K Number
K951799
Manufacturer
Date Cleared
1996-03-20

(336 days)

Product Code
Regulation Number
866.3900
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Light Diagnostics Varicella-zoster virus Direct Immunofluorescence Assay is intended for in vitro diagnostic use in the qualitative detection of VZV from vesicular smears and in cell culture viral isolation and confirmation.

Device Description

Light Diagnostics Varicella-zoster virus direct Immunofluorescence DFA) uses the standard laboratory direct Assav (VZV immunofluorescence technique for the culture confirmation and direct specimen detection of Varicella-zoster virus. The DFA is based on the principle of antigen identification using a detector monoclonal antibody conjugated to fluorescein isothiocyanate. The substrate consists of a slide prepared from a direct specimen vesicular smear or the tissue cultured cells from a clinical specimen inoculum. Anti VZV FITC labeled antibody is applied to the substrate. The antibody will bind to specific antigen, if present, in the substrate. The fluorescein conjugated monoclonal antibody allows for visualization of the antigen / antibody complex by fluorescence microscopy.

A blend of mouse monoclonal antibodies are used as detector antibodies in Light Diagnostics Varicella-zoster virus direct Immunofluorescence Assay. The monoclonal antibodies are specific for the glycoprotein gp I or the immediate early antigen of VZV. The use of monoclonal antibodies ensures increased specificity and reduced non-specific interference.

AI/ML Overview

The provided text is a 510(k) summary for the Light Diagnostics Varicella-zoster Virus Direct Immunofluorescence Assay, submitted in 1996. This summary predates the current rigorous standards for AI/ML device evaluations and thus many of the requested elements (especially those related to AI/ML specific studies, ground truth establishment, and multi-reader studies) are not applicable or not detailed in the provided document.

Here's an analysis of the available information:

1. Table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Direct Specimen Detection & Cell Culture Isolation/Confirmation:
Sensitivity (95% CI)91.3% to 100%
Specificity (95% CI)96.2% to 100%

2. Sample sized used for the test set and the data provenance

  • Sample Size for Test Set: 205 specimens
  • Data Provenance: The document does not specify the country of origin. It is a "clinical evaluation," implying the data is likely prospective at the time of the study, collected from patients presenting to a clinical setting.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The general practice for such assays is that the "ground truth" would be established by the comparator device (Meridian Diagnostics, Inc. Merifluor kit) or potentially by a combination of clinical diagnosis and culture, but the role of human experts in establishing this specific ground truth is not detailed.

4. Adjudication method for the test set

This information is not provided in the document. Given the nature of a direct immunofluorescence assay comparison, it's possible adjudication wasn't explicitly modeled in the same way as an AI/ML algorithm requiring human review. The comparison was against another established in-vitro diagnostic (IVD) kit.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not mentioned or conducted. This device is a manual laboratory assay, not an AI/ML device that assists human readers.

6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

The device itself is a standalone diagnostic assay. It is not an algorithm, and there is no "human-in-the-loop" AI component. The performance reported (sensitivity and specificity) is the standalone performance of the assay.

7. The type of ground truth used

The ground truth was established by comparison to the Meridian Diagnostics, Inc. Merifluor VZV in vitro diagnostic kit. This implies that the Merifluor kit's results were used as the reference standard for the clinical evaluation. It's also implied that culture confirmation was a part of the process, as the device is for "culture confirmation and direct specimen detection."

8. The sample size for the training set

This information is not applicable/provided. This is an immunoassay, not an AI/ML algorithm that requires a "training set" in the computational sense. The "training" of the assay involves the development and characterization of the monoclonal antibodies (as mentioned under "Nonclinical evaluation"), but this doesn't involve a distinct "training set" of patient data as understood in AI/ML.

9. How the ground truth for the training set was established

This information is not applicable/provided for the reasons stated in point 8. The development process involved "characterization for their ability to detect Varicella-zoster virus" using "reference virus strains and clinical isolates," and evaluation for "cross reactivity to a variety of viral pathogens and host cell controls." This is the process for ensuring the specificity and reactivity of the antibodies, rather than establishing ground truth for a training dataset.

{0}------------------------------------------------

K951799

510(k) SUMMARY

MAR 20 1996

    1. 510(k) Light Diagnostics 28835 Single Oak Drive Temecula, California 92590
      Contact Person: Dale Dembrow (909) 676-8080 Extension 230
    1. Device: Light Diagnostics Varicella-zoster Virus Direct Immunofluorescence Assay A direct immunofluorescence assay for the identification of Varicella-zoster virus
  • Substantial Equivalence to:

Meridian Diagnostics, Inc. Merifluor VZV

Varicella-zoster virus direct fluorescence antibody stain for VZV identification.

:

    1. Device Description:
      Light Diagnostics Varicella-zoster virus direct Immunofluorescence DFA) uses the standard laboratory direct Assav (VZV immunofluorescence technique for the culture confirmation and direct specimen detection of Varicella-zoster virus. The DFA is based on the principle of antigen identification using a detector monoclonal antibody conjugated to fluorescein isothiocyanate. The substrate consists of a slide prepared from a direct specimen vesicular smear or the tissue cultured cells from a clinical specimen inoculum. Anti VZV FITC labeled antibody is applied to the substrate. The antibody will bind to specific antigen, if present, in the substrate. The fluorescein conjugated monoclonal antibody allows for visualization of the antigen / antibody complex by fluorescence microscopy.

A blend of mouse monoclonal antibodies are used as detector antibodies in Light Diagnostics Varicella-zoster virus direct Immunofluorescence Assay. The monoclonal antibodies are specific for the glycoprotein gp I or the immediate early antigen of VZV. The use of monoclonal antibodies ensures increased specificity and reduced non-specific interference.

{1}------------------------------------------------

510(k) SUMMARY

5. Intended Use:

Light Diagnostics Varicella-zoster virus Direct Immunofluorescence Assay is intended for in vitro diagnostic use in the qualitative detection of VZV from vesicular smears and in cell culture viral isolation and confirmation.

    1. Technological Characteristics Comparison of Light Diagnostics Varicella-zoster virus Direct Immunofluorescence Assay and the Merifluor VZV kit:
    • Both kits use standard immunofluorescence assay a. Principle: technique using substrate slides prepared from vesicular smears or technique using substrate slides prep the cells from tissue culture isolation
    • Materials: Both kits incorporate monoclonal antibodies specific for b. VZV as the detector antibodies.

Performance Data for Light Diagnostics Varicella-zoster virus Direct Immunofluorescence Assay Kit

    1. Nonclinical evaluation:
      The conjugated monoclonal antibodies used in the kit were characterized for their ability to detect Varicella-zoster virus. The anti VZV antibodies reacted concordantly when tested with reference virus strains and clinical isolates.

The conjugated monoclonal antibodies were evaluated for cross reactivity to a variety of viral pathogens and host cell controls. No discordant results were obtained.

    1. Clinical Evaluation:
      Light Diagnostics Varicella-zoster virus Direct Immunofluorescence Assay was compared in clinical evaluation to the Meridian Diagnostics, Inc. Merifluor kit for direct specimen detection and cell culture isolation / confirmation.

Two hundred and five specimens were evaluated with results for 95% confidence interval sensitivity of 91.3% to 100% and specificity of 96.2% to 100%.

{2}------------------------------------------------

510(k) SUMMARY

Conclusions Drawn from Evaluations: 3.

Light Diagnostics Varicella-zoster virus Direct Immunofluorescence Assay Kit uses the standard laboratory DFA in the culture confirmation and direct detection of Varicella-zoster virus. The monoclonal antibodies used in the kit have been characterized to ensure maximum specificity and reliability. In clinical evaluations, the kit performed consistent with results obtained using the Meridian Diagnostics, Inc. Merifluor VZV in vitro diagnostic kit.

The characterization and clinical evaluation of Light Diagnostics Varicella-zoster Direct Immunofluorescence Assay verifies its safety and effectiveness when used as intended. . .

§ 866.3900 Varicella-zoster virus serological reagents.

(a)
Identification. Varicella-zoster virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to varicella-zoster in serum. The identification aids in the diagnosis of diseases caused by varicella-zoster viruses and provides epidemiological information on these diseases. Varicella (chicken pox) is a mild, highly infectious disease, chiefly of children. Zoster (shingles) is the recurrent form of the disease, occurring in adults who were previously infected with varicella-zoster viruses. Zoster is the response (characterized by a rash) of the partially immune host to a reactivation of varicella viruses present in latent form in the patient's body.(b)
Classification. Class II (performance standards).