(442 days)
The Fresenius On Line Clearance Monitor is a module that is incorporated into the Fresenius 2008H hemodialysis equipment to be used during a dialysis treatment to determine dialyzer clearance efficiency using changes in conductivity of the dialysate solution.
The Fresenius On Line Clearance Monitor is intended to be used as a module with the Fresenius 2008H hemodialysis machine for determination of dialysis treatment urea clearance efficiency using Sodium conductivity as the surrogate solute marker.
The Fresenius On Line Clearance Monitor (OLC) is a module that is incorporated into the Fresenius 2008H hemodialysis equipment. The OLC consists of hardware modifications to the 2008H dialysate fluid circuit which includes a second conductivity probe in the effluent dialysate line and software. The dialyzer clearance determination is made using changes in dialysate conductivity (Sodium concentration). The software either increases or decreases the dialysate conductivity. This change in conductivity causes a sodium concentration gradient to be formed between the blood and dialysate sides of the dialyzer. Sodium is then either diffused into the blood or out of the blood depending whether the dialysate conductivity is high or low. Accurate conductivity readings of the inlet and outlet dialysate solution can be used to calculate the flux of sodium (or change in conductivity because of changes in sodium concentration) in the dialysate. This value can then be used to calculate the dialyzer clearance of sodium. The clearance of sodium has been used as a surrogate dialyzer performance solute since the beginning of dialysis since urea and Sodium (from NaCl) pass through the dialysis membrane at essentially the same rate.
The use of the On Line Clearance Monitor is non invasive and is performed automatically by the hemodialysis equipment with no effect to the patient or action of the operator and minimizes the need for blood based urea analysis and dialyzer performance determination.
The provided text does not contain a specific table of acceptance criteria or details of a study designed to explicitly meet quantitatively defined acceptance criteria for the Fresenius On Line Clearance Monitor.
However, based on the information provided, we can infer the acceptance criteria and study as follows:
Inferred Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Primary Goal: Substantial Equivalence to predicate devices and established methods for determining dialyzer performance. Specifically: - Equivalence to equipment adaptors using conductivity for dialyzer performance (K830190, K841153) - Equivalence to manual invasive blood urea determinations. | Non-Clinical Testing: Statistically significant correlation between clearance values obtained using Sodium (conductivity) as a marker and urea. Clinical Testing: No difference in clearance determinations between the OLC (using Sodium conductivity) and identical blood-side urea measurements. |
| Operational Impact: Non-invasive, automatic operation with no effect on the patient or operator, minimizing the need for blood-based urea analysis. | The On Line Clearance Monitor is non-invasive and performed automatically by the hemodialysis equipment with no effect to the patient or action of the operator. |
| Mechanism: Utilize conductivity (Sodium concentration) to calculate dialyzer clearance. | The OLC accurately measures inlet and outlet dialysate conductivity to calculate sodium flux, which is then used to calculate dialyzer clearance. |
Study Information
-
Sample sizes used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated. The non-clinical testing involved "a number of different dialyzer membranes." The clinical testing used "a modified 2008H hemodialysis machine... clinically to evaluate dialyzer performance during dialysis." Specific patient or case numbers are not provided.
- Data Provenance: Not specified for non-clinical. For clinical testing, it was likely conducted in a clinical setting ("clinically to evaluate dialyzer performance during dialysis"), implying prospective or retrospective data collection from patients undergoing dialysis. Country of origin is not mentioned but can be inferred to be the USA, given the applicant's address.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not specified. The ground truth was established by comparing the OLC's performance to "identical blood side urea measurements" and "blood urea determinations," which are established clinical methods. It is implied that standard clinical laboratory practices and medical professionals would be involved in obtaining and interpreting these ground truth measurements, but no specific number or qualifications of experts are given for establishing the ground truth for the test set.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- No adjudication method is described. The comparison relies on direct measurements from the OLC versus established clinical methods (blood urea measurements).
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done. The device (OLC) is an automated monitor, not a diagnostic aid requiring human interpretation of output in the traditional sense of an MRMC study.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone study was primarily done. The OLC is described as functioning "automatically by the hemodialysis equipment with no effect to the patient or action of the operator." The reported "statistically significant correlation" and "no difference in the clearance determinations" are evaluations of the device's performance directly against established methods.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
-
The sample size for the training set:
- Not specified. The document does not describe a distinct "training set" in the context of machine learning model development. This device appears to be based on an algorithmic calculation from physical measurements rather than a machine learning model that would require a distinct training phase.
-
How the ground truth for the training set was established:
- As there's no mention of a traditional machine learning training set, the concept of establishing ground truth for it is not applicable here. The device's underlying principles are based on established physiological and engineering principles (sodium/urea diffusion and conductivity).
{0}------------------------------------------------
Fresenius USA, Inc. 2637 Shadelands Drive Valnut Creek, CA 94598 10) 295-0200 1) 988-1900 (Fax)
Image /page/0/Picture/1 description: The image shows the Fresenius USA logo, which is a stylized letter F. Above the logo, there is handwritten text that reads "K961465" and "pg. 1 of 3". The logo is in black and white and is positioned on the left side of the image. The handwritten text is in the upper right corner.
510 (k) Summary of Safety and Effectiveness Information for the Fresenius On Line Clearance Monitor
JUL - 3 1997
1. Applicant:
Fresenius USA, Inc 2637 Shadelands Drive Walnut Creek, CA 94598
| Contact Person: | Thomas I. FoldenDirector, Product Development |
|---|---|
| Telephone: | 1-510-295-02001-510 988-1958 (Fax) |
| Date Summary Prepared: | April 15, 1996 |
- Fresenius On Line Clearance Monitor 2. Device Name:
3. Predicate Devices
Fresenius USA, Inc. claims that the On Line Clearance Monitor is substantially equivalent to equipment adaptors for determination of dialyzer performance using conductivity that were made commercially available and found substantially equivalent under premarket notifications K830190 and K841153. Fresenius also claims substantial equivalence to manual invasive blood urea determinations of dialyzer performance.
4. Intended Use
ﻟﻤﺴﺘﻘ く
The Fresenius On Line Clearance Monitor is a module that is incorporated into the Fresenius 2008H hemodialysis equipment to be used during a dialysis treatment to determine dialyzer clearance efficiency using changes in conductivity of the dialysate solution.
K961465 - 1/3
{1}------------------------------------------------
K961465
pg. 2 of 3
Device Description 5.
The Fresenius On Line Clearance Monitor (OLC) is a module that is incorporated into the Fresenius 2008H hemodialysis equipment. The OLC consists of hardware modifications to the 2008H dialysate fluid circuit which includes a second conductivity probe in the effluent dialysate line and software. The dialyzer clearance determination is made using changes in dialysate conductivity (Sodium concentration). The software either increases or decreases the dialysate conductivity. This change in conductivity causes a sodium concentration gradient to be formed between the blood and dialysate sides of the dialyzer. Sodium is then either diffused into the blood or out of the blood depending whether the dialysate conductivity is high or low. Accurate conductivity readings of the inlet and outlet dialysate solution can be used to calculate the flux of sodium (or change in conductivity because of changes in sodium concentration) in the dialysate. This value can then be used to calculate the dialyzer clearance of sodium. The clearance of sodium has been used as a surrogate dialyzer performance solute since the beginning of dialysis since urea and Sodium (from NaCl) pass through the dialysis membrane at essentially the same rate.
The use of the On Line Clearance Monitor is non invasive and is performed automatically by the hemodialysis equipment with no effect to the patient or action of the operator and minimizes the need for blood based urea analysis and dialyzer performance determination.
Comparison of Technological Characteristics and Statement of 6. Substantial Equivalence
The Fresenius On Line Clearance Monitor utilizes conductivity as a marker to determine dialyzer performance in the same manner as the Colorado Medical equipment adaptor previously found substantially equivalent under premarket notifications K 830190 and K841153 by monitoring the inlet and outlet dialysate conductivities with standard conductivity probes and calculating a clearance value for the dialyzer. The OLC also is equivalent to blood urea clearance determinations currently used clinically to rate dialyzer efficiency.
Testing:
Non Clinical Testing: The use of Sodium (conductivity) as a clearance marker was compared to urea using a number of different dialyzer membranes and found that there was statistically significant correlation that the clearance values obtained were the same.
K961465 - 2/3
{2}------------------------------------------------
K961465
pg. 3 of 3
Clinical Testing: A modified 2008H hemodialysis machine was used clinically to evaluate dialyzer performance during dialysis comparing clearance performance from the OLC to identical blood side urea measurements. The results of the testing indicated that there was no difference in the clearance determinations between the two methods.
K961465 - 3/3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The seal is presented in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Tom Folden Director, Product Development Fresenius USA 2637 Shadelands Drive Walnut Creek, California 94598
JUL - 3 1997
Re: K961465 2008H On Line Clearance Monitor Dated: April 2, 1997 Received: April 4, 1997 Regulatory class: III 21 CFR §876.5860/Product code: 78 KDI
Dear Mr. Folden:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the inclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If vour device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/odrh/dsmamain.html"
Sincerely vours
H.J.Liau Yin
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE
510(k) Number: K961465
Device Name: . Fresenius 2008H On Line Clearance Monitor . . . . . . . . . . .
Indications for Use:
The Fresenius On Line Clearance Monitor is intended to be used as a module with the Fresenius 2008H hemodialysis machine for determination of dialysis treatment urea clearance efficiency using Sodium conductivity as the surrogate solute marker.
Robert R. Ratliff/
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices K961465 510(k) Number_
L Prescription Use _ (Per 21 CFR 801.109)
Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________
§ 876.5860 High permeability hemodialysis system.
(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”