K Number
K961490
Device Name
ELECSYS T4 ASSAY
Date Cleared
1996-06-13

(56 days)

Product Code
Regulation Number
862.1700
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunoassay for the in vitro quantitative determination of total thyroxine in human serum and plasma.

Device Description

The Elecsys® test principle is based on competition principle. Total duration of assay: 18 minutes (37°C).
• 1st incubation (9 minutes): Sample (15 µL) and a specific anti-T4 antibody labeled with a ruthenium complex (75 uL).
• 2nd incubation (9 minutes): After addition of biotinylated T4 (75 uL) and streptavidin-coated microparticles (35 uL), the still-free binding sites of the labeled antibody become occupied, with formation of an antibody-hapten complex. The entire complex is bound to the solid phase via interaction of biotin and streptavidin.
•The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier (0.4 second read frame).
•Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent bar code.

AI/ML Overview

The provided document is a 510(k) summary for the Elecsys® T4 Assay, an in vitro diagnostic device for the quantitative determination of total thyroxine in human serum and plasma. This type of document is used to demonstrate substantial equivalence to a predicate device, not necessarily to set and meet acceptance criteria in the same way a clinical study for a novel therapeutic might.

However, we can infer "acceptance criteria" from the comparative performance tables against the predicate device, the Enzymun T4 Assay (K901346). The study aims to show that the Elecsys® T4 Assay performs comparably or better than the predicate device across various metrics.

Here's an attempt to structure the information based on your request, using the available data:


Acceptance Criteria and Device Performance Study for Elecsys® T4 Assay

The study presented in the 510(k) summary compares the performance of the Elecsys® T4 Assay to its predicate device, the Enzymun T4 Assay (K901346). The "acceptance criteria" are implied by the performance characteristics of the predicate device, with the new device aiming to demonstrate comparable or improved performance.

1. Table of Acceptance Criteria (Implied) and Reported Device Performance

FeatureImplied Acceptance Criteria (Predicate: Enzymun T4 Assay)Reported Device Performance (Elecsys® T4 Assay)
PrecisionModified NCCLS (nmol/L):
(Within-Run %CV)Control 1: 2.3%Control 1: 2.3%
Control 2: 2.0%Control 2: 2.0%
Low: 2.3%Low: 2.3%
Mid: 2.4%Mid: 2.4%
High: 2.0%High: 2.0%
(Total %CV)Control 1: 3.3%Control 1: 3.3%
Control 2: 2.7%Control 2: 2.7%
Low: 3.1%Low: 3.1%
Mid: 2.9%Mid: 2.9%
High: 2.5%High: 2.5%
Lower Detection Limit0.6 µg/dL (7.7 nmol/L)0.23 µg/dL (3 nmol/L)
Linearity0.6 - 25.0 µg/dL (±10% deviation)0.23 - 24.86 µg/dL (±10% deviation)
Method ComparisonLeast Squares (vs Enzymun-Test® T4):Least Squares (vs Enzymun-Test® T4):
(Correlation)r = 0.996r = 0.947
Interfering Substances
BilirubinNo interference at 64.5 mg/dLNo interference at 25 mg/dL
HemoglobinNo interference at 1 g/dLNo interference at 1 g/dL
LipemiaNo interference at 1250 mg/dLNo interference at 1500 mg/dL
BiotinNo interference at 200 ng/mLNo interference at 30 ng/mL
Specificity (% Cross-reactivity)
L-T4100100
D-T4100100
L-T33.51.53
D-T32.91.38
3-iodo-L-tyrosine<0.10.002
3,5-diiodo-L-tyrosine<0.10.01
tetraiodo-thyroacetic acid2038.5

Analysis of Performance vs. Implied Criteria:

  • Precision: The Elecsys® T4 Assay achieved identical %CVs for both Within-Run and Total precision at comparable control levels as the predicate device.
  • Lower Detection Limit: The Elecsys® T4 Assay demonstrated a significantly lower detection limit (0.23 µg/dL) compared to the predicate (0.6 µg/dL), indicating improved sensitivity.
  • Linearity: The Elecsys® T4 Assay showed a comparable linear range, slightly wider at the lower end, with the same ±10% deviation tolerance.
  • Method Comparison: While both devices were compared against the Enzymun-Test® T4, the Elecsys® T4 had a correlation coefficient (r=0.947) against the predicate from a larger sample size (N=196), which is lower than the predicate's own correlation (r=0.996) against the same Enzymun-Test® T4 from a smaller sample size (N=78). This suggests the predicate had very high agreement with itself when tested, while the Elecsys® T4, though showing good correlation, was not as tightly correlated with the predicate as the predicate was with itself.
  • Interfering Substances: For most interfering substances, the Elecsys® T4 showed comparable or slightly different non-interference concentrations. Notably, it showed lower tolerance for Bilirubin and Biotin compared to the predicate, but higher tolerance for Lipemia.
  • Specificity (% Cross-reactivity): The Elecsys® T4 showed similar cross-reactivity for L-T4 and D-T4. It demonstrated lower cross-reactivity for L-T3, D-T3, 3-iodo-L-tyrosine, and 3,5-diiodo-L-tyrosine, indicating improved specificity for T4 over these related compounds. However, it had a higher cross-reactivity for tetraiodo-thyroacetic acid.

2. Sample Size and Data Provenance

  • Precision:
    • For the Elecsys® T4 Assay: N=60 for Control 1 and Control 2 at different levels.
    • For the Enzymun T4 Assay: N=120 for Low and Mid levels, N=118 for High level.
  • Method Comparison:
    • Elecsys® T4 Assay: N=196 samples.
    • Enzymun T4 Assay (predicate's self-comparison): N=78 samples.
  • Data Provenance: Not explicitly stated. Given it's a 510(k) submission by Boehringer Mannheim Corporation, the data would likely originate from internal studies or contract research organizations involved in the device's development and validation, likely in the US or Europe. It's a retrospective analysis of collected samples for performance evaluation.

3. Number of Experts and Qualifications for Ground Truth

Not applicable for this type of in vitro diagnostic device study. The "ground truth" for quantitative assays is typically established by reference methods or validated comparative assays, not by expert consensus on interpretations.

4. Adjudication Method for the Test Set

Not applicable. This is a quantitative assay comparison, not an imaging or clinical decision-making study requiring adjudication of interpretations.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

Not applicable. This is an in vitro diagnostic device evaluating analytical performance, not a study involving human readers' interpretation (with or without AI assistance) of diagnostic cases.

6. Standalone Performance Study

Yes, the data provided for the "Elecsys® T4" column represents the standalone performance of the algorithm (analytical assay) in determining total thyroxine levels. The comparison made to the "Enzymun T4" column establishes its performance relative to a predicate device, which is a form of standalone evaluation against an accepted standard.

7. Type of Ground Truth Used

The "ground truth" for analytical performance studies like this is established by:

  • Reference Methods: For assay standardization, "GC/MS Gas Chromatography/Mass Spectrometry" is mentioned for both devices, indicating a high-accuracy reference method.
  • Comparative Assay: The predicate device itself (Enzymun T4 Assay) serves as a comparative "ground truth" for demonstrating substantial equivalence, meaning the new device's results align closely with an established and legally marketed device. The "Enzymun-Test® T4" mentioned in method comparison is likely the predicate or a closely related assay used as the comparative standard.

8. Sample Size for the Training Set

No specific "training set" is mentioned in the context of machine learning or AI models, as this device is a traditional immunoassay. The development and optimization of the assay would involve various experiments and optimizations, but not a distinct "training set" in the AI sense.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there isn't a "training set" in the context of an AI/ML model for this device. The assay development would rely on chemical principles, reagent optimization, and calibration against known standards (e.g., those standardized by GC/MS).

{0}------------------------------------------------

JUN 1 3 1896

用 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

Image /page/0/Picture/1 description: The image contains the text "K961490 Diagnostics". The text "K961490" is written in a larger, bolder font than the word "Diagnostics". There is also a logo to the left of the word "Diagnostics", but it is too blurry to make out any details.

510(k) Summary

According to the requirements of 21 CFR 807.92, the following information Introduction provides sufficient detail to understand the basis for a determination of substantial equivalence. Boehringer Mannheim Corporation 1. Submitter 2400 Bisso Lane name, P.O. Box 4117 address, Concord, CA 94524-4117 contact (510) 674 - 0690, extension 8415 Contact Person: Mary Koning Date Prepared: April 9, 1996 2. Proprietary name: Elecsys® T4 Assay Device name Common name: Electrochemiluminescence assay for the determination of total thyroxine (T4). Classification name: Radioimmunoassay, Total Thyroxine 3. We claim substantial equivalence to the Enzymun T4 Assay( K901346). Predicate device 4. The Elecsys® test principle is based on competition principle. Total duration Device of assay: 18 minutes (37°C). Description · 1st incubation (9 minutes): Sample (15 µL) and a specific anti-T4 antibody labeled with a ruthenium complex (75 uL). ·2nd incubation (9 minutes): After addition of biotinylated T4 (75 uL) and streptavidin-coated microparticles (35 uL), the still-free binding sites of the labeled antibody become occupied, with formation of an antibody-hapten complex. The entire complex is bound to the solid phase via interaction of biotin and streptavidin. Continued on next page

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a black square on the left and the word "Diagnostics" on the right. The word is written in a serif font. Below the word "Diagnostics" are two lines of text in a different language.

,

4.DeviceDescription•The reaction mixture is aspirated into the measuring cell where themicroparticles are magnetically captured onto the surface of the electrode.Unbound substances are then removed with ProCell. Application of a voltageto the electrode then induces chemiluminescent emission which is measuredby a photomultiplier (0.4 second read frame).•Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided viathe reagent bar code.
5.Intended useImmunoassay for the in vitro quantitative determination of total thyroxine inhuman serum and plasma.
6.Comparisonto predicatedeviceThe Boehringer Mannheim Elecsys® T4 Assay is substantially equivalent toother products in commercial distribution intended for similar use. Mostnotably it is substantially equivalent to the currently marketed Enzymun T4Assay (K901346).
The following table compares the Elecsys® T4 Assay with the predicatedevice, Enzymun T4 Assay. Specific data on the performance of the testhave been incorporated into the draft labeling in attachment 5. Labeling forthe predicate device is provided in attachment 6.
Similarities:
•Intended Use: Immunoassay for the in vitro quantitative determinationof total thyroxine
•Sample type: Serum and plasma
•Antibody: Polyclonal Sheep anti-T4 antibodies
•Solid phase binding principle: Streptavidin/Biotin
•Assay standardization: GC/MS Gas Chromatography/MassSpectrometry
FeatureElecsys® T4Enzymun T4
Reaction testprincipleElectrochemiluminescenceELISA/1-step sandwich assayusing streptavidin technology
InstrumentrequiredElecsys® 2010ES 300
CalibrationStabilityA calibration is recommendedevery 7 days if kit is notconsumed; 4 weeks with samereagent lot if reagent isconsumed within 7 days.A calibration is required everyrun

Continued on next page

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "Diagnostics" in a serif font. The word is in black and is on a white background. To the left of the word is a black square.

في تعدد القرب المقابل المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخ

Comparison to predicate device cont.

Differences:

Performance Characteristics:

FeatureElecsys® T4Enzymun T4
PrecisionModified NCCLS (nmol/L):Modified NCCLS (µg/dL):
LevelSampleControl 1Control 2LowMidHigh
N606060120120118
Within-Run33.39113.10180.554.556.7915.49
%CV4.72.32.02.32.42.0
Total33.39113.10180.554.556.7915.49
%CV6.93.32.73.12.92.5
Modified NCCLS (µg/dL):
SampleControl 1Control 2
N606060
Within-Run2.608.8214.08
%CV4.72.32.0
Total2.608.8214.08
%CV6.93.32.7

Continued on next page

.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains a black square on the left side and the word "Diagnostics" on the right side. The word is written in a serif font and is slightly slanted. The black square appears to have some sort of pattern inside of it, but it is difficult to make out the details.

ﺍﻟﻤﺴﺎﻋﺪ ﺍﻟﻤ

Comparison
to predicate device, (cont.)

Performance Characteristics:

FeatureElecsys® T4Enzymun T4
LowerDetectionLimit0.23 µg/dL(3 nmol/L)0.6 µg/dL(7.7 nmol/L)
Linearity0.23 - 24.86 µg/dL (3.0 - 320 nmol/L) (with a deviation from a linear line of ±10%)0.6 - 25.0 µg/dL (with a deviation from a linear line of ±10%)
MethodComparisonVs Enzymun-Test® T4Least Squares$y =1.02x + 0.85$$r=0.947$$SEE =5.39$$N=196$Passing/Bablok$y =0.93x + 4.55$$r=0.947$$SEE =5.39$$N=196$Vs Enzymun-Test® T4Least Squares$y =0.90x - 0.45$$r=0.996$$SEE =0.648$$N=78$

Continued on next page

. .

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Diagnostics" in a serif font. The word is written in black ink on a white background. To the left of the word is a black rectangle with some white markings.

Performance Characteristics:

: 【: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :

ﺮﺏ ﺑﺎﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻘ

Comparison

to predicate
device, (cont.)

FeatureElecsys® T4Enzymun T4
InterferingsubstancesNo interference at:No interference at:
Bilirubin25 mg/dL64.5 mg/dL
Hemoglobin1 g/dL1 g/dL
Lipemia1500 mg/dL1250 mg/dL
Biotin30 ng/mL200 ng/mL
Specificity% Cross-reactivity% Cross-reactivity
L-T4100100
D-T4100100
L-T31.533.5
D-T31.382.9
3-iodo-L-tyrosine0.002<0.1
3,5-diiodo-L-tyrosine0.01<0.1
tetraiodo-thyroaceticacid38.520

。 ·

§ 862.1700 Total thyroxine test system.

(a)
Identification. A total thyroxine test system is a device intended to measure total (free and protein bound) thyroxine (thyroid hormone) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.