K Number
K953269
Date Cleared
1996-12-20

(526 days)

Product Code
Regulation Number
868.5430
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SafeCircuit Manifold/Pulmonary Unit, is intended to scavenging of waste anesthetic gases from patients recovery for general anesthesia and provide supplemental oxygen. When used with the SafeCircuit Adhesive mask and a standard breathing circuit. the unit helps to reduce the amount of anesthetic aqents and airborne contaminants released to the work environment of the healthcare worker.

Device Description

The SafeCircuit ManifoldPulmonary Unit is comprised of a single manifold unit which has been divided into two (2) halves or functional chambers which function to: 1) Deliver supplemental oxygen during inhalation and, 2) Scavenge the patient's exhaled breath. The SafeCircuit Manifold / Pulmonary Unit incorporates directional valves similar to the directional valves of the standard CO2 absorber system of anesthesia gas machines and the Dryden disposable CO absorber to establish the inhalation flow of gases. In addition, as required in the Standard Specification for Anesthetic Equipment - Scavenging Systems for Anesthetic Gases. ASTM 1343-91. there must be a means to relieve excess positive and negative pressures. The SafeCircuit Manifold utilizes relief valves, similar to the Ohmeda Waste Gas unit and the Dryden SONRA, which will - 1. "Pop-off" or release any excess positive pressure on the exhalation side into the room should there be excess positive pressures. (Labeled EPPRV) 2. Entrain room air, should there be excess negative pressure from the vacuum so that this excess negative pressure is not directed toward the patient circuit. (Labeled ENPRV) In order to provide safe and effective oxygen delivery / inhalation unit, the use of excess positive and negative pressure relief valves are also incorporated in the inhalation side of the patient circuit. In this case they function to perform when - 1. Insufficient or no flow of oxygen occurs then, the patient can entrain room air, 21% oxygen through the excess negative pressure valve. (Labeled INPRV) 2. If there is excess oxygen flow, then the excess positive pressure valve will "pop-off" and the excess flow and / or pressure relieved to the outside. (Labeled IPPRV) The method of connecting the SafeCircuit Manifold to the patient is via a standard breathing circuit which is comprised of two (2) tubes, wye piece to connect to a mask and a reservoir bag to collect excess gases and allow for manual ventilation if required. This circuit would connect to the SafeCircuit Adhesive Mask, approved K950771, to provide an effective seal between the patient's face / airway and the SafeCircuit Manifold. As on the Waste Gas Interface Systems of Ohmeda a scavenging reservoir bag, with 19 / 30 mm fitting, is utilized, the SafeCircuit Manifold/ Pulmonary Unit will also have this overflow (expiratory) reservoir bag. It is intended to utilize a 30 mm fitting for this scavenge reservoir bag as the proposed international, ISO, standard is 30 mm. The unit has also incorporated filters at the back of the unit at each excess relief condition port opening to fitter the air released to the environment or entrained when excess positive or negative oressure conditions occur. These filters are to reduce any airborne contaminants, dust, dirt from the air entrained or released during exhalation. The SafeCircuit Manifold/Pulmonary Unit is designed to be mounted on the wall or on a pole so that the device can be conveniently located next to the patient and various wall outlets - oxygen and vacuum.

AI/ML Overview

The provided text describes a medical device, the "SafeCircuit Manifold / Pulmonary Unit," and compares its features to several predicate devices. It does not contain information about acceptance criteria, study methodologies, or performance metrics in a way that allows for the completion of the requested table and detailed study description. The document is primarily a device description and a comparison table of features with predicate devices, likely for a 510(k) submission.

Therefore, I cannot extract the requested information. The text does not elaborate on:

  1. Specific acceptance criteria: While it lists features and briefly mentions ASTM standards, it doesn't define quantitative acceptance criteria for performance.
  2. Reported device performance: The table only states "Yes" or "No" for features, not actual performance values (e.g., specific pressure relief levels, flow rates, scavenging efficiency percentage).
  3. Study type, sample size, or data provenance: There is no mention of any study conducted to prove the device meets performance criteria.
  4. Expert involvement, ground truth, or adjudication methods: These details are entirely absent, as there's no mention of a study involving human-in-the-loop or standalone algorithm performance.
  5. Training set information: This is irrelevant as no algorithm or AI is mentioned.

{0}------------------------------------------------

K953269

July 12, 1995

Advanced Warming Systems, Inc.3414 - 22nd St.Lubbock, TX 79410Tel - (806) 796-1246Fax - (806) 793-8030
Official Contact:Randy Hickle, M.D., President
Proprietary or Trade name:SafeCircuit Manifold / Pulmonary Unit
Common / Usual name:Gas scavenging apparatus
Classification name:Gas scavenging apparatus
Intended device:SafeCircuit Manifold / Pulmonary Unit
Predicate devices:Ohmeda and North American Drager Waste Gas ScavengingInterfaces; Dryden SONRA, Gibeck-Dryden CO2 absorber,Ohmeda CO2 absorber

Device description:

The SafeCircuit ManifoldPulmonary Unit is comprised of a single manifold unit which has been divided into two (2) halves or functional chambers which function to:

  1. Deliver supplemental oxygen during inhalation and,
    1. Scavenge the patient's exhaled breath.
      The SafeCircuit Manifold / Pulmonary Unit incorporates directional valves similar to the directional valves of the standard CO2 absorber system of anesthesia gas machines and the Dryden disposable CO absorber to establish the inhalation flow of gases.

In addition, as required in the Standard Specification for Anesthetic Equipment - Scavenging Systems for Anesthetic Gases. ASTM 1343-91. there must be a means to relieve excess positive and negative pressures. The SafeCircuit Manifold utilizes relief valves, similar to the Ohmeda Waste Gas unit and the Dryden SONRA, which will -

  1. "Pop-off" or release any excess positive pressure on the exhalation side into the room should there be excess positive pressures. (Labeled EPPRV)
  1. Entrain room air, should there be excess negative pressure from the vacuum so that this excess negative pressure is not directed toward the patient circuit. (Labeled ENPRV)

{1}------------------------------------------------

July 12, 1995

Device description (continued):

In order to provide safe and effective oxygen delivery / inhalation unit, the use of excess positive and negative pressure relief valves are also incorporated in the inhalation side of the patient circuit. In this case they function to perform when -

  1. Insufficient or no flow of oxygen occurs then, the patient can entrain room air, 21% oxygen through the excess negative pressure valve. (Labeled INPRV)
  1. If there is excess oxygen flow, then the excess positive pressure valve will "pop-off" and the excess flow and / or pressure relieved to the outside. (Labeled IPPRV)

The method of connecting the SafeCircuit Manifold to the patient is via a standard breathing circuit which is comprised of two (2) tubes, wye piece to connect to a mask and a reservoir bag to collect excess gases and allow for manual ventilation if required. This circuit would connect to the SafeCircuit Adhesive Mask, approved K950771, to provide an effective seal between the patient's face / airway and the SafeCircuit Manifold. As on the Waste Gas Interface Systems of Ohmeda a scavenging reservoir bag, with 19 / 30 mm fitting, is utilized, the SafeCircuit Manifold/ Pulmonary Unit will also have this overflow (expiratory) reservoir bag. It is intended to utilize a 30 mm fitting for this scavenge reservoir bag as the proposed international, ISO, standard is 30 mm.

The unit has also incorporated filters at the back of the unit at each excess relief condition port opening to fitter the air released to the environment or entrained when excess positive or negative oressure conditions occur. These filters are to reduce any airborne contaminants, dust, dirt from the air entrained or released during exhalation.

The SafeCircuit Manifold/Pulmonary Unit is designed to be mounted on the wall or on a pole so that the device can be conveniently located next to the patient and various wall outlets - oxygen and vacuum.

Intended use:

The SafeCircuit Manifold/Pulmonary Unit, is intended to scavenging of waste anesthetic gases from patients recovery for general anesthesia and provide supplemental oxygen. When used with the SafeCircuit Adhesive mask and a standard breathing circuit. the unit helps to reduce the amount of anesthetic aqents and airborne contaminants released to the work environment of the healthcare worker.

Indicated population and environment:

The SafeCircuit Manifold/Pulmonary Unit, is indicated for use with adult and pediatric patients recovering from general anesthesia. The environment is in the post anesthesia recovery room which may be located in a hospital, outpatient surgery center or equivalent facility where patients are observed and monitored following anesthesia.

{2}------------------------------------------------

July 12, 1995

AttributeSafeCircuitManifoldNorth AmericanDragerWaste GasInterfaceOhmedaWaste GasInterfaceDrydenSONRAGibeck-DrydenCO2 absorber
Use
Scavenging wasteanesthetic gasesYesYesYesYesN/A
Delivers gases likeoxygen to patientsYesN/AN/AYesYes
Indicated for adult / pediatricpatientsYesYesYesYesYes
Indicated for extended useYesYesYesYesNo
Design
Directional flow valvesinhalation / exhalationYesNoNoYesYes
Excess positive pressurereliefsYesYesYesYesN/A
Excess negative pressurereliefsYesYesYesYesN/A
Can entrain room airduring inhalationYesN/AN/AYesN/A
AttributeSafeCircuit ManifoldNorth American DragerWaste Gas InterfaceOhmedaWaste Gas InterfaceDryden SONRAGibeck-Dryden CO2 absorber
Can entrain room air to satisfyexcess negative pressure fromvacuumYesYesYesYesN/A
Has patient inhalation reservoirbagYesN/AN/AYesYes
Has overflow gas scavenging(exhalation) reservoir bagYesYesYesYesN/A
Connects to vacuum forscavengingYesYesYesYesNo
Patient circuit fittings 22 mmYesN/AN/AYesYes
Scavenging fittings 19 /30 mmYesYesYesYesYes
Requires periodic maintenanceYesYesYesYesN/A
Provided non-sterileYesYesYesYesYes
Has filters at excess pressurerelief port openingsYesNoNoNoN/A
Materials
Directional valvesSR, PE, PPN/APE, PPSRSR
AttributeSafeCircuitManifoldNorth AmericanDragerWaste GasInterfaceOhmedaWaste GasInterfaceDrydenSONRAGibeck-DrydenCO2 absorber
Housing / bodyABS, PC, PPPS, metalABS, metalmetalPC,PSPS,PC,ABS
22 mm / 15 mmASTM 1054-87YesYesYesYesYes
ASTM Scavenging1343-91YesYesYesYesN/A
5.3.1 Normal negativepressure < 0.5 cmYesYesYesYesN/A
5.3.2 Positive pressureambientYesYesYesYesN/A
5.3.3 Positive w/ flushNot applicable - no flush capabilites in intended environment
5.3.4 Positive pressureoccluded < 10 cmYesYesYesYesN/A
5.3.5 Positive, occluded,flushNot applicable - no flush capabilites in intended environment
5.4 Leakage < 100ml/minYesYesYesYesN/A
ISO 8835.3.2 draft 30 mmYesNoNoNoNo

page 5 of 117

r

{3}------------------------------------------------

July 12, 1995

:

page 6 of 117

{4}------------------------------------------------

.

:

July 12, 1995

page 7 of 117

ﺴﺴ

§ 868.5430 Gas-scavenging apparatus.

(a)
Identification. A gas-scavenging apparatus is a device intended to collect excess anesthetic, analgesic, or trace gases or vapors from a patient's breathing system, ventilator, or extracorporeal pump-oxygenator, and to conduct these gases out of the area by means of an exhaust system.(b)
Classification. Class II (performance standards).