Intended Use

The Dexterity® Pneumo Sleeve is intended to provide abdominal access for the surgeon's hand while preserving pneumoperitoneum during laparoscopic surgery.

The Protector® Retractor is intended to provide incision retraction and to protect it against wound contamination during both laparoscopic and open surgery.

This device allows the surgeon to use his hand while performing corrective surgery in the abdominal cavity under pneumoperitoneum.

Device Description

The Dexterity® Pneumo Sleeve is a sterile, disposable device. It prevents leakage of CO2 and provides an extension of the pneumoperitoneum. It has an adhesive flange at one end that is placed over the inoision and adhered to the patient's abdomen. A velcro strap at the other end is used to tighten the Sleeve around the surgeon's arm.

The Protector® Retractor is a sterile, disposable device. The Retractor consists of an open-ended cylinder with a flexible ring at each end. One ring is inserted through the incision into the peritoneal cavity while the other remains outside the incision. The Retractor holds the incision open and protects it against wound contamination.

AI/ML Overview

Here's an analysis of the provided text regarding acceptance criteria and the supporting study, structured according to your request.

It's important to note that the provided 510(k) summary (K962147 for the Dexterity® Pneumo Sleeve and Protector® Retractor) describes a medical device for surgical access, not an AI/ML powered device. As such, many of your requested criteria, which are highly specific to AI/ML device validation (e.g., sample size for test set/training set, expert grounding, MRMC studies, standalone performance, data provenance for AI), are not applicable to this traditional medical device submission. I will answer those questions with "Not Applicable" and explain why.


K962147: Dexterity® Pneumo Sleeve and Protector® Retractor

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary does not explicitly list quantitative "acceptance criteria" in the way one might for an AI algorithm (e.g., target specificity, sensitivity, or AUC). Instead, the "acceptance criteria" for this traditional medical device are defined by its ability to demonstrate safety and effectiveness for its intended use, typically by being substantially equivalent to a predicate device and showing favorable outcomes in a clinical setting against established surgical practices.

Acceptance Criteria (Inferred from study goals)Reported Device Performance (Summary of Clinical Results)
Safety:
- No increased morbidity or mortality compared to routine laparoscopic surgery.Confirmed safety (comparable morbidity and mortality)
- No unexpected adverse events.Implied by "confirmed safety"
Effectiveness:
- Provides abdominal access for surgeon's hand while preserving pneumoperitoneum."validated the operating principles of the device"
- Allows surgeon to use hand during laparoscopic surgery.Confirmed effectiveness for intended use.
- Incision retraction and protection against wound contamination.(Not directly evaluated in pivotal study, but implicit for Protector® Retractor component)
- Comparable overall procedure time to routine laparoscopic surgery.Assessed in pivotal study. (Specific impact not detailed, but overall "confirmed effectiveness")
- Comparable number of instrument changes required.Assessed in pivotal study. (Specific impact not detailed, but overall "confirmed effectiveness")
- Comparable conversion rates (to open procedure).Assessed in pivotal study. (Specific impact not detailed, but overall "confirmed effectiveness")
- Comparable direct operation costs.Assessed in pivotal study. (Specific impact not detailed, but overall "confirmed effectiveness")
- Comparable time for patients to return to normal activity.Assessed in pivotal study. (Specific impact not detailed, but overall "confirmed effectiveness")

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set:
    • Pilot Study: 10 surgeons and 40 patients.
    • Pivotal Clinical Study: Not explicitly stated, but was a "prospective, randomized, controlled trial" across 10 clinical centers. The exact number of patients or surgeons is not provided in this summary.
  • Data Provenance: The pivotal clinical study was conducted at "10 clinical centers in the United States and Europe." This indicates a prospective collection of clinical data from these regions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not Applicable. This is a device for surgical access, not an AI algorithm that requires expert annotation for "ground truth" establishment in a diagnostic or interpretive sense. The "ground truth" for this device's performance is derived from clinical outcomes and surgeon's ability to perform the procedure as intended. The "experts" observing the performance would be the surgeons and clinical staff involved in the studies, assessing patient outcomes and device function.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not Applicable. Adjudication methods like 2+1 or 3+1 are typically used for resolving discrepancies in expert interpretations (e.g., image readings) to establish a definitive ground truth for AI model training or evaluation. This concept does not apply to the evaluation of a surgical access device. The clinical study would have had its own statistical analysis plan and potentially an independent data monitoring committee, but not expert adjudication in this sense.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. MRMC studies are specific to evaluating diagnostic AI systems and their impact on human reader performance. This device is a surgical instrument and has no "readers" or "AI assistance" component in that context. The pivotal clinical study was a "randomized, controlled trial" comparing the device-assisted approach to "routine laparoscopic surgery," which is a comparative effectiveness study, but not of the MRMC type.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This is a physical surgical device, not an algorithm. Standalone performance is a concept for AI models.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • The "ground truth" for this device's efficacy and safety was established by clinical outcomes data from the randomized controlled trial. This includes metrics like overall procedure time, number of instrument changes, conversion rates, direct operation costs, time to return to normal activity, and morbidity and mortality. The "operating principles" were initially validated by direct observation and feedback from surgeons in the pilot study.

8. The sample size for the training set

  • Not Applicable. This summary describes a traditional hardware medical device, not an AI/ML algorithm that requires a "training set."

9. How the ground truth for the training set was established

  • Not Applicable. As there is no AI/ML training set, there's no ground truth to establish for it.

{0}------------------------------------------------

K962147

APPENDIX H

510(k) SUMMARY

This summary is organized per 21 CFR 807.92.

Medical Creative Technologies, Inc. 2950-N Advance Lane Colmar, PA 18915

Telephone Number: 215-997-9689 Fax Number: 215-997-2371

Name of Contact: Perk Crook

Date Summary Prepared: April 3, 1996

Name of Device:

Trade Name: Dexterity® Pneumo Sleeve and Protector® Retractor

Common Name: Extended Laparoscopy Device

Classification Name: Extended Laparoscopy Device

Identification of Predicate or Legally Marketed Device(s):

    1. Trocar/Cannula by Storz
    1. Protector Retractor by Medical Creative Technologies, Inc.

The Dexterity® Pneumo Sleeve and Protector® Retractor are substantially equivalent to a Trocar/Cannula from an access to the peritoneal cavity standpoint. Both devices allow access to the peritoneal cavity while preventing CO2 leakage during insertion and withdrawal of instruments. In the case of the Trocar/Cannula, the instruments are laparoscopic instruments and in the case of

{1}------------------------------------------------

the Dexterity® Pneumo Sleeve and Protector® Retractor, the instrument is the surgeon's hand. The Trocar/Cannula protects the wound as does the Protector® Retractor from trauma and/or cross-contamination.

The Protector® Retractor which is a component of this device is the same device as the Protector® Retractor (second predicate device named above) approved by 510(k) Number K954824. The Dexterity® Pneumo Sleeve is made from the same 3 mil thick plastic film used to make the sleeve of the Protector® Retractor.

Description of the Device:

The Dexterity® Pneumo Sleeve is a sterile, disposable device. It prevents leakage of CO2 and provides an extension of the pneumoperitoneum. It has an adhesive flange at one end that is placed over the inoision and adhered to the patient's abdomen. A velcro strap at the other end is used to tighten the Sleeve around the surgeon's arm.

The Protector® Retractor is a sterile, disposable device. The Retractor consists of an open-ended cylinder with a flexible ring at each end. One ring is inserted through the incision into the peritoneal cavity while the other remains outside the incision. The Retractor holds the incision open and protects it against wound contamination .

Statement of Intended Use:

The Dexterity® Pneumo Sleeve is intended to provide abdominal access for the surgeon's hand while preserving pneumoperitoneum during laparoscopic surgery.

The Protector® Retractor is intended to provide incision retraction and to protect it against wound contamination during both laparoscopic and open surgery.

This device allows the surgeon to use his hand while performing corrective surgery in the abdominal cavity under pneumoperitoneum.

Summary of Technical Characteristics of this Device Compared to the Predicate Device:

The Dexterity® Pneumo Sleeve and Protector® Retractor are technically the same as the Trocar/Cannula (one of the predicate devices) in that the Trocar/Cannula

{2}------------------------------------------------

allows access to the abdominal cavity for laparoscopic instruments and the Dexterity® Pneumo Sleeve and Protector® Retractor allow access to the abdominal cavity for the surgeon's hand. The Trocar/Cannula is made out of stainless steel or plastic while the Dexterity® Pneumo Sleeve and Protector® Retractor are made out of a plastic film, flexible plastic and an acrylate adhesive.

The Protector® Retractor component of the Dexterity® Pneumo Sleeve and Protector® Retractor is the "Protector® Retractor" shown as the second predicate device above. The material making up the sleeve of the Protector® Retractor is the same material as used to make the Dexterity® Pneumo Sleeve.

Summary of Performance Data:

Non-clinical Test Results:

Bench testing was performed on each component of the Dexterity® Pneumo Sleeve and Protector® Retractor to assure that they would function as intended during use.

Based on the data collected in the bench testing, it was concluded that all of the materials were proper for their intended use.

Animal Test Results:

Functional testing on pigs:

Several studies were accomplished using pigs to further verify the function of the device and its components. The results of these studies were effective in fine-tuning the design and working out surgeons' preferences.

Clinical Results:

Two clinical studies were conducted with the Dexterity® Pneumo Sleeve. An initial pilot study involving ten surgeons and forty patients validated the operating principles of the device and allowed surgeons to gain additional device experience. As a result of this original study, minor device changes were implemented in the device design.

{3}------------------------------------------------

A second, pivotal clinical study comparing extended laparoscopic surgery using the Dexterity® Pneumo Sleeve and Protector® Retractor with routine laparoscopic surgery for colon resections, was conducted at 10 clinical centers in the United States and Europe. Safety and effectiveness endpoints considered in this trial included: overall procedure time, number of instrument changes required, conversion rates, i.e., the number of times a case could not be completed laparoscopically and had to be converted to an open procedure, direct operation costs, time for patients to return to normal activity, morbidity and mortality. The results of this prospective, randomized, controlled trial confirmed the safety and effectiveness of the Dexterity® Pneumo Sleeve Set for its intended use as an adjunct to laparoscopic surgery.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.