general modes of operation: 1) static, and 2) turn assist, with an optional third mode, pulsation, available only on the Therapulse II. The Display membrane panel mounted on the footboard and in each foot siderail, are used to select the static and pulsing modes. For the static mode, four individually adjustable sections of upper cushions are inflated to support the patient with minimal air pressures. For the pulsing mode, over an adjustable cycle of between 2 1/2 and 40 minutes, some cushions are inflated and deflated more or less than others in order to alternate the pressures supporting the patient. The cycle repeats continuously until the pulse mode is deactivated.
The Turn Assist mode is only available on the Home Display membrane panel mounted on the footboard. When the caregiver selects this mode to turn the patient either left or right, either the left or right bladder cushion will inflate and lean the patient approximately 15-20 degrees and stay inflated until the caregiver deactivates the turn assist mode. This feature was added to assist the caregiver when performing routine patient care, such as bathing, bedpan placement, etc.
The blower unit and Air Supply Unit are mounted on the Stryker bed frame, beneath the mattress. They are controlled to achieve the various modes of operation according to user input
{1}------------------------------------------------
510(k) Summary Page 2
through switches on the membrane panels mounted on the footboard and on the outside of the foot siderails. All user input is accomplished through these membrane panels. The membrane panels with the various settings and displays are illustrated throughout the KinAir IV and Therapulse II Quick Reference Guides.
The Air Supply Unit's warmer is provided to warm air in the mattress for patient comfort. The blower is mechanically and electrically isolated to help prevent injury and reduce The warming element itself is a conventional resistive silicon heater. The warmer noise. actually warms the air in the Air Supply Unit's manifold, and the air then passively warms the patient to the degree selected by the patient's attendant. There are three settings: 1) LOW, 2) MEDIUM and 3) HIGH. At the HIGH setting, the heating element is capable of achieving maximum temperatures of roughly ten degrees Fahrenheit above ambient. Temperature feedback from sensors adjacent to the blower manifold allows for control of such temperatures. This is the same system that was used in the predicates Therapulse and KinAir.
When first inflated, the KinAir IV and the Therapulse II are automatically set in the static mode. Pulsation therapy, available on the Therapulse II, only, when activated, begins alternating pressure between sets of cushions. When pulsation is first activated, the cushion closest to the footboard decreases its pressure, together with every other cushion along the mattress length, while the other pressures are increased. Once the first set of cushions reaches its lowest point and starts to increase, the second set of cushions starts to decrease. The second set reaches its minimum just as the first set reaches its maximum. The cycle then repeats all over again, resulting in a pulsating effect. The minimum cycle time of 2 1/2 minutes is the it takes for a pulsating cushion to partially deflate and then inflate again.
INTENDED USE OF THE KINAIR IV, THERAPULSE II
The KinAir IV and Therpulse II beds are for supporting bedridden patients in a manner that helps prevent and treat complications of immobility such as skin breakdown and decubitus ulcers. More particularly, it employs a patient supporting mattress overlay having an arrangement of transverse inflatable cushions that can be controlled to provide slowly pulsating pressures beneath the patient, as well as bladders to assist in turning the patient to one side or the other.
Prepared by: J. Harbour Kinetic Concepts, Inc. P.O. Box 659508 San Antonio, TX 78265
{2}------------------------------------------------
510(k) Summary Page 3
INDICATIONS
The KinAir IV and Therapulse II beds are indicated to assist in the prevention and treatment of complications of immobility, especially for patients who would benefit from a pressure relieving surface with the additional optional benefit of pneumatic pulsation.
CONTRAINDICATIONS
Patient conditions for which the application of therapy on the KinAir IV and Therapulse II beds are contraindicated include the following:
- Unstable spinal cord injuries .
- Cervical traction .
DIFFERENCES BETWEEN CURRENT AND PREDICATE DEVICES
The KinAir IV and Therapulse II are modifications and replacements for the predicate devices KinAir and Therapulse beds. The major modification is the Stryker MPS 3000 frame used for the platform, replacing the Hill-Rom frame.
By using the Stryker MPS 3000 frame, the air components are mounted in the base of the frame, in one enclosure, as opposed to multiple enclosures in the litter and base of the Hill-Rom frame. Other features include:
- · retractable bed frame shortens the bed frame up to 14"
- Trendelenburg/reverse Trendelenburg and Cardiac Chair are controlled from the . Head Siderail and Footboard
- Foot end control panel is located at the top of the Footboard as opposed to below the . footboard
- Head siderails have patient controls located inside and Nurse controls on the outside .
- In-bed scale is built into the frame .
- Patient exit alarm system .
- · Nurse control panels are also located outside of foot siderails.
510(k) Summary
KinAir IV, Therapulse II 510(k) Summary
Prepared by: J. Harbour Kinetic Concepts, Inc. P.O. Box 659508 San Antonio, TX 78265 Date Prepared: June 4, 1997
{3}------------------------------------------------
Page 4
The low air loss mattress replacement system is essentially the same as the predicates KinAir and Therapulse devices. Design modifications were implemented to facilitate assembly of the bed, removal of the bed after use and cleaning procedures, yet retaining the same functions as the predicate devices. The KinAir IV mattress (blue colored), or the Therpulse II mattress (burgundy colored) can be used on the same modified Stryker MPS 3000 frame.
Date Prepared: June 4, 1997
{4}------------------------------------------------
Comparative Information
| KT AIR BED | KINAIR | THERAPULSE | TRIADYNEPLUS | |
|---|---|---|---|---|
| DIMENSIONS | ||||
| Patient SupportSurface &Frame | Stryker frame | Hill-Rom Frame | Hill-Rom Frame | Stryker frame |
| Length | 93" | 93" | 93" | 93" |
| Width | 43" | 38" | 38" | 43" |
| Height to Floor - High | 47" | 47" | 47" | 47" |
| Low | 35" | 36" | 36" | 35" |
| MAX.PATIENT WEIGHT | 700 lbs. | 300 lbs. | 500 lbs. | 500 lbs. |
| PATIENT SURFACEMATERIAL | Gore-Tex | Gore-Tex | Gore-Tex | Gore-Tex |
| Low Air-Loss TherapyCushions | Yes | Yes | Yes | Yes |
| PERCUSSION/VIBRATION | No | No | No | Yes |
| ROTATION | No | Yes | No | Yes |
| Max. Turning Angle | N/A | 45 | N/A | 40 |
| TURN ASSIST | Yes | No | No | No |
| PULSATION | Optional | Yes | Yes | Yes |
| WARMED AIR | Yes | Yes | Yes | Yes |
| MAXIMUMTRENDELENBURG | 12 | 12 | 12 | 12 |
| MAXIMUM REVERSETRENDELENBURG | 12 | 12 | 12 | 12 |
| BACKUP BATTERY | Yes / Inverter | Yes / Transport Air | Yes / Transport Air | Yes / Inverter |
| ELECTRICAL DATA | ||||
| Voltage | 115 v | 120 v | 120 v | 115 v |
| Frequency | 60 Hz | 60 Hz | 60 Hz | 60 Hz |
| Max. Leakage | 100 microamps | 90 microamps | 90 microamps | 100 microamps |
" Prepared by: J. Harbour Kinetic Concepts, Inc.
P.O. Box 659508 San Antonio, TX 78265 Date Prepared:
June 4, 1997
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 5 1997
Mr. William H. Quirk ·Director of Regulatory Affairs Kinetic Concepts, Inc. .............................................................................................................................................................................. 3440 East Houston Street - -San Antonio, Texas 78265-9508
K970468 Re: KT Air Bed Regulatory Class: II Product Code: FNM Dated: June 5, 1997 Received: June 9, 1997
Dear Mr. Quirk:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2 - Mr. William H. Quirk -----
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witteń, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K970468
Device Name:_KinAir IV, Therapulse II (a.k.a. KT Air Bed)
Indications for Use:
The KinAir IV, Therapulse II Air Bed bed is indicated to assist in the prevention and treatment of the complications of immobility, especially for patients who would benefit from a pressure relieving surface with the additional optional benefit of pneumatic pulsation.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---|---|
| Division Sign-Off) |
(Division Sign-On)
Division of General Restorative Devices,
510(k) Number K970468
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
§ 880.5550 Alternating pressure air flotation mattress.
(a)
Identification. An alternating pressure air flotation mattress is a device intended for medical purposes that consists of a mattress with multiple air cells that can be filled and emptied in an alternating pattern by an associated control unit to provide regular, frequent, and automatic changes in the distribution of body pressure. The device is used to prevent and treat decubitus ulcers (bed sores).(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 880.9.