K Number
K962129
Date Cleared
1997-04-15

(316 days)

Product Code
Regulation Number
870.1210
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A continuous flush device is an accessory to a catheter-transducer system that permits a continuous flow to maintain catheter patency.

Device Description

The flush device is comprised of seven small components which, when ultrasonically welded together, constitute the finished product. The device is supplied in two configurations; adult and neonate. The adult configuration differs from the neonate configuration by flow rate and color. The adult flow specification is a nominal 3 milliliters per hour (ml/hr) with a static head pressure of 250 millimeters of mercury (mmHg). The neonate flow specification is a nominal 30 ml/hr at 250 mmHg static head pressure. The device has a fast flush feature which can be momentarily actuated by squeezing the tabs on either side of the device. The fast flush feature is used to aid in filling and debubbling the pressure monitoring lines. The device has the capability of creating a momentary pressure spike to verify the frequency response of the pressure monitoring system. The pressure spike is created by pulling up and releasing (snapping) the pull tab on top of the device. The device incorporates a pressure relief capability to provide over-pressure protection to the transducer. At pressures greater than 600 mmHg (11.6 psi), the silastic control component stretches and allows fluid to by-pass the flow restrictor thereby providing over-pressure protection.

AI/ML Overview

The provided text describes a 510(k) summary for a "Continuous Flush Device" by Merit Medical Systems, comparing it to predicate devices. It focuses on device specifications and functionalities rather than a study proving the device meets acceptance criteria related to AI or clinical effectiveness, as commonly requested in such prompts for medical AI devices.

Based on the provided text, here's a breakdown of the requested information, noting where the information is not applicable or not present:


1. Table of Acceptance Criteria and Reported Device Performance

Feature/ParameterAcceptance Criteria (Predicate Device)Merit Flush Device Performance (Reported in Submission)
Intended UseFor use in physiological pressure measurements which require continuous flow to maintain catheter patency.For use in physiological pressure measurements which require continuous flow to maintain catheter patency.
Adult Flow Rate$3 \pm 1$ cc/hr @ 300 mmHg (Spectramed) $3 \pm 1$ ml/hr (Baxter)$3 \pm 1$ ml/hr @ 250 mmHg pressure
Neonate Flow Rate30 cc/hr (Spectramed) $30 \pm 10$ ml/hr (Baxter)$30 \pm 10$ ml/hr @ 250 mmHg pressure
Fast Flush ModeYesYes
Bypass PressureUnknown bypass pressure (Spectramed) >500 mmHg (9.7 psi) (Baxter)>600 mmHg (11.6 psi)
SterilizationEthylene Oxide GasEthylene Oxide Gas
Single UseYesYes
Pyrogen FreeYesYes
PackagingTyvek™ pouchTyvek™ pouch
Step Response CapabilityYes - by bouncing lever (Spectramed) Yes - via the Snap-Tab™ (Baxter)Yes - via pull tab
Adult ColorsDifferent models for adults, neonate applications (Spectramed) Blue color (Baxter)Adult devices supplied with squeeze tabs in white, red, blue or green.
Neonate ColorsDifferent models for adults, neonate applications (Spectramed) Yellow color (Baxter)Neonate devices supplied with squeeze tabs in yellow color.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable to the provided document. The document describes a 510(k) submission for a physical medical device (flush device), not an AI or software as a medical device (SaMD). Therefore, there is no "test set" in the context of algorithm performance or data provenance. The comparison is based on device specifications and features against predicate devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable. See the explanation for point 2. The "ground truth" here would be the physical measurements of the device's performance (e.g., flow rate, pressure relief), which are established through engineering testing, not expert consensus on medical images or clinical data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable. See the explanation for point 2. Adjudication methods are relevant for expert review of data, which is not described here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable. The device is a physical medical component, not an AI or diagnostic imaging system that would involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. The device is a physical medical component, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" in this context refers to the engineering specifications and performance characteristics of the device, established through physical testing and measurement. For example, flow rates (ml/hr), bypass pressure (mmHg), and material compatibility (e.g., sterilization method). This is fundamentally different from clinical ground truth based on expert review or pathology.

8. The sample size for the training set

This information is not applicable. The device is a physical medical component, not an AI model, so there is no "training set."

9. How the ground truth for the training set was established

This information is not applicable. See the explanation for point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Merit Medical. The logo consists of a black square with three white diagonal lines inside, followed by the word "MERIT" in black and the word "MEDICAL" in white with black outlines. The logo is simple and modern, and it is likely used to represent the company's brand.

Merit Medical Systems, Summary of Safety and Effectiveness, Flush Device, Page 1

APR 1 5 1997

K962129

510(k) SUMMARY OF SAFETY and EFFECTIVENESS

Date of Summary:May 31, 1996MERIT MEDICAL
Submitter's name:Merit Medical Systems, Inc.1600 West Merit ParkwaySouth Jordan, UT 84095-2415(801) 253-1600SYSTEMS, INC.
Contact Person:Dennis (Dan) Reigle1600 WEST
Name of device:Continuous Flush DeviceMERIT PARKWAY
Class:Class IISOUTH JORDAN,
Classification name:The flush device is an accessory to an extravascularblood pressure transducer (product code 74DRS).UTAH 84095
Predicate device:Predicate devices are flush device manufactured by:
Spectramed, Inc.Critical Care Division1900 Williams BlvdOxnard, CA 93030(800) 235-5945Baxter Healthcare CorporationEdwards Critical Care Division17221 Red Hill AvenueIrvine, CA 92714(800) 424-3278801-253-1600
FAX 801-253-1651

K792141

INDICATIONS FOR USE

K873168

A continuous flush device is an accessory to a catheter-transducer system that permits a continuous flow to maintain catheter patency.

DEVICE DESCRIPTION

The flush device is comprised of seven small components which, when ultrasonically welded together, constitute the finished product. The device is supplied in two configurations; adult and neonate. The adult configuration differs from the neonate configuration by flow rate and color. The adult flow specification is a nominal 3 milliliters per hour (ml/hr) with a static head pressure of 250 millimeters of mercury (mmHg). The neonate flow specification is a nominal 30 ml/hr at 250 mmHg static head pressure.

{1}------------------------------------------------

The device has a fast flush feature which can be momentarily actuated by squeezing the tabs on either side of the device. The fast flush feature is used to aid in filling and debubbling the pressure monitoring lines.

The device has the capability of creating a momentary pressure spike to verify the frequency response of the pressure monitoring system. The pressure spike is created by pulling up and releasing (snapping) the pull tab on top of the device.

The device incorporates a pressure relief capability to provide over-pressure protection to the transducer. At pressures greater than 600 mmHg (11.6 psi), the silastic control component stretches and allows fluid to by-pass the flow restrictor thereby providing over-pressure protection.

FeatureMerit Flush DeviceSpectramed Flush DeviceBaxter Flush Device
Intended UseFor use in physiologicalpressure measurementswhich requirecontinuous flow tomaintain catheterpatency.For use in physiologicalpressure measurementswhich requirecontinuous flow tomaintain catheterpatency.For use in physiologicalpressure measurementswhich requirecontinuous flow tomaintain catheterpatency.
Adult flowrate$3\pm1$ ml/hr @ 250mmHg pressure$3\pm1$ cc/hr @ 300mmHg$3\pm1$ ml/hr (blue Snap-Tab™)
Neonate flowrate$30\pm10$ ml/hr @ 250mmHg pressure30 cc/hr$30\pm10$ ml/hr (yellowSnap-Tab™)
Fast flushmodeYesYesYes
Bypasspressure>600 mmHg (11.6 psi)Yes - unknown bypasspressureYes - >500 mmHg(9.7 psi)
SterilizationEthylene Oxide GasEthylene Oxide GasEthylene Oxide Gas
Single UseYesYesYes
Pyrogen FreeYesYesYes
PackagingTyvek™ pouchTyvek™ pouchTyvek™ pouch
StepresponsecapabilityYes - via pull tabYes - by bouncing leverYes - via the Snap-Tab™
Adult ColorsAdult devices suppliedwith squeeze tabs inwhite, red, blue orgreen.Different models foradults, neonateapplicationsBlue color
NeonateColorsNeonate devicessupplied with squeezetabs in yellow color.Different models foradults, neonateapplicationsYellow color
510kThis submissionK873168K792141

SUBSTANTIAL EQUIVALENCE COMPARISON

§ 870.1210 Continuous flush catheter.

(a)
Identification. A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.(b)
Classification. Class II (performance standards).