(192 days)
Not Found
No
The device description focuses on the hardware components and physical mechanism of action (X-ray emission) for tumor treatment. There is no mention of software algorithms, image processing, or any terms related to AI/ML. The performance studies describe clinical trials evaluating the physical effects of the radiation, not the performance of a diagnostic or analytical algorithm.
Yes
The device is described as delivering a "prescribed therapeutic radiation dose" to "destroy the tumor from the inside out," directly indicating its therapeutic purpose.
No
The device description clearly states its purpose is for "highly focused treatment of intracranial tumors" by emitting X-rays to destroy them. This indicates a therapeutic function rather than a diagnostic one.
No
The device description explicitly states that the system includes hardware components such as an X-ray Source and Control Box, an interstitial probe, and accessories for clinical treatment and quality assurance. It is a complete hardware system for delivering radiation therapy.
Based on the provided information, the Photon Radiosurgery System is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is "for the irradiation of intracranial tumors." This describes a therapeutic treatment, not a diagnostic test performed on samples taken from the body.
- Device Description: The description details a system for delivering radiation directly to a tumor within the body. It involves an X-ray source, a probe inserted into the tumor, and accessories for positioning and quality control. This is consistent with a medical device used for treatment.
- Lack of IVD Characteristics: The description does not mention any components or processes related to analyzing biological samples (blood, urine, tissue, etc.) outside of the body, which is the defining characteristic of an in vitro diagnostic device.
In summary, the Photon Radiosurgery System is a therapeutic medical device used for treating tumors, not a diagnostic device used for testing samples.
N/A
Intended Use / Indications for Use
The Photon Radiosurgery System is intended to be used for the irradiation of intracranial tumors.
Product codes
90 JAD
Device Description
The Photon Radiosurgery System (PRS) is a complete system for highly focused treatment of intracranial tumors. The PRS includes an X-ray Source and Control Box, along with accessories for clinical treatment, quality assurance, laboratory calibration and dose verification. The miniature X-ray source incorporates a 3.175 mm diameter, 10 cm long interstitial probe and is designed to be mounted on a stereotactic frame so that the probe tip, covered with a sterile sheath, can be positioned precisely into the tumor. High dose rate, low energy X-rays are emitted from the probe tip in a spherical pattern and a prescribed therapeutic radiation dose delivered in a single fraction destroys the tumor from the inside out. Voltage, beam current, and treatment time or photon count are set on the Control Box which is powered by a rechargeable battery. Accessories are provided to assist in placement of the interstitial probe and to perform quality control of the X-ray source in the clinical setting. Additional laboratory-based components of the PRS include an automated dosimetry water tank for calibration and a CCD-camera based radiochromic film reader for dose verification.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
intracranial tumors
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Preclinical Tests: Several preclinical in vitro laboratory studies were conducted to characterize performance of the PRS, especially with regard to its output parameters and their effects. Preclinical tests also included animal studies in rats and dogs which characterized the effects of in vitro PRS irradiation of liver and brain tissue. In vitro and in vivo studies were also conducted to characterize the thermal properties of the PRS X-ray source. These studies demonstrated that the PRS was able to create highly demarcated areas of tissue destruction in targeted animal tissue without evidence of deleterious effects. Preclinical tests conducted on the PRS also included testing for compliance with electromagnetic compatibility and materials biocompatibility requirements.
Clinical Tests: Two phases of clinical studies to evaluate the safety and effectiveness of the Photon Radiosurgery System (PRS) for irradiation of intracranial tumors were conducted over a four year period at multiple clinical sites. Studies included treatment of eligible subjects diagnosed with primary or metastatic intracranial tumors which were either solitary or multiple in number. Assessment methods used to evaluate safety and effectiveness included clinical, radiological, functional and quality of life measures. Autopsy data was also obtained and provided to FDA as direct, objective evidence of the effects of PRS irradiation in comparison with the effects of other radiation therapy devices. A summary and analysis, as well as complete patient data, were provided to FDA as evidence of the safety and effectiveness of the PRS for treatment of intracranial tumors.
Key Metrics
Not Found
Predicate Device(s)
The Nucletron-Oldeft microSelectron-HDR interstitial brachytherapy device and the Elekta Radiosurgery Inc. Leksell Gamma Unit Model 23004.
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.5900 X-ray radiation therapy system.
(a)
Identification. An x-ray radiation therapy system is a device intended to produce and control x-rays used for radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II.
0
JUN 20 1997
510(k) Summary for Photoelectron Corporation PHOTON RADIOSURGERY SYSTEM
Date Prepared: 1.
December 9, 1996
- Submitter's Name and Address: 2.
Photoelectron Corporation 5 Forbes Road Lexington, MA 02173 USA
- Contact Person: Thomas R. Varricchione, MBA, RRT Director. Clinical Research & Regulatory Affairs Photoelectron Corporation
Telephone: (617) 861-2069 Facsimile: (617) 259-0482
Proprietary Name: Photon Radiosurgery System Device Name: র্ব Miniature X-ray Source with Control and Common Name: Calibration Systems and Accessories Classification Name: X-ray Radiation Therapy System
5. Predicate Device:
The Photon Radiosurgery System is substantially equivalent to the Nucletron-Oldeft microSelectron-HDR interstitial brachytherapy device and the Elekta Radiosurgery Inc. Leksell Gamma Unit Model 23004.
Device Description: 6.
The Photon Radiosurgery System (PRS) is a complete system for highly focused treatment of intracranial tumors. The PRS includes an X-ray Source and Control Box, along with accessories for clinical treatment, quality assurance, laboratory calibration and dose verification. The miniature X-ray source incorporates a 3.175 mm diameter. 10 cm long interstitial probe and is designed to be mounted on a stereotactic frame so that the probe tip, covered with a sterile sheath, can be
1
positioned precisely into the tumor. High dose rate, low energy X-rays are emitted from the probe tip in a spherical pattern and a prescribed therapeutic radiation dose delivered in a single fraction destroys the tumor from the inside out. Voltage, beam current, and treatment time or photon count are set on the Control Box which is powered by a rechargeable battery. Accessories are provided to assist in placement of the interstitial probe and to perform quality control of the X-ray source in the clinical setting. Additional laboratory-based components of the PRS include an automated dosimetry water tank for calibration and a CCD-camera based radiochromic film reader for dose verification.
Intended Use: 7.
The intended use of the Photon Radiosurgery System is to irracranial tumors.
Comparison of Technological Characteristics: 8.
Technological and functional characteristics of the PRS are similar to those of interstitial brachytherapy systems with radionuclides and to external beam stereotactic radiosurgery systems (LINAC and/or Gamma Knife). Comparison of the PRS with these devices included radiation sources, methods of application, relative dose rates, dose adjustment methods, collateral radiation protection, and the exposure of healthy This discussion illustrated the PRS's similarities with tissue to radiation. technological and functional features of the predicate devices.
9. Preclinical Tests:
Several preclinical in vitro laboratory studies were conducted to characterize performance of the PRS, especially with regard to its output parameters and their Preclinical tests also included animal studies in rats and dogs which effects. characterized the effects of in vitro PRS irradiation of liver and brain tissue. In vitro and in vivo studies were also conducted to characterize the thermal properties of the PRS X-ray source. These studies demonstrated that the PRS was able to create highly demarcated areas of tissue destruction in targeted animal tissue without evidence of deleterious effects. Preclinical tests conducted on the PRS also included testing for compliance with electromagnetic compatibility and materials biocompatibility requirements.
12/9/96
2
10. Clinical Tests:
Two phases of clinical studies to evaluate the safety and effectiveness of the Photon Radiosurgery System (PRS) for irradiation of intracranial tumors were conducted over a four year period at multiple clinical sites. Studies included treatment of eligible subjects diagnosed with primary or metastatic intracranial tumors which were either solitary or multiple in number. Assessment methods used to evaluate safety and effectiveness included clinical, radiological, functional and quality of life measures. Autopsy data was also obtained and provided to FDA as direct, objective evidence of the effects of PRS irradiation in comparison with the effects of other radiation therapy devices. A summary and analysis, as well as complete patient data. were provided to FDA as evidence of the safety and effectiveness of the PRS for treatment of intracranial tumors.
11. Conclusions:
Studies performed with the Photon Radiosurgery System (PRS) demonstrate that it emits high dose rate, low energy X-ray radiation in a spherical pattern from the tip of an X-ray Source probe which can be placed inside the body. Targeted turnors can be irradiated and destroyed in a precise and controllable manner. Data provided in this 510(k) Premarket Notification demonstrates that the Photon Radiosurgery System has the same intended use, has similar technological and functional features, employs similar construction methods and materials, and is as safe and effective as legally marketed predicate devices for the irradiation of intracranial tumors.
3
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is a stylized emblem featuring three abstract human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 20 1997
Thomas R. Varricchione, MBA, RRT Director, Clinical Research and Regulatory Affairs PeC Photoelectron Corporation 5 Forbes Road Lexington, MA 02173
Re: K964947
Photon Radiosurgery System (PRS) Dated: April 8, 1996 Received: April 9, 1997 Regulatory class: II 21 CFR 892.5900/Procode: 90 JAD
Dear Mr. Varrichione:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have decemined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivaliation assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations,
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591 for Radiology devices, Nose and Throat devices. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
William Yi
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
510(k) Number (if Known):_K964947
Device Name: Photon Radiosurgery System
Indications For Use:
The Photon Radiosurgery System is intended to be used for the irradiation of intracranial tumors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
OR
510(k) Number K964947
Prescription Use V (Per 21 CFR 801.109)
Photoelectron Corporation Photon Radiosurgery System 510(k)
Page vii
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________