K Number
K963738

Validate with FDA (Live)

Manufacturer
Date Cleared
1997-05-23

(248 days)

Product Code
Regulation Number
870.3680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ELA Medical endocardial leads are designed to be used with implantable cardiac pacemakers. ELA Medical Stela ™ Model BJ44 and BJ45 bipolar pacing leads are intended for atrial implantation only.

Device Description

Stela™ BJ44 and BJ45 models are silicone rubber, tined, J-shaped, bipolar transvenous leads that provide a permanent electrical pathway between a pacemaker and the atrium. They are similar in design and construction to other such leads in commercial distribution.

AI/ML Overview

The provided text describes the Stela™ Models BJ44 and BJ45 pacing leads. Here's an analysis of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Test/Performance MetricPass/Fail CriterionReported Device Performance
In-Vitro Functional TestingStylet Insertion/RemovalMet established criterionAll test results met the established pass/fail criterion.
Stylet BottomingMet established criterionAll test results met the established pass/fail criterion.
Electrical ResistanceMet established criterionAll test results met the established pass/fail criterion.
IS-1 Connector ConformityMet established criterionAll test results met the established pass/fail criterion.
Leak ResistanceMet established criterionAll test results met the established pass/fail criterion.
Distal Tip Electrode/Conductor Coil Tensile StrengthMet established criterionAll test results met the established pass/fail criterion.
Insulation IntegrityMet established criterionAll test results met the established pass/fail criterion.
Conductor Coil FatigueMet established criterionAll test results met the established pass/fail criterion.
Tip FlexibilityMet established criterionAll test results met the established pass/fail criterion.
BiocompatibilityBiocompatibilityNot applicable (successful history of materials)Not performed due to successful history with same materials in other lead products.
Sterilization & PackagingSterilization TestingMet established criterionPerformed, and all test results met the established pass/fail criterion.
Mechanical/Environmental Packaging ValidationMet established criterionPerformed, and all test results met the established pass/fail criterion.
Clinical PerformancePacing CharacteristicsAppropriate pacing characteristicsExhibit appropriate pacing characteristics.
Sensing CharacteristicsAppropriate sensing characteristicsExhibit appropriate sensing characteristics.
ComplicationsNo complications observedNo complications were observed.

2. Sample Size Used for the Test Set and Data Provenance:

  • In-vitro functional testing, sterilization, and packaging validation: The text states "All test results demonstrated that the established pass / fail criterion was met in all cases." However, it does not specify the sample size for these in-vitro tests.
  • Clinical study: Described as a "limited clinical study." The exact sample size is not provided, only that it involved the Model BJ45 lead.
  • Data Provenance:
    • In-vitro tests: Likely conducted in a lab environment by the manufacturer (ELA Medical).
    • Clinical study: The text indicates the device was "recently introduced into commercial distribution outside the U.S." and that "No unanticipated adverse device effects have been reported for these leads" from this OUS experience. The limited clinical study mentioned is likely related to this OUS experience or a precursor to it. It is a prospective study for the clinical data from the "limited clinical study."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

  • Not Applicable. The described studies are primarily in-vitro engineering tests and a limited clinical study evaluating device performance, not diagnostic accuracy requiring expert panel review for ground truth.

4. Adjudication Method:

  • Not Applicable. See point 3.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No. An MRMC study was not described. This device is a medical lead, not an imaging or diagnostic AI tool that would typically undergo an MRMC study.

6. Standalone Performance:

  • Partially. The in-vitro functional testing, sterilization, and packaging validation can be considered standalone performance assessments of the device itself (algorithm only is not applicable here as it's a hardware device).
  • The "limited clinical study" also assesses the device's performance in patients in a standalone manner (without another human-in-the-loop comparison).

7. Type of Ground Truth Used:

  • In-vitro functional testing: The ground truth is based on engineering specifications and established pass/fail criteria for physical and electrical properties.
  • Clinical study: The ground truth for pacing and sensing characteristics would be derived from physiological measurements and clinical observations in patients. The absence of complications serves as a negative ground truth.

8. Sample Size for the Training Set:

  • Not Applicable. The description does not involve a "training set" in the context of machine learning or AI. This is a hardware device.

9. How the Ground Truth for the Training Set Was Established:

  • Not Applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Figure/0 description: The image shows a sequence of handwritten digits and a letter. The sequence starts with the letter 'K', followed by the digits '9', '6', '3', '7', '2', and '8'. The digits are written in a slightly irregular style, with some variations in thickness and spacing. The overall impression is that of a quickly written or somewhat stylized numerical code.

510(k) Notification for Stela leads, Models BJ44/BJ45

2. Summary & Certification

MAY 23 1997

2.1 Summary of safety and effectiveness information

2.1.1 General Information

Device Generic Name: Atrial tined pacing lead.

Device Trade Name: Stela™ Model BJ44 and BJ45 Pacing Leads

Applicant's Name and Address: ELA Medical, Inc., 15245 Minnetonka Boulevard, Minnetonka, MN 55345, Tel. (612) 935-2033

Date of Summary Preparation: September 16th 1996

Contact Person: Cathy G. Goble

510(k) Number: K963698

Date of Judgment of Substantial Equivalence Sent to Applicant:

Predicate Devices: ELA Medical Focus® Model T44F bipolar tined silicone rubber pacing lead (Document Control Number K904255) and Focus® Model J83F tined silicone rubber pacing lead (Document Control Number K.896742).

2.1.2 Description of Conditions for Which the Devices are Indicated

The Stela™ Model BJ44 and BJ45 leads are indicated for cardiac pacing and sensing, which is the same as other transvenous tined leads.

2.1.3 Device Description

Stela™ BJ44 and BJ45 models are silicone rubber, tined, J-shaped, bipolar transvenous leads that provide a permanent electrical pathway between a pacemaker and the atrium. They are similar in design and construction to other such leads in commercial distribution.

2.1.4 Alternatives

The alternatives for Stela™ Model BJ44 and BJ45 leads are other commercially available transvenous pacing leads.

ELA Medical Incorporated

Confidential and proprietary

{1}------------------------------------------------

2.1.5 Marketing History

Stela™ Model BJ44 and BJ45 leads are not in commercial distribution in the U.S. They were recently introduced into commercial distribution outside the U.S. No unanticipated adverse device effects have been reported for these leads.

2.1.6 Potential Adverse Effects

The potential adverse effects of Stela™ Model BJ44 and BJ45 leads are the same as those for tined, J-shape, transvenous leads in commercial distribution. Lead-related complications are described in the ELA Medical generic lead manual.

2.1.7 Summary of Studies

The following in-vitro functional testing was performed on the Stela™ Model BJ44 and/or BJ45 leads: stylet insertion/removal, stylet bottoming, electrical resistance, IS-1 connector conformity, leak resistance, distal tip electrode/conductor coil tensile strength, insulation integrity, conductor coil fatigue, and tip flexibility. Biocompatiblity testing was not performed, due to the successful history with the same materials in other lead products. Sterilization testing and mechanical / environmental packaging validation were performed, because the lead package design and sterilization method changed. All test results demonstrated that the established pass / fail criterion was met in all cases.

A limited clinical study demonstrated that the Model BJ45 lead exhibits appropriate pacing and sensing characteristics. No complications were observed in this study.

2.1.8 Conclusion

The information presented in this submission provides reasonable assurance that the Stela™ Model BJ44 and/or BJ45 leads will perform in a safe and effective manner.

ELA Medical Incorporated

Confidential and proprietary

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 3 1997

Ms. Catherine G. Goble Regulatory Affairs Manager ELA Medical, Inc. 2950 Xenium Lane North Plymouth, Minnesota 55441

Re: K963738 Stela Models BJ44 and BJ45 Pacing Leads Regulatory Class: III (three) Product Code: DTB Dated: February 21, 1997 Received: February 25, 1997

Dear Ms. Goble:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (act). The general controls provisions of the act include requirements for reqistration, listing of devices, good manufacturing practices, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Fevices: General (GMP) regulation (21 CFR Part 620) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlation.

{3}------------------------------------------------

Page 2 - Ms. Catherine G. Goble

新闻网 - 1

Under Section 522(a) of the act, manufacturers of certain types of devices identified by the Act or designated by FDA are required to conduct postmarket surveillance studies. FDA has identified under Section 522(a) (1) (A) the device cleared for marketing by this letter as requiring postmarket surveillance.

Within thirty (30) days of first introduction or delivery for introduction of this device into interstate commerce you are required to submit to FDA certification of the date of introduction into interstate commerce, a detailed protocol which describes the postmarket surveillance study, and a detailed profile of the study's principal investigator that clearly establishes the qualifications and experience of the individual to conduct the proposed study. For your information, general guidance on preparing a protocol for a postmarket surveillance study is attached.

Submit five (5) copies to:

Center for Devices and Radiological Health Postmarket Surveillance Studies Document Center Room 3083 (HFZ-544) 1350 Piccard Drive Rockville, Maryland 20850

Within sixty (60) days of receipt of your protocol, FDA will either approve or disapprove it and notify you of the Agency's action in writing. You should not begin your postmarket surveillance study of this device until the protocol has been approved. Data generated under an unapproved protocol may not satisfy your obligation under section 522. Please note that you must continue to collect and report data needed to maintain compliance with Medical Device Reporting requlations (21 CFR 803).

Failure to certify accurately the date of initial introduction of your device into interstate commerce, to submit timely an acceptable protocol, or to undertake and complete and FDA approved postmarket surveillance study consistent with the protocol will be considered violations of section 522. In accordance with the Medical Device Amendments of 1992, failure of a manufacturer to meet its obligations under section 522 is a probibited act under section 301(q) (1) (C) of the Act (21 U.S.C. 331 (q)(1)(C). Further, under section 502(t)(3) of the act (21 U.S.C. 352(t)(3)), a device is misbranded if there is a failure or refusal to comply with any requirement under section 522 of the act. Violations of sections 301 or 502 may lead to requlatory actions including seizure of your product, injunction, prosecution, or civil money penalties.

{4}------------------------------------------------

Page 3 - Ms. Catherine G. Goble

If you have questions specifically concerning postmarket surveillance study requirements, contact the Postmarket Surveillance Studies Branch at (301) 594-0639.

In addition, on August 16, 1993, the Final Rule for Device Tracking was published in the Federal Reqister, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1996, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of trackinq that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirement of the requlation are found in 21 CFR 821 as described in the August 16, 1993 Federal Register beginning on page 43447.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Wolf Saporta mo

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{5}------------------------------------------------

1.10 Indications for Use Statement

510 (k) Number:K963738
Device Name:Stela™ Model BJ44 and BJ45 pacing leads.

Indication for Use:

ELA Medical endocardial leads are designed to be used with implantable cardiac pacemakers.

ELA Medical Stela ™ Model BJ44 and BJ45 bipolar pacing leads are intended for atrial implantation only.

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number. K963735

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use _

(Per 21 CFR 801.109)

ELA Medical Incorporated

Confidential and proprietary

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.