(127 days)
The IMMAGE Immunochemistry System Gentamicin (GEN) reagent, when used in coniunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 3 Plus, is intended for the quantitative determination of cgentamicin in human serum or plasma by rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Tobramycin (TOB) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 3 Plus, is intended for the quantitative determination of tobramycin in human serum or plasma by rate nephelometric inhibition immunoassay.
Becknan Drug Calibrator 3 Plus, used in conjunction with IMMAGE reagents, is intended for use on Beckman's IMMAGE™ Immunochemistry Systems for the calibration of Gentamicin and Tobramycin test systems.
The IMMAGE Immunochemistry System (TDM) GEN and TOB Reagents in conjunction with Beckman Drug Calibrator 3 Plus, are intended for use in the quantitative determination of gentamicin and tobramycin concentrations respectively in human serum and plasma samples on Beckman's IMMAGE Immunochemistry System.
The provided document describes the IMMAGE™ Immunochemistry System Therapeutic Drug Monitoring Reagents for Gentamicin (GEN) and Tobramycin (TOB). This document is a "Summary of Safety & Effectiveness" submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92. It is a 510(k) premarket notification.
However, it does not contain specific "acceptance criteria" for performance that would typically be described with explicit thresholds (e.g., "sensitivity must be >X%", "specificity must be >Y%"). Instead, it presents performance data (method comparison, stability, and imprecision) and states that this data supports "substantial equivalence to chemistry test systems already in commercial distribution." The "acceptance criteria" here appear to be implied by successful demonstration of equivalence to the predicate device (Abbott TDx Gentamicin and Tobramycin Reagents).
Here's an attempt to extract and interpret the information based on the provided text, while acknowledging the limitations:
1. Table of Acceptance Criteria and Reported Device Performance
As stated above, explicit acceptance criteria are not provided in the typical sense of quantitative minimums/maximums. The goal is "substantial equivalence" to the predicate device. The performance data presented are used to demonstrate this equivalence.
| Performance Metric | Implied Acceptance Criteria (Demonstrated Substantial Equivalence to Predicate) | Reported Device Performance (IMMAGE System Reagents) |
|---|---|---|
| Method Comparison (Gentamicin) | Slope close to 1, Intercept close to 0, High correlation (r) to Abbott TDx Gentamicin | Slope: 0.994, Intercept: -0.05, r: 0.998 |
| Method Comparison (Tobramycin) | Slope close to 1, Intercept close to 0, High correlation (r) to Abbott TDx Tobramycin | Slope: 1.009, Intercept: -0.12, r: 0.996 |
| Stability | Performance within acceptable limits over time (specific criteria not detailed, but implied as sufficient for substantial equivalence) | "Stability Study Results" table is present but contains garbled text, so specific numeric results cannot be extracted. However, the text states stability experiments "relate results obtained from the Abbott TDx Reagents to the IMMAGE System Reagents," suggesting satisfactory stability. |
| Imprecision (Within-Run) | Low variability; specific coefficient of variation (CV) or standard deviation (SD) criteria not detailed, but implied as sufficient for substantial equivalence. | "Estimated Within-Run Imprecision" table is present but contains garbled text, so specific numeric results cannot be extracted. The text indicates that imprecision experiments "relate results obtained from the Abbott TDx Reagents to the IMMAGE System Reagents," implying acceptable imprecision. |
2. Sample Size Used for the Test Set and Data Provenance
- Gentamicin: 135 samples
- Tobramycin: 94 samples
- Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). However, the context of a 510(k) submission for a clinical diagnostic device implies these would be human serum/plasma samples, likely from a clinical setting. The specific nature (e.g., healthy vs. disease, patient demographics) is not detailed. The term "method comparison" suggests these were patient samples run on both the new device and the predicate device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of device (therapeutic drug monitoring reagent) does not typically involve "experts" establishing a ground truth in the way an imaging AI device would. The "ground truth" for method comparison is the measurement obtained from the predicate device, the Abbott TDx system, which is an already marketed and accepted device for these analytes. There are no radiologists or similar experts involved in this type of ground truth establishment for a chemical assay.
4. Adjudication Method for the Test Set
Not applicable. As described above, the "ground truth" is the predicate device's measurement. There is no human adjudication process involved in comparing two quantitative chemical assays in this manner.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a diagnostic reagent for an automated immunoassay system, not an AI-assisted diagnostic tool that involves human readers interpreting results in a comparative study against an AI. The device itself performs the quantitative determination.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This is essentially the case. The IMMAGE system (including its reagents and internal algorithms for signal processing to determine concentration) operates as a standalone system to provide quantitative results for gentamicin and tobramycin. The data presented ("Method Comparison Study Results") reflects the performance of this system directly against the predicate.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the method comparison study was the measurements obtained from the predicate device, the Abbott TDx Gentamicin and Tobramycin Reagents systems. This is a common and accepted "ground truth" for demonstrating substantial equivalence of a new diagnostic assay to an established one.
8. The Sample Size for the Training Set
The document does not provide information about a "training set" in the context of machine learning. This device is a chemical immunoassay system, not an AI/ML algorithm that undergoes explicit training in the way typically discussed for image recognition or predictive models. The system's "performance" is based on its chemical reactions and optical measurement principles. If there were any internal calibrators or controls used for system calibration or quality control, these are inherent to the system's operation but not a separate "training set" as in AI.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there's no mention or indication of an AI/ML "training set" in the document. The "ground truth" inherent to the system's operation would be defined by the known concentrations of calibrated reference materials and controls used to ensure accurate measurement by the immunoassay.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "BECKMAN" in bold, black letters. The letters are large and fill the frame. The font is sans-serif and appears to be a standard typeface.
FEB | | 1997
Summary of Safety & Effectiveness IMMAGE™ Immunochemistry System Therapeutic Drug Monitoring Reagents [Gentamicin (GEN) and Tobramycin (TOB)]
1.0 Submitted By:
Annette Hellie Regulatory Specialist, Product Submissions Beckman Instruments, Inc. 200 S. Kraemer Blvd., W-337 Brea, California 92822-8000 Telephone: (714) 993-8767 FAX: (714) 961-4457
2.0 Date Submitted:
03 October 1996
3.0 Device Name(s):
3.1 Proprietary Names
IMMAGE™ Immunochemistry System Gentamicin (GEN) Reagent IMMAGE™ Immunochemistry System Tobramycin (TOB) Reagent Beckman Drug Calibrator 3 Plus
3.2 Classification Names
Gentamicin Test System (21 CFR §862.3450) Tobramycin Test System (21 CFR §862.3900) Calibrator (21 CFR §862.1150)
4.0 Predicate Device(s):
| IMMAGE SystemReagent | Predicate | Manufacturer | DocketNumber |
|---|---|---|---|
| Gentamicin (GEN)Reagent | TDx ® ** GentamicinReagent | Abbott* Laboratories,Inc. | K904226 |
| Tobramycin (TOB)Reagent | TDx TobramycinReagent | Abbott Laboratories,Inc. | K904226 |
*Abbott Laboratories, Abbott Park, IL 60064
** Trademark of Abbott Laboratories
Beckman Instruments, Inc.
{1}------------------------------------------------
Beckman Instruments, Inc., Section 510(k) Notification IMMAGE™ Immunochemistry System TDM (GEN & TOB) Reagents Summary of Safety & Effectiveness
5.0 Description:
The IMMAGE Immunochemistry System (TDM) GEN and TOB Reagents in conjunction with Beckman Drug Calibrator 3 Plus, are intended for use in the quantitative determination of gentamicin and tobramycin concentrations respectively in human serum and plasma samples on Beckman's IMMAGE Immunochemistry System.
6.0 Intended Use:
The IMMAGE Immunochemistry System Gentamicin (GEN) reagent, when used in coniunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 3 Plus, is intended for the quantitative determination of cgentamicin in human serum or plasma by rate nephelometric inhibition immunoassay.
The IMMAGE Immunochemistry System Tobramycin (TOB) reagent, when used in conjunction with the Beckman IMMAGE™ Immunochemistry Systems and Beckman Drug Calibrator 3 Plus, is intended for the quantitative determination of tobramycin in human serum or plasma by rate nephelometric inhibition immunoassay.
Becknan Drug Calibrator 3 Plus, used in conjunction with IMMAGE reagents, is intended for use on Beckman's IMMAGE™ Immunochemistry Systems for the calibration of Gentamicin and Tobramycin test systems.
7.0 Comparison to Predicate(s):
The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.
| Reagent | Aspect/Characteristic | Comments |
|---|---|---|
| IMMAGE System(GEN & TOB)Reagents | Intended use. | Same as Abbott TDxReagents |
| Liquid stable reagents. | ||
| IMMAGE System(GEN & TOB)Reagents | IMMAGE utilizes nephelometricinhibition immunoassay | Abbott TDx reagents utilizefluorescence polarizationimmunoassay |
| IMMAGE System(GEN & TOB)Reagents | Antibody source for IMMAGE GEN& TOB reagents is mouse. | Antisera source for TDxGentamicin and Tobramycinis sheep. |
{2}------------------------------------------------
8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method companson, stability, and imprecision experiments that relate results obtained from the Abbott TDx Reagents to the IMMAGE System Reagents.
Method Comparison* Study Results IMMAGE System GEN, TOB, PHY, and THE Redgents
| Analyte | Slope | Intercept | r | n | Predicate Method |
|---|---|---|---|---|---|
| Gentamicin | 0.994 | -0.05 | 0.998 | 135 | Abbott TDxGentamicin |
| Tobramycin | 1.009 | -0.12 | 0.996 | 94 | Abbott TDxTobramycin |
- Deming Regression statistics used
Stability Study Results
| 1888888889944111488288888889999999998888891188811987******************************************************************************************************************************************************************************Управленавания серьеснавания в веседние в вередния в семента в седерение вVARACTRANSATES ADARABEREERERERESSESSERVERATO . CONSE. Каралинания саландар максинания качата карактивные | .TARABARANARESERESSERSENSENENEARAGASSAULER PARABARET..TALLEGELEAADAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAACHILIPLE CHILIS COLLECTION COLLECTION CARDENTICAL CONSULTATION CONSULTATION COLLECTION COLLECTION CONTENTERS CONTAINING |
|---|---|
Estimated Within-Run Imprecision
| ITTATTATTTTTTACKATTTTTTTTTTTTTTTTTTTATATTATTTTTAAATTAATTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTTand dans a partificas Mar Me F Land Le M P M P S S S S M P S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S S.Last Product of Acades an accounts mater un to Bar veness to approvent.ABARRESERATEDERESbelangenben behambes annus spanness and assesses and中学生学科研学学 SpaceALARSERESERVERSED DE ProdSeptember 1 1 - Auto LaborPhoneseeeeeARRA EXEREESLodennances*******- Land and 1. ANNAVA. And P PAnenanascooobbabbbSand T Subde School States of Children Come of Children Come of Children Come of Children Come of Children Come of Children Come of Children Come of Children Come of ChildrenALASSE LEAD.PRACESCORPPPPBAREAREEEEES..ANRARENERARRESSERVEARRESSESSESSESSESSEA P R P S P Sиффеверенния в в приведения в приведения в видеосимер варианнаринансандария и кали кала как***********************The rate the state and separeA.S. BREPPERSERRESERVECALERSESSESSESSMENTES | ||||||
|---|---|---|---|---|---|---|
| .илатья визвичника в му масса сама в руди и развидив в в развиди в причина тинчевствования извественния полика в аниципаравания правич... с Рода 1 дванфувания как в в седенTherensshhann a Bebenaum gaber Banker BerrearenNEERESERESERESEREEARER------------------------------------------------------------------------------------------------------------------------------------------------------------------------------And a bank be be me a pr14888AR48888888888888814 BREND & BOUNDAPERSERARRESSEDFARSESSERSESSESSES PA.BEAR BARREA FREE FREE FREE LAND LET COLLECT LEE SERVER LA LEEDACRASSESSMENTCarlos Concession1 90 100 100 100440000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000..And and the first of the firstAngedanagakabalaana; acceptad - accares - accareeres and esta and andA freel of the comments = E b =------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |||||
|---|---|---|---|---|---|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |||||
| I Land Con come come come couple a | -------------------------------------------------------------------- | ||||
| AND I WAS CONCE L CONS OF | |||||
| . |
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
..
§ 862.3450 Gentamicin test system.
(a)
Identification. A gentamicin test system is a device intended to measure gentamicin, an antibiotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.(b)
Classification. Class II.