K Number
K961848
Date Cleared
1997-09-26

(501 days)

Product Code
Regulation Number
870.1875
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3M Littmann™ Electronic Stethoscope is a diagnostic aid and used as part of a physical assessment of a patient by a Health Care Professional. It can be used for the amplification of faint heart, lung and other body sounds as well as normal auscultation and selective frequency filtering. The extended range mode amplifies sounds in a broad frequency range, including a range higher than the traditional diaphragm mode, where some lung and heart sounds may occur, as well as providing low frequency response at a slightly reduced level compared to the bell mode.

Device Description

The 3M Littmann™ Electronic Stethoscope is a diagnostic aid and used as part of a physical assessment of a patient by a Health Care Professional. It can be used for the amplification of faint heart, lung and other body sounds as well as normal auscultation and selective frequency filtering. The extended range mode amplifies sounds in a broad frequency range, including a range higher than the traditional diaphragm mode, where some lung and heart sounds may occur, as well as providing low frequency response at a slightly reduced level compared to the bell mode.

AI/ML Overview

This 510(k) submission for the 3M Littmann™ Electronic Stethoscope does not provide the detailed acceptance criteria and study data in the format requested. The document describes non-clinical testing for frequency response and gain to establish substantial equivalence to a predicate device, but it does not specify quantitative acceptance criteria or report performance against such criteria in a structured manner.

Here's an breakdown of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance:

No explicit table of acceptance criteria or quantitative performance metrics against these criteria is provided in the document. The submission focuses on demonstrating "substantial equivalence" to a predicate device through non-clinical testing.

2. Sample Size Used for the Test Set and Data Provenance:

The document describes non-clinical testing using a "computerized audio analysis system" and a "transducer." This indicates a laboratory setup, not a clinical test set with patient data. Therefore, there's no information about:

  • Sample size: Not applicable as it's not a clinical test involving a "test set" of patient data.
  • Data provenance: Not applicable. The testing is described as lab-based.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Since this was a non-clinical, lab-based test for frequency response and gain, there were no human "experts" establishing ground truth in the clinical sense (e.g., radiologists interpreting images). The ground truth for the frequency response test would be the known input frequencies and amplitudes generated by the swept frequency oscillator.

4. Adjudication Method for the Test Set:

Not applicable. There was no human interpretation or adjudication involved, as it was a non-clinical, instrumental test.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, an MRMC comparative effectiveness study was not done. The testing described is non-clinical, focusing on the device's physical properties (frequency response and gain) compared to a predicate device. It does not involve human readers using the device with or without AI assistance.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop):

The device itself is an electronic stethoscope, designed to be used by a human healthcare professional. The "non-clinical testing" described is effectively a standalone performance test of the device's technical specifications (frequency response and gain) in a controlled lab environment. However, the document doesn't present it as an "algorithm only" study in the modern sense of AI device evaluation. It's a characterization of the physical device.

7. Type of Ground Truth Used:

The ground truth for the non-clinical testing was the known, controlled audio frequencies and amplitudes generated by the "frequency oscillator" and presented to the stethoscope via a transducer. This is objective, physical measurement data, not expert consensus, pathology, or outcomes data.

8. Sample Size for the Training Set:

Not applicable. This device is an electronic stethoscope, not an AI/ML algorithm that requires a "training set" in the context of machine learning model development. The document describes a physical product.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for an AI/ML algorithm.


Summary of what is available from the provided text:

The provided 510(k) summary focuses on demonstrating substantial equivalence of the 3M Littmann™ Electronic Stethoscope to a predicate electronic stethoscope through non-clinical testing of its physical characteristics (frequency response and gain).

  • Acceptance Criteria (Implicit): The implicit acceptance criteria were that the frequency response and gain of the new device should be "substantially equivalent" to the predicate, with intentional differences designed for improvement (e.g., improved bell mode, reduced noise in diaphragm mode, reduced masking in extended range mode).
  • Study: A non-clinical study using a computerized audio analysis system.
    • A frequency oscillator swept frequencies from 10 to 5000 Hz.
    • The stethoscope was placed on a transducer receiving these frequencies.
    • A calibrated microphone in the eartip measured the stethoscope's response.
    • The computer recorded and stored this response.
    • This was performed for each of the three frequency response modes.
  • Conclusion: The manufacturer concluded that the device performs "as well as, and in many cases substantially better than the predicate electronic stethoscope" based on these measurements, justifying the intentional design differences.

{0}------------------------------------------------

رن 10007

510(k) Summary

3M Littmann™ Electronic Stethoscope

  • Name and address of Device Manufacturer submitting 510(k) Notification: 1.
    3M 3M Health Care 3M Center St. Paul, MN 55144-1000

  • Regulatory Correspondent of Device Manufacturer: 2. Linda Johnsen, Senior Requlatory Affairs Associate 612 737- 4376

  • Date Summary was prepared: May 10, 1996 3.

  • Name of Device: 3M Littmann™ Electronic Stethoscope 4.

  • Predicate Device to which 3M is claiming Substantial Equivalence: 5. 3M Littmann Brand Electronic Stethoscope 510(k) K771653.

6. Intended Use:

The 3M Littmann™ Electronic Stethoscope is a diagnostic aid and used as part of a physical assessment of a patient by a Health Care Professional. It can be used for the amplification of faint heart, lung and other body sounds as well as normal auscultation and selective frequency filtering. The extended range mode amplifies sounds in a broad frequency range, including a range higher than the traditional diaphragm mode, where some lung and heart sounds may occur, as well as providing low frequency response at a slightly reduced level compared to the bell mode.

Non Clinical Testing Brief Description: 8.

Testing to determine that the frequency response and gain are substantially equivalent to the predicate electronic stethoscope was performed with a computerized audio analysis system that sweeps a frequency oscillator from 10 to 5000 Hz. This frequency sweep is presented to a transducer on which the stethoscope is placed. The computer measures the response from the stethoscope with a small calibrated microphone placed in the eartip. The signal from the microphone is fed back to the computer where it is recorded, stored and later can be printed out. This test was done in each of the three frequency responses modes.

റ്റ. Conclusive Statement:

While there are some differences in the three modes, these differences are intentional and designed to beneficially provide improved frequency response in the bell mode, reduce noise pick-up in the diaphragm mode, and reduce masking of the higher frequency sounds in the extended range mode. In this way, we have determined that the present Littmann Electronic Stethoscope performs as well as, and in many cases substantially better than the predicate electronic stethoscope.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle or other bird-like figure with three wing-like shapes.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 26 1997

Ms. Linda Johnsen 3M Health Care 3M Center, Building 275-3E-08 St. Paul, Minnesota 55144-1000

Re: K961848 3M Littmann Electronic Stethoscope Requlatory Class: II (two) Product Code: 74 DQD Dated: July 2, 1997 Received: July 3, 1997

Dear Ms. Johnsen:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2 - Ms. Linda Johnsen

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) 3M Littmann™ Electronic Stethoscope

510(k) Number (if known):

Device Name:__________________________________________________________________________________________________________________________________________________________________

Indications For Use:

Intended Use:

The 3M Littmann™ Electronic Stethoscope is a diagnostic aid and used as part of a physical assessment of a patient by a Health Care Professional. It can be used for the amplification of faint heart, lung and other body sounds as well as normal auscultation and selective frequency filtering. The extended rage mode amplifies sound over a broad range of frequencies, including the range where some lung and prosthetic heart valve sounds occur.

Copy of labeling which includes instructions for use and maintenance is included in attachment 2 to the 510(k) submission.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

chat i'do La AAC

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number K961848

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 870.1875 Stethoscope.

(a)
Manual stethoscope —(1)Identification. A manual stethoscope is a mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.(b)
Electronic stethoscope —(1)Identification. An electronic stethoscope is an electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.(2)
Classification. Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.