K Number
K955324
Manufacturer
Date Cleared
1996-12-20

(396 days)

Product Code
Regulation Number
868.5895
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BIPAP S/T-D 30 System is intended for hospital or institutional use for treatment of obstructive sleep apnea, respiratory failure, or respiratory insufficiency.

Device Description

The BiPAP S/T-D 30 System is a ventilation support system designed to augment a patient's ability to breathe on a spontaneous basis. It is not intended for life support situations. The BiPAP S/T-D 30 System is intended for hospital or institutional use for treatment of obstructive sleep apnea, respiratory failure, or respiratory insufficiency. The BiPAP S/T- D offers four modes of operation: Continuous Positive Airway Pressure (CPAP), Spontaneous (S) mode, Spontaneous/Timed (S/T) mode, and Timed (T) mode. The BiPAP S/T-D 30 System provides ventilation support by applying a positive pressure (IPAP) during the inspiratory segment of a breathing cycle, and subsequently cycling to a lower pressure level (EPAP) during the expiratory segment of the cycle for spontaneous breaths. The transition point between inspiration and expiration is sensed by the BiPAP System as a function of a decrease in patient flow demand. The BiPAP S/T-D 30 System also senses the transition point between expiration and inspiration. As the patient begins to inhale, the BiPAP System senses the increase in flow and transitions from EPAP to IPAP. These points are called the trigger thresholds. The BiPAP S/T-D 30 System monitors delivered pressure and adjusts flow as applied to the patient circuit to maintain set IPAP and EPAP pressures. The System is also capable of compensating for most leaks around the patient interface.

AI/ML Overview

The provided text describes the BiPAP S/T-D 30 System and its performance, but it does not contain information about the acceptance criteria or a study proving that an AI device meets those criteria. The document is a 510(k) summary for a medical device (a ventilator), not an AI/ML diagnostic tool. Therefore, I cannot provide the requested table and information about AI model validation.

Here's a breakdown of the information that is present, organized to show what could be extracted if this were an AI device, and what is missing because it's a traditional medical device:

1. Table of Acceptance Criteria and Reported Device Performance

If this were an AI device, this table would typically show metrics like sensitivity, specificity, AUC, F1-score, etc., alongside their acceptance thresholds.

Acceptance Criteria (Not AI-specific)Reported Device Performance (Non-AI)
Pressure Regulation/Flow Levels: (Acceptance criteria not explicitly stated as a pass/fail threshold, but implied to be met by the given performance)+.8, -1.5 cm H₂O above setpoint over flow rates of -60 to 100 LPM
Control Accuracy, IPAP and EPAP Settings: (Acceptance criteria not explicitly stated as a pass/fail threshold, but implied to be met by the given performance)±2 cm H2O of setting in the range of 4 to 30 cm H₂O
Control Setting Accuracy, Breaths Per Minute (BPM): (Acceptance criteria not explicitly stated as a pass/fail threshold, but implied to be met by the given performance)± 2 BPM over range of 4 to 30 BPM, or ± 10% of setpoint, whichever is greater
Control Setting Accuracy %IPAP: (Acceptance criteria not explicitly stated as a pass/fail threshold, but implied to be met by the given performance)± 10% of setpoint measured at 90% and 10% IPAP settings
Electromagnetic Environmental Testing: (Acceptance criteria are compliance with listed standards)Compliant with CSA C22.2 125, UL544, German Postal Law Requirements per VFG 234/1991, IEC 601-1-2 (including IEC 801-1, -2, -3 and CISPIR 14), MIL-STD-461D, MIL-STD-462D, FCC Class B Part 15.J Emissions.
Shock / Vibration Testing: (Acceptance criteria are compliance with listed standards)Compliant with IEC 68-2-6, IEC 68-2-27 Shock, IEC 68-34 Random Wide Band Vibration, National Safe Transit Association Test Labs standards.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set:
    • 31 adult patients for prospective, randomized, controlled clinical studies.
    • 81 patients for prospective, non-randomized studies.
  • Data Provenance: The document does not specify the country of origin of the data. The studies were described as "prospective."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable/provided as the study described is a clinical study for a medical device (ventilator) and not an AI model requiring ground truth established by experts for image interpretation or similar. The "ground truth" for this device is the patient's respiratory condition and clinical outcomes.

4. Adjudication Method for the Test Set

This information is not applicable/provided as the study described is a clinical study for a medical device and not an AI model requiring adjudication for ground truth establishment.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No, an MRMC comparative effectiveness study was not done. This type of study is specifically relevant for diagnostic AI tools where human readers (e.g., radiologists) interpret cases with and without AI assistance. The document describes clinical studies comparing the device's effectiveness in treating respiratory conditions.
  • Effect Size of Human Readers Improve with AI vs. without AI Assistance: Not applicable, as no AI assistance was involved in this study.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not applicable. The "algorithm" in this context is the control system of a physical ventilator, not a standalone AI diagnostic algorithm. The device itself operates in a standalone manner as a ventilator, but this is distinct from an AI algorithm's standalone performance.

7. The Type of Ground Truth Used

  • Clinical Outcomes/Patient Condition: The "ground truth" for this study was the patient's acute respiratory condition and whether the device "successfully treat[ed] these patients." This is derived from clinical observation and patient health status, not expert consensus on an image or pathology. The primary outcome measure was whether noninvasive ventilation using the BiPAP System could successfully treat patients with acute respiratory failure.

8. The Sample Size for the Training Set

  • Not applicable/provided. The document describes a traditional medical device, not an AI model that undergoes a "training set" phase.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable/provided. As it's not an AI model, there isn't a "training set" with established ground truth in the AI/ML sense.

{0}------------------------------------------------

Section 12 - 510(k) Summary

Image /page/0/Picture/3 description: The image shows the logo for Respironics Inc. The logo features a stylized human figure inside a triangle. Below the logo, the text "RESPIRONICS INC." is printed in a bold font, followed by the address "1001 Murry Ridge Drive, Murrysville, PA 15668".

K955324

DEC 20 1988

Date:

September 3, 1996

Respironics Inc. Manufacturer: 1001 Murry Ridge Drive Murrysville, PA 15668 Phone: (412) 733-0200

Francis X. Dobscha Official Contact: Manager, Regulatory Affairs

Device Name: BiPAP® S/T-D 30

Common Name: Continuous Ventilator, Passive Exhalation Port, Critical Care

Classification: 868.5895, 73 MNT

  • Predicate Device: Respironics BiPAP® (K883825)
    Respironics BiPAP® S/T-D System (K951264)

Detachable Control Panel (K905540)

Siemens Servo 900 (K811102)

Healthdyne Quantum™ Pressure Support Ventilator (K950331)

{1}------------------------------------------------

Description of the BiPAP S/T-D 30 System

The BiPAP S/T-D 30 System is a ventilation support system designed to augment a patient's ability to breathe on a spontaneous basis. It is not intended for life support situations. The BiPAP S/T-D 30 System is intended for hospital or institutional use for treatment of obstructive sleep apnea, respiratory failure, or respiratory insufficiency.

The BiPAP S/T- D offers four modes of operation:

  • Continuous Positive Airway Pressure (CPAP) .
  • Spontaneous (S) mode. The unit cycles between Inspiratory ● Positive Airway Pressure (IPAP) and Expiratory Positive Airway Pressure (EPAP) in response to the patient rate. The difference between IPAP and EPAP is the delivered pressure support level.
  • Spontaneous/Timed (S/T) mode. The unit cycles between . Inspiratory Positive Airway Pressure (IPAP) and Expiratory Positive Airway Pressure (EPAP) in response to the patient rate. The difference between IPAP and EPAP is the delivered pressure support level.
  • · Timed (T) mode. The unit cycles between Inspiratory Positive Airway Pressure (IPAP) and Expiratory Positive Airway Pressure (EPAP) levels based solely on set breaths per minute (BPM) and %IPAP Time Controls.

The BiPAP S/T-D 30 System provides ventilation support by applying a positive pressure (IPAP) during the inspiratory segment of a breathing cycle, and subsequently cycling to a lower pressure level (EPAP) during the expiratory segment of the cycle for spontaneous breaths. The transition point between inspiration and expiration is sensed by the BiPAP System as a function of a decrease in patient flow demand. The BiPAP S/T-D 30 System also senses the transition point between expiration and inspiration. As the patient begins to inhale, the BiPAP System senses the increase in flow and transitions from EPAP to IPAP. These points are called the trigger thresholds.

Levels of IPAP and EPAP can be adjusted by the clinician for individual patient requirements. The IPAP control setting is prohibited from being greater than EPAP by design of the control system. IPAP settings are always set equal to or greater than EPAP. By setting IPAP and EPAP levels the same, CPAP therapy is applied. If the EPAP level is set greater than the ambient atmospheric pressure (greater than 0), then a level of positive end expiratory pressure (PEEP) is applied to the patient's airway. The BiPAP S/T-D 30 System is a prescription device. IPAP and EPAP levels are determined by a physician or by a clinician under the direction of a physician.

The BiPAP S/T-D 30 System monitors delivered pressure and adjusts flow as applied to the patient circuit to maintain set IPAP and EPAP pressures. The System is also capable of compensating for most leaks around the patient

{2}------------------------------------------------

interface. Patient interfaces include specified nasal masks, full face masks covering both the mouth and nose, and a mouthpiece that is used to apply pressures orally.

Each patient interface system uses an exhalation leak to direct the patient's exhaled gasses to the atmosphere. This leak is incorporated either as an accessory valve or bleed port.

Intended Uses of the BiPAP S/T-D 30 System

The BIPAP S/T-D 30 System is intended for hospital or institutional use for treatment of obstructive sleep apnea, respiratory failure, or respiratory insufficiency.

The BiPAP S/T-D 30 System can be used to administer oxygen by using an external oxygen tank and bleeding oxygen into the patient circuit. This therapy is not metered or otherwise regulated. Patient monitoring must be performed to ensure effective administration of oxygen therapy.

The BiPAP S/T-D 30 System is similar to the BiPAP described in K883825, and other continuous ventilators, such as the Siemens Servo 900 and the Puritan Bennett 7200. However, there are specific differences between the BiPAP SIT-D 30 and these traditional life-support ventilators as noted below.

Comparison of the BiPAP to Predicate Devices

The Respironics BiPAP S/T-D 30 System is equivalent to the current BiPAP S/T-D System as described in K951264. The major difference between these two devices is that the BIPAP S/T-D System is limited to a maximum working pressure of 20cm H2O while the BiPAP S/T-D 30 System can be set for a maximum pressure of 30cm H2O.

The major technical difference between the BiPAP S/T-D 30 System and most continuous ventilators is in the technical approach used to achieve patient support ventilation. Like the Healthdyne Quantum PSV, the BiPAP S/T-D 30 System has a self-contained, internal blower assembly to generate airwav pressure. Most comparable predicate devices utilize an external pressure source.

The BiPAP S/T-D 30 System is an electromechanically controlled system that controls the pressure applied to the patient based on patient demand, while the flow is adjusted to maintain a set inspiratory and expiratory pressure. Changes between inspiratory phases and expiratory phases are determined by changes in patient flow patterns.

The BiPAP S/T-D 30 System has an electronic mechanism for compensating for minor leaks around the patient's mask or in the patient circuit.

{3}------------------------------------------------

The pressure supplied by the BiPAP S/T-D 30 System is constant within +0.8 to -1.5 cm H2O in the flow ranges of -60 to 100 Liters per Minute (LPM). (The -60 LPM designation represents patient exhalation into the BiPAP S/T-D 30 patient circuit.) Most comparable continuous ventilators require a manual operator adjustment to the flow rate control to compensate for leaks in the masks and other areas of the patient circuit.

There are differences in operational characteristics between the BiPAP S/T-D 30 system and predicate devices that are fully compliant with ASTM F 1100-90, the voluntary standard for critical care ventilators:

  • . The BiPAP S/T-D 30 System provides continuous positive airway pressure (CPAP) and positive pressure ventilation and is indicated for assisted ventilation of adults. This system does not provide mandatory ventilation with guaranteed tidal volume delivery. Patients requiring mandatory ventilation at predetermined tidal volumes are not candidates for pressure support or pressure-limited ventilation.
  • . The BiPAP S/T-D 30 System requires an intentional leak port, instead of an actively controlled exhalation valve to remove exhaled air from the circuit. Therefore, specific masks and circuits using an intentional leak port are required for normal operation. The pressurized air from the BiPAP S/T-D 30 System causes a continuous flow of air to exhaust from the leak port to flush the exhaled air from the circuit. The machine should be turned on and the intentional leak port should be checked before the mask is applied.
  • Pressurized air from the BiPAP S/T-D 30 System causes a continuous flow . of air to exhaust from the leak port to flush exhaled air from the circuit. The ability to completely exhaust exhaled air from the circuit is dependent upon the EPAP setting and I:E ratio. At low EPAP settings or with short expiratory times (e.g., high breathing rates) the leak rate through the intentional leak port may be inadequate to clear all exhaled gas from the circuit. Some rebreathing may occur.
  • A controlled method for setting and measuring inspired oxygen . concentrations cannot be accomplished while using the BiPAP S/T-D 30 System. Instead, oxygen administration is accomplished by using an external low flow oxygen source that is titrated into the mask. At a fixed flow rate of supplemental oxygen, the inspired oxygen concentration will vary, depending on the IPAP and EPAP settings, patient breathing pattern, mask fit and the leak rate. Continuous patient monitoring is recommended while administering oxygen.
  • The Airway Pressure Monitor is recommended for use while operating the . BiPAP S/T-D 30 System. The Airway Pressure Monitor will monitor proximal airway pressures and provide audible and visual alerts for high and low pressure conditions, and alert the user to a low internal battery.

{4}------------------------------------------------

and incidents when the monitor's power switch is inadvertently turned OFF while the BiPAP S/T-D 30 System is in use. Each alarm function should be checked before use by performing the Performance Verification procedure described in Chapter 8 of the Clinical Manual.

  • The BiPAP alarm system differs from some ventilators in the following . respects:
    • Although the low pressure alarm on the Airway Pressure Monitor will Λ sound in the event of a power failure, the BiPAP Svstem does not have an independent loss-of-power alarm.
    • The Airway Pressure Monitor cannot detect complete or partial Λ obstruction of the exhalation port.
    • The Airway Pressure Monitor will sound if pressures exceed the > user-set threshold, but there is no mechanism to relieve excess pressures generated in the event of a ventilator control system failure.
    • The BiPAP's alarms provide indication only while the alarm condition exists. There is no mechanism to retain the alarm message after a transient alarm condition returns to normal.

  • Unlike most ventilators, the oxygen inlet (located on the mask) will . accept general purpose tubing. Caution must be exercised to ensure that only oxygen is connected to the oxygen inlet.
  • . The BiPAP System can be set to extreme inverse I:E ratios. Use of such ratios may have adverse physiological consequences and are not recommended.

The BiPAP Systems have been functionally tested to meet their published specifications for applied pressure, patient airflow, under varying environmental conditions. Specifications that were met include:

BiPAP S/T-D 30 SystemEnvironmentalConditions
Pressure Regulation/Flow Levels+.8, -1.5 cm H₂O abovesetpoint over flow rates of -60to 100 LPMNominal 115 VAC132 VAC. 100 VAC,72-75° F Nominal105° F High
Control Accuracy, IPAPand EPAP Settings±2 cm H2O of setting in therange of 4 to 30 cm H₂ONominal temperatureand input voltage

{5}------------------------------------------------

Control SettingAccuracy, Breaths PerMinute (BPM)$\pm$ 2 BPM over range of 4 to 30BPM, or $\pm$ 10 of setpoint,whichever is greaterNominal temperatureand input voltage
Control Setting Accuracy%IPAP$\pm$ 10% of setpoint measured at90% and 10% IPAP settingsNominal temperatureand input voltage

Electromagnetic environmental testing and shock / vibration testing were conducted by independent laboratories to the following National and International Standards for electrical safety for medical devices:

  • · Canadian Standards Association (CSA) C22.2 125, as tested by CSA.
  • · UL544 Medical and Dental Equipment Standards, as tested by the City of LA Test Labs.
  • · German Postal Law Requirements for Conducted and Radiated Emissions per VFG 234/1991, as tested by EMACO for TüV.
  • · Shock and Vibration testing per IEC 68-2-6 (sinusoidal vibration), IEC 68-2-27 Shock, and IEC 68-34 Random Wide Band Vibration, as tested by East West Labs.
  • · Shock and vibration for the final packaged configuration per National Safe Transit Association Test Labs.
  • · Electromagnetic compatibility testing to IEC 601-1-2 for medical electrical equipment. This is a standard that includes IEC 801-1, -2, -3 and CISPIR 14 requirements identified in the FDA Reviewer's Guidance Document. The BiPAP S/T-D 30 System was also tested to MIL-STD-461D and MIL-STD-462D for magnetic field susceptibility. EMI and magnetic field testing was conducted by Canadian Standards Association.
  • · FCC Class B, Part 15.J Emissions.

Results of the performance testing indicate that the BiPAP S/T-D 30 System will operate within its specifications.

Results of environmental testing indicate that the BiPAP S/T-D 30 System will continue to function after being subjected to shock and vibration, and electromagnetic interference.

Radiation emissions meet the requirement of FCC Class B. Part 15.J.

{6}------------------------------------------------

Clinical Study Summary

Clinical data are presented that verify the suitability of BiPAP Systems for use in treating adult patients with obstructive sleep apnea, respiratory insufficiency, or respiratory failure. These data indicate that the BiPAP S/T-D 30 System can be used in critical care situations with these patients to manage their acute respiratory condition. Data and analysis indicate with a high confidence (p<0.05) that there is a statistical basis to show that the BiPAP S/T-D 30 system can be used as an alternative to intubation in selected patients.

The clinical data includes 31 adult patients with acute respiratory failure who were enrolled in prospective, randomized, controlled clinical studies to determine, as a primary outcome measure, if noninvasive ventilation using the BiPAP System can successfully treat these patients. Additional data concerning 81 patients enrolled in prospective, non-randomized studies were included to substantiate the conclusions drawn from the controlled studies.

Adverse Effects

Adverse effects identified during the clinical studies include abrasions on the bridge of the nose due to tight fitting masks. These were resolved by readjusting the mask, using foam spacers, or applying a skin barrier.

There was one report of a patient who had respiratory failure as a result of being taken off of the BiPAP S/T-D 30 System to enable transport to another procedure. It is advised that an alternate method of ventilation be available when patients are transported.

Conclusions

The technical data presented indicate that the BiPAP S/T-D System is compatible with typical clinical or institutional environments. The BiPAP S/T-D 30 has been tested to meet technical safety and performance specifications for medical equipment as established by Canadian Standards Association and IEC.

Clinical data demonstrate that the BiPAP S/T-D 30 System can be used in the hospital environment and other institutional settings to provide ventilation to spontaneously breathing adult patients who suffer from obstructive sleep apnea, are in acute respiratory failure, or have acute respiratory insufficiency.

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).