K Number
K954932

Validate with FDA (Live)

Date Cleared
1996-08-21

(299 days)

Product Code
Regulation Number
866.2560
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BBL® MGIT™ Mycobacteria Growth Indicator Tube and BBL® MGIT™ supplemented with BBL® MGIT™ OADC and BBL® MGIT™ supplemented with BBL® MGIT™ PANTA antibiotic mixture, is intended for the detection and recovery of mycobacteria from decontaminated clinical specimens (except urine) and sterile body fluids (except blood). Acceptable clinical specimen types are digested and decontaminated.

Device Description

The BBL® MGIT™ Mycobacteria Growth Indicator Tube contains 4 mL of modified Middlebrook 7H9 Broth Base. The complete medium is supplemented with 0.5 mL OADC enrichment and 0.1 mL of PANTA antibiotic mixture, when necessary. The medium components are substances essential for the growth of mycobacteria. A fluorescent compound is embedded in the bottom of the BBL® MGIT™ tube which is sensitive to the presence of oxygen dissolved in the broth. Specimens are inoculated (0.5 mL) into the BBL® MGIT™ tube and incubated at the appropriate temperature. Tubes are read daily from the second day of inoculation. Actively respiring microorganisms consume the oxygen and allow the compound to fluoresce. Each inoculated BBL® MGIT™ tube is compared to the fluorescence Positive Control and Negative Control tube to assist in the interpretation of a positive signal from the BBL® MGIT™ tube. A BBL® MGIT™ tube which exhibits fluorescence comparable to the Positive Control tube is considered presumptively positive for growth.

AI/ML Overview

The provided text describes the BBL® MGIT™ Mycobacteria Growth Indicator Tube, intended for the detection and recovery of mycobacteria from clinical specimens. The device relies on an oxygen-sensitive fluorescent sensor that detects oxygen consumption by actively respiring microorganisms.

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity, specificity, or false positive rates. However, the study aims to show comparable performance to predicate devices. Therefore, the "acceptance criteria" appear to be implicit: the BBL® MGIT™ tubes should perform similarly to or better than the BACTEC® 460TB System and the BBL® SEPTI-CHEK® AFB Mycobacteria Culture System in terms of recovery rate and time to detection, while maintaining an acceptable false positive rate.

MetricAcceptance Criteria (Implicit: Comparable to Predicate)Reported BBL® MGIT™ Performance
Mycobacterial Recovery RateComparable to BACTEC® 460TB System (79%) and BBL® SEPTI-CHEK® AFB (79%)77% (253 of 330 isolates)
False Positive RateAcceptably low (no explicit numerical criterion, but likely aiming for minimal
                               false positives relative to clinical impact) | 0.5%                            |

| Time-to-Detection | Comparable to predicate devices (not explicitly stated, but implies being efficient | 14.1 days (average for all mycobacteria) |
| Unrecovered Isolates | Minimally higher than predicate devices (with further recovery via recommended second medium) | 3.7% (isolates recovered by reference, but not MGIT) |
| Breakthrough Contamination Rate | Acceptably low | 9.7% |

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set: A combined total of 2801 clinical specimens were evaluated.
  • Data Provenance:
    • Country of origin: Not explicitly mentioned, but the submitting company is Becton Dickinson Microbiology Systems, located in Sparks, MD, USA. It's highly probable the studies were conducted in the USA.
    • Retrospective or Prospective: Not explicitly stated. However, external evaluations are typically conducted prospectively to compare a new device against existing methods in real-world clinical settings. The description of "specimens were evaluated" and "isolates recovered during the study" suggests a prospective collection and testing over a period.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document does not specify the number of experts or their qualifications for establishing the ground truth. The ground truth was established by culture results from reference methods:

  • BACTEC® 460TB System
  • BBL® SEPTI-CHEK® AFB Mycobacteria Culture System
  • Conventional solid agar media

Presumably, trained laboratory personnel followed standard protocols for these reference systems to determine the presence or absence of mycobacteria and to identify isolates.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe an adjudication method involving human experts. The ground truth for mycobacterial presence was determined directly by the growth observed in the reference culture systems. If there was discordance between the BBL® MGIT™ and the reference, it was noted as an unrecovered isolate by the MGIT, not subjected to a separate expert adjudication process to determine a "true" positive or negative outside of the reference method's result.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No, this was not a multi-reader, multi-case (MRMC) comparative effectiveness study.
  • AI (Artificial Intelligence): The BBL® MGIT™ device is a culture-based diagnostic system, not an AI-assisted diagnostic tool. Therefore, there's no discussion of AI assistance or its effect size on human readers. The growth detection is described as "Manual observation of fluorescence via excitation with longwave UV light."

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The BBL® MGIT™ device itself is a standalone diagnostic tool in the sense that its detection mechanism (fluorescence) is independent of human interpretation for the initial signal. However, the final interpretation of a "presumptively positive" signal and subsequent identification of the mycobacteria would involve human intervention. The "manual observation of fluorescence" indicates human involvement in reading the fluorescence.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth used was culture-based detection and recovery of mycobacteria using established reference methods:

  • BACTEC® 460TB System
  • BBL® SEPTI-CHEK® AFB Mycobacteria Culture System
  • Conventional solid agar media

The presence of mycobacteria, and their specific isolates, as determined by growth in one or more of these reference systems, formed the ground truth against which the BBL® MGIT™ performance was compared.

8. The sample size for the training set

The document does not specify a separate "training set" in the context of machine learning or AI models. This device is a culture medium system, not a software algorithm that requires training data in that sense. The "internal testing" mentioned likely involved R&D and optimization of the medium components and detection method, rather than training a predictive model.

9. How the ground truth for the training set was established

As there was no explicit "training set" for an algorithm, the concept of establishing ground truth for it doesn't apply directly in this context. The "internal testing" would have relied on known mycobacterial strains and clinical samples where the presence/absence of mycobacteria was confirmed by standard laboratory methods.

{0}------------------------------------------------

Becton Dickinson Microbiology Systems 7 Loveton Circle Sparks, Maryland 21,152-0999 (410) 316-4000

K954932

Image /page/0/Picture/2 description: The image shows the words "BECTON DICKINSON" stacked on top of each other. The words are in a bold, sans-serif font and are black in color. The word "BECTON" is on the top line, and the word "DICKINSON" is on the bottom line. A line is drawn through the middle of the words.

510(k) SUMMARY

AUG 21 1996

BECTON DICKINSON MICROBIOLOGY SYSTEMS SUBMITTED BY: 7 Loveton Circle Sparks, MD 21152

Dennis R. Mertz, Manager of Regulatory Affairs CONTACT: (410)316-4099 TELEPHONE:

August 13, 1996 PREPARED:

BBL® MGIT™ Products TRADE NAME:

Selective Culture Medium COMMON NAME:

CLASSIFICATION Selective Culture Medium NAME:

BACTEC® 460TB System PREDICATE DEVICES: BACTEC 460TB System AFB Mycobacteria Culture System

The BBL® MGIT™ Mycobacteria Growth Indicator Tube INTENDED USE: The BBL MGTT™ Mycobaction Crown and BBL® MGIT™
supplemented with BBL® MGIT™ OADC and BBL® MGIT™ supplemented with BBL "IMST" - Greater is intended for PANTA antibiotic, mixture, whole provery of mycobacteria. Acceptable
the detection and recovery of mycobacterial clinical the detection and recovery of the seconaminated clinical specimen types are uigested and doesnament.
specimens (except urine) and sterile body fluids (except blood).

{1}------------------------------------------------

DEVICE DESCRIPTION:

The BBL® MGIT™ Mycobacteria Growth Indicator Tube contains 4 mL of modified Middlebrook 7H9 Broth Base. The complete medium is supplemented with 0.5 mL OADC enrichment and 0.1 mL of PANTA antibiotic mixture, when necessary. The medium components are substances essential for the growth of mycobacteria. A fluorescent compound is embedded in the bottom of the BBL® MGIT™ tube which is sensitive to the presence of oxygen dissolved in the broth.

Specimens are inoculated (0.5 mL) into the BBL® MGIT™ tube and incubated at the appropriate temperature. Tubes are read daily from the second day of inoculation.

Actively respiring microorganisms consume the oxygen and allow the compound to fluoresce. Each inoculated BBL® MGIT™ tube is compared to the fluorescence Positive Control and Negative Control tube to assist in the interpretation of a positive signal from the BBL® MGIT™ tube.

A BBL® MGIT™ tube which exhibits fluorescence comparable to the Positive Control tube is considered presumptively positive for growth.

{2}------------------------------------------------

DEVICE TECHNOLOGICAL CHARACTERISTICS:

Tables 1 & 2 summarize the similarities and differences between the BBL® MGIT™ tube and the predicate devices.

BBL® MGIT™BACTEC® 460TB
Intended UseGrowth and detection of mycobacteria from clinical specimens (excluding blood and urine).Growth and detection of mycobacteria from clinical specimens.
Sample TypePrimary sample type - respiratory; other body fluids acceptable (excluding blood and urine).Primary sample type - respiratory; other body fluids (excluding blood) acceptable.
Sample Volume0.5 mL0.5 - 1.0 mL
Reactive IngredientConcentrations ofGrowth MediumAPPROX. COMPOSITION/1000 mLAPPROX. COMPOSITION/1000 mL
Na2HPO4L-AsparagineKH2PO4Sodium glutamate(NH4)2SO4Sodium citrateMgSO4•7H2OFerric ammonium citrateCuSO4•5H2OPyroxidineZnSO4•7H2OBiotinCaCl2•2H2OCasein PeptoneGlycerolNa2HPO4KH2PO4Sodium glutamate(NH4)2SO4Sodium citrateMgSO4•7H2OFerric ammonium citrateCuSO4•5H2OPyroxidineZnSO4•7H2OBiotinCaCl2•2H2OCasein Hydrolysate
2.5 g1.251.00.50.50.10.50.041.0 mg1.01.00.50.51.25 g3.1 mL2.5 g1.00.50.50.10.050.041.0 mg1.01.00.50.50.1 g
Additional MediumGrowth FactorsBBL® MGIT™ OADC enrichment:Oleic Acid, Albumin, Dextrose, CatalaseAlbumin, Catalase
DetectorO2 sensitive fluorescent sensor in silicone rubber base.14C labeled fatty acid present in the medium.
AntimicrobialSupplementBBL® MGIT™ PANTA™ antibiotic mixture:Polymixin B, amphotericin B, nalidixic acid, trimethoprim & azlocillin.BACTEC® PANTA™ PLUS:Polymixin B, amphotericin B, nalidixic acid, trimethoprim & azlocillin.
Growth DetectionManual observation of fluorescence via excitation with longwave UV light.Fluorescence results from O2 consumption by mycobacterial growth.Radiometric detection of 14CO2 liberated by mycobacterial growth.
IncubationTemperature37° C ± 1° C37° C ± 1° C

Table 1: Comparison of BBL® MGIT™ to the BACTEC® 460TB System

{3}------------------------------------------------

BBL® MGITTMBBL® SEPTI-CHEK® AFB
Intended UseGrowth and detection of mycobacteria from clinical specimens.Growth and detection of mycobacteria from clinical specimens.
Sample TypePrimary sample type - respiratory; other body fluids acceptable (excluding blood and urine).Primary sample type - respiratory; other body fluids acceptable (excluding blood).
Sample Volume0.5 mL0.5 - 1.0 mL
Reactive IngredientConcentrations ofGrowth MediumAPPROX. COMPOSITION/1000 mLAPPROX. COMPOSITION/1000 mL
Na2HPO42.5 g1.579 g
L-Asparagine1.25
KH2PO41.01.579
Sodium glutamate0.5L-Glutamic acid 0.526
(NH4)2SO40.5(NH4)2SO4 0.526
Sodium citrate0.1Sodium citrate 0.421
MgSO4•7H2O0.5MgSO2•7H2O 0.053
Ferric ammonium citrate0.04Ferric ammonium citrate 0.042
CuSO4•5H2O1.0 mg
Pyroxidine1.0
ZnSO4•7H2O1.0
Biotin0.5Biotin 5.0 mg
CaCl2•2H2O0.5CaCl2•2H2O 0.5
Casein Peptone1.25 gCasein hydrolysate 1.053 g
Glycerol3.1 mL
NaCl 0.895
Malachite Green 1.0 mg
Additional MediumGrowth FactorsBBL® MGITTM OADC enrichment:Oleic Acid, Albumin, Dextrose, CatalaseBBL® SEPTI-CHEK® AFB MycobacterialCulture Supplement:Oleic Acid, Albumin, Dextrose, Catalase
Other Growth Mediaas Part of SystemNoneBBL® SEPTI-CHEK® AFB Slide:Middlebrook 7H11 AgarEgg based agar mediumChocolate Agar
DetectorO2 sensitive fluorescent sensor in siliconerubber base.None
AntimicrobialSupplementBBL® MGITTM PANTATM antibiotic mixture:Polymixin B, amphotericin B, nalidixic acid,trimethoprim & azlocillin.BBL® SEPTI-CHEK® AFB MycobacterialCulture Supplement:Polymixin B, amphotericin B, nalidixic acid,trimethoprim & azlocillin.
Growth DetectionManual observation of fluorescence viaexcitation with longwave UV light.Fluorescence results from O2 consumption bymycobacterial growth.Macroscopic observance of growth on agarand/or in broth.
IncubationTemperature37° C ± 1° C37° C ± 1° C

Table 2: Comparison of BBL® MGIT™ to BBL® SEPTI-CHEK® AFB

4

{4}------------------------------------------------

SUMMARY OF DEVICE TESTING:

Internal testing of the BBL® MGIT™ tubes demonstrated the ability to recover a wide variety of mycobacterial species. Additionally, internal testing showed comparable recovery between the BACTEC® 460TB System and the BBL® MGIT™ tubes.

An external evaluation was performed at six (6) sites. Four (4) sites used the B&CTEC® 460TB System and conventional solid agar media as reference methods for comparison to the BBL® MGIT™ tubes. Two (2) sites used the BBL® SEPTI-CHEK® AFB Mycobacteria Culture System as the reference method for comparison to the BBL® MGIT™ tubes. A combined total of 2801 specimens were evaluated. The distribution of specimens tested by source was respiratory (78%), gastric (0.4%), body fluid (9.8%), tissue (7.0%), stool (2.5%) and other (2.4%). A total of 318 specimens were positive which represented 330 isolates recovered during the study

Of these 330 isolates, 253 (77%) were recovered by the BBL® MGIT™ tubes, 260 (79%) were recovered by the BACTEC® 460TB System and the BBL® SEPTI-CHEK® AFB and 219 (66%) were recovered by conventional solid media. The average time-to-detection for the BBL® MGIT™ tubes for All mycobacteria was 14.1 days. The BBL® MGIT™ tubes demonstrated a 0.5% false positive rate (MGIT fluorescent, no AFB present). The BBL® MGIT™ tubes failed to recover 3.7% of the isolates which were recovered in one or more of the reference systems (BACTEC ® 460TB, BBL® SEPTI-CHEK® AFB or conventional solid media). While this percentage represents a potential loss of recovery, it is not indicative of an actual false negative determination (refer to "Limitations of the Procedure" section). Use of a second medium, as recommended, will increase the probability of recovery of mycobacterial organisms. The average breakthrough contamination rate for the BBL® MGIT™ tubes was 9.7%.

CONCLUSIONS

Based on the internal and external evaluation of the BBL® MGIT™ tubes, the overall performance of the BBL® MGIT™ tubes is comparable to the BACTEC® 460TB System and the BBL® SEPTI-CHEK® AFB Mycobacterial Culture System; therefore, we believe the BBL® MGIT™ tube to be substantially equivalent to these devices.

§ 866.2560 Microbial growth monitor.

(a)
Identification. A microbial growth monitor is a device intended for medical purposes that measures the concentration of bacteria suspended in a liquid medium by measuring changes in light scattering properties, optical density, electrical impedance, or by making direct bacterial counts. The device aids in the diagnosis of disease caused by pathogenic microorganisms.(b)
Classification. Class I. With the exception of automated blood culturing system devices that are used in testing for bacteria, fungi, and other microorganisms in blood and other normally sterile body fluids, this device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter.