K Number
K960451

Validate with FDA (Live)

Manufacturer
Date Cleared
1996-04-18

(79 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Reduction and stabilization of grades III & IV spondylolisthesis of the fifth lumbar to first sacral vertebral interspace only. The Webb-Morley Spine System is for temporary use and must be accompanied by the application of autogenous source bone graft only. Once solid fusion is achieved, surgical removal of the device system is recommended. Planned removal of the device occurs within one year of the original surgery in adults and within two years of reaching skeletal maturity for adolescents.

Device Description

The Webb-Morley Spine System is made from surgical implant grade Stainless steel meeting ASTM F-138-86 requirements. The Webb-Morley Spine System has the same indications for use as the recently cleared TSRH® and PLSA systems. The use of the Webb-Morley Spine System as a pedicle screw system involves screw attachment at the sacrum, the L5 vertebral pedicles, the L4 vertebral pedicles and in isolated cases, the L3 vertebral pedicles. The inclusion of additional pedicular levels may sometimes be necessary in order to reduce the spondylolisthetic deformity at L5 - S1 and maintain adequate correction. The Screws of this system are limited to L3-S1 or iliac screw fixation.

AI/ML Overview

Acceptance Criteria and Device Performance for Webb-Morley Spine System

This document summarizes the acceptance criteria and study proving the Webb-Morley Spine System meets these criteria, based on the provided text.

1. Table of Acceptance Criteria and Reported Device Performance

The Webb-Morley Spine System's acceptance criteria are implicitly defined by its claim of "Substantial Equivalence" to two predicate devices: the TSRH® Posterior Spine System (K932029) and the Posterior Lumbar Sacral Attachment (PLSA) Spine System (K945064). The acceptance criteria, therefore, are that the Webb-Morley system performs equivalently or superiorly to these predicate devices in terms of mechanical strength, fatigue, and corrosion resistance, and shares the same indications for use, contraindications, and other design features.

Acceptance Criterion (Implicit, based on Substantial Equivalence claims)Reported Device Performance (Webb-Morley Spine System)
Functional Equivalence/Superiority in Strength and Fatigue"Static testing showed that the Webb-Morley Spine System has sufficient strength to perform in a functionally equivalent manner to the comparison device systems. The Webb-Morley Spine System demonstrates excellent fatigue and strength characteristics. Based on the testing, the Webb-Morley Spine System is equal or superior in performance to the TSRH® Spine System."
Material CompositionMade from surgical implant grade Stainless steel (ASTM F-138-86 requirements), same as or equivalent to predicate devices. (TSRH® uses Titanium Alloy & 316 LVM Stainless Steel, PLSA uses Titanium Alloy. Webb-Morley specifically uses 316 LVM stainless steel as per Feature Comparison Table, demonstrating equivalence to the stainless steel component of TSRH®).
Indications for Use"Reduction and stabilization of grades III & IV spondylolisthesis of the fifth lumbar to first sacral vertebral interspace only." - Stated to have the "same indications for use as the recently cleared TSRH® and PLSA systems."
ContraindicationsSame general contraindications as the predicate devices (spinal infection, morbid obesity, mental illness, pregnancy, metal sensitivity, inadequate tissue coverage, open wounds local to the operative area).
Surgical ApproachPosterior Surgical Approach (same as predicate devices)
Design Elements (e.g., rods, pedicle screws, hooks, transverse interconnection)"Design of the rods and pedicle screws... do not represent a significant departure from long established designs... The Webb-Morley Spine System utilizes identical implant materials... The design features... are equivalent to those offered by the TSRH® and PLSA systems." "Components of the comparable systems are all made from the same or equivalent implant grade materials." Specific details in the Feature Comparison Table corroborate this for hooks, rods, screws, and transverse interconnection.
Sterilization MethodSteam autoclave sterilization (recommended as gravity steam sterilization for 45 minutes at 121°C), with a Sterility Assurance Level (SAL) of 10⁻⁶. "Essentially the same as that used for the TSRH® and the PLSA systems."
Compliance with StandardsMeets BS-7252, ISO-5832-1 & ASTM F-138-86 (comparable or equivalent to predicate devices ASTM F-138-86, Gr. 2 & F-136-92).
Performance in Clinical Context"Performance of the Webb-Morley Spine System for severe spondylolisthesis of L5/S1 is expected to be comparable to the TSRH® and PLSA devices."

2. Sample size used for the test set and the data provenance

The study primarily relies on physical testing of the device components and a comparison against predicate devices.

  • Test Set Sample Size: Not explicitly stated as a number of devices or components. The testing appears to have been conducted on an unspecified number of Webb-Morley Spine System implants.
  • Data Provenance: The data comes from in-house engineering review and commercial laboratories that performed static and fatigue testing. This is likely prospective testing specifically conducted for this submission. The origin of the data is the company (Electro-Biology, Inc.) and the commercial laboratories they contracted. No patient data or clinical trial data is mentioned for this specific device. The clinical experience listed for the Webb-Morley system is "European," suggesting that some observational data or initial use might have occurred there, but this is not part of the described "physical testing."

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The concept of "ground truth" as typically applied to diagnostic algorithms (e.g., expert consensus on medical images) is not applicable here. The "ground truth" for this device's performance acceptance is based on:

  • Engineering Standards: Established ASTM, ISO, and BS standards for implant materials and design.
  • Performance of Predicate Devices: The established safe and effective performance of the TSRH® and PLSA systems.
  • Qualified Commercial Laboratories: These labs performed the physical testing (static and fatigue) to objective engineering criteria. Their qualifications are mentioned as "qualified commercial laboratories," but specific expert numbers or individual qualifications (e.g., "bioengineers with 10 years of experience") are not provided.

4. Adjudication method for the test set

Not applicable. The evaluation is based on objective physical testing results against established engineering criteria and direct comparison with predicate device performance, not on subjective expert review requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted and is not applicable to this type of device (spinal implant system). This device is a physical medical implant, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm performance study was not conducted. This device is a physical medical implant.

7. The type of ground truth used

The ground truth for this submission is based on:

  • Objective Engineering Test Results: Data from static and fatigue testing.
  • Compliance with Recognized Standards: Adherence to ASTM, ISO, and BS material and performance standards.
  • Predicate Device Performance: The established safety and effectiveness of the legally marketed predicate devices (TSRH® and PLSA systems), against which the Webb-Morley system demonstrated equivalence or superiority.

8. The sample size for the training set

Not applicable. This submission concerns a physical medical device, not an AI or machine learning algorithm requiring a training set.

9. How the ground truth for the training set was established

Not applicable. This submission concerns a physical medical device, not an AI or machine learning algorithm.

{0}------------------------------------------------

K960451

APR 1 8 1996

Summary of Safety and Effectiveness Information (releasable upon request only)

Regulatory Authority:

Safe Medical Devices Act of 1990, 21 CFR 807.92

Company Name/Contact

  • Electro-Biology, Inc. Company: 6 Upper Pond Road Parsippany, NJ 07054-1079
  • Ms. Sharon A. Starowicz Contact: Director, Regulatory Affairs Electro-Biology, Inc. 6 Upper Pond Road Parsippany, NJ 07054-1079 (201) 299-9022 (201) 299-0906 - Fax

Establishment Registration Number: 2242816

Device Name:

Trade Name:

Webb-Morley Spine System

Common Name(s): Pedicular Spine System

Classification Spondylolisthesis Spinal Fixation Device System Name(s):

Classification Code(s): 87MNH

Substantially Equivalent Device(s)

  • TSRH® Posterior Spine System 1. Sofamor/Danek K932029
    1. Posterior Lumbar Sacral Attachment (PLSA) Spine System Advanced Spine Fixation Systems K945064

{1}------------------------------------------------

Device Description

Background. This document summarizes safety and effectiveness information about the Webb-Morley Spine System. The full information is contained in the 510(k) Premarket Notification submitted to FDA in support of EBIs' request for a finding of Substantial Equivalence.

The Webb-Morley Spine System is made from surgical implant grade Stainless Stainless steel of this type is steel meeting ASTM F-138-86 requirements. particularly well suited for use as surgical implants. In addition to excellent strength, fatigue and corrosion resistance characteristics, stainless steel has a very long history of use in the human body as an implant material.

The Webb-Morley Spine System has the same indications for use as the recently cleared TSRH® and PLSA systems. This indication for use is:

Reduction and stabilization of grades III & IV spondylolisthesis of the fifth lumbar to first sacral vertebral interspace only. The Webb-Morley Spine System is for temporary use and must be accompanied by the application of autogenous source bone graft only. Once solid fusion is achieved, surgical removal of the device system is recommended. Planned removal of the device occurs within one year of the original surgery in adults and within two years of reaching skeletal maturity for adolescents.

The use of the Webb-Morley Spine System as a pedicle screw system involves screw attachment at the sacrum, the L5 vertebral pedicles, the L4 vertebral pedicles and in isolated cases, the L3 vertebral pedicles. The inclusion of additional pedicular levels may sometimes be necessary in order to reduce the spondylolisthetic deformity at L5 - S1 and maintain adequate correction.

The Screws of this system are limited to L3-S1 or iliac screw fixation.

Contraindications. The existence of the following diseases or medical conditions generally excludes candidates from treatment with this type of spinal implant system.

  • A. Patients with spinal infection or inflammation
  • B. Morbid obesity
  • C. Mental illness, alcoholism, drug abuse
  • D. Pregnancy
  • E. Metal sensitivity or foreign body sensitivity
  • F. Patients with inadequate tissue coverage at the operative site
  • G. Open wounds local to the operative area

As an additional precaution, the Webb-Morley Spine System should only be implanted by surgeons fully experienced in the use of such implants and the required specialized spinal surgery techniques. Instructional course demonstrations and/or bio-skills workshops demonstrating the use of the Webb-Morley Spine

{2}------------------------------------------------

System implants may be offered following a finding of substantial equivalence by the FDA.

A comprehensive list of indications, contraindications, adverse effects, warnings and precautions may be reviewed in the package insert supplied with the system. Only the package insert is to be considered definitive or accurate in terms of system applications, uses or restrictions.

Packaging:

All packaging is commercially available industry standard items and is sufficient in design and material quality to provide durable identification and protection from physical damage during transportation and storage. The type of packaging used is essentially the same as that used for the TSRH® and the PLSA systems.

Sterilization/Re-sterilization:

All instruments and implants are supplied Non-Sterile. Device packages clearly indicate the sterility status of the contained component. Steam autoclave sterilization is the only processing method recommended for the Webb-Morley Spine System. A cleaning and validated sterilization process recommendation is located in the product package insert. Please obtain and review the package insert before using the system. The recommended sterilization method, time and temperature for the implants is gravity steam sterilization for 45 minutes at 1219 C (250° F).

The Sterility Assurance Level (SAL) of the recommended sterilization cycle is 10° (SAL 10°). Validation of the recommended cycle has been conducted by qualified The validation method used is known as the overkill commercial laboratories. method.

Summary of Physical Testing:

Device and system testing consisted of static testing and comparison fatigue testing against the TSRH® system. The results of both modes of testing revealed that the Webb-Morley Spine System is suitable for the cleared indications and the anticipated physiologic loads imposed on the system. Comparison test values for the Webb-Morley system and the TSRH® system implants are equivalent.

Static testing showed that the Webb-Morley Spine System has sufficient strength to perform in a functionally equivalent manner to the comparison device systems. The Webb-Morley Spine System demonstrates excellent fatigue and strength characteristics. Based on the testing, the Webb-Morley Spine System is equal or superior in performance to the TSRH® Spine System.

{3}------------------------------------------------

Equivalence:

The table concluding this section summarizes available information for the comparison devices. The Substantially Equivalent devices and the Webb-Morley Spine System are all posterior surgical approach systems. All are used to treat similar or the same conditions. All have essentially the same cautions and contraindications for use. All have the same labeling warnings imposed by FDA. All are basic spinal rod, transverse interconnection and pedicle/sacral bone screw posterior spinal systems.

Design of the rods and pedicle screws of the Webb-Morley Spine System do not represent a significant departure from long established designs in terms of safety or effectiveness. The Webb-Morley Spine System utilizes identical implant materials (stainless steel) meeting BS, ISO and ASTM standards.

Conclusion:

The design features of the Webb-Morley Spine System are equivalent to those offered by the TSRH® and PLSA systems. Components of the comparable systems are all made from the same or equivalent implant grade materials. Device testing, engineering review, and comparison of the system components to the referenced devices demonstrates their functional engineering equivalence .

The Webb-Morley Spine System is currently distributed in the US, Europe and elsewhere in the world. Performance of the Webb-Morley Spine System for severe spondylolisthesis of L5/S1 is expected to be comparable to the TSRH® and PLSA devices.

Since the indications for use, system limitations and surgical techniques are well established and understood, spinal implant systems using such basic design elements may reasonably be expected to perform predictably. EBI uses current Good Manufacturing Practice regulations, in-house SOPs and appropriate ASTM standards to assure that the Webb-Morley Spine System performs predictably and is Substantially Equivalent. A summary feature comparison table follows.

{4}------------------------------------------------

FEATUREWebb-Morley®TSRH® SystemPLSA SpinalSystemSubstantialEquivalence
SurgicalApproach:PosteriorPosteriorPosteriorYes
Materials:316 LVM stainless steelTitanium Alloy & 316LVM Stainless SteelTitanium AlloyYes
Indications forUse:Severe Spondylolisthesis (Gr.III & IV) L5-S1,Degenerative Disc Disease,Fracture DislocationSame<SameYes
ClinicalExperience:EuropeanUS & EuropeanUS & EuropeanYes
Hooks:Laminar hooks, pedicle hooks,transverse process hooksSame TypesSame TypesYes
Rods4.0mm round & 4.0mm X6.0mm oval1/4" & 3/16" rods1/4" & 3/16"Yes
Pedicular &Sacral Screws:Cancellous threads5.5mm, 6.5mm & 7.5mmSameSameYes
Meets BS, ISO &ASTM Standards:BS-7252, ISO-5832-1 & ASTMF-138-86F-138-86, Gr. 2 & F-136-92F-136-92Yes
TransverseInterconnection:YesYesYesYes
SterilizationMethod:Steam Sterilization perHIMA/AORNSameSameYes
Manufacturer:Electro-Biology, Inc.Sofamor/DanekAdvanced SpineFixation Systems,Inc.Yes
ClassificationCode:87KWP & 87MNH87KWP & 87MNH87KWP & 87MNHYes
K-NumberNAK932029K945064NA

FEATURE COMPARISON TABLE

:

N/A