K Number
K960441
Date Cleared
1997-05-08

(462 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Steritec Integraph is designed to provide an integrated response to steam sterilization at 132 C for sterilizers operating for 3 1/2 minutes or longer.

Device Description

The Steritec Integraph CI 101 is a chemical indicator/integrator equivalent in construction and operation to the ATT Steam Close (predicate device).

AI/ML Overview

This is a summary of the provided text regarding the 'SteriTec Integraph' device and its performance testing.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied)Reported Device Performance
Proper indication of sterilization conditions (Pass/Fail)"In all cases they indicated properly whether the sterilization conditions had been met or not in the sterilizer chamber as verified by biological spore strips."
No false "PASS" indications when sterilization conditions were NOT met"There were NO instances in which the SteriTec strips indicated that sterilization conditions were met when they were not (NO FAI SI PASS RESULTS)."
Margin of safety (no "PASS" result if biological spores survive)"No SteriTec strip tested provided a "PASS" test result in tests when there were any biological spore survivors."
Stability after storage in high temperature"Independent laboratory testing showed the SteriTec Integraph is stable after being subjected to extremes of heat..."
Stability after storage in high humidity"...and high humidity."
Stability after 2.5 years of storage"For stability after storage for two and one half years years."
Stability after being sterilized and then stored for 2 years"For stability after being sterilized and then stored for two years."
Maintenance of original color (purple) during storage"All strips maintained their purple color after storage testing."
Proper color change (purple to green) after sterilization"After exposure to sterilization conditions, the indicators properly changed color from purple to green."

2. Sample Size for the Test Set and Data Provenance

  • Sample Size: The document does not specify the exact sample size (number of Integraph strips) used for the performance testing. It refers to "tests" and "strips tested" in plural.
  • Data Provenance: The document does not explicitly state the country of origin. It indicates that "Independent laboratory testing" was conducted, but the location of these labs is not specified. The studies appear to be prospective, as the device was actively subjected to various sterilization conditions and storage studies.

3. Number of Experts and Qualifications for Ground Truth

  • The document does not mention the use of human experts to establish ground truth.

4. Adjudication Method for the Test Set

  • No human adjudication method (e.g., 2+1, 3+1) was used. The performance was verified directly against "biological spore strips."

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed. The device is a chemical indicator, not an AI system requiring human interpretation comparison.

6. Standalone Performance

  • Yes, a standalone performance assessment was conducted. The SteriTec Integraph's performance was evaluated based on its own chemical reaction and color change in response to sterilization conditions, as verified by biological indicators. No human-in-the-loop performance is described.

7. Type of Ground Truth Used

  • The primary ground truth used was biological indicators (biological spore strips). These strips are the accepted standard for verifying the effectiveness of sterilization processes. The device's color change was compared directly against the survival/kill of spores on these biological indicators.

8. Sample Size for the Training Set

  • The concept of a "training set" is not applicable here as the SteriTec Integraph is a chemical indicator, not a machine learning or AI algorithm that requires training data. Its performance is based on its inherent chemical properties, not learned through data.

9. How Ground Truth for the Training Set Was Established

  • This question is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

K960441

SteriTec Products Mfg. Co,, Inc. Steritec Integraph Integrators FDA 510(k) Application-K960441 Original Submission-January 24, 1996 Current Date-May 2, 1997

510(k) SUMMARY

,

MAY - 8 1997

510(k) PREMARKET NOTIFICATION SUMMARY

510(k) Premarket Notification Summary

SUBMITTER:

  • STERITEC PRODUCTS MFG. CO,, INC. ● 680 Atchison Way - Suite 600 Castle Rock, CO 80104 (303) 660-4201 Establishment Registration Number: 2028456 ● Date _ May 2, 1997 ● C TOM ROLL Printed name of person required to submit 510(k) Signature of person required to submit 510(k) PRESIDENT ● Title of person submitting 510(k) Proprietary Name: Integraph Common/Usual Name: Steam Sterilization Indicator/Integrator Classification Name: Sterilization Indicator
    Classification:

FDA has classified Physical/Chemical Sterilization Indicators in Class II under Classification Number 80JOJ, Regulation Number 880.2800.

{1}------------------------------------------------

PB2

SteriTec Products Mfg. Co., Inc. Steritec Integraph Integrators FDA 510(k) Application-K960441 Original Submission-January 24, 1996 Current Date-May 2, 1997

510(k) SUMMARY

Identification of Predicate devices

The Steritec Integraph CI 101 is a chemical indicator/integrator equivalent in construction and operation to the ATT Steam Close (predicate device). Since the Steritec Integraph is equivalent in construction (o the predicate device, it will perform equivalently .

Description of 510(k) submission device:

Intended use:

The Steritec Integraph is designed to provide an integrated response to steam sterilization at 132 C for sterilizers operating for 3 1/2 minutes or longer.

Comparison to Predicate device:

Compared to the predicate device our performance data shows the Integraph Integrator to he substantially coulvalem

Performance Testing:

To demonstrate the performance of the SteriTec Integraph the product was tested:

  • At multiple sterilization temperatures and times against biological (1) indicators .
  • For stability after storage in high temperature conditions. (2)
  • For stability after storage in high humidity conditions. (3)
  • For stability after storage for two and one half years years. (4)
  • For stability after being sterilized and then stored for two years. (ર)

No false "PASS" Indications

In all cases they indicated properly whether the sterifization conditions had been met or not in the sterilizer chamber as verified by biological spore strips. There were NO instances in which the SteriTec strips indicated that sterilization conditions were met when they were not (NO FAI SI PASS RESULTS).

Margin of Safety

Test results show the SteriTce strips provide a "murgin of safely". During these tests, No SteriTec strip tested provided a "PASS" test result in tests when there were any biological spore survivors.

{2}------------------------------------------------

SteriTec Products Mfg. Co., Inc. Steritec Integraph Integrators FDA 510(k) Application-K960441 Original Submission-January 24, 1996 Current Date-May 2, 1997

510(k) SUMMARY

18

were any biological spore survivors.

Stability Testing:

Independent laboratory testing showed the SteriTec Integraph is stable after being subjected to extremes of heat and high humidity. In addition, under room temperarture storage of both exposed and unexposed strips for 2 years the strips also performed satisfactorily. All strips maintained their purple color after storage testing. After exposure to sterilization conditions, the indicators properly changed color from purple to green.

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).