K Number
K252105

Validate with FDA (Live)

Device Name
Ligence Heart
Manufacturer
Date Cleared
2025-09-26

(85 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi).

Device Description

Ligence Heart is an image post-processing software used for viewing and quantifying adult cardiac ultrasound DICOM studies. The device is intended to generate structured measurement reports for echocardiography analysis and aid qualified sonographers and cardiologists in their decision-making process.

Ligence Heart automatically identifies standard transthoracic echo views with machine-learning–based view classification, cardiac cycle selection, and border detection, then generates reproducible quantitative left-ventricular volumetric and functional measurements. The results are inserted into a PACS-compatible report that the reviewing cardiologist or sonographer can accept, edit, supplement with additional manual measurements, or entirely replace with manual measurements. The software also organizes, displays, and compares each measurement with reference-guideline ranges. Completed reports are exported in PDF, streamlining routine echocardiography workflow while leaving final diagnostic responsibility with the clinician.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving Ligence Heart meets them, based on the provided FDA 510(k) clearance letter:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Primary Clinical Performance Endpoint)Reported Device Performance
Individual Equivalence Coefficient (IEC): Upper level of the 95% confidence interval for agreement between Ligence Heart and expert sonographers < 0.25 (non-inferiority margin).All four parameters met the non-inferiority criterion, implying IEC < 0.25.
Secondary Agreement Metric (for consistency)
Intraclass Correlation Coefficients (ICC) for all four parameters.ICCs were $\ge$ 0.90 for all four parameters.

Study Details

1. Sample size used for the test set and the data provenance:

  • Sample Size: 524 echocardiographic studies. (Initially, 600 were acquired, but 76 were excluded).
  • Data Provenance: Retrospective, acquired from a U.S.-based independent Echocardiography Core Laboratory.

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Number of Experts: Three independent expert sonographers.
  • Qualifications of Experts: The document explicitly states "expert sonographers." No further specific details (e.g., years of experience, board certification) are provided in this excerpt.

3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • The document states, "524 echocardiographic studies were assessed by all three human readers..." This implies that all three contributed to the ground truth, but it doesn't explicitly state an adjudication method like 2+1 or 3+1. Given the use of "agreement between Ligence Heart and three independent expert sonographers...quantified using the reference-scaled IEC," it suggests that each expert's measurements were compared against the device, rather than a consensus ground truth being established before comparison with the device.

4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not explicitly done to measure human reader improvement with AI assistance. The study focused on the standalone performance of the Ligence Heart device and its interchangeability with expert human measurements. It did not directly assess how human readers' performance would change when using the AI as an assistant.

5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, a standalone performance study was done. The primary clinical performance endpoint measured the agreement between "Ligence Heart" (the automated measurements) and "three independent expert sonographers." This evaluates the device's output independently against expert human measurements.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Expert measurements/readings. The ground truth for the comparison was established by the measurements from "three independent expert sonographers."

7. The sample size for the training set:

  • The document explicitly states, "Test datasets were strictly segregated from algorithm training datasets, as they are from completely separate cohorts." However, the sample size for the training set is not provided in this excerpt.

8. How the ground truth for the training set was established:

  • The document states that the test datasets were "strictly segregated from algorithm training datasets." However, how the ground truth for the training set was established is not described in this excerpt. It can be inferred that it likely involved expert annotations or measurements, similar to the test set, but the details are not available here.

FDA 510(k) Clearance Letter - Ligence Heart

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.08.00

September 26, 2025

Ligence, UAB
℅ Raymond Kelly
Consultant
Arazy Group Consultants Inc.
3422 Leonardo Lane
New Smyrna Beach, Florida 32168

Re: K252105
Trade/Device Name: Ligence Heart
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: September 5, 2025
Received: September 5, 2025

Dear Raymond Kelly:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Page 2

K252105 - Raymond Kelly Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

Page 3

K252105 - Raymond Kelly Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K252105

Please provide the device trade name(s).

Ligence Heart

Please provide your Indications for Use below.

Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi).

Limitations:

  • Poor image capture will lead to poor annotations and subsequent measurements.
  • Multiple image quality algorithms are used to filter out images of poor quality.
  • Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions.
  • If no gender is present, female referenced guideline values will be used for conclusions.
  • If Body Surface Area (BSA) is not present, indexed values cannot be provided.
  • During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software.
  • Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves, implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated measurement accuracy.

Please select the types of uses (select one or both, as applicable).

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

Page 5

510(k) Summary

21 CFR 807.92

K252105

I. SUBMITTER (21 CFR 807.92(a)(1))

Submitter: Ligence, UAB
Taikos pr. 54 Kaunas,
51305, Lithuania

Phone: +37061144345

Contact Person: Alvaro Perez-Moreno, Ph.D.

Date Prepared: July 2, 2025

II. DEVICE (21 CFR 807.92(a)(2))

Name of Device: Ligence Heart
Common or Usual Name: Automated Radiological Image Processing Software
Classification Name: Medical Image Management and Processing System (§892.2050)
Regulatory Class: II
Product Code: QIH

III. PREDICATE DEVICE (21 CFR 807.92(a)(3))

Us2.v1, K210791

This predicate has not been subject to a design-related recall.

No reference devices were used in this submission.

IV. DEVICE DESCRIPTION (21 CFR 807.92(a)(4))

Ligence Heart is an image post-processing software used for viewing and quantifying adult cardiac ultrasound DICOM studies. The device is intended to generate structured measurement reports for echocardiography analysis and aid qualified sonographers and cardiologists in their decision-making process.

Ligence Heart automatically identifies standard transthoracic echo views with machine-learning–based view classification, cardiac cycle selection, and border detection, then

Page 6

generates reproducible quantitative left-ventricular volumetric and functional measurements. The results are inserted into a PACS-compatible report that the reviewing cardiologist or sonographer can accept, edit, supplement with additional manual measurements, or entirely replace with manual measurements. The software also organizes, displays, and compares each measurement with reference-guideline ranges. Completed reports are exported in PDF, streamlining routine echocardiography workflow while leaving final diagnostic responsibility with the clinician.

V. INDICATIONS FOR USE (21 CFR 807.92(a)(5))

Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi).

Limitations:

  • Poor image capture will lead to poor annotations and subsequent measurements.
  • Multiple image quality algorithms are used to filter out images of poor quality.
  • Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions.
  • If no gender is present, female referenced guideline values will be used for conclusions.
  • If Body Surface Area (BSA) is not present, indexed values cannot be provided.
  • During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software.
  • Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves, implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated measurement accuracy.

Page 7

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE (21 CFR 807.92(a)(5) & 21 CFR 807.92(a)(6))

Both the subject and predicate devices share the following technological characteristics:

Variations in automated-measurement implementation:

Ligence Heart automatically generates four core left-ventricular measurements essential for efficient and accurate routine cardiac assessments: lateral mitral-annular early-diastolic velocity (lateral e', Le), left-ventricular end-diastolic volume by the biplane Method of Disks (LVEDV MOD), left-ventricular ejection fraction by the biplane Method of Disks (LVEF MOD), and left-ventricular end-systolic volume by the biplane Method of Disks (LVESV MOD).

AttributePredicate Device: K210791Subject Device
Product CodeQIHQIH
Regulation892.2050892.2050
ClassificationIIII
Rx/OTCRxRx
Physical CharacteristicsSoftware package that operates utilizing off-the-shelf hardwareSoftware package that operates utilizing off-the-shelf hardware
User interfaceThe software is designed for use on a personal computer that has received images from a compatible PACSThe software is designed for use on a personal computer that has received images from a compatible PACS
Automation levelFully automated, including clip selectionFully automated, including clip selection
User confirmation / rejection of resultYesYes
Manual editing of automated result by userYesYes
Input DataEchocardiographic imagesEchocardiographic images
Communication protocolDICOM Standard ComplianceDICOM Standard Compliance
Patient PopulationsAdultsAdults

Page 8

VII. PERFORMANCE DATA (21 CFR 807.92(b))

Ligence Heart was developed and tested in accordance with Ligence Heart's life cycle development processes. The device has been subject to extensive safety and performance testing. Verification and validation test results established that the device meets its design requirements and intended use. Specifically, software verification was conducted at the unit, integration, and system levels. Risk management analysis generated multiple risk mitigation measures and verification activities. A cybersecurity analysis and testing were conducted to verify that security measures are properly implemented in Ligence Heart to protect Ligence Heart's data, including patient health information (PHI). A summative human factors study was performed to validate that Ligence Heart can be used safely and effectively by the intended user population.

A formal retrospective non-interventional validation study was conducted at a U.S. based independent Echocardiography Core Laboratory using 600 previously acquired echocardiograms to evaluate the performance of the device for each claimed measurement. Test datasets were strictly segregated from algorithm training datasets, as they are from completely separate cohorts.

A cumulative total of 76 studies were excluded because of missing views or corrupt metadata. Ultimately, 524 echocardiographic studies were assessed by all three human readers, satisfied the predefined quality criteria, and were included in the final analysis. Baseline demographic characteristics of the cohort are summarized in the table below.

Participants characteristics
Age, years49.48 ± 13.56
Male, n (%)319 (60.88%)
Female, n (%)205 (39.12%)
Height, cm173.36 ± 10.52
Weight, kg88.92 ± 21.11
LVEF, %34.45% ± 15.78%
Normal LVEF (>50%), n (%)120 (22.90%)
Reduced LVEF (<50%), n (%)404 (77.10%)

Page 9

Participants characteristics
Mildly reduced LVEF (40-50%), n (%)64 (12.21%)
Moderately reduced LVEF (30-40%), n (%)85 (16.22%)
Severely reduced LVEF (<30%), n (%)255 (48.66%)

Primary Clinical Performance Endpoint was Individual Equivalence Coefficient (IEC). Agreement between Ligence Heart and three independent expert sonographers was quantified using the reference-scaled IEC. Automated measurements are deemed interchangeable when the upper level of the IEC 95% confidence interval was < 0.25, a non-inferiority margin derived from individual-bioequivalence principles.

All four parameters met the non-inferiority criterion, confirming that Ligence Heart measurements are interchangeable with expert human measurements. Secondary agreement metrics were consistent with the IEC result. Intraclass Correlation Coefficients (ICC) were ≥ 0.90 for all four parameters. No device-related adverse events, malfunctions, or unmediated issues occurred during the study.

These combined non-clinical and clinical data demonstrate that Ligence Heart satisfies its specified performance criteria and support determination of substantial equivalence.

VIII. CONCLUSIONS (21 CFR 807.92(b))

Ligence Heart is substantially equivalent to the predicate device Us2.v1 (K210791). The subject and predicate share the same intended use, core technological characteristics, and fundamental principles of operation. No distinctions between the two systems do not introduce new questions of safety or effectiveness.

Comprehensive non-clinical testing, including software verification and validation, cybersecurity assessment, and summative human-factors validation, demonstrated that the software platforms as intended and can be used safely under the specified conditions. Complementary clinical validation confirmed that automated measurements of LVEF, LVEDV, LVESV, and lateral e´ are interchangeable with expert manual measurements and remain within normal inter-observer variability.

Taken together, the non-clinical and clinical data show that Ligence Heart is at least as safe and effective as the predicative device for post processing analysis of adult transthoracic echocardiograms.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).