K Number
K960446
Device Name
MODEL 5471 & 5471L STERILE DISPOSABLE PATIENT CABLE ASSEMBLIES
Manufacturer
Date Cleared
1996-05-22

(111 days)

Product Code
Regulation Number
870.2900
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The intended use of the patient cables is identical to previous patient cables.
Device Description
The proposed patient cables are identical to the commercially-approved Models 5455, 5455L, 5455S, and 5455SL surgical cables, with the exception of a modified connector for heart wires and leads.
More Information

5455, 5455L, 5455S, 5455SL

Not Found

No
The summary describes patient cables with a modified connector and focuses on mechanical, electrical, and environmental testing, with no mention of AI or ML.

No
The device described is a patient cable, which is an accessory used with other medical devices. It does not directly provide therapy. The "surgical cables" mentioned as predicates further suggest a connection to surgical equipment rather than therapeutic intervention itself.

No

Explanation: The device is described as "patient cables" specifically "surgical cables" with a modified connector for heart wires and leads. The summary of performance studies focuses on mechanical, electrical, environmental, and biocompatibility tests of the cables, not on diagnostic capabilities. The text does not mention any function related to identifying a disease or condition.

No

The device description explicitly states it is a "patient cable" and describes physical modifications and performance studies related to mechanical, electrical, and environmental properties, indicating it is a hardware device.

Based on the provided text, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • No mention of in vitro testing: The description focuses on patient cables used in a surgical setting, connecting to heart wires and leads. There is no indication of the device being used to examine specimens (like blood, urine, or tissue) outside of the body.
  • Focus on electrical and mechanical performance: The performance studies described are related to the physical and electrical properties of the cables, not the analysis of biological samples.
  • Intended use in a surgical setting: The device is described as "surgical cables," which are used directly on or with a patient during a procedure, not for laboratory analysis.

IVD devices are specifically designed to perform tests on samples taken from the human body to provide information for diagnosis, monitoring, or screening. This description does not align with that purpose.

N/A

Intended Use / Indications for Use

The intended use of the patient cables is identical to previous patient cables.

Product codes

Not Found

Device Description

The proposed patient cables are identical to the commercially-approved Models 5455, 5455L, 5455S, and 5455SL surgical cables, with the exception of a modified connector for heart wires and leads.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Identical

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

Mechanical, electrical, and environmental/packaging tests show that the proposed patient cable assemblies are qualified and raised no new questions with regard to safety and efficacy. Mechanical test results for cable retaining force, lead, and heartwire connection withdrawal forces, and button depression all met requirements. Electrical tests for current, resistance, and dielectric withstanding voltage also met requirements. Environmental/packaging tests demonstrated that the packaging can withstand various shock, vibration, temperature, and humidity conditions. Biocompatibility tests showed that the connector does not produce cytotoxicity or dermal irritation. All tests met their respective requirements.

Key Metrics

Not Found

Predicate Device(s)

5455, 5455L, 5455S, and 5455SL

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.2900 Patient transducer and electrode cable (including connector).

(a)
Identification. A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.(b)
Classification. Class II (performance standards).

0

MAY 2 2 1996

K960446

510(k) SUMMARY [As required by 21 CFR 807.92(a)]

A. Submitter Information

Submitter's Name:Medtronic, Inc.
Address:7000 Central Ave. N.E.
Minneapolis, MN 55432-3576
Telephone Number:612/574-4654
Fax Number:612/574-6424
Contact Person:Valerie K. Stuhr
Submission
Preparation Date:January 31, 1996
Device Information
Trade Name:Models 5471 and 5471L Sterile Disposable Patient Cable
Assemblies
Common Name:Patient Cable
Classification Name:Patient transducer and electrode cable (including
connector), Class II (21 CFR 870.2900)
  • This 510(k) summary iclentifies the Models 5455, 5455L, Predicate Device: 5455S, and 5455SL surgical cables as substantially equivalent to the proposed patient cable assemblies.

1

  • Device Description: The proposed patient cables are identical to the commercially-approved Models 5455, 5455L, 5455S, and 5455SL surgical cables, with the exception of a modified connector for heart wires and leads.
  • Intended Use: The intended use of the patient cables is identical to previous patient cables.
  • C. Comparison of Required Technological Characteristics

The proposed patient cables are substantially equivalent to the commerciallyavailable Models 5455, 5455L, 5455S, and 5455SL surgical cables. The proposed cables are identical to the current cables except that the proposed cables include a modified connector.

CharacteristicsCurrent and Proposed
a. Product LabelingSubstantially equivalent
b. Intended UseIdentical
c. Physical CharacteristicsSubstantially equivalent
d. Anatomical SitesIdentical
e. Target PopulationIdentical
f. Performance TestingSubstantially equivalent
g. Safety CharacteristicsSubstantially equivalent
  • D. Summary of Nonclinical Tests
    Mechanical, electrical, and environmental/packaging tests show that the proposed patient cable assemblies are qualified and raised no new questions with regard to safety and efficacy. Mechanical test results for cable retaining force, lead, and heartwire connection withdrawal forces, and button depression all met requirements. Electrical tests for current, resistance, and dielectric withstanding voltage also met requirements. Environmental/packaging tests demonstrated that the packaging can withstand various shock, vibration, temperature, and humidity conditions. Biocompatibility tests showed that the connector does not produce cytotoxicity or dermal irritation. All tests met their respective requirements.