(56 days)
The ISE Reagents are intended for use on the ISE Module of the BMD Hitachi Clinical Chemistry Systems. The ISE High and Low Calibrators are intended to provide calibration points for the Na +, K+, and Cl- Electrodes on the ISE Subsystem. The IN KCI Solutions are intended to provide stable reference potential for the Na +, K+, and Cl- Electrodes. The ISE Diluent dilutes all measured patient samples for the quantitative determination of Na +, K+, and Cl- in Serum, Plasma and Urine by ISE. The ISE Internal Reference Solutions are intended as a means of compensating for calibration drift in the quantitative determination of Na +, K+, and Cl- in Serum, Plasma and Urine samples on the BMD Hitachi Clinical Chemistry Systems.
All product(s) encompassed by this 510(k) submission are Class I (75JJG) and Class II (75JIX) In Vitro Diagnostic Solutions manufactured by ALKO Diagnostic Corporation. The ISE Reagents are intended for use on the ISE Module of the BMD Hitachi Clinical Chemistry Systems. The ISE High and Low Calibrators are intended to provide calibration points for the Na +, K+, and Cl- Electrodes on the ISE Subsystem. The IN KCI Solutions are intended to provide stable reference potential for the Na +, K+, and Cl- Electrodes. The ISE Diluent dilutes all measured patient samples for the quantitative determination of Na +, K+, and Cl- in Serum, Plasma and Urine by ISE. The ISE Internal Reference Solutions are intended as a means of compensating for calibration drift in the quantitative determination of Na +, K+, and Cl- in Serum, Plasma and Urine samples on the BMD Hitachi Clinical Chemistry Systems.
Here's an analysis of the acceptance criteria and study detailed in the provided text:
Device: ALKO Diagnostic Corporation ISE Reagents for BMD Hitachi Clinical Chemistry 700 Series and 911 Systems
Description of Device: ISE Reagents (Diluent, Internal Reference Solution, 1N KCI Solution, Low Standard, High Standard) intended for use on the ISE Module of BMD Hitachi Clinical Chemistry Systems for the quantitative determination of Na+, K+, and Cl- in Serum, Plasma, and Urine. They are designed as direct replacements for Boehringer Mannheim Diagnostic (BMD) products.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the sponsor's claim of "performance equivalence" to the predicate BMD reagents. The study aims to demonstrate that results obtained with ALKO reagents are comparable to those obtained with BMD reagents, and that ALKO reagents exhibit acceptable precision.
Performance Equivalence (Correlation Study):
The acceptance criteria are not explicitly stated as numerical thresholds (e.g., "slope must be between X and Y"). Instead, the document presents linear regression results (Slope, Intercept, R-squared) between ALKO and BMD reagents, implying that a slope close to 1, an intercept close to 0, and a high R-squared value (close to 1) indicate equivalence. For the purpose of this table, we will highlight the reported statistics.
Precision Study:
The acceptance criteria for precision are also not explicitly stated in numerical terms (e.g., "CV% must be < X%"). The reported data includes Total Standard Deviation (Ttl SD) and Total Coefficient of Variation (Ttl CV%), as well as Within-Run Standard Deviation (WR SD) and Within-Run Coefficient of Variation (WR CV%). The expectation is that these values should be within generally accepted limits for clinical chemistry assays, and the sponsor presents them as evidence of acceptable precision.
| Test Type | Analyte | Sample Type | System | Acceptance Criteria (Implicit) | Reported Device Performance (ALKO Reagents) |
|---|---|---|---|---|---|
| Correlation | Slope ≈ 1, Intercept ≈ 0, | ||||
| (ALKO vs. BMD) | R. Sq ≈ 1 | ||||
| Na | Urine | 736 | Slope: 1.0076, Intercept: 1.5483, R. Sq: 0.9995 | ||
| K | Urine | 736 | Slope: 1.0028, Intercept: 0.4064, R. Sq: 0.9995 | ||
| Cl | Urine | 736 | Slope: 1.0419, Intercept: -2.5008, R. Sq: 0.9986 | ||
| Na | Urine | 704 | Slope: 1.0798, Intercept: -5.1321, R. Sq: 0.9985 | ||
| K | Urine | 704 | Slope: 1.0436, Intercept: -0.4654, R. Sq: 0.9904 | ||
| Cl | Urine | 704 | Slope: 1.0769, Intercept: -5.9423, R. Sq: 0.9988 | ||
| Na | Urine | 737 | Slope: 1.0239, Intercept: -2.0167, R. Sq: 0.9992 | ||
| K | Urine | 737 | Slope: 0.9828, Intercept: -0.0362, R. Sq: 0.9996 | ||
| Cl | Urine | 737 | Slope: 1.0243, Intercept: -2.0894, R. Sq: 0.9978 | ||
| Na | Urine | 911 | Slope: 1.0368, Intercept: -4.2943, R. Sq: 0.9997 | ||
| K | Urine | 911 | Slope: 0.9574, Intercept: 0.1334, R. Sq: 0.9992 | ||
| Cl | Urine | 911 | Slope: 1.0295, Intercept: -1.5562, R. Sq: 0.9993 | ||
| Na | Plasma | 736 | Slope: 1.0141, Intercept: -1.6476, R. Sq: 0.9955 | ||
| K | Plasma | 736 | Slope: 0.9747, Intercept: 0.1266, R. Sq: 0.9990 | ||
| Cl | Plasma | 736 | Slope: 0.9968, Intercept: -0.4270, R. Sq: 0.9955 | ||
| Na | Plasma | 704 | Slope: 0.9705, Intercept: 3.64833, R. Sq: 0.9913 | ||
| K | Plasma | 704 | Slope: 0.9704, Intercept: 0.1084, R. Sq: 0.9982 | ||
| Cl | Plasma | 704 | Slope: 1.0023, Intercept: -1.3518, R. Sq: 0.9946 | ||
| Na | Plasma | 737 | Slope: 1.0331, Intercept: -7.4335, R. Sq: 0.9949 | ||
| K | Plasma | 737 | Slope: 0.9843, Intercept: -0.04915, R. Sq: 0.9991 | ||
| Cl | Plasma | 737 | Slope: 1.0185, Intercept: -4.5429, R. Sq: 0.9976 | ||
| Na | Plasma | 911 | Slope: 1.0096, Intercept: -1.2435, R. Sq: 0.9930 | ||
| K | Plasma | 911 | Slope: 0.9553, Intercept: -0.2228, R. Sq: 0.9992 | ||
| Cl | Plasma | 911 | Slope: 0.9574, Intercept: 5.8746, R. Sq: 0.9917 | ||
| Precision | Acceptable CV% for clinical | ||||
| (ALKO Reagents only) | chemistry (context dependent) | ||||
| Na (L1) | Urine | 704 | Mean: 88.43, Ttl SD: 2.38, Ttl CV%: 2.69, WR SD: 0.42, WR CV%: 0.47 | ||
| K (L1) | Urine | 704 | Mean: 28.73, Ttl SD: 0.48, Ttl CV%: 1.67, WR SD: 0.34, WR CV%: 1.19 | ||
| Cl (L1) | Urine | 704 | Mean: 80.28, Ttl SD: 3.05, Ttl CV%: 3.80, WR SD: 0.32, WR CV%: 0.39 | ||
| Na (L2) | Urine | 704 | Mean: 188.39, Ttl SD: 2.32, Ttl CV%: 1.23, WR SD: 0.73, WR CV%: 0.39 | ||
| K (L2) | Urine | 704 | Mean: 73.29, Ttl SD: 4.01, Ttl CV%: 5.47, WR SD: 3.02, WR CV%: 4.12 | ||
| Cl (L2) | Urine | 704 | Mean: 190.09, Ttl SD: 5.58, Ttl CV%: 2.94, WR SD: 1.86, WR CV%: 0.98 | ||
| Na (L1) | Urine | 717 | Mean: 83.28, Ttl SD: 1.77, Ttl CV%: 2.13, WR SD: 0.57, WR CV%: 0.68 | ||
| K (L1) | Urine | 717 | Mean: 29.64, Ttl SD: 0.59, Ttl CV%: 1.98, WR SD: 0.55, WR CV%: 1.85 | ||
| Cl (L1) | Urine | 717 | Mean: 77.53, Ttl SD: 3.91, Ttl CV%: 5.04, WR SD: 1.13, WR CV%: 1.46 | ||
| Na (L2) | Urine | 717 | Mean: 180.05, Ttl SD: 2.26, Ttl CV%: 1.26, WR SD: 1.00, WR CV%: 0.56 | ||
| K (L2) | Urine | 717 | Mean: 77.73, Ttl SD: 3.91, Ttl CV%: 5.04, WR SD: 1.13, WR CV%: 1.46 | ||
| Cl (L2) | Urine | 717 | Mean: 182.78, Ttl SD: 6.71, Ttl CV%: 3.67, WR SD: 1.82, WR CV%: 1.00 | ||
| Na (L1) | Urine | 736 | Mean: 81.25, Ttl SD: 2.00, Ttl CV%: 2.47, WR SD: 0.87, WR CV%: 1.07 | ||
| K (L1) | Urine | 736 | Mean: 29.25, Ttl SD: 0.46, Ttl CV%: 1.46, WR SD: 0.35, WR CV%: 1.20 | ||
| Cl (L1) | Urine | 736 | Mean: 80.38, Ttl SD: 4.32, Ttl CV%: 5.38, WR SD: 2.37, WR CV%: 2.95 | ||
| Na (L2) | Urine | 736 | Mean: 179.05, Ttl SD: 2.63, Ttl CV%: 1.47, WR SD: 2.53, WR CV%: 1.41 | ||
| K (L2) | Urine | 736 | Mean: 94.90, Ttl SD: 1.99, Ttl CV%: 2.09, WR SD: 0.44, WR CV%: 0.46 | ||
| Cl (L2) | Urine | 736 | Mean: 159.48, Ttl SD: 6.44, Ttl CV%: 4.04, WR SD: 1.75, WR CV%: 1.10 | ||
| Na (L1) | Urine | 737 | Mean: 27.83, Ttl SD: 1.39, Ttl CV%: 4.99, WR SD: 0.69, WR CV%: 2.48 | ||
| K (L1) | Urine | 737 | Mean: 7.57, Ttl SD: 0.11, Ttl CV%: 1.45, WR SD: 0.09, WR CV%: 1.24 | ||
| Cl (L1) | Urine | 737 | Mean: 26.75, Ttl SD: 4.03, Ttl CV%: 15.08, WR SD: 1.83, WR CV%: 6.84 | ||
| Na (L2) | Urine | 737 | Mean: 57.33, Ttl SD: 1.59, Ttl CV%: 2.78, WR SD: 0.52, WR CV%: 0.91 | ||
| K (L2) | Urine | 737 | Mean: 15.85, Ttl SD: 0.17, Ttl CV%: 1.07, WR SD: 0.04, WR CV%: 0.28 | ||
| Cl (L2) | Urine | 737 | Mean: 56.55, Ttl SD: 2.29, Ttl CV%: 4.05, WR SD: 0.59, WR CV%: 1.05 |
Study to Prove Device Meets Acceptance Criteria
The studies conducted are described as:
- Comparative Analysis (Correlation Study): Comparing results from BMD Hitachi Chemistry Systems calibrated with ALKO products versus BMD products.
- Precision Study: Measuring samples over 20 runs on BMD Hitachi Clinical Chemistry Systems with ALKO products.
2. Sample Size Used for the Test Set and Data Provenance
- Correlation Study (Urine Samples):
- Hitachi 736 System: N=40
- Hitachi 704 System: N=40
- Hitachi 737 System: N=42
- Hitachi 911 System: N=42
- Correlation Study (Plasma Samples):
- Hitachi 736 System: N=59
- Hitachi 704 System: N=57 (Na, K); N=59 (Cl)
- Hitachi 737 System: N=59
- Hitachi 911 System: N=59
- Precision Study (Urine Samples): For each analyte (Na, K, Cl) at two levels (Level 1, Level 2) on systems 704, 717, 736, and 737, the sample size (N) is 20.
Data Provenance: Not explicitly stated in the provided text. It is a retrospective analysis comparing against an existing predicate. The document refers to "samples measured on Hitachi..." suggesting patient samples were collected and analyzed. No mention of country of origin.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
N/A. This is a study comparing analytical reagent performance, not an AI or diagnostic imaging device requiring expert interpretation for ground truth. The "ground truth" for the correlation study is the measurement obtained using the predicate (BMD) reagents.
4. Adjudication Method for the Test Set
N/A. Not applicable for this type of analytical performance study. The adjudication is the agreement between the ALKO and BMD reagent measurements.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
N/A. This is not an MRMC study. It is an analytical performance study of reagents.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done
N/A. This is not an AI or algorithm-based device. It's a chemical reagent. The performance described is the standalone performance of the ALKO reagents in terms of their analytical output on the specified instruments.
7. The Type of Ground Truth Used
For the correlation study, the ground truth is the measurements obtained using the Boehringer Mannheim Diagnostic (BMD) predicate reagents on the same Hitachi Clinical Chemistry Systems. This is a comparative type of ground truth where the new device's performance is gauged against an established, accepted method.
For the precision study, the ground truth is implicitly the statistical expectation of minimal variability around the mean value of repeated measurements of the same sample. There isn't an external "ground truth" measurement, but rather an assessment of the consistency of the device's own measurements.
8. The Sample Size for the Training Set
N/A. This is not a machine learning or AI device that requires a training set. The reagents are chemical formulations.
9. How the Ground Truth for the Training Set Was Established
N/A. Not applicable.
{0}------------------------------------------------
ALKO Diagnostic Corporation
Image /page/0/Picture/2 description: The image shows a logo with a thick, black triangle at the top. Inside the triangle is a white space that also forms a triangle. Below the triangle is the word "ALKO" in bold, black capital letters. To the right of the top triangle is a small circle with an R inside, indicating a registered trademark.
A Life Sciences International Company
510(k) Summary
OCT 2 4 1996
ISE Reagents
for BMD Hitachi Clinical Chemistry 700 Series and 911 Systems
All product(s) encompassed by this 510(k) submission are Class I (75JJG) and Class II (75JIX) In Vitto Diagnostic Solutions manufactured by ALKO Diagnostic Corporation, 333 Fiske Street, Holliston, MA 01746. The ISE Reagents are intended for use on the ISE Module of the BMD Hitachi Clinical Chemistry Systems. The ISE High and Low Calibrators are intended to provide calibration points for the Na +, K+, and Cl- Electrodes on the ISE Subsystem. The IN KCI Solutions are intended to provide stable reference potential for the Na +, K+, and Cl- Electrodes. The ISE Diluent dilutes all measured patient samples for the quantitative determination of Na +, K+, and Cl- in Serum, Plasma and Urine by ISE. The ISE Internal Reference Solutions are intended as a means of compensating for calibration drift in the quantitative determination of Na +, K+, and Cl- in Serum, Plasma and Urine samples on the BMD Hitachi Clinical Chemistry Systems. Boehringer Mannheim Diagnostics is the original equipment manufacturer (OEM ) of the analyzer(s) and of predicate ISE Reagents which are necessary for the continued operation and use of ISE Subsystem on the analyzer.
The ISE Hitachi Reagents are intended to serve as direct replacement to like named product manufactured by Boehringer Mannheim Diagnostic. Listed below are ALKO products and their OEM equivalents.
| ALKO # | OEM | Description | Models | Class |
|---|---|---|---|---|
| A858-044 | 820637 | ISE Diluent | 704 | 1 |
| A820-636 | 820638 | ISE Diluent | 717,736,737,747 | 1 |
| A450-043 | 450043 | ISE Diluent | 911 | 1 |
| A820-644 | 836246 | ISE Internal Reference Solution | 704,717,737,747 | 1 |
| A836-244 | 836245 | ISE Internal Reference Solution | 736 | 1 |
| A450-042 | 450042 | ISE Internal Reference Solution | 911 | 1 |
| A820-652 | 820639 | ISE 1N KCI Solution | 704,717,736,737,747 | 1 |
| A450-041 | 450041 | ISE 1N KCI Solution | 911 | 1 |
| A620-428 | 620428 | ISE Low Standard | 704,717,736,737,747,911 | 2 |
| A620-427 | 620427 | ISE High Standard | 704,717,736,737,747,911 | 2 |
ALKO uses a similar composition, description and packaging design as that used by Boehringer Mannheim in its products. ALKO has shown performance equivalence of its products to Boehringer Mannheim Company's products in the following manner:
- O Through a comparative analysis where results obtained on BMD Hitachi Chemistry Systems calibrated with ALKO products were compared with results obtained on the same system calibrated with BMD Hitachi products; and
- O Through a precision study where ALKO products were installed on BMD Hitachi Clinical Chemistry Systems and samples were measured over 20 runs.
A summary of the results these studies follows: Worldwide Supplier of Blood Gas and ISE
Image /page/0/Picture/14 description: The image shows the text "SHEET 53 OF/62" in a bold, sans-serif font. The numbers "53" and "62" are larger than the words "SHEET" and "OF". The text appears to be part of a document or a page numbering system.
{1}------------------------------------------------
PERFORMANCE CHARACTERISTICS
Correlation with BMD Reagents - Urine Samples
- -1. Correlation data collected from urine samples measured on Hitachi 736, 704, 737 and 911 System, calibrated with ALKO and BMD products separately. The study compares values obtained using ALKO products as compared BMD products wherein Linear Regression Analyses were performed using ALKO Data as the Dependent Y Variable and BMD Data as the Independent X Variable in the equation Y = a + bX.
Hitachi 736 System
| Analyte | (N) | Slope | Intercept | R. Sq. | Range |
|---|---|---|---|---|---|
| Na | 40 | 1.0076 | 1.5483 | 0.9995 | 9-253 |
| K | 40 | 1.0028 | 0.4064 | 0.9995 | 5.7-81.1 |
| Cl | 40 | 1.0419 | -2.5008 | 0.9986 | 13-214 |
Hitachi 704 System
| Analyte | (N) | Slope | Intercept | R Sq | Range |
|---|---|---|---|---|---|
| Na | 40 | 1.0798 | -5.1321 | 0.9985 | 10.0-258.8 |
| K | 40 | 1.0436 | -0.4654 | 0.9904 | 5.1-83.2 |
| Cl | 40 | 1.0769 | -5.9423 | 0.9988 | 10.8-237.7 |
$
Hitachi 737 System
| Analyte | (N) | Slope | Intercept | R Sq | Range |
|---|---|---|---|---|---|
| Na | 42 | 1.0239 | -2.0167 | 0.9992 | 6-190 |
| K | 42 | 0.9828 | -0.0362 | 0.9996 | 3.3-102.5 |
| Cl | 42 | 1.0243 | -2.0894 | 0.9978 | 7-235 |
Hitachi 911 System
| Analyte | (N) | Slope | Intercept | R Sq | Range |
|---|---|---|---|---|---|
| Na | 42 | 1.0368 | -4.2943 | 0.9997 | 2-187 |
| K | 42 | 0.9574 | 0.1334 | 0.9992 | 3.2-101.5 |
| Cl | 42 | 1.0295 | -1.5562 | 0.9993 | 14-231 |
R Sq = Correlation Coefficient Squared
SHEET54 OF/
{2}------------------------------------------------
Page 3/ ÀLKO 510 (k) Summary for ISE Hitachi Reagents
Correlation with BMD Reagents - Plasma Samples
-
- Correlation data collected from plasma samples measured on Hitachi 736, 704, 737 and 911 Clinical Chemistry Systems, calibrated with ALKO and BMD products separately. The study compares values obtained using ALKO products as compared BMD products wherein Linear Regression Analyses were performed using ALKO Data as the Dependent Y Variable and BMD Data as the Independent X Variable in the equation Y = a + bX.
| Analyte | (N) | Slope | Intercept | R Sq | Range |
|---|---|---|---|---|---|
| Na | 59 | 1.0141 | -1.6476 | 0.9955 | 86-178 |
| K | 59 | 0.9747 | 0.1266 | 0.9990 | 3.0-11.9 |
| Cl | 59 | 0.9968 | -0.4270 | 0.9955 | 64-154 |
Hitachi 736 System
Hitachi 704 System
| Analyte | (N) | Slope | Intercept | R Sq | Range |
|---|---|---|---|---|---|
| Na | 57 | 0.9705 | 3.64833 | 0.9913 | 84.6-180.9 |
| K | 57 | 0.9704 | 0.1084 | 0.9982 | 2.9-12.2 |
| Cl | 59 | 1.0023 | -1.3518 | 0.9946 | 63.6-150.0 |
Hitachi 737 System
| Analyte | (N) | Slope | Intercept | R Sq | Range |
|---|---|---|---|---|---|
| Na | 59 | 1.0331 | -7.4335 | 0.9949 | 85-180 |
| K | 59 | 0.9843 | -0.04915 | 0.9991 | 3.0-12.2 |
| Cl | 59 | 1.0185 | -4.5429 | 0.9976 | 62-156 |
Hitachi 911 System
| Analyte | (N) | Slope | Intercept | R Sq | Range |
|---|---|---|---|---|---|
| Na | 59 | 1.0096 | -1.2435 | 0.9930 | 86-179 |
| K | 59 | 0.9553 | -0.2228 | 0.9992 | 3.0-12.2 |
| Cl | 59 | 0.9574 | 5.8746 | 0.9917 | 67-154 |
R Sq = Correlation Coefficient Squared
SHEET 55 OF 62
{3}------------------------------------------------
Page 4/ ALKO 510 (k) Summary for ISE Hitachi Reagents
Precision of ALKO Reagents - Urine Samples
-
- Precision data collected from urine samples measured on Hitachi 704, 717, 736 and 737 Clinical Chemistry Systems, over a period of four to six days using all ALKO reagents.
| Level | Analyte | (N) | Mean | Ttl SD | Ttl CV% | WR SD | WR CV% |
|---|---|---|---|---|---|---|---|
| Level 1 | Na | 20 | 88.43 | 2.38 | 2.69 | 0.42 | 0.47 |
| K | 20 | 28.73 | 0.48 | 1.67 | 0.34 | 1.19 | |
| Cl | 20 | 80.28 | 3.05 | 3.80 | 0.32 | 0.39 | |
| Level 2 | Na | 20 | 188.39 | 2.32 | 1.23 | 0.73 | 0.39 |
| K | 20 | 73.29 | 4.01 | 5.47 | 3.02 | 4.12 | |
| Cl | 20 | 190.09 | 5.58 | 2.94 | 1.86 | 0.98 |
Hitachi 704 System
Hitachi 717 System
| Level | Analyte | (N) | Mean | Ttl SD | Ttl CV% | WR SD | WR CV% |
|---|---|---|---|---|---|---|---|
| Level 1 | Na | 20 | 83.28 | 1.77 | 2.13 | 0.57 | 0.68 |
| Level 1 | K | 20 | 29.64 | 0.59 | 1.98 | 0.55 | 1.85 |
| Level 1 | Cl | 20 | 77.53 | 3.91 | 5.04 | 1.13 | 1.46 |
| Level 2 | Na | 20 | 180.05 | 2.26 | 1.26 | 1.00 | 0.56 |
| Level 2 | K | 20 | 77.73 | 3.91 | 5.04 | 1.13 | 1.46 |
| Level 2 | Cl | 20 | 182.78 | 6.71 | 3.67 | 1.82 | 1.00 |
Hitachi 736 System
| Level | Analyte | (N) | Mean | Ttl SD | Ttl CV% | WR SD | WR CV% |
|---|---|---|---|---|---|---|---|
| Level 1 | Na | 20 | 81.25 | 2.00 | 2.47 | 0.87 | 1.07 |
| K | 20 | 29.25 | 0.46 | 1.46 | 0.35 | 1.20 | |
| Cl | 20 | 80.38 | 4.32 | 5.38 | 2.37 | 2.95 | |
| Level 2 | Na | 20 | 179.05 | 2.63 | 1.47 | 2.53 | 1.41 |
| K | 20 | 94.90 | 1.99 | 2.09 | 0.44 | 0.46 | |
| Cl | 20 | 159.48 | 6.44 | 4.04 | 1.75 | 1.10 |
{4}------------------------------------------------
Page 5/ ALKO 510 (k) Summary for ISE Hitachi Reagents
| Level | Analyte | (N) | Mean | Ttl SD | Ttl CV% | WR SD | WR CV% |
|---|---|---|---|---|---|---|---|
| Level 1 | Na | 20 | 27.83 | 1.39 | 4.99 | 0.69 | 2.48 |
| K | 20 | 7.57 | 0.11 | 1.45 | 0.09 | 1.24 | |
| Cl | 20 | 26.75 | 4.03 | 15.08 | 1.83 | 6.84 | |
| Level 2 | Na | 20 | 57.33 | 1.59 | 2.78 | 0.52 | 0.91 |
| K | 20 | 15.85 | 0.17 | 1.07 | 0.04 | 0.28 | |
| Cl | 20 | 56.55 | 2.29 | 4.05 | 0.59 | 1.05 |
Hitachi 737 System
Conclusion
ALKO manufactures products that serve as functional equivalents to consumable distributed by Original Equipment Manufacturer. All such products are manufactured with the intent of complying with the Good Manufacturing Practice Regulations.
As previously stated the product encompassed by this request serves as a functional equivalent to the product currently manufactured and marketed by Boehringer Mannheim Diagnostics. We believe that BMD has received 510 (k) market approval given that they are currently distributing the instruments nd related consumable domestically. The previously described labeling and performance verification "chould provide adequate documentation as to the equivalence of the ALKO Product.
I hope you find the documentation provided informative and helpful.
Janet A. McGrath Regulatory Affairs
SHEET 5 OF 1/2
§ 862.1170 Chloride test system.
(a)
Identification. A chloride test system is a device intended to measure the level of chloride in plasma, serum, sweat, and urine. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.(b)
Classification. Class II.