K Number
K960280
Manufacturer
Date Cleared
1996-04-04

(76 days)

Product Code
Regulation Number
880.5440
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use for the IVAC Needle Free Valve administration sets is to allow IV fluid delivery of a wide range of fluids. The needle free valve allows the user to add medication into the primary line without the use of a needle.

Device Description

The IVAC Needle Free Valve administration sets incorporate standard IVAC set components using standard materials. The sets include a needle free valve that allows the user to add medication into the primary line without the use of a needle. When the valve is in the closed position it has a flat smooth surface for cleaning. When the male connector of a syringe or secondary line is pushed into the valve the rubber piston inside the valve is compressed. The top of the rubber piston is then allowed to expand into the increased diameter inside the valve. Fluid path is achieved through the male connector into the center of the rubber piston and the valve housing. When the male connector is removed from the valve, the bellows section of the piston provides a spring force to return the piston to the top of the valve and to maintain a sealed fluid path. A cap is not required to maintain sterility.

AI/ML Overview

This 510(k) premarket notification for the IVAC Needle Free Valve Administration Sets describes a medical device and its substantial equivalence to a predicate device, the ICU Medical, Inc. Clave Connector. However, it does not contain information about specific acceptance criteria, device performance metrics, or any studies designed to prove the device meets such criteria.

The submission focuses on demonstrating substantial equivalence based on technological characteristics, intended use, and materials. It does not provide the kind of detailed performance study information typically required for describing acceptance criteria and a study proving their fulfillment in the context of device performance, as outlined in your request.

Therefore, I cannot extract the following information from the provided text:

  1. A table of acceptance criteria and the reported device performance: No such table or performance data is presented.
  2. Sample sized used for the test set and the data provenance: No test sets or study data are described.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No ground truth establishment or expert involvement is mentioned.
  4. Adjudication method for the test set: No test sets or adjudication are described.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This device is not an AI-driven diagnostic tool, and no MRMC study is mentioned.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: This is not an algorithmic device.
  7. The type of ground truth used: No ground truth is discussed.
  8. The sample size for the training set: This is not an AI/machine learning device; no training set exists.
  9. How the ground truth for the training set was established: Not applicable.

The document is a regulatory submission for substantial equivalence based on material and functional similarity to an existing device, not a performance study report with detailed acceptance criteria and testing results.

{0}------------------------------------------------

K960280

APR - 4 1996

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

    1. SUBMITTER'S NAME: IVAC CORPORATION 10300 Campus Point Drive San Diego, CA 92121-1579 (619) 458-7825
    • CONTACT PERSON: Jennifer S. Hankard Sr. Manager of QA/RA

DATE PREPARED: January 17, 1996

    1. DEVICE NAME: Proprietary Name: IVAC Needle Free Valve Administration Sets
      Common Name: Intravascular Administration Set

Classification Name: Intravascular Administration Set

    1. PREDICATE DEVICE: ICU Medical, Inc. Clave Connector
  • ব . DEVICE DESCRIPTION: The IVAC Needle Free Valve administration sets incorporate standard IVAC set components using standard materials. The sets include a needle free valve that allows the user to add medication into the primary line without the use of a needle. When the valve is in the closed position it has a flat smooth surface for cleaning. When the male connector of a syringe or secondary line is pushed into the valve the rubber piston inside the valve is compressed. The top of the rubber piston is then allowed to expand into the increased diameter inside the valve. Fluid path is achieved through the male connector into the center of the rubber piston and the valve housing. When the male connector is removed from the valve, the bellows section of the piston provides a spring force to return the piston to the top of the valve and to maintain a sealed fluid path. A cap is not required to maintain sterility.
    1. INTENDED USE: The intended use for the IVAC Needle Free Valve administration sets is to allow IV fluid delivery of a wide range of fluids. The needle free valve allows the user to add medication into the primary line without the use of a needle.

{1}------------------------------------------------

6. TECHNOLOGICAL CHARACTERISTICS:

The IVAC Needle Free Valve administration sets and the ICU Medical, Inc. Clave Connector use the same method of manufacture, incorporate standard set components, and use standard materials. The primary function of the IVAC Needle Free Valve administration sets is the same as the ICU Medical, Inc. Clave Connector: to allow the user to add medication into the primary line without the use of a needle. The technological characteristics of the devices are the same and raise no new questions of safety and effectiveness. IVAC Corporation concludes that the IVAC Needle Free Valve administration sets are substantially equivalent to the ICU Medical, Inc. Clave Connector.

§ 880.5440 Intravascular administration set.

(a)
Identification. An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.(b)
Classification. Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.