K Number
K964506
Date Cleared
1997-05-21

(194 days)

Product Code
Regulation Number
862.2170
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The COBAS AMPLICOR Analyzer is an in vitro diagnostic device intended for use in clinical laboratories for the automation of the AMPLICOR™ Polymerase Chain Reaction test procedures.

Device Description

The COBAS AMPLICOR is a flexible, automated bench top batch analyzer that automates the amplification and detection steps of the Polymerase Chain Reaction (PCR) process. The COBAS AMPLICOR combines the operations of automated sample handling, reagents deliver, thermal cycling, controlled temperature incubation, photometric detection and result reporting into a single automated analyzer. The instrument consists of five major sub-components: (1) a thermal cycler module; (2) an automated pipeting station; (3) an incubation station; (4) a wash station; and (5) a photometer. An internal computer controls and monitors the major components includingsystem and run control, input/output, communication, results calculaton, and system diagnostic tests.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the COBAS AMPLICOR™ Analyzer, based on the provided document:

Acceptance Criteria and Device Performance

The document does not explicitly state formal "acceptance criteria" with numerical targets (e.g., minimum sensitivity or specificity values). Instead, the primary acceptance criteria focused on demonstrating substantial equivalence to an existing, legally marketed device, particularly the Roche AMPLICOR Chlamydia trachomatis Test performed with separate components. The performance characteristics were evaluated through a precision study and a clinical comparison study.

The core claim is that the COBAS AMPLICOR Analyzer, when used with the COBAS AMPLICOR Chlamydia trachomatis Test, performs equivalently to the manual AMPLICOR Chlamydia trachomatis Test using a Perkin-Elmer 9600 thermal cycler and a Bio-Tek EL800 plate reader.

Table of Acceptance Criteria (Implied) and Reported Device Performance:

Acceptance Criterion (Implied)Reported Device Performance
Precision/ReproducibilityCOBAS AMPLICOR Chlamydia trachomatis Test Precision Analysis (See Table 2 in source document):- Chlamydia trachomatis Spiked STM (IFU/PCR): - 0 IFU/PCR: Mean Absorbance 0.004, Total CV 117.6% (high due to very low mean value) - 1 IFU/PCR: Mean Absorbance 3.856, Total CV 3.7% - 10 IFU/PCR: Mean Absorbance 3.685, Total CV 5.9% - 50 IFU/PCR: Mean Absorbance 3.639, Total CV 5.4%- Kit Controls: - Negative: Mean Absorbance 0.004, Total CV 97.0% (high due to very low mean value) - Positive: Mean Absorbance 3.831, Total CV 7.0%Interpretation: Shows good reproducibility for positive samples across different operators and days, while very low absorbance negative samples show higher CV due to proximity to zero.
Clinical PerformanceComparative Clinical Performance vs. Predicate Device (AMPLICOR MWP) (See Table 3 in source document):- Cervical Specimens (266 total): - COBAS AMPLICOR Pos & AMPLICOR MWP Pos: 34 - COBAS AMPLICOR Pos & AMPLICOR MWP Neg: 1 - COBAS AMPLICOR Neg & AMPLICOR MWP Pos: 1 - COBAS AMPLICOR Neg & AMPLICOR MWP Neg: 230- Male Urethral Specimens (93 total): - COBAS AMPLICOR Pos & AMPLICOR MWP Pos: 16 - COBAS AMPLICOR Pos & AMPLICOR MWP Neg: 1 - COBAS AMPLICOR Neg & AMPLICOR MWP Pos: 0 - COBAS AMPLICOR Neg & AMPLICOR MWP Neg: 76- Male Urine Specimens (280 total): - COBAS AMPLICOR Pos & AMPLICOR MWP Pos: 63 - COBAS AMPLICOR Pos & AMPLICOR MWP Neg: 4 - COBAS AMPLICOR Neg & AMPLICOR MWP Pos: 5 - COBAS AMPLICOR Neg & AMPLICOR MWP Neg: 208Statistical analysis (McNemar's Test) found P > 0.4 for each specimen type, providing strong evidence of no significant difference in positive and negative results between the two test procedures.
Substantial EquivalenceThe document concludes that the COBAS AMPLICOR Analyzer (with the COBAS AMPLICOR Chlamydia trachomatis Test) provides substantially equivalent analytical and clinical performance to the predicate device. This is based on the precision data and the non-inferiority demonstrated in the clinical comparison.

Study Details

  1. Sample Size used for the Test Set and the Data Provenance:

    • Precision Study (Non-Clinical):
      • Sample Size: 27 replicates for each of the 6 sample types (0, 1, 10, 50 IFU/PCR spiked samples, negative kit control, positive kit control). Total of 162 tests.
      • Data Provenance: Not explicitly stated, but likely laboratory-generated and prospective (performed specifically for the study).
    • Clinical Performance Study:
      • Sample Size: 639 clinical specimens.
        • 266 endocervical swabs
        • 93 male urethral swabs
        • 280 male urine specimens
      • Data Provenance: Not explicitly stated, but implies clinical samples (from patients). It's specified as "The clinical data reported here were obtained using the AMPLICOR Chlamydia trachomatis Test" implying retrospective use of existing clinical data, or prospective collection of samples that were then tested by both methods. Given the direct comparison side-by-side, it's more likely a prospective collection but doesn't explicitly state "prospective."
  2. Number of Experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Precision Study: No external "ground truth" experts were used. Ground truth was defined by the spiked concentrations of C. trachomatis IFU/PCR in the samples and the kit controls.
    • Clinical Performance Study: The "ground truth" was established by the AMPLICOR microwell assay (predicate device), which served as the comparator. No specific number or qualifications of experts for reading "ground truth" are mentioned, as the comparison was between two assays. However, it's noted that "Two samples were culture positive/MOMP positive; three samples were culture negative/MOMP positive" in the footnotes of Table 3. This indicates that independent methods (culture, MOMP) were used as additional references for discordant results, but not as the primary ground truth for the overall comparison.
  3. Adjudication method for the test set:

    • No explicit adjudication method (like 2+1 or 3+1) is mentioned, as the studies did not involve subjective human reader interpretation of images or complex data needing consensus. The studies compared quantitative measurements (absorbance) and qualitative results (positive/negative) from an automated system with a predicate device.
    • For the discordant results in the clinical comparison (e.g., COBAS AMPLICOR Pos/AMPLICOR MWP Neg), the footnote suggests culture and MOMP positivity were used for reference, but it doesn't describe a formal adjudication process.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device (COBAS AMPLICOR Analyzer) is an automated in vitro diagnostic (IVD) instrument, not an AI-powered image analysis tool or decision support system intended to assist human readers. Its purpose is to perform laboratory tests automatically, replacing manual laboratory procedures. Therefore, the concept of "human readers improve with AI vs without AI assistance" does not apply to this submission.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the studies focused on standalone performance. The COBAS AMPLICOR Analyzer is itself a standalone automated system.
      • The precision study evaluated the intrinsic reproducibility of the automated system.
      • The clinical performance study evaluated the output (qualitative result) of the COBAS AMPLICOR system against the output of a manually-performed predicate test. Both are "algorithm only" in the sense that they produce a result based on a defined protocol, though the COBAS AMPLICOR automates more steps.
  6. The type of ground truth used:

    • Precision Study: Defined concentrations of C. trachomatis (IFU/PCR) spiked into transport medium, and certified kit controls.
    • Clinical Performance Study: The predicate device (AMPLICOR microwell assay) was used as the comparator, effectively serving as the "ground truth" for the comparison of substantial equivalence. For specific discordant cases, culture positivity and MOMP (Major Outer Membrane Protein) positivity were used as additional, higher-tier references.
  7. The sample size for the training set:

    • Not applicable/Not mentioned. The COBAS AMPLICOR Analyzer is an automated instrument performing a defined chemical/molecular assay, not a machine learning or AI model that requires a "training set" in the conventional sense of algorithm development. The development of the instrument itself would involve engineering validation, but not a data-based "training set" like an AI product.
  8. How the ground truth for the training set was established:

    • Not applicable, as there was no "training set" for an AI or machine learning model.

{0}------------------------------------------------

K964506 MAY
21 1997

Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Somerville, New Jersey 08876

510(k) Summary

COBAS AMPLICOR™ Analyzer

In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.

Identification of 510(k) Sponsor: I.

Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Somerville, New Jersey 08876-1760

Roche Molacular Syste

subsidiary of Hoffmann-La Roche

510(k) submission dated November, 1996

  • Contact: Alex Wesolowski Telephone: (908) 253-7540 Facsimile: (908) 253-7547

II. Device Name:

COBAS AMPLICOR™ Analyzer Trade/Proprietary Name -Automated batch analyzer for nucleic acid amplification Common or Usual Name and detection

None applicable Classification Name -

Identification of legally marketed device to which the 510(k) sponsor claims III. equivalence:

The primary operational components of the COBAS AMPLICOR analyzer are the pipettor, thermal cycler, wash stations, and the photometric detection system. As independent components, each is categorized as Class I devices under 21 CFR 862 (Table 1). The COBAS AMPLICOR combines these operational functions into a single automated instrument that provides an interpretative qualitative result. The performance of the COBAS AMPLICOR is substantially equivalent to several currently marketed device used independently to perform the same functions.

{1}------------------------------------------------

Predicate DeviceRegulatory ClassClassificationNumber .Predicate ProductNamePredicate510(k)Number
Automated PipettorClass 1862.2750Matrix TechnologiesImpact PipettorK940390
Thermal cycler (incubator)Class I exempt862.2750Perkin-Elmer 9600Thermal cyclerNA
IncubatorClass I exempt862.2050Fisher Dry HeatIncubator Model 630DNA
Automated MicroplateWasherClass 1 exempt866.2500Bio-Tek InstrumentsAutomated MicroplateWasher Model ModelELP40NA
Automated MicroplateReaderClass 1862.2300Bio-Tek InstrumentsAuomated EAIMicroplate ReaderModels Elx800/EL800K950104

TABLE 1 Predicate Devices for the COBAS AMPLICOR Analyzer

The COBAS AMPLICOR analyzer is designed for exclusive use with COBAS AMPLICOR The Roche AMPLICOR Chlamydia trachomatis Test (microwell plate) and the reagents. COBAS AMPLICOR Chlamydia trachomatis Test are both intended for the qualitative determination of Chlamydia trachomatis plasmid DNA in female endocervical, male urethral, and male urine specimens from symptomatic and asymptomatic patients. The Roche COBAS AMPLICOR Chlamydia trachomatis Test performed on the Roche COBAS AMPLICOR analyzer is substantially equivalent to the Roche AMPLICOR Chlamydia trachomatis Test (K922906/C, June 15,1993) performed on the Perkin-Elmer 9600 thermal cycler and the Bio-Tek EL800 (K950104) microtiter plate reader.

IV. Description of the Device:

The COBAS AMPLICOR is a flexible, automated bench top batch analyzer that automates the amplification and detection steps of the Polymerase Chain Reaction (PCR) process. The COBAS AMPLICOR combines the operations of automated sample handling, reagents deliver, thermal cycling, controlled temperature incubation, photometric detection and result reporting into a single automated analyzer. The instrument consists of five major sub-components: (1) a thermal cycler module; (2) an automated pipeting station; (3) an incubation station; (4) a wash station; and (5) a photometer. An internal computer controls and monitors the major components includingsystem and run control, input/output, communication, results calculaton, and system diagnostic tests.

Specimens are prepared off-line with AMPLICOR Specimen Preparation reagents before being transferred to amplification tubes (A-tubes) placed on the analyzer. Twelve A-tubes with caps are provided as a molded assembly called an A-ring. The COBAS AMPLICOR automatically performs test specific thermal cycling operations for PCR amplification. Amplified product is automatically transferred to polystyrene specimen cups (D cups) and primer coated magnetic beads are added. Following incubation, unbound material is removed by serial wash steps. Hybridized amplicon is measured by a enzymatically catalyzed color reaction. The intensity of the color is proportional to the amount of infectious organisms present in the specimen and qualitative results are determined based on a specified absorbance cut-off.

The amplification and detection operations can be run in parallel or sequentially depending on the number of samples in the run and user preference. In sequential mode, the analyzer automatically proceeds to sample detection after amplification, and the pipettor transfers amplified products from the A-ring in the thermal cycling segments to the D-cups for hybridization, wash and detection.

{2}------------------------------------------------

In parallel operation, a second set of samples is amplified as the first set is detected. After amplification of the first set of specimens is complete, the system pauses for transfer of the first set of amplified A-rings to the ambient temperature chambers. A second set of A-rings are then added to the thermal cycler for amplification, and the automated pipettor transfers amplified product from each of the first set of A-tubes in the ambient temperature segments to D-cups for hybridization, wash and detection.

Statement of Intended Use: V.

The COBAS AMPLICOR is an in vitro diagnostic device intended for use in the amplification and detection steps of the Polymerase Chain Reaction (PCR) process.

Summary of the technological characteristics of the new device in comparison to VI. those of the predicate:

The COBAS AMPLICOR Analyzer is a flexible, automated, bench top analyzer that automates the amplification and detection steps of the Polymerase Chain Reaction (PCR) process. The principles of operation of the COBAS AMPLICOR are substantially equivalent to manual invitro diagnostic PCR testing using general laboratory equipment. The COBAS AMPLICOR Analyzer consists of five major sub-components: (1) a thermal cycler module; (2) an automated pipeting station; (3) an incubation station; (4) a wash station; and (5) a photometer. An internal computer controls and monitors the major components includingsystem and run control, input/output, communication, results calculaton, and system diagnostic tests. The COBAS AMPLICOR Analyzer operating as a system performing automatically various steps such as pipetting, heating and measuring color intensity is substantially equivalent to several currently marketed devices used independently to perform the same functions. The COBAS AMPLICOR is designed to be used in conjunction with AMPLICOR Specimen Preparation Kits (for off-line specimen prepration), AMPLICOR Amplification Kits and COBAS AMPLICOR Detection Kits.

  • The AMPLICOR Test for Chlamydia trachomatis used with the Perkin-Elmer 9600 thermal cvcler and Bio-Tek plate reader, and the COBAS AMPLICOR Chlamydia trachomatis are both intended for the qualitative determination of Chlamydia trachomatis from urogenital swab specimens in specimen transport medium and male wine. The COBAS AMPLICOR Chiamydia trachomatis Test is substantially equivalent to the AMPLICOR Chlamydia trachomatis Test (K922906/C, June 15,1993) performed on the Perkin-Elmer 9600 thermal cycler and the Bio-Tek EEL800 (K950104) microtiter plate reader.

Similarities and Differences to Comparable Commercial Products

The specimen preparation, amplification and detection methods used in the COBAS AMPLICOR system are similar to that previously described for the microwell plate AMPLICOR Test. The notable similarities and differences are as follows:

Similarities

  1. The COBAS AMPLICOR Analyzer provides an automated method for performing the same basic procedures used in manual molecular diagnostic testing. These include, sample handling, reagent preparation, sample incubation, thermal cycling washing and photometric measurement

The procedures used specimen preparation, amplification and detection portions of the COBAS AMPLICOR procedure utilize identical reagent formulations as the AMPLICOR microplate assay. Identical reagent formulations include specimen diluent, master mix, free nucleotides, nucleotide primers and probes, TMB chromagen, peroxide substrate, and wash buffers.

  1. The COBAS AMPLICOR analyzer performs thermal cycling to tolerances that meet or exceed operational characteristics of conventional general laboratory equipment (e.g. the Perkin Elmer 9600).

{3}------------------------------------------------

  1. The COBAS AMPLICOR analyzer measures absorbance using a fixed wavelength photometer with comparable performance to other conventional general laboratory equipment and to automated clinical instrumentation such as the Roche COBAS MIRA.

Differences

  1. The COBAS AMPLICOR system automates all sample handling, washing and pipetting steps for test amplification and detection compared to manual requirements with the AMPLICOR microplate assay.

  2. PCR amplified products are captured by hybridization to probe-coated paramagnetic microparticles on the COBAS AMPLICOR compared to probe-coated microwell plates in the AMPLICOR test format.

  3. The COBAS AMPLICOR method does not require a stop reagent following the substrate incubation as does the AMPLICOR because the automated system precisely times each portion of the hybridization and detection procedures.

The optical density of the detection system is measured at 660 nm with the COBAS 4. AMPLICOR system compared to 450 nm in the AMPLICOR system. The stop reagent used in the AMPLICOR microplate assay modifies the oxidation state and thus, the absorbance maximum, of the chromophore.

VI. A brief discussion of the nonclinical and clinical performance data

Non-Clinical Performance

Precision

A multi-operator study was performed to determine the reproducibility of the COBAS AMPLICOR Chlamydia trachomatis Test. The study design was based upon the study suggested in the NCCLS document EP5-T2. Three independent operators tested a series of samples once a day for three days. A test run consisted of the amplification and detection of the following samples in triplicate: COBAS AMPLICOR Chlamydia trachomatis Test CT (+) and CT (-) kit controls and specimens that were prepared by spiking AMPLICOR Specimen Transport Medium with infected McCoy cell suspensions to obtain levels of 0, 1, 10 and 50 C. trachomatis IFUIPCR. Calculations were performed using the Analysis of Variance model suggested in the NCCLS document to derive estimates of between-day, between-operator, within-operator, and total variance for the Test at each sample concentration. The results of this study are presented in Table 2.

{4}------------------------------------------------

SAMPLE
Chlamydia trachomatis Spiked STM (IFU/PCR)Kit Controls
011050NegativePositive
Total Replicates272727272727
Mean Absorbance0.0043.8563.6853.6390.0043.831
Minimum0.0003.6463.2203.2200.0003.279
Maximum0.0174.0004.0004.0000.0114.000
Between Day Variance0.00000.00000.00400.00370.00000.0022
Standard Deviation0.00130.00000.06320.06100.00000.0468
CV (%)32.00.01.71.70.01.2
Between Operator Variance0.00000.01450.02710.02300.00000.0587
Standard Deviation0.00290.12060.16470.15170.00270.2422
CV (%)72.43.14.54.269.76.3
Within Operator Variance0.00000.00590.01620.01250.00000.0109
Standard Deviation0.00350.07670.12740.11180.00260.1045
CV (%)87.02.03.53.167.42.7
Total Variance0.00000.02040.04740.03920.00000.0718
Standard Deviation0.00470.14290.21760.19800.00370.2679
CV (%)117.63.75.95.497.07.0

TABLE 2 COBAS AMPLICOR Chlamydia trachomatis Test Precision Analysis of Variance Test Results

Clinical Performance

The clinical data reported here were obtained using the AMPLICOR Chlamydia trachomatis Test. The AMPLICOR microwell assay and the COBAS AMPLICOR Chlamydia trachomatis Test procedures are highly conserved and differ only in the method of detection. The COBAS AMPLICOR utilizes a magnetic microparticle based solid support compared to microwell plates and the automated analyzer enables the elimination of a endpoint hydrolysis Stop reagent that changes the absorbance maximum of the chromophore. The COBAS AMPLICOR and AMPLICOR Tests use the exact same specimen preparation, detection and amplification reagents, including identical amplification primer (CP24 and CP27) and detection probe (CP35) sequences. Therefore, the equivalent clinical performance is expected from the microwell and automated COBAS Tests.

Clinical Results - COBAS AMPLICOR

To ensure that equivalent clinical performance of the COBAS AMPLICOR Chiamydia trachomatis Test and the AMPLICOR Chlamydia Trachomatis Test are obtained, a comparison of the two test procedures was performed in which 639 clinical specimens were tested. The specimens included 266 endocervical swabs, 93 male urethral swabs, and 280 male urine specimens. Table 3 shows the comparative clinical performance obtained when these specimens were tested by the two test procedures. The results for each specimen type were evaluated for statistical difference in comparative clinical performance using McNemar's Test. For each of the three specimen types tested, the McNemar's Test found a value of P > 0.4. This finding provides strong evidence that there are no differences in the positive and negative results obtained when testing these specimens by either the COBAS AMPLICOR Chlamydia Trachomatis Test or the AMPLICOR Chlamydia Trachomatis Test.

{5}------------------------------------------------

Table 3 Comparative Clinical Performance of the COBAS AMPLICOR Chlamydia Trachomatis Test versus the AMPLICOR Chlamydia Trachomatis Test

CervicalMale UrethralMale Urine
PosNegPosNegPosNeg
COBASAMPLICOR
Pos341*161+634+
Neg1"2300765#208

キキ

AMPLICOR MWP

Two samples were culture positive/MOMP positive; three

samples were culture negative/MOMP positive

VIII. Conclusions

The COBAS AMPLICOR is an automated bench-top analyzer for the in vitro diagnostic analysis of clinical laboratory specimens using PCR based nucleotide amplification methods. The instrument gives substantially equivalent analytical performance to other general laboratory equipment for thermal cycling and photometric detection of colormetric clinical assays. Based on the clinical performance data, substantially equivalent results were obtained for the qualitative COBAS AMPLICOR Chlamydia trachomatis Test on the COBAS AMPLICOR and the Roche AMPLICOR Chlamydia trachomatis Test (microwell plate) performed using the Perkin-Elmer 9600 thermal cycler and a conventional photometric microwell plate reader.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The logo is black and white.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

MAY 21 1997

Alex Wesolowski

Director, Regulatory and Clinical Affairs Roche Molecular Systems, Inc. . . . 1080 U.S. Highway 202 Branchburg, New Jersey 08876-1760

K964506 Re : COBAS AMPLICOR™ Analyzer Regulatory Class: I Product Code: JJF Dated: March 24, 1997 Received: March 25, 1997

Dear Mr. Wesolowski:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include reguirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

Paqe 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as - described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Page__________________________________________________________________________________________________________________________________________________________________________

510(k) Number (if known)

Device Name: COBAS AMPLICOR™ Analyzer

Indications for Use:

The COBAS AMPLICOR Analyzer is an in vitro diagnostic device intended for use in clinical laboratories for the automation of the AMPLICOR™ Polymerase Chain Reaction test procedures.

LINE-BELOW THIS (PLEASE DO NOT WRITE CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) ાર

Over-The-Counter Use (Optional Format 1-2-96)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K964506

§ 862.2170 Micro chemistry analyzer for clinical use.

(a)
Identification. A micro chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. The distinguishing characteristic of the device is that it requires only micro volume samples obtainable from pediatric patients. This device is intended for use in conjunction with certain materials to measure a variety of analytes.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.