K Number
K963947
Date Cleared
1997-03-05

(154 days)

Product Code
Regulation Number
878.4810
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Epilaser" Normal Mode Ruby Laser is Intended Use: to hair removal intended effect of patients with skin types 1 - 4 through photothermolysis hair selective of in dermatology and plastic follicles surgery. This results in a prolonged growth delay.

Device Description

The Epilaser" Normal Mode Ruby Laser operates 694.3 at nanometers with pulse durations of 0.2 - 3.0 msec. uhe delivered energy 18 to the treatment site by an articulated arm. A water-cooled handpiece (7 mm or 10 mm) is firmly held against the skin for 2-3 seconds and the laser is then fired. The Bpilaser Normal Mode Ruby Laser is capable of producing energy fluences of 10-75 J/cm4.

AI/ML Overview

Acceptance Criteria and Study for Epilaser™ Normal Mode Ruby Laser (K963947)

The provided document describes the Epilaser™ Normal Mode Ruby Laser (K963947) for hair removal. The information available centers on demonstrating substantial equivalence to predicate devices rather than setting specific quantitative acceptance criteria with corresponding performance data. The device's intended effect is "prolonged growth delay" of hair through selective photothermolysis.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
No scarring in any subjectAchieved: "There was no scarring in any subject."
Effective selective photothermolysis of hair follicles for prolonged growth delayAchieved: "The initial study demonstrated that selective photothermolysis targeting melanin in the human hair follicle is an effective tool for hair removal, resulting in a prolonged growth delay."
No alterations to follicular epithelium or dermal collagen in control skinAchieved: "Biopsies of control skin showed no alterations the follicular epithelium or dermal collagen."
Damage to follicular epithelium in irradiated skinAchieved: "Examination of irradiated skin showed damage to the follicular epithelium."
Focal collagen damage in reticular dermis immediately adjacent to hair follicle in irradiated skinAchieved: "Focal collagen reticular dermas occurred damage in immediately adjacent to the hair follicle."
Observed effects increase with fluence levelsAchieved: "Observed effects increased with fluence levels."

Note: The document does not provide specific numerical targets (e.g., a percentage reduction in hair growth, or a score on a scarring scale) that would typically constitute explicit acceptance criteria. The criteria are inferred from the safety and effectiveness claims made.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document mentions "an initial study" and "subsequent studies, including patients treated for facial hair." However, specific sample sizes for these clinical studies are not provided.
  • Data Provenance: The studies were prospective clinical studies conducted on human subjects ("Observations in an initial study were recorded prior to treatment and at 1, 3 and 6-months after treatment," and "Subsequent studies, including patients treated for facial hair, confirmed the results of the initial study."). The country of origin of the data is not specified, but the applicant's address is Lexington, MA, USA, which suggests the studies may have been conducted in the USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not detail how ground truth was established for the clinical studies. It mentions "Observations in an initial study were recorded," implying clinical assessments, but it does not specify the number of experts or their qualifications. For animal studies, "biopsies" and "examination of irradiated skin" suggest histological analysis, which would typically be performed by veterinary pathologists, but this is not explicitly stated.

4. Adjudication Method for the Test Set

The document does not provide information on any adjudication method used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted (or at least not reported) in the provided information. The study focuses on the device's standalone performance and its substantial equivalence to predicate devices, not on comparing human readers with and without AI assistance.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

This is a physical device, a laser, not an algorithm. Therefore, the concept of a "standalone (i.e. algorithm only without human-in-the-loop performance)" study does not apply in the usual sense. The clinical studies described assessed the performance of the device when operated by trained personnel on human subjects.

7. Type of Ground Truth Used

  • Clinical Studies: The ground truth for the clinical studies appears to be based on clinical observation of hair reduction and the absence of scarring over 1, 3, and 6 months post-treatment. This is a form of direct clinical outcome assessment.
  • Animal Studies: The ground truth for the animal studies was based on histopathological examination of biopsies from control and irradiated skin, assessing follicular epithelium, dermal collagen, and damage.

8. Sample Size for the Training Set

The concept of a "training set" typically applies to machine learning algorithms. Since the Epilaser™ is a physical device and not an AI algorithm, there is no "training set" in this context. The animal and clinical studies are performance evaluation studies, not training.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" for a physical medical device, this question is not applicable. The ground truth for the performance studies was established through clinical observations and histopathological examinations as described in point 7.

{0}------------------------------------------------

K963947

MAR - 5 1997

ATTACHMENT I (Rev. 12/96)

510(K) SUMMARY

Bpilaser™ Normal Mode Ruby Laser

This 510(k) Summary of safety and effectiveness for the Epilaser™ Normal Mode Ruby Laser is submitted in accordance with the requirements of SMDA 1990 and follows Office of Device Evaluation guidance concerning the organization and content of 510 (k) a summary.

45 Hartwell Avenue

Technologies, Inc.

617-676-7330 (Fax)

December 26, 1996

Lexington, MA

617-676-7300

Ruby Laser

( K955612 )

(K950019)

Mr. Sanford Lane

Applicant :

  • 1

Address :

Contact Person:

Telephone :

Preparation Date: (of the Revised Summary)

Epilaser™ Normal Mode Ruby Laser

President, Spectrum Medical

Spectrum Medical Technologies, Inc.

02173

Common Name:

Classification Name:

Device Trade Name:

Laser surgical instrument for use in general and plastic surgery and in dermatology (see 21 CFR 878-4810) Product Code: ઉંઝર્ Panel: 79

Epilaser™ Normal Mode Ruby Laser

Thermolase LT-100 Nd:YAG Laser

Legally Marketed Predicate Device(s):

Description of the Device:

The Epilaser" Normal Mode Ruby Laser operates 694.3 at nanometers with pulse durations of 0.2 - 3.0 msec. uhe delivered energy 18 to the treatment site by an articulated arm. A water-cooled handpiece (7 mm or 10 mm) is firmly held against the skin for 2-3 seconds and the laser is then fired.

{1}------------------------------------------------

The Bpilaser Normal Mode Ruby Laser is capable of producing energy fluences of 10-75 J/cm4.

  • The Epilaser" Normal Mode Ruby Laser is Intended Use: to hair removal intended effect of patients with skin types 1 - 4 through photothermolysis hair selective of in dermatology and plastic follicles surgery. This results in a prolonged growth delay.
  • The Epilaser" Normal Mode Ruby Laser for Comparison : hair removal has the same technological characteristics as the predicate Epilaser (1955612) . It differs from the Thermolase LT-100 Nd:YAG Laser predicate device (K950019) in that the Thermolase applies laser energy at 1064 nanometers in 10 nsec pulses onto a cream containing carbon. The devices operate by selective photothermolysis.
  • Performance Data: Animal and clinical studies were conducted to provide assurance that differences in the specifications of the Epilaser™ Normal Mode Ruby Laser and the claimed predicate device for hair removal did not result in different performance during use.
  • Animal Studies: Biopsies of control skin showed no alterations the follicular epithelium or dermal collagen. Examination of irradiated skin showed damage to the follicular epithelium. Focal collagen reticular dermas occurred damage in immediately adjacent to the hair follicle.

Observed effects increased with fluence levels.

  • Clinical Studies: Observations in an initial study were recorded prior to treatment and at 1, ా and 6-months after treatment. There was no scarring in any subject.
    The initial study demonstrated that selective photothermolysis targeting melanin in the human hair follicle is an effective tool for hair removal, resulting in a prolonged growth delay.

2

{2}------------------------------------------------

Subsequent studies, including patients
treated for facial hair, confirmed the results of the initial study.

CONCLUSION : on the foregoing, the Epilaser" Based Normal Mode Ruby Laser is substantially equivalent to legally marketed claimed
predicate devices, i.e., the Bpilaser™ for dermatology and plastic surgery and the Thermolase LT-100 Nd:YAG laser for hair removal.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.